1、 Tentative Interim Amendment NFPA99 Health Care Facilities Code 2015 Edition Reference: 14.2.9.1.1 TIA 15-3 (SC 16-11-4 / TIA Log #1241) Pursuant to Section 5 of the NFPA Regulations Governing the Development of NFPA Standards, the National Fire Protection Association has issued the following Tentat
2、ive Interim Amendment to NFPA 99, Health Care Facilities Code, 2015 edition. The TIA was processed by the Technical Committee on Hyperbaric and Hypobaric Facilities and the Correlating Committee on the Health Care Facilities Code, and was issued by the Standards Council on December 1, 2016, with an
3、effective date of December 21, 2016. A Tentative Interim Amendment is tentative because it has not been processed through the entire standards-making procedures. It is interim because it is effective only between editions of the standard. A TIA automatically becomes a public input of the proponent f
4、or the next edition of the standard; as such, it then is subject to all of the procedures of the standards-making process. 1. Revise paragraph 14.2.9.1.1 to read as follows: 14.2.9.1.1 Detectors, sensors, transducers, and communications equipment located inside the chamber shall meet the requirement
5、s of paragraph 14.2.8.3.16 for Class A chambers and paragraph 14.2.8.6 for Class B chambers. Issue Date: December 1, 2016 Effective Date: December 21, 2016 (Note: For further information on NFPA Codes and Standards, please see www.nfpa.org/docinfo) Copyright 2016 All Rights Reserved NATIONAL FIRE PROTECTION ASSOCIATION