1、ANSI/ASABE S613-4 AUG2017 Tractors and self-propelled machinery for agriculture Air quality systems for cabs Part 4: Performance test of a cab American Society of Agricultural and Biological Engineers ASABE is a professional and technical organization, of members worldwide, who are dedicated to adva
2、ncement of engineering applicable to agricultural, food, and biological systems. ASABE Standards are consensus documents developed and adopted by the American Society of Agricultural and Biological Engineers to meet standardization needs within the scope of the Society; principally agricultural fiel
3、d equipment, farmstead equipment, structures, soil and water resource management, turf and landscape equipment, forest engineering, food and process engineering, electric power applications, plant and animal environment, and waste management. NOTE: ASABE Standards, Engineering Practices, and Data ar
4、e informational and advisory only. Their use by anyone engaged in industry or trade is entirely voluntary. The ASABE assumes no responsibility for results attributable to the application of ASABE Standards, Engineering Practices, and Data. Conformity does not ensure compliance with applicable ordina
5、nces, laws and regulations. Prospective users are responsible for protecting themselves against liability for infringement of patents. ASABE Standards, Engineering Practices, and Data initially approved prior to the society name change in July of 2005 are designated as “ASAE“, regardless of the revi
6、sion approval date. Newly developed Standards, Engineering Practices and Data approved after July of 2005 are designated as “ASABE“. Standards designated as “ANSI“ are American National Standards as are all ISO adoptions published by ASABE. Adoption as an American National Standard requires verifica
7、tion by ANSI that the requirements for due process, consensus, and other criteria for approval have been met by ASABE. Consensus is established when, in the judgment of the ANSI Board of Standards Review, substantial agreement has been reached by directly and materially affected interests. Substanti
8、al agreement means much more than a simple majority, but not necessarily unanimity. Consensus requires that all views and objections be considered, and that a concerted effort be made toward their resolution. CAUTION NOTICE: ASABE and ANSI standards may be revised or withdrawn at any time. Additiona
9、lly, procedures of ASABE require that action be taken periodically to reaffirm, revise, or withdraw each standard. Copyright American Society of Agricultural and Biological Engineers. All rights reserved. ASABE, 2950 Niles Road, St. Joseph, Ml 49085-9659, USA, phone 269-429-0300, fax 269-429-3852, h
10、qasabe.org ANSI/ASABE S613-4 AUG2017 Copyright American Society of Agricultural and Biological Engineers 1 ANSI/ASABE S613-4 AUG2017 Approved August 2017 as an American National Standard Tractors and self-propelled machinery for agriculture Air quality systems for cabs Part 4: Performance test of a
11、cab Proposed by the ASABE Environment within Ag Vehicle Enclosures subcommittee; approved by the Power and Machinery Community and approved as an American National Standard August 2017. Keywords: Air quality, Cabs, Engineering control, Environment, Filters, Hazard Procedures, HVAC, Safety, Terminolo
12、gy, Test, Tractor, Ventilation 1 Scope 1.1 This part of the S613 standard series defines a performance test for a cab for use in a risk management program in contaminated environments as part of an Occupational Health and Safety Management System (OHSMS). This document is intended to be a guide for
13、engineers and field technicians who are responsible for cabs in agricultural applications. 1.2 While this standard provides guidance for use of these systems and defines a way to test the level of protection provided, as defined by the OHSMS. It is expected that other tests may provide cabs that are
14、 as well suited for this purpose; it is up to the user to demonstrate equivalency. 1.3 The S613 standard series is not intended to qualify cabs for use in poison gas or oxygen depleted environments, such as where an oxygen supply respirator is required per the product label and therefore are not cov
15、ered by this standard. 2 Normative References The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies unless noted. For undated references, the latest approved edition of the referenced document (including any am
16、endments) applies. ANSI/ASABE S613-1; Tractors and self-propelled machinery for agriculture Air quality systems for cabs Terminology and Overview ANSI/ASABE S613-2; Tractors and self-propelled machinery for agriculture Air quality systems for cabs Part 2: Cab Tractors and self-propelled machinery fo
17、r agriculture Air quality systems for cabs Part 3: Filters for environmental cab HVAC systems ISO 29463-1; High efficiency filters and filter media for removing particles from air Part 1 Classification, performance testing and marking ISO 14269-5:1997; Tractors and self-propelled machines for agricu
18、lture and forestry Operators enclosure environment Part 5: Pressurization system test method ANSI/ASABE S613-4 AUG2017 Copyright American Society of Agricultural and Biological Engineers 2 3 Maintenance Cab to be maintained per the manufacturers recommendations. 4 Cab Pressurization Any cab to be us
19、ed in contaminated environments shall have the cab pressurized per ISO 14269-5, with the exception of the operating conditions, in order to assure the operator is protected during use. The manufacturer shall specify the operating conditions at which the cab can meet this requirement. Following secti
20、ons renumbered due to addition of section 3 maintenance. 5 Particle Counting Test Method for Measuring Cab Integrity This tests the cab to measure the level of air leakage into the cab. When doing this procedure, check the outside of the cab for apparent leaks. 5.1 Replace the inlet filter with a HE
21、PA filter; one that is 99.95% efficient, or equivalent, reference ISO 29463, class 35H. As an option, a better performing filter may be used. 5.2 Remove any recirculation or secondary filters from the system 5.3 Test methods and conditions The testing shall be conducted using particle counters 5.3.1
22、 The cab to be tested shall be mounted on an end use machine 5.3.2 The testing shall be conducted when the background wind is less than 16 km/hr, 5.3.3 Tests shall be conducted when the relative humidity is below 80% 5.3.4 The unit shall be operated with the fan(s) on the specified blower setting pe
23、r the manufacturers instructions in the operators manual and with the engine at rated speed. NOTE: If the cab manufacturers operating condition (fan speed and cab pressure) without the recirculation filter is not met, the recirculation duct may need to be restricted so that the test condition matche
24、s the cab operating specifications. 5.3.5 Two particle counters shall be used, one measuring the concentration outside the cab, Co, at the cab air inlet area, the other measuring the concentration inside the cab, Ci, approximately at the drivers breathing area. The particles to be counted shall rang
25、e from 0.3 to 0.5 m (optical particle size diameter), as operated in test procedure section 6.1.1. NOTE: If problems occur due to a lack of an adequate number of particles in this range it may be necessary to expand the range of particles counted to include more particles in the range; i.e. a range
26、of 0.3 to 1.0 m. 5.3.6 Static resistive sampling tubes from the counters for the inside and outside measurements shall be the same material, inside diameter, and length. 5.3.7 A minimum of four 10-min. tests, after equilibrium has been reached, shall be conducted. 5.3.7.1 The test equipment (particl
27、e counters) measuring source shall be switched between inside and outside measurement after each minimum 5-min. measuring period during each of the above tests. ANSI/ASABE S613-4 AUG2017 Copyright American Society of Agricultural and Biological Engineers 3 5.3.7.2 The measuring equipment shall be op
28、erated at least one minute after start up for each test period to assure the lines are properly purged before any readings are taken. NOTE: The particle counters are switched to assure that any instrument measuring bias is nullified in the results. 5.3.8 Test Particle Concentration 5.3.8.1 Measure t
29、he particle concentration in the ambient air outside the cab, particles in the 0.3 to 0.5 m size range shall be of sufficient concentration so that the reduction factor, 50:1 or as necessary for the application, can be verified; the challenge particles shall be introduced at a level so that the inst
30、ruments are in their optimum operating range, as defined in section 7.2.1. NOTE: Generally at least 20,000 to 30,000 particles per liter are needed for repeatable testing. 5.3.8.2 Particle generation to supplement ambient conditions may be done by using an appropriate particulate generator to create
31、 particles which can be measured. Warning: Some materials may cause allergic type reactions. 5.4 Reduction factor The ratio of the outside concentration to the inside concentration (Co/Ci) is the reduction factor and shall be determined for each test. The four test results shall be averaged. The red
32、uction factor shall be defined by the managed program with a minimum reduction factor of 50:1 recommended for cab use in contaminated environments. If this test shows that desired reduction factor is not reached, the cab is not acceptable for use in contaminated environments and repairs shall be mad
33、e and be retested prior to use. NOTE: If not enough particles are present in the atmosphere to conduct this test conduct this test in an enclosure that isolates the machine from the atmospheric conditions adverse to the test so challenge particles can be introduced into the enclosure (see Section 6.
34、3 for the procedure.) 6 Particle counting test method for measuring cab filtration system performance This tests the entire operator protection system to verify the level of protection provided to the operator. 6.1 Test environment 6.1.1. Testing in ambient conditions The testing shall be conducted
35、using particle counters. 6.1.1.1 The cab to be tested shall be mounted on an end use machine and all filtration components in place. 6.1.1.2 The testing shall be conducted when the background wind is less than 16 km/hr. 6.1.1.3 Tests shall be conducted when the relative humidity is below 80%. 6.1.1.
36、4 The unit shall be operated with the fan(s) on the specified blower setting and pressurization level per the manufacturers instructions in the operators manual and with the engine at rated speed. 6.1.1.5 Two particle counters shall be used, one measuring the concentration outside the cab, Co, at th
37、e cab air inlet area, the other measuring the concentration inside the cab, Ci, approximately at the drivers breathing area. The particles to be counted shall range from 0.3 to 0.5 m (optical particle size diameter) 6.1.1.6 A minimum of four 10-min. tests, after equilibrium has been reached, shall b
38、e conducted. ANSI/ASABE S613-4 AUG2017 Copyright American Society of Agricultural and Biological Engineers 4 6.1.1.6.1 The test equipment (particle counters) measuring source shall be switched between inside and outside measurement after each minimum 5-min. measuring period during each of the above
39、tests. 6.1.1.6.2 The measuring equipment shall be operated at least one minute after start up for each test period to assure the lines are properly purged before any readings are taken. NOTE: The particle counters are switched to assure that any instrument measuring bias is nullified in the results.
40、 6.2 Test Particle Concentration 6.2.1 Measure the particle concentration in the ambient air outside the cab, particles in the 0.3 to 0.5 m size range shall be of sufficient concentration so that the reduction factor, 50:1 or as necessary for the application, can be verified; the challenge particles
41、 shall be introduced at a level so that the instruments are in their optimum operating range, as defined in section 7.2.1. NOTE: Generally at least 20,000 to 30,000 particles per liter are needed for repeatable testing. If problems occur due to a lack of an adequate number of particles in this range
42、 it may be necessary to expand the range of particles counted to include more particles in the range; i.e. a range of 0.3 to 1.0 m. 6.2.2 Particle generation to supplement ambient conditions may be done by using an appropriate particulate generator to create particles which can be measured. Warning:
43、 Some materials may cause allergic type reactions. 6.3 Testing in an enclosure 6.3.1 The test may be performed in a suitable enclosure, when there is not enough background contaminant in the ambient air, where outside interference such as wind, rain or other weather related phenomenon will not inter
44、fere with the testing. 6.3.1.1 The test enclosure shall have the engine exhaust vented outside if it effects test personnel and/or test results. 6.3.1.2 The test enclosure shall be supplied with suitable air moving equipment so that challenge particles introduced will be mixed with the air in the en
45、closure and a stable concentration of the challenge aerosols is maintained during the tests. 6.3.2 Perform the test as per 6.1.1. 7 Performance criteria 7.1 Reduction factor The ratio of the outside concentration to the inside concentration (Co/Ci) is the reduction factor and shall be determined for
46、 each test. The four test results shall be averaged. The reduction factor shall be defined by the managed program with a minimum reduction factor of 50:1 recommended for cab use in contaminated environments. 7.2 Statistical Verification Statistical analysis shall be used to examine testing validity.
47、 7.2.1 Test for concentration and test equipment The concentration of particles and volume of particle counts sensed shall be such that the loss of particle counts due to coincidence is less than 5%. Verify using the following formula relating true and indicated concentration: ANSI/ASABE S613-4 AUG2
48、017 Copyright American Society of Agricultural and Biological Engineers 5 where: Ni /Nt = coincidence error as specified for the instrument Ni is indicated particle concentration (particles/liter) Nt is true particle concentration (particles/liter) at the instrument manufacturers specified 5% coinci
49、dence error Criterion: Ni 0.95 Nt If this test is not passed, it indicates that there were too many particles in the test environment to run the test or the test equipment was encountering coincidence. 7.2.2 Test for experimental error The following calculations examine the experimental error and evaluates the acceptance criteria. For the desired Reduction Factor (RFdesired) reference ASABE S613 Part 1 Terminology) RFdesired RF stn 1,0.05 n is the acceptance criteria 1 n RF RFj n j 1 where: RF is the mean reduction factor and RFj are the
copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
备案/许可证编号:苏ICP备17064731号-1