1、UL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUCTION OR DISTRIBUTION WITHOUT PERMISSION FROM UL UL 61010-2-081 (IEC 61010-2-081:2015) Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use Part 2-081: Particular Requirements for Automatic and Semi-Automat
2、ic Laboratory Equipment for Analysis and Other PurposesUL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUCTION OR DISTRIBUTION WITHOUT PERMISSION FROM ULUL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUCTION OR DISTRIBUTION WITHOUT PERMISSION FROM UL UL Standard for Safety for Safet
3、y Requirements for Electrical Equipment for Measurement, Control and Laboratory Use Part 2-081: Particular Requirements for Automatic and Semi-Automatic Laboratory Equipment for Analysis and Other Purposes, UL 61010-2-081 (IEC 61010-2-081:2015) Second Edition, Dated August 14, 2015 Summary of Topics
4、 Adoption of IEC 61010-2-081, Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use Part 2-081: Particular Requirements for Automatic and Semi- Automatic Laboratory Equipment for Analysis and Other Purposes (second edition issued January 2015) as a new IEC-based UL
5、 standard, UL 61010-2-081 (IEC 61010-2-081:2015), with no National Differences. Although this is the rst-time publication of this Standard by UL, it is being published as the Second edition in order to align ULs edition number with that of the IEC Standard. As a result, one or more UL edition number
6、s have been skipped to match that of the IEC edition number. The new requirements are substantially in accordance with Proposal(s) on this subject dated June 12, 2015 . All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form by any
7、 means, electronic, mechanical photocopying, recording, or otherwise without prior permission of UL. UL provides this Standard as is without warranty of any kind, either expressed or implied, including but not limited to, the implied warranties of merchantability or tness for any purpose. In no even
8、t will UL be liable for any special, incidental, consequential, indirect or similar damages, including loss of prots, lost savings, loss of data, or any other damages arising out of the use of or the inability to use this Standard, even if UL or an authorized UL representative has been advised of th
9、e possibility of such damage. In no event shall ULs liability for any damage ever exceed the price paid for this Standard, regardless of the form of the claim. Users of the electronic versions of ULs Standards for Safety agree to defend, indemnify, and hold UL harmless from and against any loss, exp
10、ense, liability, damage, claim, or judgment (including reasonable attorneys fees) resulting from any error or deviation introduced while purchaser is storing an electronic Standard on the purchasers computer system. The requirements in this Standard are now in effect, except for those paragraphs, se
11、ctions, tables, gures, and/or other elements of the Standard having future effective dates as indicated in the preface. The prior text for requirements that have been revised and that have a future effective date are located after the Standard, and are preceded by a SUPERSEDED REQUIREMENTS notice. A
12、UGUST 14, 2015 UL 61010-2-081 (IEC 61010-2-081:2015) tr1UL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUCTION OR DISTRIBUTION WITHOUT PERMISSION FROM UL AUGUST 14, 2015 UL 61010-2-081 (IEC 61010-2-081:2015) tr2 No Text on This PageUL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUC
13、TION OR DISTRIBUTION WITHOUT PERMISSION FROM UL AUGUST 14, 2015 1 UL 61010-2-081 (IEC 61010-2-081:2015) Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use Part 2-081: Particular Requirements for Automatic and Semi-Automatic Laboratory Equipment for Analysis and
14、Other Purposes Second Edition August 14, 2015 This ANSI/UL Standard for Safety consists of the Second Edition dated August 14, 2015. The most recent designation of ANSI/UL 61010-2-081 (IEC 61010-2-081:2015) as an American National Standard (ANSI) occurred on August 14, 2015. ANSI approval for a stan
15、dard does not include the Cover Page, Transmittal Pages, Title Page, Preface, or effective date information. The IEC Foreword is also excluded from the ANSI approval of IEC-based standards. Comments or proposals for revisions on any part of the Standard may be submitted to UL at any time. Proposals
16、should be submitted via a Proposal Request in ULs On-Line Collaborative Standards Development System (CSDS) at http:/. ULs Standards for Safety are copyrighted by UL. Neither a printed nor electronic copy of a Standard should be altered in any way. All of ULs Standards and all copyrights, ownerships
17、, and rights regarding those Standards shall remain the sole and exclusive property of UL. COPYRIGHT 2015 UNDERWRITERS LABORATORIES INC. ANSI/UL 61010-2-081-2015UL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUCTION OR DISTRIBUTION WITHOUT PERMISSION FROM UL AUGUST 14, 2015 UL 61010-2-081 (
18、IEC 61010-2-081:2015) 2 No Text on This PageUL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUCTION OR DISTRIBUTION WITHOUT PERMISSION FROM UL CONTENTS Preface (UL) 4 FOREWORD 5 1 Scope and object .8 1.2 Object .9 2 Normative references 9 3 Terms and denitions 9 4 Tests 9 5 Marking and docum
19、entation 9 5.2 Warning markings .11 5.3 Durability of markings 11 6 Protection against electric shock .12 7 Protection against mechanical HAZARDS 12 8 Resistance to mechanical stresses .13 8.1 General 13 9 Protection against the spread of re 13 10 Equipment temperature limits and resistance to heat
20、13 11 Protection against HAZARDS from uids .13 12 Protection against radiation, including laser sources, and against sonic and ultrasonic pressure .13 13 Protection against liberated gases and substances, explosion and implosion .13 13.101 Biohazardous substances .13 14 Components and subassemblies
21、.14 15 Protection by interlocks 14 15.1 General .14 16 HAZARDS resulting from application .14 17 RISK Assessment 14 Annexes Bibliography AUGUST 14, 2015 UL 61010-2-081 (IEC 61010-2-081:2015) 3UL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUCTION OR DISTRIBUTION WITHOUT PERMISSION FROM UL P
22、reface (UL) This UL Standard is based on IEC Publication 61010-2-081: second edition Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use Part 2-081: Particular Requirements for Automatic and Semi-Automatic Laboratory Equipment for Analysis and Other Purposes. IEC
23、 publication 61010-2-081 is copyrighted by the IEC. Efforts have been made to synchronize the UL edition number with that of the corresponding IEC standard with which this standard is harmonized. As a result, one or more UL edition numbers have been skipped to match that of the IEC edition number. T
24、his UL Standard 61010-2-081 Standard for Safety for Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use Part 2-081: Particular Requirements for Automatic and Semi-Automatic Laboratory Equipment for Analysis and Other Purposes, is to be used in conjunction with th
25、e third edition of UL 61010-1. The requirements for laboratory equipment for the heating of materials are contained in this Part 2 Standard and UL 61010-2-081. Requirements of this Part 2 Standard, where stated, amend the requirements of UL 61010-1. Where a particular subclause of UL 61010-1 is not
26、mentioned in UL 61010-2-081, the UL 61010-1 subclause applies. The text, gures and tables of IEC Publication Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use Part 2-081: Particular Requirements for Automatic and Semi-Automatic Laboratory Equipment for Analysis
27、 and Other Purposes, IEC 61010-2-081, copyright January, 2015 are used in this Standard with the consent of the IEC and the American National Standards Institute (ANSI). The IEC copyrighted material has been reproduced with permission from ANSI. ANSI should be contacted regarding the reproduction of
28、 any portion of the IEC material. The IEC Foreword and Introduction are not a part of the requirements of this Standard but are included for information purposes only. Copies of IEC Publication 61010-2-101 may be purchased from ANSI, 11 West 42nd Street, New York, New York, 10036, (212) 642-4900. No
29、te Although the intended primary application of this Standard is stated in its Scope, it is important to note that it remains the responsibility of the users of the Standard to judge its suitability for their particular purpose. UL Effective Date The effective date for UL 61010-2-081 (IEC 61010-2-08
30、1:2015), Second edition is August 14, 2015. A UL effective date is one established by Underwriters Laboratories Inc. and is not part of the ANSI approved standard. AUGUST 14, 2015 UL 61010-2-081 (IEC 61010-2-081:2015) 4UL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUCTION OR DISTRIBUTION W
31、ITHOUT PERMISSION FROM UL INTERNATIONAL ELECTROTECHNICAL COMMISSION SAFETY REQUIREMENTS FOR ELECTRICAL EQUIPMENT FOR MEASUREMENT, CONTROL AND LABORATORY USE Part 2-081: Particular Requirements for Automatic and Semi-Automatic Laboratory Equipment for Analysis and Other Purposes FOREWORD 1) The Inter
32、national Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising all national electrotechnical committees (IEC National Committees). The object of IEC is to promote international co-operation on all questions concerning standardization in the electrical and elect
33、ronic elds. To this end and in addition to other activities, IEC publishes International Standards, Technical Specications, Technical Reports, Publicly Available Specications (PAS) and Guides (hereafter referred to as “IEC Publication(s)”). Their preparation is entrusted to technical committees; any
34、 IEC National Committee interested in the subject dealt with may participate in this preparatory work. International, governmental and nongovernmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely with the International Organization for Standardiz
35、ation (ISO) in accordance with conditions determined by agreement between the two organizations. 2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international consensus of opinion on the relevant subjects since each technical committee has represe
36、ntation from all interested IEC National Committees. 3) IEC Publications have the form of recommendations for international use and are accepted by IEC National Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC Publications is accurate, IEC c
37、annot be held responsible for the way in which they are used or for any misinterpretation by any end user. 4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications transparently to the maximum extent possible in their national and regional publicat
38、ions. Any divergence between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in the latter. 5) IEC itself does not provide any attestation of conformity. Independent certication bodies provide conformity assessment services and, in some areas, ac
39、cess to IEC marks of conformity. IEC is not responsible for any services carried out by independent certication bodies. 6) All users should ensure that they have the latest edition of this publication. 7) No liability shall attach to IEC or its directors, employees, servants or agents including indi
40、vidual experts and members of its technical committees and IEC National Committees for any personal injury, property damage or other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and expenses arising out of the publication, use of, or reliance upon,
41、 this IEC Publication or any other IEC Publications. 8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is indispensable for the correct application of this publication. 9) Attention is drawn to the possibility that some of the elements of
42、 this IEC Publication may be the subject of patent rights. IEC shall not be held responsible for identifying any or all such patent rights. International Standard IEC 61010-2-081 has been prepared by IEC technical committee 66: Safety of measuring, control and laboratory equipment. It has the status
43、 of a group safety publication in accordance with IEC Guide 104. AUGUST 14, 2015 UL 61010-2-081 (IEC 61010-2-081:2015) 5UL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUCTION OR DISTRIBUTION WITHOUT PERMISSION FROM UL This second edition cancels and replaces the rst edition published in 200
44、1 and its Amendment 1 (2003). It constitutes a technical revision and includes the following signicant changes from the rst edition, as well as numerous other changes: excluded IEC 61010-2-101 (IVD Equipment) from scope. This separates IEC 61010-2-081 and IEC 61010-2-101 equipment; added biological
45、risks symbol to Table 1 in Clause 5; added requirement for statement on hazardous substances and gases in Instructions for Use to Clause 5; added marking requirement for ow direction to Clause 5; added requirement for OPERATORmaintenance instructions to Clause 7; excluded equipment whose size and we
46、ight make unintentional movement unlikely from drop test in Clause 8; added normative reference ISO 18113-5 for instructions for use of self-test IVD medical equipment in Clause 5; added requirement for OPERATOR maintenance instructions to Clause 7; added requirements for sample zones and loading zo
47、nes to Clause 7; excluded equipment whose size and weight make unintentional movement unlikely from drop test in Clause 8; added requirement for biohazard marking to Clause 13; added requirement for interlock systems containing electric/electronic or programmable components to Clause 15; The text of
48、 this standard is based on the following documents: FDIS Report on voting 66/544/FDIS 66/559/RVD Full information on the voting for the approval of this standard can be found in the report on voting indicated in the above table. This publication has been drafted in accordance with the ISO/IEC Direct
49、ives, Part 2. A list of all parts of the IEC 61010 series, under the general title: Safety requirements for electrical equipment for measurement, control, and laboratory use, may be found on the IEC website. This Part 2-081 is intended to be used in conjunction with IEC 61010-1. It was established on the basis of the third edition (2010). AUGUST 14, 2015 UL 61010-2-081 (IEC 61010-2-081:2015) 6UL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUCTION OR DISTRIBUTION WITHOUT PER