1、Designation: D4178 82 (Reapproved 2012)Standard Practice forCalibrating Moisture Analyzers1This standard is issued under the fixed designation D4178; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A numb
2、er in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers a calibration technique based on thepreparation of standards of known water content. This tech-nique is applicable to
3、the production of standards between 20and 2000 cm3/m3water.1.2 The values stated in SI units are to be regarded asstandard. The values given in parentheses are for informationonly.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is therespo
4、nsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Summary of Practice22.1 The practice is based on the principle that ice has avapor pressure of 0.611 kPa at 0C. Therefore, when a ca
5、rriergas at a constant gauge pressure of 207 kPa (30 psig) is passedthrough a molecular sieve saturated with water and held at 0C,the total pressure is equal to 207 kPa plus 98 kPa (oneatmosphere) and the water concentration of the gas leaving themolecular sieve is 0.611/(207 + 98) 3 106ppm or 2000
6、ppmvolume water, regardless of flow.2.1.1 A carrier gas at a constant gauge pressure of 207 kPais passed through a molecular sieve drier and then routed overa molecular sieve support saturated with water and equilibratedat 0C.2.1.2 By blending different volumes of the wet carrier gaswith dried carri
7、er gas, gas standards of known water concen-tration can be prepared.2.2 The moisture analyzer to be calibrated is then connectedto the source of the gas standard of known water concentration.3. Significance and Use3.1 This practice is intended to provide a method to cali-brate moisture analyzers use
8、d on-stream or in the laboratory.4. Apparatus4.1 Ice Bath Primary Standard Moisture ApparatusFig. 1illustrates a typical system.4.2 Wet Test Meter, 1-L divisions.4.3 Bubble Meter, graduated in cubic centimetres.4.4 Pressure GaugeA Bourdon-type spring gauge of testgauge quality, 100 to 250 mm in diam
9、eter, with a scale rangefrom 0 to 400 kPa (0 to 60 psi), maximum intermediategraduations of 1.5 kPa (0.25 psi).4.5 Wet Mole Sieve 5A, 60/80 mesh (other mesh sizes maybe used, except powder).5. Preparation of Apparatus5.1 Saturate mole sieve with water.5.1.1 Cover in beaker enough mole sieve to fill
10、the wet molesieve container. Add water to cover mole sieve. Let standovernight.5.1.2 Drain the excess water by pouring the mole sieveslurry into a filtering funnel and letting all the free water drainout.NOTE 1The useful lifetime of the wet mole sieve is not known. It isrecommended that the procedur
11、e described in 5.1.1 and 5.1.2 be followedprior to each calibration.6. Procedure6.1 Fill the ice bath bucket for the primary standard appa-ratus one third full of water; then add ice to bring the level tofull.6.2 Close the wet flow and the diluent flow needle values.6.3 Turn on the carrier gas suppl
12、y (nitrogen or air) andadjust the regulator to 207 kPa (30 psig).6.4 Open the diluent flow needle valve to obtain a carriergas flow of 1 L/min, as measured with a wet test meter.6.5 Allow the apparatus to equilibrate for 1 h. This allowsthe wet mole sieve to equilibrate at 0C.6.6 After 1 h, attach t
13、he moisture analyzer to be calibratedand check that the blank moisture content of the carrier gas is6 ppm or less.1This practice is under the jurisdiction of ASTM Committee D02 on PetroleumProducts and Lubricants and is the direct responsibility of Subcommittee D02.D0 onHydrocarbons for Chemical and
14、 Special Uses.Current edition approved April 15, 2012. Published May 2012. Originallyapproved in 1982. Last previous edition approved in 2005 as D417882(2005).DOI: 10.1520/D4178-82R12.2For a more complete discussion of this procedure, see Mator, R. T., “TraceMoisture Analyzers and Their Calibration,
15、” Proceedings of the 20th Annual ISAAnalysis Instrumentation Symposium, May 12-15, 1974, Pittsburgh, PA, “Session:Sampling and Calibration Systems,” 1974, p. 125.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.6.7 Turn off the diluen
16、t flow, cm3/m3.6.8 Attach a bubble meter to the exit of the apparatus andcheck that there is no flow through the system.6.9 Adjust the wet flow needle valve to obtain a flow of 40cm3/min, as measured with a bubble meter.6.10 In a similar manner, adjust the diluent flow needlevalve to obtain a total
17、flow of 200 cm3/min, as measured witha bubble meter.6.11 Connect the moisture analyzer to be calibrated to theexit of the “primary standard” moisture apparatus and allow itsprobe to equilibrate (10 to 15 min).6.12 The analyzer should read the calculated value of 400ppm for a carrier gas pressure of
18、207 kPa gauge. If it does not,adjust the necessary analyzer controls to obtain the correctreading.6.13 Turn off both the wet flow and the diluent flow needlesvalves.6.14 Attach a bubble meter to the exit of the apparatus andcheck that there is no flow through the system.6.15 Adjust the wet flow need
19、le valve to obtain a flow of 10cm3/min using a bubble meter, and then adjust the diluent flowneedle valve to obtain a total flow of 1 L/min. Use a wet testmeter to check the total flow.6.16 After the analyzer equilibrates (10 to 15 min), check tosee that it reads the correct calculated value of 20 c
20、m3/m3plusthe blank moisture content found in 6.6 for 207 kPa gaugecarrier gas supply.6.17 When the analyzer reads correctly at the 400-ppm levelbut is not within 62 ppm of the 20 ppm plus the blank moisturecontent value, then the sensor may be bad and needs replacing.7. Calculation7.1 Calculate wate
21、r concentration in saturated gas (wetflow) as follows:PH2O0C 5 0.611 kPa (1)Ptotal5 A kPa 1 98 kPa gaugeA kPa 5 carrier gas pressure should be 2076 2 kPa!CH2O5 PH2O!106!Ptotal50.611!106!207 1 985 2000 cm3/m3where:PH2O 0C = vapor pressure of water at 0C,Ptotal= total pressure applied to the water on
22、themolecular sieve, andCH2O=cm3/m3volume moisture in wet flow.7.2 Calculate water concentration in blended gas (total flow)as follows:CTH2O5 wet flow! CH2O!total flow1 ppm H2O blank (2)where:CH2O=cm3/m3moisture in wet flow,CTH2O=cm3/m3moisture in total flow, andtotal flow = wet flow plus diluent flo
23、w.8. Keywords8.1 calibration; carrier gas; moisture analyzer; water contentASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentionedin this standard. Users of this standard are expressly advised that determination of the validity
24、 of any such patent rights, and the riskof infringement of such rights, are entirely their own responsibility.This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years andif not revised, either reapproved or withdrawn. Your comments
25、 are invited either for revision of this standard or for additional standardsand should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of theresponsible technical committee, which you may attend. If you feel that your comments have not
26、received a fair hearing you shouldmake your views known to the ASTM Committee on Standards, at the address shown below.This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,United States. Individual reprints (single or multiple copie
27、s) of this standard may be obtained by contacting ASTM at the aboveaddress or at 610-832-9585 (phone), 610-832-9555 (fax), or serviceastm.org (e-mail); or through the ASTM website(www.astm.org). Permission rights to photocopy the standard may also be secured from the ASTM website (www.astm.org/COPYRIGHT/).FIG. 1 Schematic of Ice Bath “Primary Standard” Moisture ApparatusD4178 82 (2012)2