1、Designation: D7961 17Standard Practice forCalibrating U-tube Density Cells over Large Ranges ofTemperature and Pressure1This standard is issued under the fixed designation D7961; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the
2、year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This practice outlines procedures for the calibration ofU-tube density cells. It is applicable to instruments
3、 capable ofdetermining fluid density at temperatures in the range 10 Cto 200 C and pressures from just greater than the saturationpressure to 140 MPa. The practice refers to density cells asthey are utilized to make measurements of fluids primarily inthe compressed-liquid state. Examples of substanc
4、es for whichthe density can be determined with a calibrated U-tube densitymeter include: crude oils, gasoline and gasoline-oxygenateblends, diesel and jet fuels, hydraulic fluids, and lubricatingoils.1.2 This practice specifies a procedure for the determinationof the expanded uncertainty of the dens
5、ity measurement.1.3 This practice pertains to fluids with viscosities 1 Pas(1000 centipoise) at ambient conditions.1.4 4 The values listed in SI units are regarded as thestandard, unless otherwise stated. The SI unit for mass densityis kilograms per cubic metre (kgm-3) and can be given asgrams per c
6、ubic centimetre (gcm-3).1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prio
7、r to use.2. Referenced Documents2.1 ASTM Standards:2D1193 Specification for Reagent WaterD4052 Test Method for Density, Relative Density, and APIGravity of Liquids by Digital Density MeterD5002 Test Method for Density and Relative Density ofCrude Oils by Digital Density AnalyzerD6792 Practice for Qu
8、ality System in Petroleum Productsand Lubricants Testing LaboratoriesD7483 Test Method for Determination of Dynamic Viscosityand Derived Kinematic Viscosity of Liquids by Oscillat-ing Piston ViscometerD7578 Guide for Calibration Requirements for ElementalAnalysis of Petroleum Products and Lubricants
9、D7740 Practice for Optimization, Calibration, and Valida-tion of Atomic Absorption Spectrometry for Metal Analy-sis of Petroleum Products and Lubricants3. Terminology3.1 Definitions:3.1.1 calibration, nset of operations that establishes therelationship between the reference density of standards and
10、thecorresponding density reading of the instrument. D40523.1.2 certified reference material (CRM), nreference ma-terial one or more of whose property values are certified by atechnically valid procedure, accompanied by a traceable cer-tificate or other documentation that is issued by a certifyingbod
11、y. D02.94 D6792, D02.03 D75783.1.3 density (), nmass per unit volume at a specifiedtemperature. D02.07 D74833.1.4 standard reference material (SRM), ntrademark forreference materials certified by NIST. D02.03 D77404. Summary of Practice4.1 This practice details the considerations and proceduresneces
12、sary in order to complete a calibration of a U-tube densitymeter and an associated uncertainty analysis. The principalobjective of this practice is to provide the user with informationas to how different aspects of the calibration procedurecontribute to the overall uncertainty of related density mea
13、-surements obtained with a U-tube density meter.5. Significance and Use5.1 This practice covers a series of methods offered to aidusers in calibrating U-tube density meters to provide a measureof density and an associated expanded uncertainty. The refer-ence density, as obtained from either an equat
14、ion of state1This practice is under the jurisdiction of ASTM Committee D02 on PetroleumProducts, Liquid Fuels, and Lubricants and is the direct responsibility of Subcom-mittee D02.04 on Hydrocarbon Analysis.Current edition approved Jan. 1, 2017. Published January 2017. Originallyapproved in 2015. La
15、st previous edition approved in 2015 as D7961 15. DOI:10.1520/D7961-17.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe AST
16、M website.*A Summary of Changes section appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United StatesThis international standard was developed in accordance with internationally recognized principles on standardiz
17、ation established in the Decision on Principles for theDevelopment of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.1(EOS) or CRM has an uncertainty that arises from theuncertainty of the measurements of tempera
18、ture, pressure, andalso the chemical purity of the substance studied (origin) or forthat matter of the certified reference material. This uncertaintyresults in an additional uncertainty for the density of thesesamples. Because the measurements made with U-tube densitymeters are not absolute, the unc
19、ertainty with which theinstrument calibration is determined is directly related to theuncertainty of the density obtained.6. Apparatus6.1 This practice is applicable to U-tube density meterscapable of operating at temperatures of 10 C to 200 C andpressures to 140 MPa. Such instruments are commercial
20、lyavailable and the measurement technique is well understood.Generally, the U-tube is electronically excited at a constantamplitude and a frequency meter is used to record thefrequency of the oscillation of the U-tube. This frequency isdependent upon the density of the fluid in the U-tube. As thetec
21、hnique is not a direct measurement of density the instrumentis calibrated with a fluid or fluids, the densities of which areknown accurately over a range of temperature and pressure. Acorrelation of density can then be formulated based upontemperature, pressure, and period of oscillation of the U-tu
22、becollected during the calibration process.6.2 Additional equipment that is mandatory to carry out theprocedures described in this practice includes:6.2.1 Vacuum pump;6.2.2 Reagents to be used as calibration fluids;6.2.3 Reagents for cleaning the system; and6.2.4 Sample containers; stainless steel c
23、ylinders or glassflasks that can be evacuated.6.3 Additional equipment recommended for adherence tothis practice includes:6.3.1 High-accuracy temperature measurement device witha calibration traceable to a national metrology institute (NMI);6.3.2 High-accuracy pressure measurement device with acalib
24、ration traceable to a national metrology institute (NMI);and6.3.3 Computer and software for automated data acquisi-tion.7. Reagents7.1 The best calibration fluids are those referred to asCertified Reference Materials (CRM) which have correlationsfor density over the temperature and pressure range wh
25、eremeasurements will be made. If CRMs are not available, obtainfluids or gases in the highest purity available, preferably witha mole fraction greater than 0.999 as cited by the manufactureranalysis.7.2 Water shall conform to Specification D1193 Type II orbetter. Recommended fluids, depending on the
26、 measuringrange (pressure, p, density, , and temperature, T) to be coveredin the calibration of the U-tube density meter include, but arenot limited to: water, methane, ethane, propane, butane,nitrogen, methylbenzene decane, and dichlorotoluene. Equa-tions of state exist for all of these fluids and
27、can be found inNIST Standard Reference Database 23 or in the references forthe specific fluid equations found in the reference section ofthis document. (1-9)37.3 Recommended reagents for cleaning the instrumentinclude, but are not limited to methylbenzene, ethanol, acetone,white spirit, and quinolin
28、e.8. Calibration Procedure8.1 Choose one or more calibration fluids from those listedin Section 7. The fluids shall meet the following criteria:8.1.1 The fluid is readily available in high purity (molefractions greater than 0.999 as cited by manufacturer analysis).8.1.2 The fluid(s) selected for the
29、 calibration shall havepressure, density, temperature, (p, , T) surfaces which boundthose of the fluids to be studied. As several of the suggestedfluids have a somewhat limited density range in the tempera-ture and pressure boundaries of the U-tube instruments, it isoften useful to use two fluids in
30、 addition to vacuum for thecalibration in order for a larger p, , T surface to be covered bythe calibration equation. A good example of this is to usepropane and toluene. In contrast, selecting just one calibrationfluid which has a small p, , T surface (that is, water) greatlylimits the density rang
31、e that can be determined from thecalibration equation.8.1.3 The fluid is well described by an equation of state, andthe uncertainty associated with density predictions from thatequation is less than the desired overall uncertainty for densitymeasurements resulting from the calibration.8.2 Decide whi
32、ch calibration scheme is suitable for themeasurements.8.2.1 There are two schema that can be used to calibrate aU-tube density meter, and these are as follow:8.2.1.1 Determine a calibration equation through the fit ofcalibration data measured over a range of temperatures andpressures; and8.2.1.2 Cal
33、ibrate the instrument at a single point (deter-mined by a pressure and temperature) where measurementswill be performed. This method minimizes the influences ofpressure and temperature on the calibration that is performed atthe same temperature and pressure.8.3 Prepare the Sample(s):8.3.1 Degas samp
34、les prior to measurement. Those withboiling points below ambient temperature are ideally housed instainless steel cylinders and degassed through vacuum subli-mation; at least three cycles of freezing with liquid nitrogen,evacuation, thawing, and then either heating or ultrasound.Liquid samples can b
35、e degassed by vacuum distillation into aglass flask (with a valve of PTFE (Teflon) or some other inertmaterial) or degassed in the glass flask by the previouslymentioned procedure of freezing and evacuation.8.4 Clean the Instrument:8.4.1 Prior to beginning measurements, and after a samplehas been me
36、asured, the instrument is cleaned. This is done byfilling the measuring system with a cleaning agent. The best3The boldface numbers in parentheses refer to the list of references at the end ofthis standard.D7961 172cleaning agents are those which will readily dissolve theresidue of the most recently
37、 measured sample. Examples ofcommon cleaning agents include methylbenzene, ethanol,acetone, white spirit, and quinoline. Take caution to use onlycleaning agents that have normal boiling point temperaturesabove that of room temperature.8.4.2 When the system is thought to be clean, remove thecleaning
38、agent by flushing the system with a more volatilesolvent. Then use a stream of dry air or dry gas to dry thesystem. Evacuate the system and set the temperature to one atwhich you have previously measured the period of oscillationunder vacuum. Compare past and present results. If the currentFIG. 1 Fl
39、ow Chart of How to Choose a Calibration FluidD7961 173period of oscillation is greater than the past, further cleaningmay be necessary. The period of oscillation may shift particu-larly in a new instrument and especially as it is cycled overlarge ranges of temperature and pressure. Thus, an increase
40、 inthe period of oscillation is not necessarily indicative of theinstrument needing further cleaning however, this is recom-mended. If, after a second cleaning the period of oscillationremains high (greater than 0.02 s above the previous reading),it is likely the period has shifted and no further cl
41、eaning isnecessary. Once the instrument is clean, fill the system with thenext sample to be measured.8.5 Establish Convergence Criteria:8.5.1 Determine Instrument Temperature Stability:8.5.1.1 The temperature and pressure stability (and as aresult the period of oscillation of the U-tube) is greatlyd
42、ependent upon the mechanisms for temperature and pressurecontrol put in place by the user. As density is more strongly afunction of temperature than pressure, poor temperature con-trol will have a more adverse effect on measurement repeat-ability and thus uncertainty. As such, temperature is commonl
43、ythe first of the convergence criteria that is met. Tests areconducted to determine the temperature stability of the instru-ment. These tests can be run with the system evacuated or filledwith a fluid. Set the system to a desired temperature. Once thegiven temperature setpoint has been reached, reco
44、rd at least 2 hof temperature data at regular intervals of 1 min or less. Use astatistical test such as the Mean Square Successive Difference(MSSD) technique (10) to make certain there is no overallupward or downward trend in the data. If there is a significanttrend, thermal equilibrium most likely
45、had not been reachedwhen the measurements were made, and this should be noted.It can take several additional hours after the setpoint has beenachieved for the instrument to reach thermal equilibrium. Thisis particularly true at temperatures greater than 50 K aboveambient, or below ambient temperatur
46、e. Calculate the standarddeviation of the temperatures measured. This is the stability ofthat temperature setpoint and can be used as the convergencecriteria for temperature. Repeat this procedure at 10 K to 20 Kintervals throughout the temperature range in which measure-ments will be conducted.8.5.
47、1.2 During measurements, periodically calculate thestandard deviation of, at a minimum, the ten most recentlyrecorded temperature measurements. When the standard devia-tion is less than or equal to the convergence criteria, theinstrument can be considered to be at thermal equilibrium.8.5.2 Establish
48、 Pressure and Period of Oscillation Conver-gence Criteria:8.5.2.1 Convergence criteria for the pressure and period ofoscillation of the U-tube are determined in much the same wayas that for the temperature. Data is recorded at regularintervals, that is, every 30 s for at least 2 h after the temperat
49、ureconvergence criteria has been met. Check to make certain thereis no overall upward or downward trend in that data. If there issuch a trend, equilibrium had most likely not been reachedwhen the measurements were recorded. The standard deviationof at least ten consecutive measurements in that set can then beconsidered as the convergence criteria. During densitymeasurements, the system is considered to be at equilibriumwhen the standard deviations of the most recent ten measure-ments of temperature, pressure, and period of oscillation of theU-tube, are a
copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
备案/许可证编号:苏ICP备17064731号-1