1、Designation: E1046 85 (Reapproved 2015)Standard Specification forGlass Westergren Tube, Disposable1This standard is issued under the fixed designation E1046; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision
2、. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This specification covers a disposable tube used formeasuring the erythrocyte sedimentation rate, ESR (the sus-pension stability of
3、 red cells in diluted, anti-coagulated humanblood).1.2 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.2. Referenced Documents2.1 ASTM Standards:2E438 Specification for Glasses in Laboratory Apparatus3. Terminology3.1 Definition
4、s of Terms Specific to This Standard:3.1.1 disposable, nin accordance with this specificationand the expected product performance expressed in thisstandard, those tubes which are to be used one time only. Anyinstitution or individual who resuses a disposable tube mustbear full responsibility for its
5、 safety and effectiveness.3.1.2 Westergren3, nsurname of the individual responsiblefor the design of the reusable Westergren tube and the methodof use for both reusable and disposable tubes.4. Classification4.1 This specification covers a tube that is intended to beused one time only. It is not to b
6、e confused with a reusable tubethat is described in other published standards. The tubes shallbe of the following types:4.1.1 Type Istandard Westergren tube to be used in con-ventional racks only, and4.1.2 Type IIself-zeroing tubes to be used with racksdesigned with convenient disposable tube featur
7、es.5. Materials5.1 GlassThe tubes made to this specification shall befabricated from borosilicate glass, Type I, Class B, or soda limeglass, Type II, in accordance with Specification E438.6. Physical Requirements6.1 Design, Type I and IIThe Westergren tube shall bemade of one-piece construction from
8、 tubing that is straight andof uniform bore. The ends of the tube shall be cut at rightangles to the axis of the tube and fire polished.6.1.1 Type II onlyEach tube shall have an absorbant cottonplug inserted in the tube with the bottom edge of the plugaligned with the 0 mm calibration line. The alig
9、nment of theplug to the calibration line should not vary by more than 1 mL.The plug must be capable of supporting a 200 mm verticalcolumn of water when the plug has been thoroughly soaked.NOTE 1With possible printing scale variation and cotton plugalignment to the calibration line variation, the bot
10、tom of the cotton plug isto be located 200 6 1 mm from the bottom of the tube.6.2 WorkmanshipThe tube shall be free as possible fromvisible defects which would detract from its appearance orimpair its serviceability when viewed by the human eye undernormal room lighting. Self-zeroing tubes are to be
11、 used withracks designed with convenient disposable tube features.6.3 DimensionsDimensions and tolerances shall be inaccordance with Fig. 1. The uniformity of the bore shall be60.15 mm throughout the tube. The tube shall have aninscribed graduated scale extending over the lower 200 6 1.0mm of the tu
12、be. The markers or vendors name or mark shallbe inscribed on the tube in a selected location. Inscription ofthe word “Westergren” on the tube is optional. Type II tubesmay be made of varied lengths from 230 to 243 mm long withthe selected length within this range confined to a 61mmtolerance.6.4 Grad
13、uation LinesThickness of graduation lines mayvary in uniformity but shall not exceed 0.4 mm in thickness.They shall lie at right angles to the axis of the tube with amaximum tolerance between two adjacent markings of 0.25mm. Maximum tolerance for the total 200 mm scale shall notexceed 1.0 mm.1This s
14、pecification is under the jurisdiction of ASTM Committee E41 onLaboratory Apparatus and is the direct responsibility of Subcommittee E41.01 onLaboratory Ware and Supplies.Current edition approved June 1, 2015. Published July 2015. Originally approvedin 1985. Last previous edition approved in 2010 as
15、 E1046 85 (2010). DOI:10.1520/E1046-85R15.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Westergren, A., “S
16、tudies of the Suspension Stability of the Blood in PilmonaryTuberculosis,” Acta Medica Scandinavia, 54, 19201921, pp. 247282.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States16.5 Graduation Line NumberingThe tube shall be gradu-ated in
17、millimetres with a scale of 200 mm from the tip of thetube. The tube shall be numbered every 10 or 20 graduationlines starting with a numerical zero (0) and downward to amaximum value of 180 or 190 mm. The numerical markingsshall appear at the right side of the graduated scale when heldvertically wi
18、th the scale facing the viewer.7. Keywords7.1 disposable; glass; tube ; WestergrenASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentionedin this standard. Users of this standard are expressly advised that determination of the v
19、alidity of any such patent rights, and the riskof infringement of such rights, are entirely their own responsibility.This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years andif not revised, either reapproved or withdrawn. Your c
20、omments are invited either for revision of this standard or for additional standardsand should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of theresponsible technical committee, which you may attend. If you feel that your comments ha
21、ve not received a fair hearing you shouldmake your views known to the ASTM Committee on Standards, at the address shown below.This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,United States. Individual reprints (single or multipl
22、e copies) of this standard may be obtained by contacting ASTM at the aboveaddress or at 610-832-9585 (phone), 610-832-9555 (fax), or serviceastm.org (e-mail); or through the ASTM website(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http:/ 1E1046 85 (2015)2