1、Designation: E788 97 (Reapproved 2013)Standard Specification forPipet, Blood Diluting1This standard is issued under the fixed designation E788; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in
2、parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This specification covers requirements for glass reusableblood diluting pipets that are used for performing red and whitecell corpuscle determina
3、tions.1.2 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.2. Referenced Documents2.1 ASTM Standards:2E694 Specification for Laboratory Glass Volumetric Appa-ratusE920 Specification for Commercially Packaged LaboratoryApparatusE9
4、21 Specification for Export Packaged Laboratory Appa-ratusE1133 Practice for Performance Testing of Packaged Labo-ratory Apparatus for United States Government Procure-mentsE1157 Specification for Sampling and Testing of ReusableLaboratory Glassware3. Classification3.1 This specification covers two
5、different pipet designs andpermissible alternatives.3.1.1 Red and white cell blood diluting pipets (see Fig. 1).3.1.2 Permissible alternative designs (see Fig. 2).4. Materials4.1 The pipets shall be made of common spirit bore whiteback tubing or clear glass with a white stripe applied to theouter su
6、rface of the tubing.4.2 The beads shall be made of glass or ceramic composi-tion and shall be of red, clear, or white coloring.5. Dimensions and Permissible Variations5.1 DesignThe red and white cell blood dilution pipetsshall consist of a small uniform bore glass tube which shallhave a bulb of prop
7、er size near the proximal end (see Fig. 1).As an alternative, the pipets may be constructed of three piecesof glass that are fused together to form a one-piece pipet (seeFig. 2). The bulb shall contain a nonspherical glass or ceramicbead sufficiently large enough to prevent its being impacted inthe
8、constriction portion of the bulb. The cross section of thepipet at any point shall be circular. The distal end of the whitecell pipet shall be pulled to a point and then ground andpolished to a tapered tip. The distal end of the red cell pipetshall be ground and polished to a tapered tip. The inside
9、diameter (ID) of the pipet bore at the tip end of the white cellpipet shall be 0.2 to 0.5 mm. The external diameter of theground and polished tip shall not exceed 2.0 mm. The overalllength of both pipets shall be 104 to 121 mm. The proximal endmay be of funnel design, with an exterior taper, or grou
10、nd andpolished with a taper of sufficient angle to permit application ofa rubber suction tube or other suction device (see Fig. 2).5.2 CapacityThe capacity of the red cell pipet bulb shallbe 0.8 to 1.2 cm3and the capacity of the white cell bulb shall be0.2 to 0.4 cm3. The capacity of the red cell pi
11、pet stem (bulb todistal end) shall be 0.008 to 0.012 cm3and the capacity of thewhite cell pipet stem shall be 0.02 to 0.04 cm3.5.3 Capacity Markings:5.3.1 Pipet Bulb MarkingsThe red and white cell pipetbulb capacity shall be confined to and indicated by calibrationlines on the pipet. These lines sha
12、ll be located on both sides ofthe bulb and shall be within 2 mm to 6 mm above the bulb(proximal end) and within 3 mm to 6 mm below the bulb (distalend). The exact manner for measuring these line placements isspecified in Fig. 1.5.3.2 Pipet Stem MarkingsThe graduation lines on thestem of the red and
13、white cell pipets shall be equally dividedinto two or ten divisions. If there are two divisions, the twocalibration lines on the stem will be numbered 0.5 and 1. Ifthere are ten divisions, the ten calibration lines will benumbered 0.5 on the fifth line and 1 on the tenth linerespectively. The top li
14、ne (above bulb at the proximal end)shall be numbered 101 on the red cell pipet and 11 on the whitecell pipet, representing stem and bulb volume collectively. At1This specification is under the jurisdiction of ASTM Committee E41 onLaboratory Apparatusand is the direct responsibility of Subcommittee E
15、41.01 onApparatus.Current edition approved Nov. 1, 2013. Published December 2013. Originallyapproved in 1981. Last previous edition approved in 2008 as E788 97(2008). DOI:10.1520/E0788-97R13.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serv
16、iceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1the option of the manufacturer, pipet stems may be gradu
17、atedwith split lines or partial length lines (see Fig. 2).5.4 Graduation LinesAll graduation lines on the red andwhite cell pipets shall be at right angles to the pipet axis andparallel to each other. The thickness of the graduation linesshall not exceed 0.4 mm.5.5 AccuracyThe red and white cell pip
18、et bulb volumeshall be a volumetric ratio to the stem volume. The red cellpipet bulb to stem ratio shall be 100:1 and the white cell pipetbulb to stem ratio shall be 10:1 when tested in accordance with6.1 and 6.2.5.6 Workmanship:5.6.1 The pipets shall be free of defects that detract fromtheir appear
19、ance or impair their serviceability.5.6.2 Construction shall be such that the mechanicalstrength is provided to withstand the rigors of normal use. Thepipets shall be free of strain when tested in accordance with6.4.5.7 IdentificationEach pipet shall have the name or regis-tered trademark of the man
20、ufacturer. These markings are to belocated on the stem of the pipet directly opposite the graduationlines. The manufacturer has the option to state the pipetvolumetric tolerances on the pipet. These tolerances may belocated on the stem or the proximal end of the pipet.5.7.1 The white cell pipet shal
21、l have a clear or white glassor ceramic bead sealed within the bulb to identify it readily asa white cell pipet.5.7.2 The red cell pipet shall have a red glass or ceramicbead sealed within the bulb to identify it readily as a red cellpipet.5.8 PigmentationAll markings shall be permanently fusedin or
22、 on the pipet. The markings shall be dark amber or blackin color. When tested in accordance with 6.3, the pigmentationshall not discolor the solution. The appearance of the markings,when viewed by the eye under normal room lighting, shall bethe same after testing as before testing.6. Testing6.1 Accu
23、racyAccuracy for these pipets shall be 65 % forred cell pipets and 63.5 % for white cell pipets. If V representsthe volume between the calibration lines immediately aboveand below the bulb, V1represents the volume of the capillarystem between the pipet tip and the calibration line numbered 1,and V2r
24、epresents the volume of the capillary stem between theFIG. 1 Pipet DimensionsE788 97 (2013)2tip and the calibration line numbered 0.5, the volumetrictolerances shall be as stated in Specification E694 and ex-pressed as follows:6.1.1 Red Cell PipetsNominal ratio 100:1V/V1shall not be less than 95 or
25、more than 105V/V2shall not be less than 190 or more than 2106.1.2 White Cell PipetsNominal ratio 10:1V/V1shall not be less than 9.65 or more than 10.35V/V2shall not be less than 19.30 or more than 20.706.2 Accuracy Determination V2, V1, and V shall bedetermined by use of distilled water and a weighi
26、ng devicehaving a weight sensitivity no less than 0.001 mg.6.2.1 The weight of a dry red or white cell pipet shall betaken first and recorded. This indication shall be subtractedfrom the weight indications of the red or white cell pipet filledwith distilled water to the lines numbered 0.5, 1, 101, o
27、r 11respectively. The weight of the water at temperature T,istobeconverted to volume at 20C. The net weight indicationsrepresent V2, V1, and V factors given in 6.1.1 and 6.1.2.6.3 Pigmentation TestFreshly prepare a chromic acidcleaning solution by combining 200 g of sodium dichromate(Na2Cr2O72H2O),
28、1000 mL of water, and 1500 mL ofsulfuric acid (H2SO4, ACS Reagent 95 to 98 %). Immerse thepipets in the chromic acid solution. Let stand at room tempera-ture (20 to 25C) for 15 min. Remove the pipets from thesolution and thoroughly rinse in distilled water. Dry the pipetsby rubbing vigorously, 5 to
29、10 strokes, with a laboratory clothor tissue. The appearance of the markings should be the sameas before the test, when judged by the eye under normal roomlighting.6.4 Strain-Free TestThe pipets shall be free from strainwhen viewed under a polariscope.6.5 For additional sampling and testing data, se
30、e Specifica-tion E1157.7. Packaging7.1 For packaging, select from either Specification E920,Specification E921, or Practice E1133.8. Keywords8.1 blood; glass; pipetsFIG. 2 Permissible AlternativesE788 97 (2013)3ASTM International takes no position respecting the validity of any patent rights asserte
31、d in connection with any item mentionedin this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the riskof infringement of such rights, are entirely their own responsibility.This standard is subject to revision at any time by th
32、e responsible technical committee and must be reviewed every five years andif not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standardsand should be addressed to ASTM International Headquarters. Your comments will receive
33、careful consideration at a meeting of theresponsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you shouldmake your views known to the ASTM Committee on Standards, at the address shown below.This standard is copyrighted by ASTM Internat
34、ional, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the aboveaddress or at 610-832-9585 (phone), 610-832-9555 (fax), or serviceastm.org (e-mail); or through the ASTM website(www.astm.org). Permission rights to photocopy the standard may also be secured from the ASTM website (www.astm.org/COPYRIGHT/).E788 97 (2013)4
copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
备案/许可证编号:苏ICP备17064731号-1