1、BS 8480:2006Medical devices Chairs with electrically operated support surfaces RequirementsICS 97.140NO COPYING WITHOUT BSI PERMISSION EXCEPT AS PERMITTED BY COPYRIGHT LAWBRITISH STANDARDPublishing and copyright informationThe BSI copyright notice displayed in this document indicates when the docume
2、nt was last issued. BSI 2006ISBN 0 580 48651 6The following BSI references relate to the work on this standard:Committee reference FW/1Draft for comment 05/30140467 DCPublication historyFirst published July 2006Amendments issued since publicationAmd. no. Date Text affectedBS 8480:2006 BSI 2006 iBS 8
3、480:2006ContentsForeword iii1 Scope 12 Normative references 13 Terms and definitions 14 Requirements 2Bibliography 4List of tablesTable 1 Requirements specified in BS EN 12182 2Summary of pagesThis document comprises a front cover, an inside front cover, pages i to iv, pages 1 to 4, an inside back c
4、over and a back cover.ii BSI 2006 This page deliberately left blankBS 8480:2006 BSI 2006 iiiBS 8480:2006ForewordPublishing informationThis British Standard was published by BSI and came into effect on 31 July 2006. It was prepared by Technical Committee FW/1, Common test methods for furniture. A lis
5、t of organizations represented on this committee can be obtained on request to its secretary.Information about this documentManufacturers and importers of chairs with electrically operated support surfaces should be aware that if any part of the marketing of their products can be interpreted as clai
6、ming that a product is:“intended by the manufacturer to be used for human beings for the purpose of:treatment or alleviation of disease;treatment, alleviation of or compensation for an injury or handicap;”such chairs are considered to be medical devices and therefore subject to the provisions of EC
7、Directive 93/42 EEC (the Medical Devices Directive) 1.This British Standard is applicable to chairs that are intended by the manufacturer to be used as medical devices as defined by EC Directive 93/42 (the Medical Devices Directive) 1. However, the standard is not applicable to chairs that are inten
8、ded for use during the administration of treatment by a medical practitioner, for example dentists chairs.Manufacturers of medical devices are required by law, under the Medical Devices Regulations 2002 2, to comply with the requirements of the Medical Devices Directive, including being registered w
9、ith a regulatory body in the EU known as a “competent authority”, as being responsible for placing medical devices on the market.If a company is importing medical devices into the EU and placing them on the market, and the manufacturer, or another importer within the EU, is not registered with a com
10、petent authority as being responsible for placing medical devices on the market, that company could acquire all the responsibilities of the manufacturer under the legislation.All aspects of marketing may be considered when deciding if a product is a medical device, including point of sale claims. Ma
11、nufacturers who do not have control over claims made at the point of sale are advised to consider whether they should take legal advice regarding their position.UK manufacturers and importers of medical devices can obtain guidance from the UK competent authority, which at the time of publication of
12、this British Standard is the Medicines and Healthcare products Regulatory Agency (MHRA)1).Useful guidance on the Medical Devices Directive is available from the MHRA web site at www.mhra.gov.uk which includes guidance notes for manufacturers of Class 1 medical devices and guidance notes for the regi
13、stration of persons responsible for placing devices on the market.1)Medicines and Healthcare products Regulatory Agency, Market Towers, 1 Nine Elms Lane, London SW8 5NQ. Tel: 020 7084 2000.iv BSI 2006BS 8480:2006Hazard warningsWARNING. This British Standard calls for the use of procedures that can b
14、e injurious to health if adequate precautions are not taken. It refers only to technical suitability and does not absolve the user from legal obligations relating to health and safety at any stage.Presentational conventionsThe provisions of this standard are presented in roman (i.e. upright) type. I
15、ts requirements are presented in sentences in which the principal auxiliary verb is “shall”.Commentary, explanation and general informative material is presented in smaller italic type, and does not constitute a normative element.Contractual and legal considerationsThis publication does not purport
16、to include all the necessary provisions of a contract. Users are responsible for its correct application.Compliance with a British Standard cannot confer immunity from legal obligations.BS 8480:20061 ScopeThis British Standard specifies safety requirements for chairs with support surfaces the positi
17、on of which can be adjusted electrically by the seated person and/or an attendant, including reclining chairs and riser chairs. It is applicable to chairs for use by a person weighing up to 100 kg. The standard is not applicable to multiple seating units.In the text of this standard the term “chairs
18、” is used to refer to chairs with electrically operated support surfaces.This British Standard is applicable to chairs that are intended by the manufacturer to be used as medical devices as defined by EC Directive 93/42 (the Medical Devices Directive) 1.The standard is not applicable to chairs that
19、are intended for use during the administration of treatment by a medical practitioner, for example dentists chairs.NOTE Chairs with electrically operated support surfaces that are not intended by the manufacturer to be used as medical devices are specified in BS 8474.2 Normative referencesThe follow
20、ing referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.BS 8474:2006, Furniture Chairs with electrically operated su
21、pport surfaces RequirementsBS EN 14971:2001, Medical devices Application of risk management to medical devicesBS EN 12182:1999, Technical aids for disabled persons General requirements and test methods3 Terms and definitionsFor the purposes of this British Standard the terms and definitions given in
22、 BS 8474:2006 and the following apply.3.1 manufacturernatural or legal person with responsibility for the design, manufacture, packaging and labelling of a chair before it is placed on the marketand/or put into service, regardless of whether these operations are carried out by that person himself or
23、 on his behalf by a third party European Council Directive 93/42 of 14 June 1993 (the Medical Devices Directive) Article 1.2 (f) Paragraph one 1 BSI 2006 1BS 8480:20064 Requirements4.1 Risk management A risk analysis shall be carried out in accordance with BS EN 14971 and shall include the hazards a
24、nd risks given in BS EN 12182:1999, 5.1, 5.2, 5.4.2, 5.5, Clause 6, 8.2.1, 9.4, Clause 10, Clause 21 and Clause 24.NOTE 1 The risk analysis should include consideration of hazards associated with the following:a) the user, who might have one or more disabilities;b) carers, who might be elderly and/o
25、r have disabilities;c) healthcare personnel, who might not have been trained in the use of the chair and who might need to operate the chair adjustments whilst treating or caring for the user;d) children, particularly young children, who might play around the chair out of the sight of the user and/o
26、r carer;e) pets, particularly those that might access the inner mechanism of the chair;f) other persons living in or visiting the users environment, who might not have been trained in the use of the chair.NOTE 2 If, at the conclusion of the overall residual risk evaluation (as specified in BS EN 149
27、71:2001, Clause 7), the manufacturer concludes that the intended benefits outweigh the residual risk, he has an obligation to ensure that appropriate warnings and safety instructions are clearly, permanently and prominently applied to the chair and associated literature.4.2 General requirements4.2.1
28、 Chairs shall conform to the requirements specified in BS EN 12182:1999 for the parameters listed in Table 1.Table 1 Requirements specified in BS EN 12182Parameter Requirement specified in BS EN 12182:1999 clauseIntended performance and technical documentation4.2Clinical evaluation 4.3Aids that can
29、be dismantled 4.4Single use fasteners 4.5Flammability 5.1Biocompatibility and toxicity 5.2Contaminants and residues 5.3Infection and microbiological contamination5.4Electromagnetic compatibility Clause 7Electrical safety 8.1Overflow, spillage, leakage and ingress of liquidsClause 9Information suppli
30、ed by the manufacturer Clause 232 BSI 2006BS 8480:20064.2.2 In addition, chairs shall conform to the requirements specified in BS 8474:2006, with the exception of the requirement for information to be supplied by the manufacturer, as specified in BS 8474:2006, Clause 9.4.3 Back-up power supplyChairs
31、 shall be fitted with an internal power source capable of operating all the electrically powered support surfaces through the full range of movement, including return to the starting position, not less than six times, in the event of failure of the electrical supply mains. BSI 2006 3BS 8480:2006Bibl
32、iography1 EUROPEAN COMMUNITIES. Council Directive 93/42/EEC of 14 June 1993 concerning medical devices. Luxembourg: Office for Official Publications of the European Communities, 1993.2 GREAT BRITAIN. The Medical Devices Regulations 2002, SI 2002 No. 618 (as amended). London: The Stationery Office.4
33、BSI 2006This page deliberately left blankBS 8480:2006BSI British Standards InstitutionBSI is the independent national body responsible for preparing British Standards. It presents the UK view on standards in Europe and at the international level. It is incorporated by Royal Charter.RevisionsBritish
34、Standards are updated by amendment or revision. Users of British Standards should make sure that they possess the latest amendments or editions.It is the constant aim of BSI to improve the quality of our products and services. We would be grateful if anyone finding an inaccuracy or ambiguity while u
35、sing this British Standard would inform the Secretary of the technical committee responsible, the identity of which can be found on the inside front cover. Tel: +44 (0)20 8996 9000. Fax: +44 (0)20 8996 7400.BSI offers members an individual updating service called PLUS which ensures that subscribers
36、automatically receive the latest editions of standards.Buying standardsOrders for all BSI, international and foreign standards publications should be addressed to Customer Services. Tel: +44 (0)20 8996 9001. Fax: +44 (0)20 8996 7001. Email: ordersbsi-. Standards are also available from the BSI websi
37、te at http:/www.bsi-.In response to orders for international standards, it is BSI policy to supply the BSI implementation of those that have been published as British Standards, unless otherwise requested.Information on standardsBSI provides a wide range of information on national, European and inte
38、rnational standards through its Library and its Technical Help to Exporters Service. Various BSI electronic information services are also available which give details on all its products and services. Contact the Information Centre. Tel: +44 (0)20 8996 7111. Fax: +44 (0)20 8996 7048. Email: infobsi-
39、.Subscribing members of BSI are kept up to date with standards developments and receive substantial discounts on the purchase price of standards. For details of these and other benefits contact Membership Administration. Tel: +44 (0)20 8996 7002. Fax: +44 (0)20 8996 7001. Email: membershipbsi-.Infor
40、mation regarding online access to British Standards via British Standards Online can be found at http:/www.bsi- information about BSI is available on the BSI website at http:/www.bsi-.CopyrightCopyright subsists in all BSI publications. BSI also holds the copyright, in the UK, of the publications of
41、 the international standardization bodies. Except as permitted under the Copyright, Designs and Patents Act 1988 no extract may be reproduced, stored in a retrieval system or transmitted in any form or by any means electronic, photocopying, recording or otherwise without prior written permission fro
42、m BSI.This does not preclude the free use, in the course of implementing the standard, of necessary details such as symbols, and size, type or grade designations. If these details are to be used for any other purpose than implementation then the prior written permission of BSI must be obtained.Details and advice can be obtained from the Copyright & Licensing Manager.Tel: +44 (0)20 8996 7070. Fax: +44 (0)20 8996 7553.Email: copyrightbsi-.BS 8480:2006389 Chiswick High RoadLondonW4 4AL