1、flIL-fl-272Ll1 65 U 9977906 02593h2 7 MIL-M-29241 10 MAY 1985 MILITARY SPECIFICATTON MATERIAL HANDLING UNITS FOR MEDICAL FACILITIES, PERFORMANCE SPECIFICATIONS FOR This specification is approved for use by all Departments and Agencies of the Department of Defense. 1. SCOPE 1.1 Scope. This performanc
2、e specification covers modular wall hung material handling units for medical facilities. 1.2 Classification. The material handling units specified here include storage units, distribution carts, and rail support systems. 1.2.1 Type I, storage and distribution units. Units shall have the following us
3、age. a. Clean linen storage and distribution unit, including four shelves b. and one deep drawer. Medical and surgical supplies storage and distribution unit including three shelves, three deep drawers and five shallow drawers. Pharmaceutical and biologicals storage and distribution units, including
4、 80 cassettes. Non-medical supplies storage and distribution unit including three shelves, three deep drawers and five shallow drawers. c. d. 1.2.1.1 Dl.mensions. The Type I storage units shall have the following external limiting dimensions. Width: Minimum of 22 inches and a maximum of 24 inches. D
5、epth: Minimum of 19 inches and a maximum of 24 inches. Height: Minimum of 60 inches and a maximum af 66 inches, Beneficial comments (recommendations, additions, deletions) and any pertinent data which may be of use in improving this document should be addressed to: Commanding Officer (Code 156), Nav
6、al Construction Battalion Center, Port Hueneme, CA 93043, by using the self-addressed Standardization pocument Improvement Proposal (DD Form 1426) appearing at the end of this dacument or by letter. a l I - h i FSC 6530 THIS DOCUMENT CONTAINS 12 PAGES Provided by IHSNot for ResaleNo reproduction or
7、networking permitted without license from IHS-,-,-MIL-M-272Ll1 65 = 9779906 0257363 9 W MIL-M-29241 1.2.2 Type II. distribution carts. Will be used to move and hang wall hung storage and distribution units. 1.2.2.1 Capability. Carts shall be capable of raising fully loaded units onto and lowering th
8、em from rails and transporting them to and from point of use. 1.2.3 Type III, support system. Shall include: a, Rails b. Rail hangers or fasteners 1.2.3.1 Rails and rail hangers or fasteners. Storage and distribution unit support rails shall be capable of being installed at any height above the floo
9、r and shall be shaped to securely engage the support hooks on the back of Type I storage and distribution units. 2. APPLICABLE DOCUMENTS 2.1 Government documents. 2.1.1 Specification and standards. Unless otherwise specified, the following specifications and standards of the issue listed in that iss
10、ue of the Department of Defense Index of Specifications and Standards (DODISS) specified in the solicitation form a part of this specification to the extent specified herein. STANDARDS MIL IT ARY MIL-STD-129 - Marking for Shipment and Storage, 2.2 Other publications. The following documents form a p
11、art of this specification to the extent specified herein. AMERICAN NATIONAL STANDARDS INSTITUTE, INC. (ANSI) A156.11 - Auxiliary Locks. (Application for copies should be addressed to the American National Standards Institute, Inc., 1430 Broadway, New York, NY 10018.) NATIONAL ELECTRICAL MANUFACTURER
12、S ASSOCIATION (NEMA) LD3-1980 - High Pressure Decorative Laminates. (Application for copies should be addressed to the National Electrical Manufacturers Association, 2101 L Street, N.W., Washington, DC 20037.) 2 Provided by IHSNot for ResaleNo reproduction or networking permitted without license fro
13、m IHS-,-,-MIL-M-2724L 65 9 777770b 0257364 O m MIL-M-29241 NATIONAL SANITATION FOUNDATION (NSF) a 3 No. 30 - Cabinetry and Laboratory Furniture, 1980. (Application for copies should be addressed to National Sanikary Foundation, 3475 Plymouth Road, P. O. Box 1468, Ann Arbor, MI 48106.) (Industry asso
14、ciation specifications and standards are generally available from reference libraries. Federal agencies.) They are also distributed among technical groups and 2.3 Order of precedence. If there is a conflict between the text of this specification and the references cited, the text of this specificati
15、on shall take precedence. - 3. REQUIREMENTS 3.1 Description. The modular material handling and storage units and the distribution carts, and rail support system to be provided under this performance specification shall be so designed and constructed as to meet the requirements of this section of the
16、 specification. shown in paragraph 1.2.1.1. Units shall be sized as 3.2 Manufacturers performance requirements. The manufacturer shall provide units that are standard commercial products conforming to the design and installation requirements. 3.2.1 Standard commercial product. The material handling
17、units shall be in accordance with the requirements of this specification and shall be the manufacturers standard commercial product. Additional or better features that are not prohibited by this specification but are a part of the manufacturers standard commercial product shall be included in the un
18、its. A standard commercial product is a product that has been sold or is being offered for sale on the commercial market through advertisements, manufacturers catalogs, or brochures and represents the latest production model. 3.2.2 Design. Minor deviations from the specified dimensions will not be c
19、onsidered cause for rejection provided the item is suitable for the purpose intended and is the manufacturers cataloged stock product substantially equivalent to the item described. The manufacturer shall have no design prerogatives or deviations differing from the requirements and options specified
20、 unless such deviations are numbered, notated, and described by the manufacturer and approved by the contracting officer. 3.2.3 Installation. The manufacturer shall provide the Type I storage units and the Type II distribution carts. fasteners of the Type III support system shall be provided and ins
21、talled in the specific locations indicated by the contracting officer or his representative. The rails and rail hangers or 3 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-_- MIL-M-27241 65 W 9797906 0257365 2 M HIL-H-24S41 3.3 Modularity. The Type
22、I storage units shall be modular in that they shall be interchangeable in position on the support rails and shall occupy a linear space along the support rails of no more than 24 inches. be interchangeable with drawers. 12 foot lengths that can be cut or spliced with simple hand tools. Shelves shall
23、 The support rails shall be of modular 3.4 Movability. Wall hung Type I units shall be moveable from one area to another on distribution carts that are designed to position units on the rails or remove them from the rails. placed, and secured by no more than one unskilled person without tools. Type
24、I units shall be designed to be moved, 3.5 Adjustabilitp. Rail mounting shall be at any height above the floor so that Type I storage units may be adjustable vertically. units shall have vertical adjustability of shelves, drawers, or trays throughout the height of the unit in increments no greater t
25、han 3 inches. Type I storage 3.6 Materials, Materials used shall be free from defects that would adversely affect performance or maintainability. Exterior and interior surfaces of Type I units shall be free of crevices, cracks, pits, or inclusions, and shall have rounded corners and rounded or bevel
26、ed edges. There shall be no protrusions that could cause bodily harm upon contact. 3.6.1 Cleanability. The finish surfaces shall have a cleanability at le -qual to a No. 4 finish (150 grit) on 300 series stainless steel, which onL square inch. (NSF No. 30, paragraph 3.2.1) -lean, can demonstrate a r
27、esidual soil of no more than 70 micro-grams per 3.6.2 Resistance to chemicals. When exposed to the chemicals or gases listed in NSF No. 30, Table III, Hospital Cabinetry, the surface of Type I units shall show no signs of deterioration other than a slight change of gloss, slight discoloration, or a
28、temporary slight softening of the finish with no loss of adhesion or film protection. 3.6.3 Resistance to heat. Type I units shall show no blisters, cracks, or breakdown of the surface and no perceivable change in color or surface texture when exposed to temperatures up to 356 degrees Fahrenheit (OF
29、) for 20 minutes. (NEMA LD3-1980, Sect. LD3-3.06). 3.6.4 Fire resistance. Type I units shall posed of materials that are non-combustible at temperatures of 356F or less. 3.7 Color. Color shall be as selected by the contracting officer or his representative from the complete range of the manufacturer
30、s standard colors. 3.8 Load capacity. The load capacity of Type I units shall not be less than 20 pounds per cubic foot. The lifting and load capacity of Type II, distribution carts, shall not be less than 300 pounds. Type III rail support system shall not be less than 200 pounds per linear foot. Th
31、e load capacity of 3.9 Closure and security. Type I storage and distribution units shall be provided with a roll front, lift front, hinge front, or other such closure that has no opening of more than 1/32 inch in width when closed (NSF No. 30, Provided by IHSNot for ResaleNo reproduction or networki
32、ng permitted without license from IHS-,-,-MIL-M-29241 Sect. 1, paragraph 2.3). Closures shall be designed so they may be fitted with locks that comply with ANSI A156.11. be specified (see 6.2.1). Locks and keying instructions shall 4. QUALITY ASSURANCE PROVISIONS. 4.1 Responsibility for inspection a
33、nd testinq. Unless otherwise specified in the contract or purchase order, the contractor is responsible for the performance of all inspection and testing requirements as specified herein. Except as otherwise specified in the contract or purchase order, the contractor may use his own or any other fac
34、ilities suitable for the performance of the inspection and testing requirements unless disapproved by the Government. inspections set forth in the specification where inspections are deemed necessary to assure conformance to requirements. The Government reserves the right to perform any of the 4.1.1
35、 Component and material inspection. Types I, II, and III units will be inspected and shall be tested in accordance with all the requirements specified herein and in applicable referenced documents. 4.2 First article inspection. The first article inspection and testing will be performed by the Govern
36、ment on each Type of unit to be acquired. item to be inspected and tested shall be an installed standard production item from the contractors current inventory. Unless otherwise specified, the first article shall consist of one unit of each of the items to be acquired (see 6.2.1). The 4.3 Inspection
37、. Each Type of material handling unit will be inspected by the Government on delivery. The inspection will encompass all visual and tactile aspects and dimensions that apply to the units and final assembly of the units. Failure-of a unit to meet any requirement shall constitute cause for nullificati
38、on of all acquisition documents or purchase orders for that item. 4.4 Tests. The following test shall apply to all Type I units in the acquisition documents. 4.4.1 Materials. There shall be a certificate of compliance for Type I storage and distribution units from the manufacturer stating that the m
39、aterials fulfill the requirements of paragraphs 3.6, 3.6.2, 3.6.3, and 3.6.4 and the cleanability requirements of paragraph 3.6.1. 4.5 Security tests. Type I storage and distribution units provided with locks shall be able, upon delivery, to show compliance with specifications and tests of ANSI A156
40、.11 and with paragraph 3.9. 4.6 Packaging inspection. The preservation, packaging, and marking of material handling units shall be inspected by the contractor to verify conformance with the requirements of section 5 of this specification. 5 Provided by IHSNot for ResaleNo reproduction or networking
41、permitted without license from IHS-,-,-MIL-M-2924L 65 9979706 0259367 6 MIL-M-29241 5. PACKAGING 5.1 Preservation and packaging. Unless otherwise specified, material handling units, Types I, II, and III shall be preserved and packaged in an assembled condition in a manner that will insure protection
42、 against deterioration and damage during shipment from the supply source to the first receiving activity. Drawers and doors shall be closed and secured with tape or by other means. components and small attachments, such as screws, nuts and bolts, and brackets, shall be packaged in commercial cartons
43、 or bags and secured to the module in a protected location. applicable carrier rules and regulations. (see 6.2.1) Surfaces shall be protected to prevent damage. Loose Containers and packaging shall comply with 5.2 Harkinq. Unless otherwise specified, marking shall be in accordance with MIL-STD-i29 (
44、see 6.2.1). 6. NOTES 6.1 Intended use. The material handling units are intended for use in medical facilities. 6.2 Ordering data. 6.2.1 Acquisition requirements. Acquisition documents should specify: a. Title, number, and date of this specification. b. Number of Type I, II, and III units (see Append
45、ix A). c. Locks required and keying instructions (see 3.9). d. First articles required for inspection and approval (see 4.2). e, Level of preservation and level of packing required (see 5.1). f. Marking other than as specified (see 5.2). 6,2.2 Specifying material handling units. It shall be the resp
46、onsibility of the specifier to determine the requirements for material handling units for the medical facility. 6.2.3 Guidelines. The specifier should use the guidelines and the matrix that follows to determine number and size of units required. 6.2.3.1 Assumptions. a. Point-of-use managed system. b
47、. c. d. Average distribution of commodities (for specialty units, add e. Occupancy of 100 percent or full clinic schedules, where applicable. Quantities are for 24-hour point-of-use. 15 percent to volumetric requirements). The unit of volume used in the following tables in calculating the number of
48、material handling units required for each designated space is 9 cubic feet, For space/room designations marked with an asterisk in Appendix A, the number of units provided may vary as long as the total capacity is met based on 9 cubic feet per unit. 6 , . Provided by IHSNot for ResaleNo reproduction
49、 or networking permitted without license from IHS-,-,-MIL-M-27241 65 7777906 0257368 8 MIL-M-29241 6.2.3.2 Units by space designation. units required for each space designation. for every six Type I units. Appendix A shows number of Type I There should be one Type II unit O Custodians: Army - MD Navy - YD Air Force - 03 Preparing Acticity: Navy - YD (Project 6530-1709) Provided by IHSNot for ResaleNo reproduction or networking permitted without license fr