ANSI ASTM D7739-2011 Standard Practice for Thermal Oxidative Stability Measurement via Quartz Crystal Microbalance《通过石英晶体微量天平测量热氧化稳定性的操作规程》.pdf

上传人:visitstep340 文档编号:432486 上传时间:2018-11-11 格式:PDF 页数:6 大小:315.45KB
下载 相关 举报
ANSI ASTM D7739-2011 Standard Practice for Thermal Oxidative Stability Measurement via Quartz Crystal Microbalance《通过石英晶体微量天平测量热氧化稳定性的操作规程》.pdf_第1页
第1页 / 共6页
ANSI ASTM D7739-2011 Standard Practice for Thermal Oxidative Stability Measurement via Quartz Crystal Microbalance《通过石英晶体微量天平测量热氧化稳定性的操作规程》.pdf_第2页
第2页 / 共6页
ANSI ASTM D7739-2011 Standard Practice for Thermal Oxidative Stability Measurement via Quartz Crystal Microbalance《通过石英晶体微量天平测量热氧化稳定性的操作规程》.pdf_第3页
第3页 / 共6页
ANSI ASTM D7739-2011 Standard Practice for Thermal Oxidative Stability Measurement via Quartz Crystal Microbalance《通过石英晶体微量天平测量热氧化稳定性的操作规程》.pdf_第4页
第4页 / 共6页
ANSI ASTM D7739-2011 Standard Practice for Thermal Oxidative Stability Measurement via Quartz Crystal Microbalance《通过石英晶体微量天平测量热氧化稳定性的操作规程》.pdf_第5页
第5页 / 共6页
点击查看更多>>
资源描述

1、Designation: D7739 11 (Reapproved 2016) An American National StandardStandard Practice forThermal Oxidative Stability Measurement via Quartz CrystalMicrobalance1This standard is issued under the fixed designation D7739; the number immediately following the designation indicates the year oforiginal a

2、doption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This laboratory practice covers the quantitative determi-nation of su

3、rface deposits produced during the thermal oxida-tion of gas turbine fuels by monitoring the oscillation fre-quency of a quartz crystal during thermal exposure. In thispractice, “thermal oxidative stability” refers to the tendency ofa fuel to resist surface deposit formation during heating.1.2 The v

4、alues stated in SI units are to be regarded as thestandard. The values given in parentheses are for informationonly.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate

5、 safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Summary of Practice2.1 A quartz crystal, fitted with gold electrodes, is fullyimmersed in test fuel contained within a reactor. An oscillatorcircuit, connected to the crystal, supplies energy to e

6、xcite thequartz crystal and monitors its resonant frequency (nominally5 MHz) over time via a computer interface. The reactor isequipped with a magnetic stir bar, pressure gauge/transducer,oxygen sensor (not recommended for certain test conditions,see 4.11), and thermocouple to monitor and control te

7、stconditions. Prior to testing, the fuel is bubbled with the test gasfor 30 min to equilibrate. After equilibration, the reactor vesselis isolated and raised to test temperature and pressure. Asdeposits accumulate on the crystal surface during the run, thecrystal frequency decreases. The shift in re

8、sonance frequencycan be quantitatively related, in real time, to surface depositaccumulation via a variation of the Sauerbrey equation.23. Significance and Use3.1 The tendency of a jet fuel to resist the formation ofdeposits at elevated temperature is indicative of its oxidativethermal stability. Th

9、is practice provides a technique for thesimultaneous determination of deposit formation and oxygenconsumption during the thermal oxidation of jet fuels and otherhydrocarbon liquids. The practice can be used to evaluate thethermal stability of fuels and to determine the efficacy ofadditives in inhibi

10、ting deposition or slowing oxidation, or both.Atest temperature of 140 C and run length up to 16 h has beenfound to be effective for the relative evaluation of fuels andfuel additives. This practice has also been employed for otherhydrocarbon liquids, such as gasoline and diesel fuels, butadditional

11、 safety issues may need to be addressed by the user.4. Apparatus4.1 All dimensions without tolerance limits are nominalvalues.4.2 ReactorA T316, 100 mL stainless steel reactor cylin-der with an internal diameter of 5.23 cm (2.06 in.) and a depthof 4.93 cm (1.94 in.).3,4A T316 stainless steel reactor

12、 headwith several openings (for example, gas inlet via dip tube, gasrelease fitted with a dial gauge or pressure transducer,thermocouple, safety rupture disk, frequency signalconnection, sleeve for oxygen concentration probe). A0.952 cm (38 in.) hole is drilled in the center of the reactor headto ac

13、commodate the frequency signal connectors. This holeshall have a 0.952 cm (38 in.) clearance from any adjacentopening.4.3 SMA Coaxial Connector AssemblyThis assembly pro-vides the electronic connection through the reactor head to thequartz crystal and consists of several key parts (see Fig. 1). Thec

14、able from the oscillator (see 4.6) connects to a subminiature1This practice is under the jurisdiction of ASTM Committee D02 on PetroleumProducts, Liquid Fuels, and Lubricants and is the direct responsibility of Subcom-mittee D02.J0.03 on Combustion and Thermal Properties.Current edition approved Apr

15、il 1, 2016. Published May 2016. Originallyapproved in 2011. Last previous edition approved in 2011 as D7739 11. DOI:10.1520/D7739-11R16.2Klavetter, E. A., Martin, S. J., and Wessendorf, K. O., “Monitoring Jet FuelThermal Stability Using a Quartz Crystal Microbalance,” Energy overtightening can cause

16、 the quartzcrystal to fracture, while under-tightening will not create asuitable electrical connection.7.1.6 Lock the quartz crystal per the oscillator manufactur-ers instructions.21If locked properly, frequency will vary no15The sole source of supply of the apparatus (combination of a Cole-Parmer p

17、art#04660-40 stir plate and a Fisher Scientific part#14-513-98 stir bar) known to thecommittee at this time is Cole-Parmer Instrument Company, 625 East Bunker Court,Vernon Hills, IL 60061-1844. and Thermo Fisher Scientific, 81 Wyman St.,Waltham, MA 02454.16The sole source of supply of the apparatus

18、(Mettler Toledo InPro 6800 sensorand 4100e transmitter) known to the committee at this time is Mettler-Toledo Inc.,1900 Polaris Parkway, Columbus, OH 43240.17The sole source of supply of the apparatus (Honeywell TJE pressuretransducer with a GM display unit) known to the committee at this time isHon

19、eywell International, 101 Columbia Rd., Mailstop - M6/LM, Morristown, NJ07962.18The sole source of supply of the apparatus (Inficon Model SC-501-1, Part#149211-1) known to the committee at this time is Inficon, Two Technology Place,East Syracuse, NY 13057.19The sole source of supply of the apparatus

20、 (Aldrich Catalog #26,406-7)known to the committee at this time is Sigma-Aldrich, 3050 Spruce St., St. Louis,MO 63103.20The sole source of supply of the apparatus (Alfa Aesar #14728) known to thecommittee at this time is Alfa Aesar, 26 Parkridge Rd., Ward Hill, MA 01835.21See Section 4.1, PLO-10 Ser

21、ies Phase Lock Oscillator, Operation and ServiceManual, Inficon IPN 605800, Rev. G.FIG. 2 Quartz Crystal (Front) Showing Proper Location of IndiumWireD7739 11 (2016)3more than approximately 5 Hz and voltage will vary no morethan approximately 5 mV. If the readings are not steady, thecrystal may not

22、be secure.7.1.7 An optional best practice is to record the voltage (alsoknown as conductance) of the quartz crystal in air, mV, prior toinstalling the reactor head (typically 3215 mV).7.2 Reactor:7.2.1 Insert the magnetic stir bar in the bottom of the reactorvessel.7.2.2 Charge the vessel with 60 mL

23、 of test fuel, ensuring thequartz crystal is full immersed.7.2.3 An optional best practice is to carefully remove any airbubbles from the upper surface of the test fuel. A disposablepipette has been found effective for this task. The proper settingfor the magnetic stir bar should be established duri

24、ng the initialtest setup with the reactor head off. The bar should create aslight vortex on the surface of the test fuel. Note this setting forfuture runs.7.2.4 Mate the reactor head to the reactor vessel and looselyattach the split ring assembly.7.2.5 Loosely attach the outer locking ring.7.2.6 Ori

25、ent the split ring assembly so the tightening boltswill not contact any part of the head assembly. Tighten thebolts on the split ring assembly to 33.9 N-m (25 lbf-ft) incre-mentally and in an alternating pattern to gradually and evenlysecure the head to the vessel.7.2.7 Tighten the outer locking rin

26、g.7.2.8 Insert the reactor into the heater assembly.7.2.9 Ensure the green “LOCK” indicator on the oscillator islit. If not, adjust as necessary per the oscillator manufacturersinstructions.7.2.10 An optional best practice is to record the voltage(also known as conductance) of the quartz crystal imm

27、ersed inthe test fuel, mV (typically 150 mV).7.2.11 Adjust the magnetic stirrer to the proper setting (see7.2.3).7.2.12 Attach the thermocouple, flexible pressure relieftubing, gas inlet and release tubing, to the reactor head. Thethermocouple should be placed just off the bottom of thevessel. Compl

28、ete connections to the pressure transducer andoxygen concentration transmitter, if so equipped.7.2.13 Tighten the heating band securely to the outside ofthe reactor vessel and attach the skin thermocouple if soequipped (see Fig. 4 and Fig. 5).7.2.14 Secure any safety shielding around the reactor if

29、soequipped.7.2.15 A best practice is to ensure the reactor is sealedproperly via the following, optional steps.7.2.15.1 Slowly pressurize the reactor vessel with nitrogento 689 kPa gauge (100 psig).7.2.15.2 Isolate the reactor and ensure there is no significantpressure drop for 30 min. Some minor pr

30、essure drop due tosaturation of the test fuel may be observed and is acceptable.7.2.15.3 After 30 min, slowly release the nitrogen pressure.Releasing too quickly could disturb or possibly expel the testfuel.FIG. 3 Attachment of Quartz Crystal to AdapterD7739 11 (2016)47.2.16 Slowly introduce the tes

31、t gas to saturate the test fuel.Saturation may be conducted at either atmospheric or elevatedpressure. If saturation is to be conducted at atmosphericpressure, ensure the reactor exhaust valve is open and slowlyincrease test gas flow to 50 mL min 6 10 mL min. If satura-tion is to be conducted at ele

32、vated pressure, close the reactorexhaust valve, slowly introduce the test gas and allow pressureto build to the required pressure. Use the reactor exhaust valveto then regulate test gas flow to 50 mL min 6 10 mL min.7.2.17 While saturating the test fuel, frequency stabilizationshould be verified.7.2

33、.18 After 30 min, achieve test pressure by adjusting thetest gas inlet and exhaust valves. Once test pressure has beenachieved, ensure both the test gas inlet and exhaust valves aresecurely closed, isolating the reactor.8. Procedure8.1 Energize the heater controller and activate data loggingon the c

34、ontrol software if so equipped. A data logging intervalbetween 15 s and 60 s has been found informative for this test.8.2 Initiate the test program via the control software.8.3 After the test run, turn the heater off and slowlydepressurize the reactor.8.4 Ensure the vessel has cooled to near ambient

35、 conditionsprior to opening.8.5 Remove the reactor head and dispose of the test fuelproperly.8.6 Clean all hardware that came in contact with test fuelusing wipes or swabs with the TAM solution applied (see 5.2).Perform a subsequent rinse with acetone and allow the hard-ware to air dry.9. Calculatio

36、n or Interpretation of Results9.1 Calculate the surface mass per unit area, s, per Eq 1:s5 22.21 3105!*fo2 f!/fo2(1)where:s= surface mass per unit area, g/cm2,fo= initial quartz crystal resonant frequency after hightemperature stabilization, MHz, andf = quartz crystal resonant frequency at some reac

37、tion time,t, after fo, MHz.10. Report10.1 Report test temperature, C.10.2 Report test gas utilized (for example, ultra-zero air, or60 % oxygen and 40 % nitrogen).10.3 Report initial total gas pressure, kPa.10.4 Report the calculated surface mass, s, versus test time.10.5 If recorded, report oxygen c

38、oncentration versus testtime.10.6 If recorded, report total pressure versus test time.11. Precision and Bias11.1 Due to the limited number of installations, precisionand bias has not been established for this practice.12. Keywords12.1 gas turbine fuels; quartz crystal microbalance; thermaloxidative

39、stabilityFIG. 4 Fully Assembled ReactorD7739 11 (2016)5ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentionedin this standard. Users of this standard are expressly advised that determination of the validity of any such patent

40、rights, and the riskof infringement of such rights, are entirely their own responsibility.This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years andif not revised, either reapproved or withdrawn. Your comments are invited either

41、for revision of this standard or for additional standardsand should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of theresponsible technical committee, which you may attend. If you feel that your comments have not received a fair hear

42、ing you shouldmake your views known to the ASTM Committee on Standards, at the address shown below.This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,United States. Individual reprints (single or multiple copies) of this standard

43、may be obtained by contacting ASTM at the aboveaddress or at 610-832-9585 (phone), 610-832-9555 (fax), or serviceastm.org (e-mail); or through the ASTM website(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http:/ 5 Close-Up of Reactor (Skin Thermocouple Shown Below Dial Pressure Gauge)D7739 11 (2016)6

展开阅读全文
相关资源
  • ANSI Z97 1-2009 American National Standard for Safety Glazing Materials used in Buildings - Safety Performance Specifications and Methods of Test《建筑物中窗用玻璃材料安全性用.pdfANSI Z97 1-2009 American National Standard for Safety Glazing Materials used in Buildings - Safety Performance Specifications and Methods of Test《建筑物中窗用玻璃材料安全性用.pdf
  • ANSI Z97 1 ERTA-2010 Re ANSI Z97 1 - 2009 Errata《修订版 美国国家标准学会Z97 1-2009标准的勘误表》.pdfANSI Z97 1 ERTA-2010 Re ANSI Z97 1 - 2009 Errata《修订版 美国国家标准学会Z97 1-2009标准的勘误表》.pdf
  • ANSI Z21 40 2a-1997 Gas-Fired Work Activated Air-Conditioning and Heat Pump Appliances (Same as CGA 2 92a)《燃气、工作激活空气调节和热泵器具(同 CGA 2 92a)》.pdfANSI Z21 40 2a-1997 Gas-Fired Work Activated Air-Conditioning and Heat Pump Appliances (Same as CGA 2 92a)《燃气、工作激活空气调节和热泵器具(同 CGA 2 92a)》.pdf
  • ANSI Z124 9-2004 American National Standard for Plastic Urinal Fixtures《塑料小便器用美国国家标准》.pdfANSI Z124 9-2004 American National Standard for Plastic Urinal Fixtures《塑料小便器用美国国家标准》.pdf
  • ANSI Z124 4-2006 American National Standard for Plastic Water Closet Bowls and Tanks《塑料抽水马桶和水箱用美国国家标准》.pdfANSI Z124 4-2006 American National Standard for Plastic Water Closet Bowls and Tanks《塑料抽水马桶和水箱用美国国家标准》.pdf
  • ANSI Z124 3-2005 American National Standard for Plastic Lavatories《塑料洗脸盆用美国国家标准》.pdfANSI Z124 3-2005 American National Standard for Plastic Lavatories《塑料洗脸盆用美国国家标准》.pdf
  • ANSI T1 659-1996 Telecommunications - Mobility Management Application Protocol (MMAP) RCF-RACF Operations《电信 可移动管理应用协议(MMAP) RCF-RACF操作》.pdfANSI T1 659-1996 Telecommunications - Mobility Management Application Protocol (MMAP) RCF-RACF Operations《电信 可移动管理应用协议(MMAP) RCF-RACF操作》.pdf
  • ANSI T1 651-1996 Telecommunications – Mobility Management Application Protocol (MMAP)《电信 可移动性管理应用协议》.pdfANSI T1 651-1996 Telecommunications – Mobility Management Application Protocol (MMAP)《电信 可移动性管理应用协议》.pdf
  • ANSI T1 609-1999 Interworking between the ISDN User-Network Interface Protocol and the Signalling System Number 7 ISDN User Part《电信 ISDN用户间网络接口协议和7号信令系统ISDN用户部分.pdfANSI T1 609-1999 Interworking between the ISDN User-Network Interface Protocol and the Signalling System Number 7 ISDN User Part《电信 ISDN用户间网络接口协议和7号信令系统ISDN用户部分.pdf
  • ANSI T1 605-1991 Integrated Services Digital Network (ISDN) - Basic Access Interface for S and T Reference Points (Layer 1 Specification)《综合服务数字网络(ISDN) S和T基准点的.pdfANSI T1 605-1991 Integrated Services Digital Network (ISDN) - Basic Access Interface for S and T Reference Points (Layer 1 Specification)《综合服务数字网络(ISDN) S和T基准点的.pdf
  • 猜你喜欢
    相关搜索

    当前位置:首页 > 标准规范 > 国际标准 > ANSI

    copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
    备案/许可证编号:苏ICP备17064731号-1