1、T 1217 sp-12 FORMERLY TIP 0804-06 STANDARD PRACTICE 1998 REVISED 2002 REAFFIRMED 2007 REAFFIRMED 2012 2012 TAPPI The information and data contained in this document were prepared by a technical committee of the Association. The committee and the Association assume no liability or responsibility in c

2、onnection with the use of such information or data, including but not limited to any liability under patent, copyright, or trade secret laws. The user is responsible for determining that this document is the most recent edition published. Approved by the Standard Specific Interest Group for this Tes

3、t Method TAPPI CAUTION: This Test Method may include safety precautions which are believed to be appropriate at the time of publication of the method. The intent of these is to alert the user of the method to safety issues related to such use. The user is responsible for determining that the safety

4、precautions are complete and are appropriate to their use of the method, and for ensuring that suitable safety practices have not changed since publication of the method. This method may require the use, disposal, or both, of chemicals which may present serious health hazards to humans. Procedures f

5、or the handling of such substances are set forth on Material Safety Data Sheets which must be developed by all manufacturers and importers of potentially hazardous chemicals and maintained by all distributors of potentially hazardous chemicals. Prior to the use of this method, the user must determin

6、e whether any of the chemicals to be used or disposed of are potentially hazardous and, if so, must follow strictly the procedures specified by both the manufacturer, as well as local, state, and federal authorities for safe use and disposal of these chemicals. Photometric linearity of optical prope

7、rties instruments 1. Scope 1.1 This standard practice describes a test for linearity required by the following TAPPI optical methods: T 425 Opacity T 452, 525, 534, 646 Brightness T 480, 653 Gloss T 524, 527 Color T 560, 562 Whiteness 1.2 This standard practice is normally used by instrument manufac

8、turers as the procedure for correction of photometric linearity errors. 2. Significance This method is based on the rather simple principle that, if one can measure a radiant flux corresponding to A and another radiant flux corresponding to B, then when A and B are combined, one should have an indic

9、ation corresponding to A + B. This is termed the flux addition method and implies that the current or voltage output of a photocell, or photocell plus electronic processor, should be strictly proportional to the flux upon the active surface of the cell. As an example, a linear optoelectronic system

10、generating 1 A output for a flux of 1 W (a responsivity of 1 A/W) should produce exactly 2 A if the flux is doubled to 2W, 1.24 A for 1.24W, etc. More generally, KW should generate CKA (or volts), where C is an arbitrary constant. 3. Safety precautions There are no specific safety precautions associ

11、ated with this standard practice. T 1217 sp-12 Photometric linearity of optical properties instruments / 2 4. Description A simple expression of this principle may be realized in the darkroom, using two small incandescent lamps each contained in a separate enclosure with exit aperture, blackened mat

12、te inside and out. A rectangular blackened cardboard mask is prepared so that either lamp may be shuttered from an external photoreceptor, either separately or simultaneously (see Figure 1). A 0.01% regulated variable DC supply may power both lamps, although two independent supplies provide greater

13、flexibility. A concentrating lens system is recommended for each lamp. Each source must also contain a variable iris or spectrally non-selective graded-density neutral wedge, plus optical filtering to restrict energy falling upon the photoreceptor to the spectral range of interest. (If insensitive p

14、hotodetectors are used, filtering must include a good infrared-blocking component for tests in the visual region.) To minimize error from multiple reflections between the sources and receptor assembly, the receptor should be separated from them by a minimum of 60 mm, and any specular interreflection

15、s avoided by slight tilting of components. The two sources should subtend an angle of not more than 6at the active surface of a flat photodetector. (For strongly curved photodetector surfaces, such as those of many vacuum phototubes, it is necessary to eliminate shadowing by eliminating the angular

16、difference between the two beams by means of a beam-combining pellicle or semi-transparent mirror.) Distribution of flux over the active surface should be as uniform as practicable. The receptor photocell should be shielded from stray light. NOTE 1: This method is not suitable for use (without modif

17、ication) where an integrating sphere, used in a substitution-type measurement mode, forms a part of the optical system. 5. Procedure 5.1 Set each lamp to its rated voltage, or somewhat below this for increased life. Using the variable iris or neutral-density wedge in conjunction with the blackened m

18、ask, first set the readout to zero with both beams obscured; then set each beam to register 50.00. Recheck zero. When the mask is entirely removed, the readout should become 50.00 + 50.00 = 100.00. Any deviation from the arithmetic sum of the two beams may be taken as an indication of the nonlineari

19、ty of the system. Make several determinations and compute the average delta from 100.00. 5.2 Measurements at other irradiance levels (5X, 25X, 1/5, 1/25) may be performed for extended range linearity determinations. Alternatively, several intermediate linearity steps (such as 0, 20, 40, 60, 80, 100)

20、 are available by setting the first (“stepping”) lamp to 20, shuttering it and matching the output with the second (“matching”) lamp, exposing both lamps for a combined reading corresponding to 40, reshuttering the stepping lamp and adjusting the matching lamp to this reading, reexposing the steppin

21、g lamp, etc., until full-scale is achieved (1). In this case it is useful to normalize the result so that the average full-scale reading becomes 100. 6. Keywords Optical properties, Optical instruments, Photometry, Linearity 3 / Photometric linearity of optical properties instruments T 1217 sp-12 7.

22、 Additional information 7.1 Effective date of issue: April 24, 2012. 7.2 This standard practice formerly was TIP 0804-06 and was revised in 1998. Literature cited 1. Hfert H.J., and Loof, H., “Calibration of the Photometric Scale of a Reflectance Photometer,” Zeitschrift fr Instrumentenkunde 72:139 (1964); communication in English, Carl Zeiss, Oberkochen. Your comments and suggestions on this procedure are earnestly requested and should be sent to the TAPPI Standards Department. g

展开阅读全文
相关资源
  • ANSI Z97 1-2009 American National Standard for Safety Glazing Materials used in Buildings - Safety Performance Specifications and Methods of Test《建筑物中窗用玻璃材料安全性用.pdfANSI Z97 1-2009 American National Standard for Safety Glazing Materials used in Buildings - Safety Performance Specifications and Methods of Test《建筑物中窗用玻璃材料安全性用.pdf
  • ANSI Z97 1 ERTA-2010 Re ANSI Z97 1 - 2009 Errata《修订版 美国国家标准学会Z97 1-2009标准的勘误表》.pdfANSI Z97 1 ERTA-2010 Re ANSI Z97 1 - 2009 Errata《修订版 美国国家标准学会Z97 1-2009标准的勘误表》.pdf
  • ANSI Z21 40 2a-1997 Gas-Fired Work Activated Air-Conditioning and Heat Pump Appliances (Same as CGA 2 92a)《燃气、工作激活空气调节和热泵器具(同 CGA 2 92a)》.pdfANSI Z21 40 2a-1997 Gas-Fired Work Activated Air-Conditioning and Heat Pump Appliances (Same as CGA 2 92a)《燃气、工作激活空气调节和热泵器具(同 CGA 2 92a)》.pdf
  • ANSI Z124 9-2004 American National Standard for Plastic Urinal Fixtures《塑料小便器用美国国家标准》.pdfANSI Z124 9-2004 American National Standard for Plastic Urinal Fixtures《塑料小便器用美国国家标准》.pdf
  • ANSI Z124 4-2006 American National Standard for Plastic Water Closet Bowls and Tanks《塑料抽水马桶和水箱用美国国家标准》.pdfANSI Z124 4-2006 American National Standard for Plastic Water Closet Bowls and Tanks《塑料抽水马桶和水箱用美国国家标准》.pdf
  • ANSI Z124 3-2005 American National Standard for Plastic Lavatories《塑料洗脸盆用美国国家标准》.pdfANSI Z124 3-2005 American National Standard for Plastic Lavatories《塑料洗脸盆用美国国家标准》.pdf
  • ANSI T1 659-1996 Telecommunications - Mobility Management Application Protocol (MMAP) RCF-RACF Operations《电信 可移动管理应用协议(MMAP) RCF-RACF操作》.pdfANSI T1 659-1996 Telecommunications - Mobility Management Application Protocol (MMAP) RCF-RACF Operations《电信 可移动管理应用协议(MMAP) RCF-RACF操作》.pdf
  • ANSI T1 651-1996 Telecommunications – Mobility Management Application Protocol (MMAP)《电信 可移动性管理应用协议》.pdfANSI T1 651-1996 Telecommunications – Mobility Management Application Protocol (MMAP)《电信 可移动性管理应用协议》.pdf
  • ANSI T1 609-1999 Interworking between the ISDN User-Network Interface Protocol and the Signalling System Number 7 ISDN User Part《电信 ISDN用户间网络接口协议和7号信令系统ISDN用户部分.pdfANSI T1 609-1999 Interworking between the ISDN User-Network Interface Protocol and the Signalling System Number 7 ISDN User Part《电信 ISDN用户间网络接口协议和7号信令系统ISDN用户部分.pdf
  • ANSI T1 605-1991 Integrated Services Digital Network (ISDN) - Basic Access Interface for S and T Reference Points (Layer 1 Specification)《综合服务数字网络(ISDN) S和T基准点的.pdfANSI T1 605-1991 Integrated Services Digital Network (ISDN) - Basic Access Interface for S and T Reference Points (Layer 1 Specification)《综合服务数字网络(ISDN) S和T基准点的.pdf
  • 猜你喜欢
    相关搜索

    当前位置:首页 > 标准规范 > 国际标准 > ANSI

    copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
    备案/许可证编号:苏ICP备17064731号-1