资源描述

1、 FORD LABORATORY TEST METHOD AK 116-01 Date Action Revisions 2001 01 22 Revised Editorial no technical change A. Cockman 2000 03 23 Revised Editorial no technical change A. Cockman 1990 06 12 Printed copies are uncontrolled Page 1 of 4 Copyright 2001, Ford Global Technologies, Inc. QUANTITATIVE ANAL

2、YSIS OF SURFACTANTS AND WETTING AGENTS IN CLEANERS Application This test method is used to determine the percent surfactants and wetting agents in a cleaner. The procedure is intended to be a simplified version of the applicable ASTM standard (e.g. D 501 or D 800) as well as being more appropriate f

3、or dealing with the many formulations of cleaners that are commonly encountered in plant operations. Apparatus Required Infrared Analyzer Suitable for use with cleaners. Magnetic Stirrer Hot Plate Analytical Balance Precision +/- 0.1 mg Oven Mechanical convection; capable of maintaining 105 +/- 2 C

4、for 1 h Steam Bath Desiccator Flask (with reflux condenser) 250 mL Beakers 250 mL Funnel Suitable for use with filter paper Graduated Cylinder 100 mL FORD LABORATORY TEST METHOD AK 116-01 Page 2 of 4 Copyright 2001, Ford Global Technologies, Inc. Glass Rod The above apparatus is available from any r

5、eputable laboratory equipment supply house. Materials Required Hydrochloric Acid Concentrated, Reagent grade; CAUTION: Corrosive Chloroform Reagent grade; CAUTION: Good ventilation required Methyl Alcohol Reagent grade; CAUTION: Low flash point Acetone Reagent grade; CAUTION: Low flash point; good v

6、entilation required Distilled or Deionized Water 20 micromhos/cm at 25 C maximum conductivity (ASTM D 1125) Filter Paper Whatman No. 541 Solutions Required Hydrochloric Acid Solution 0.5 N Conditioning and Test Conditions All test values indicated herein are based on material conditioned in a contro

7、lled atmosphere of 23 +/- 2 C and 50 +/- 5 % relative humidity for not less than 24 h prior to testing and tested under the same conditions unless otherwise specified. FORD LABORATORY TEST METHOD AK 116-01 Page 3 of 4 Copyright 2001, Ford Global Technologies, Inc. Procedure 1. Take a 1 to 5 g sample

8、 of the material in a 250 mL flask. A. Powders Start extractions B. Liquids Dry at 100 C for 16 h to remove moisture prior to extractions C. If caustic is present Neutralize with 0.5 N HCl. Dry at 100 C to remove moisture. Then run the extractions. 2. Extract the sample from 1A, 1B or 1C with chloro

9、form as follows: (A) Add 30 to 40 mL chloroform to the sample in the 250 mL flask (with reflux condenser). 1. Heat to boiling. 2. Remove from hot plate. 3. Agitate using magnetic stirrer. 4. Let settle. 5. Filter by decantation into a tared 250 mL beaker through Whatman 541 filter paper. 6. Repeat S

10、teps 2A1 through 2A5 for a total of 4 extractions. 7. Gently wash out with residual chloroform in the beaker onto the filter. Note: Make sure that a minimum of insolubles present is washed onto the filter. 8. Wash the filter for 5 to 6 times to remove any chloroform solubles into the tared beaker. (

11、B) Add 30 to 40 mL acetone to the residue in the 250 mL flask. 1. Repeat steps in 2A using acetone instead of chloroform. (C) Add 30 to 40 mL methyl alcohol to the residue in the 250 mL flask. 1. Repeat steps in 2A using methyl alcohol instead of chloroform. (D) Evaporate the solvents from the extra

12、ctions on a steam bath. (E) After solvent removal, dry in oven at 105 C for 1 h. FORD LABORATORY TEST METHOD AK 116-01 Page 4 of 4 Copyright 2001, Ford Global Technologies, Inc. (F) Cool in desiccator. (G) Weigh. (H) Calculate, % of recovered material in each beaker. (I) Check recovery by infrared.

13、Combine percentage if necessary. Interferences: 1. If nitrites are present, they will usually be found in the alcohol soluble extract. 2. If chelating agents are present, they may also be soluble in the chloroform, acetone or alcohol. 3. Nitrites and chelating agents are not surfactants or wetting agents. Chemicals, materials, parts, and equipment referenced in this document must be used and handled properly. Each party is responsible for determining proper use and handling in its facilities.

展开阅读全文
相关资源
猜你喜欢
  • BS 7882-2017 Method for calibration and classification of torque measuring devices .pdf BS 7882-2017 Method for calibration and classification of torque measuring devices .pdf
  • BS 7883-2005 Code of practice for the design selection installation use andmainten ance of anchor devices conforming to BS EN 795《符合BS EN 795标准的固定装置的设计、选择安装、使用和维护的实用规程》.pdf BS 7883-2005 Code of practice for the design selection installation use andmainten ance of anchor devices conforming to BS EN 795《符合BS EN 795标准的固定装置的设计、选择安装、使用和维护的实用规程》.pdf
  • BS 7884-1997 Specification for copper and copper-cadmium stranded conductors for overhead electric traction and power transmission systems《架空电力牵引和传输系统用铜和铜镉绞捻线规范》.pdf BS 7884-1997 Specification for copper and copper-cadmium stranded conductors for overhead electric traction and power transmission systems《架空电力牵引和传输系统用铜和铜镉绞捻线规范》.pdf
  • BS 7885-1997 Code of practice for safe entry into silos《筒仓安全通道实用规程》.pdf BS 7885-1997 Code of practice for safe entry into silos《筒仓安全通道实用规程》.pdf
  • BS 7887-1997 Ultrasonics - Real-time pulse-echo systems - Guide for test procedures to determine performance specifications《超声波学 实时脉冲回波系统 测定性能规范的试验程序指南》.pdf BS 7887-1997 Ultrasonics - Real-time pulse-echo systems - Guide for test procedures to determine performance specifications《超声波学 实时脉冲回波系统 测定性能规范的试验程序指南》.pdf
  • BS 7888-4 1-2006+A1-2008 Test requirements on naccessories for use non power cables of nrated voltage from n3 6 6(7 2) kV up to n20 8 36(42) kV — nPart 1 Cables with extruded insul.pdf BS 7888-4 1-2006+A1-2008 Test requirements on naccessories for use non power cables of nrated voltage from n3 6 6(7 2) kV up to n20 8 36(42) kV — nPart 1 Cables with extruded insul.pdf
  • BS 7888-4 2-2006+A1-2008 Test requirements on naccessories for use non power cables of nrated voltage from n3 6 6(7 2) kV up to n20 8 36(42) kV — nPart 2 Cables with impregnated pa.pdf BS 7888-4 2-2006+A1-2008 Test requirements on naccessories for use non power cables of nrated voltage from n3 6 6(7 2) kV up to n20 8 36(42) kV — nPart 2 Cables with impregnated pa.pdf
  • BS 7889-2012 Electric cables Thermosetting insulated non-armoured cables with a voltage of 600 1000 V for fixed installations《电缆 固定装置用600 1 000V电压热固绝缘性非铠装电缆》.pdf BS 7889-2012 Electric cables Thermosetting insulated non-armoured cables with a voltage of 600 1000 V for fixed installations《电缆 固定装置用600 1 000V电压热固绝缘性非铠装电缆》.pdf
  • BS 7895-1997 Specification for bayonet lampholders with enhanced safety《增强式安全卡口灯座规范》.pdf BS 7895-1997 Specification for bayonet lampholders with enhanced safety《增强式安全卡口灯座规范》.pdf
  • 相关搜索

    当前位置:首页 > 标准规范 > 国际标准 > 其他

    copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
    备案/许可证编号:苏ICP备17064731号-1