1、BSI Standards Publication Ultrasonics Pulse-echo scanners Low-echo sphere phantoms and method for performance testing of gray-scale medical ultrasound scanners applicable to a broad range of transducer types PD IEC/TS 62791:2015National foreword This Published Document is the UK implementation of IE
2、C/TS 62791:2015. The UK participation in its preparation was entrusted to Technical Committee EPL/87, Ultrasonics. A list of organizations represented on this committee can be obtained on request to its secretary. This publication does not purport to include all the necessary provisions of a contrac
3、t. Users are responsible for its correct application. The British Standards Institution 2015. Published by BSI Standards Limited 2015 ISBN 978 0 580 80391 8 ICS 11.040.50; 17.140.50 Compliance with a British Standard cannot confer immunity from legal obligations. This Published Document was publishe
4、d under the authority of the Standards Policy and Strategy Committee on 30 September 2015. Amendments/corrigenda issued since publication Date Text affected PUBLISHED DOCUMENT PD IEC/TS 62791:2015 IEC TS 62791 Edition 1.0 2015-09 TECHNICAL SPECIFICATION Ultrasonics Pulse-echo scanners Low-echo spher
5、e phantoms and method for performance testing of gray-scale medical ultrasound scanners applicable to a broad range of transducer types INTERNATIONAL ELECTROTECHNICAL COMMISSION ICS 11.040.50; 17.140.50 ISBN 978-2-8322-2902-6 Registered trademark of the International Electrotechnical Commission Warn
6、ing! Make sure that you obtained this publication from an authorized distributor. colour inside PD IEC/TS 62791:201 2 IEC TS 62791:2015 IEC 2015 CONTENTS FOREWORD . 6 INTRODUCTION . 8 1 Scope 10 2 Normative references 10 3 Terms and definitions 10 4 Symbols 12 5 General and environmental conditions
7、. 13 6 Equipment required . 14 6.1 General . 14 6.2 Phantom geometries . 14 6.2.1 Phantoms for use in the frequency range 2 MHz to 7 MHz . 14 6.2.2 Phantoms for use in the frequency range 7 MHz to 15 MHz including “micro-convex“ arrays 14 6.2.3 Total internal-reflection surfaces 15 6.2.4 Spatially r
8、andom distribution of low-echo spheres 15 6.3 Ultrasonic properties of the tissue-mimicking (TM) phantoms 15 7 Data acquisition assuming a spatially random distribution of low-echo spheres . 16 7.1 Methodology . 16 7.2 Storage of digitized image data . 17 7.3 Digital image files available from the s
9、canner itself . 18 7.4 Image archiving systems . 18 8 Automated data analysis for quantifying low-echo sphere detectability 18 8.1 General . 18 8.2 Computation of mean pixel values (MPVs) 18 8.3 Determination of the LSNR m -value for a given depth interval . 21 8.3.1 Preliminaries . 21 8.3.2 Computa
10、tion of the LSNR n -values and LSNR m -value in a given depth interval 21 8.3.3 Standard error corresponding to each LSNR n -value 21 Annex A (informative) Example of a phantom for performance testing in the 2 MHz to 7 MHz frequency range . 22 Annex B (informative) Illustrations of the computation o
11、f LSNR m -values as a function of depth 24 Annex C (informative) Sufficient number of data images to assure reproducibility of results 29 C.1 General . 29 C.2 Phantom with low-echo sphere diameter 3,2 mm, having 2 spheres per millilitre . 29 C.3 Phantom with 2 mm-diameter, low-echo spheres and 8 sph
12、eres per millilitre 32 Annex D (informative) Example of a phantom for performance testing in the 7 MHz to 15 MHz frequency range . 36 Annex E (informative) Determination of low-echo sphere positions to within D/8 in x, y and z Cartesian coordinates 39 E.1 Procedure . 39 E.2 Argument for the choice o
13、f seven MPV nearest-neighbour sites for determining the centres of low-echo spheres 40 PD IEC/TS 62791:2015IEC TS 62791:2015 IEC 2015 3 Annex F (informative) Test of total internal reflection produced by alumina and plate- glass, plane reflectors . 41 Annex G (informative) Results of a test of repro
14、ducibility of LSNR mversus depth for a phantom with 4 mm-diameter low-echo spheres and 2 spheres per millilitre 48 Annex H (informative) Results for low-echo sphere-concentration dependence of LSNR mversus depth for phantoms with 4 mm-diameter spheres . 50 Annex I (informative) Results for low-echo
15、sphere-concentration dependence of LSNR m versus depth for phantoms with 3,2 mm-diameter spheres 53 Annex J (informative) Comparison of two different makes of scanner with similar transducers and console settings 57 Annex K (informative) Special considerations for 3-D probes 59 K.1 3-D probes operat
16、ing in 2-D imaging mode . 59 K.2 2-D arrays operating in 3-D imaging mode for determining LSNR m -values as a function of depth for reconstructed images . 59 K.3 Mechanically driven 3-D probes operating in 3-D imaging mode 59 Bibliography 60 Figure 1 Flow chart 17 Figure 2 Schematic of an image plan
17、e . 20 Figure A.1 End view of the phantom applicable for 2 MHz to 7 MHz showing the spatially random distribution of 4-mm diameter low-echo spheres . 22 Figure A.2 Top view of phantom with 4 mm-diameter, low-echo spheres 23 Figure B.1 Convex-array image of a prototype 4 mm-diameter low-echo sphere p
18、hantom for use in the 2 MHz to 7 MHz frequency range 24 Figure B.2 Auxiliary figures relating to Figure B.1 25 Figure B.3 Results corresponding to Figures B.1 and B.2, demonstrating reproducibility . 25 Figure B.4 Results corresponding to Figures B.1, B.2 and B.3 . 26 Figure B.5 One of 80 parallel l
19、inear-array images of the phantom containing 4 mm- diameter, low-echo spheres, at 4 MHz with focus at 3 cm . 26 Figure B.6 Three successive images of the set of 80, separated by D/4 equal to 1 mm 27 Figure B.7 Results for the 4 cm-wide, 3 cm-focus, linear array addressed in Figures B.5 and B.6 . 27
20、Figure B.8 Results for the 4 cm-wide, 3 cm-focus, linear array addressed in Figures B.5, B.6 and B.7, using all 80 image frames corresponding to Figure B.7 . 28 Figure C.1 One image obtained from a phantom containing 3,2 mm-diameter, low- echo spheres by using a 4 MHz linear array focused at 3 cm 29
21、 Figure C.2 Reproducibility result for two independent sets of 70 images with a mean number of low-echo sphere centres that is about 15 per 5 mm-depth interval 30 Figure C.3 Results obtained by using both sets of 70 independent images corresponding to Figure C.2 30 Figure C.4 Sector image (curved ar
22、ray) at 4,5 MHz with multiple foci at 4 cm, 8 cm and 12 cm depths; the low-echo spheres are 3,2 mm in diameter . 31 Figure C.5 Reproducibility results for a multiple-lateral-focus (4 cm, 8 cm and 12 cm) case corresponding to Figure C.4 . 31 Figure C.6 Reproducibility results for the case correspondi
23、ng to Figure C.5, except that there is a single focus at 10 cm depth 32 Figure C.7 Reproducibility results for the case corresponding to Figure C.5, except that there is a single focus at 4 cm depth 32 PD IEC/TS 62791:2015 4 IEC TS 62791:2015 IEC 2015 Figure C.8 Image of the phantom containing 2 mm-
24、diameter, low-echo spheres, made with a curved array having 1,5 cm radius of curvature, with its focus at 3 cm 33 Figure C.9 Reproducibility results corresponding to Figure C.8 33 Figure C.10 Results using all 100 images in the image set that gave rise to Figure C.9 . 34 Figure C.11 Image of the pha
25、ntom containing 2 mm-diameter, low-echo spheres, made with a high-frequency (15 MHz) linear array, laterally focused at 4 cm. 34 Figure C.12 Reproducibility results corresponding to Figure C.11 35 Figure C.13 Results using all 200 images in the image set that gave rise to Figure C.12 35 Figure D.1 E
26、nd- and top-view diagrams of the phantom containing 2 mm-diameter, low-echo spheres for use in the 7 MHz to 15 MHz frequency range. 37 Figure D.2 Image obtained by using the phantom containing 2 mm-diameter, low- echo spheres and a pediatric transducer with a radius of curvature of about 1,5 cm . 38
27、 Figure F.1 Average of 10 images obtained by using a phased array . 42 Figure F.2 Plot of the data with blue data computed in the left rectangle in Figure F.1 and red data computed in the right rectangle 42 Figure F.3 Plot of the data when the reflector is on the right side with blue computed in the
28、 left rectangle and red computed in the right rectangle . 43 Figure F.4 The percentage by which the mean pixel values resulting from reflections differ from the mean pixel values not involving reflections . 44 Figure F.5 Wide sector (153), 1 cm-radius-of-curvature transducer with alumina reflector o
29、n the left 45 Figure F.6 Plot of the data with blue computed in the left rectangle in Figure F.5 and red computed in the right rectangle . 45 Figure F.7 Plot of the data when the reflector is on the right side with blue computed in the left rectangle and red computed in the right rectangle . 46 Figu
30、re F.8 The percentage by which the mean pixel values resulting from reflections differ from the mean pixel values not involving reflections . 46 Figure G.1 Example image of the phantom with a 4,2 MHz curved array and two low- echo spheres per millilitre . 48 Figure G.2 Reproducibility results corres
31、ponding to the image set, one of which is shown in Figure G.1 49 Figure H.1 Example of an image from the image set giving rise to the results in Figure H.2; the phantom contained an average of one 4 mm-diameter, low-echo sphere per millilitre . 50 Figure H.2 Results corresponding to an image set, on
32、e of which is shown in Figure H.1 51 Figure H.3 Example of an image from the data set giving rise to the results in Figure H.4; the phantom contained an average of two 4 mm-diameter, low-echo spheres per millilitre . 51 Figure H.4 Results corresponding to an image set, one of which is shown in Figur
33、e H.3 52 Figure I.1 Example of an image from the 4 ml 1data set producing the results shown in Figure I.2 53 Figure I.2 Results for the phantom containing four 3,2 mm-diameter, low-echo spheres per millilitre 54 Figure I.3 Example of an image from the 2 ml 1data set producing the results shown in Fi
34、gure I.4 54 Figure I.4 Results for the phantom containing two 3,2 mm-diameter, low-echo spheres per millilitre 55 PD IEC/TS 62791:2015IEC TS 62791:2015 IEC 2015 5 Figure I.5 Example of an image from the 1 ml 1data set producing the results shown in Figure I.6 55 Figure I.6 Results for the phantom co
35、ntaining one 3,2 mm-diameter, low-echo sphere per millilitre . 56 Figure J.1 Results for System A scanner and 7CF2 3-D (swept convex array) transducer focused at 4 cm and operated at 4,5 MHz in 2-D mode 57 Figure J.2 Results for System B scanner with a 4DC7-3 3-D (convex array) transducer, operated
36、at 4 MHz in 2-D mode and focused at 4 cm. The sector angle and all other console settings mimicked those for the System A case (Figure J.1) . 57 PD IEC/TS 62791:2015 6 IEC TS 62791:2015 IEC 2015 INTERNATIONAL ELECTROTECHNICAL COMMISSION _ ULTRASONICS PULSE-ECHO SCANNERS LOW-ECHO SPHERE PHANTOMS AND
37、METHOD FOR PERFORMANCE TESTING OF GRAY-SCALE MEDICAL ULTRASOUND SCANNERS APPLICABLE TO A BROAD RANGE OF TRANSDUCER TYPES FOREWORD 1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising all national electrotechnical committees (IEC National C
38、ommittees). The object of IEC is to promote international co-operation on all questions concerning standardization in the electrical and electronic fields. To this end and in addition to other activities, IEC publishes International Standards, Technical Specifications, Technical Reports, Publicly Av
39、ailable Specifications (PAS) and Guides (hereafter referred to as “IEC Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested in the subject dealt with may participate in this preparatory work. International, governmental and non- governmental
40、 organizations liaising with the IEC also participate in this preparation. IEC collaborates closely with the International Organization for Standardization (ISO) in accordance with conditions determined by agreement between the two organizations. 2) The formal decisions or agreements of IEC on techn
41、ical matters express, as nearly as possible, an international consensus of opinion on the relevant subjects since each technical committee has representation from all interested IEC National Committees. 3) IEC Publications have the form of recommendations for international use and are accepted by IE
42、C National Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any misinterpretation by any end user. 4) In order to promote international unifor
43、mity, IEC National Committees undertake to apply IEC Publications transparently to the maximum extent possible in their national and regional publications. Any divergence between any IEC Publication and the corresponding national or regional publication should be clearly indicated in the latter. 5)
44、IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any services carried out by independent certification bodies. 6) All users should ensur
45、e that they have the latest edition of this publication. 7) No liability should attach to IEC or its directors, employees, servants or agents including individual experts and members of its technical committees and IEC National Committees for any personal injury, property damage or other damage of a
46、ny nature whatsoever, whether direct or indirect, or for costs (including legal fees) and expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC Publications. 8) Attention is drawn to the Normative references cited in this publication. Use of the ref
47、erenced publications is indispensable for the correct application of this publication. The main task of IEC technical committees is to prepare International Standards. In exceptional circumstances, a technical committee may propose the publication of a Technical Specification when the required suppo
48、rt cannot be obtained for the publication of an International Standard, despite repeated efforts, or the subject is still under technical development or where, for any other reason, there is the future but no immediate possibility of an agreement on an International Standard. Technical Specification
49、s are subject to review within three years of publication to decide whether they can be transformed into International Standards. Technical Specification IEC TS 62791 has been prepared by IEC technical committee 87 Ultrasonics. PD IEC/TS 62791:2015IEC TS 62791:2015 IEC 2015 7 The text of this Technical Specification is based on the following documents: DTS Report on voting 87/554/DTS 87/570/RVC Full information on the vo