1、Medical electrical equipment Deployment, implementation and operational guidelines for identifying febrile humans using a screening thermograph PD ISO/TR 13154:2017 BSI Standards Publication WB11885_BSI_StandardCovs_2013_AW.indd 1 15/05/2013 15:06National foreword This Published Document is the UK i
2、mplementation of ISO/TR 13154:2017. It supersedes PD ISO/TR 13154:2009 which is withdrawn. The UK participation in its preparation was entrusted by Technical Committee CH/121, Anaesthetic and respiratory equipment, to Subcommittee CH/121/9, Lung Ventilators multiple 21 During the outbreaks of pandem
3、ics, internationally agreed measures designed to restrict the movement of possibly infected people were instituted and were assessed by WHO to have greatly contributed to bringing the disease under control. Possible measures to delay or slow the transmission of infectious diseases include: providing
4、 travel advice on travel to and from affected countries; providing health information for exiting and returning travellers; providing health screening at ports of entry and exit. 1112182027282930313237404142 434445 In a severe pandemic, absenteeism attributable to illness, the need to care for ill f
5、amily members and fear of infection can reach 40 % during the peak weeks of a community outbreak, with lower rates of absenteeism during the weeks before and after the peak. Certain public health measures (closing schools, quarantining household contacts of infected individuals) are likely to increa
6、se rates of absence from the workplace. Actions that reduce the likelihood of disease exposure and limit transmission, assure the public of the ability to maintain domestic safety and security, advise the public to curtail non-essential travel and communal activities while preparing for implementati
7、on of community disease containment measures as the epidemic spreads, are important public policy objectives. 91011To support the objective of pandemic prevention, a screening thermograph with appropriate follow-up of febrile persons can be useful to separate potentially infectious individuals from
8、others in locations such as: 20 entrances to hospitals and clinics, including emergency rooms; entrances to critical infrastructure facilities; entrances to workplaces; entrances to schools; entrances to government buildings, including police and fire stations; entrances to other communal locations;
9、 public transportation. A screening thermograph should be an element of the layered screening process for those diseases specifically associated with elevated fever. It can also play an important epidemiological role in defining vi ISO 2017 All rights reserved PD ISO/TR 13154:2017 ISO/TR 13154:2017(
10、E) the geographical boundaries of an outbreak. A screening thermograph complying with IEC 80601-2- 59 is a non-contact, accurate and repeatable means of quickly screening individuals for fever when proper procedures are followed. International experience has demonstrated numerous instances of noncom
11、pliance with the International Standard and the recommendations of the first edition of this document, as well as Internet postings documenting inappropriate use of radiometry and infrared cameras. It should emphasized that the efficacy and utility of this technology is contingent on the use of both
12、 equipment and protocols meeting the recommendations of these committees. 45 NOTE The requirements for a screening thermograph are found in IEC 806012-59. In this document, the following print types are used: Text and definitions: roman type. Informative material appearing outside of tables, such as
13、 notes, examples and references: in smaller type. Normative text of tables is also in a smaller type. Terms defined in clause 3 or as noted: small capitals. In this document, the conjunctive “or” is used as an “inclusive or” so a statement is true if any combination of the conditions is true. ISO 20
14、17 All rights reserved vii PD ISO/TR 13154:2017 Medical electrical equipment Deployment, implementation and operational guidelines for identifying febrile humans using a screening thermograph 1 Scope This document provides general guidelines for the deployment, implementation and operation of a scre
15、ening thermograph intended to be used for non-invasive febrile temperature screening of individuals under indoor environmental conditions to prevent the spread of infection. NOTE The equipment standard for screening thermographs is found in IEC 806012-59. 2 Normative references There are no normativ
16、e references in this document. 3 T erms a nd definiti ons For the purposes of this document, the following terms and definitions apply. ISO and IEC maintain terminological databases for use in standardization at the following addresses: IEC Electropedia: available at h t t p :/ www .electropedia .or
17、g/ ISO Online browsing platform: available at h t t p :/ www .iso .org/ obp NOTE For convenience, an alphabetized index of defined terms used in this document is given in Annex C. 3.1 a c cesso r y additional part for use with equipment in order to achieve the intended use; adapt it to some special
18、use; facilitate its use; enhance its performance; enable its functions to be integrated with those of other equipment. SOURCE: IEC 60601-1:2005, 3.3 3.2 a c c o m p a n y i n g d o c u m e n t document accompanying me equipment, an me system, equipment or an accessory and containing information for the responsible organization or operator , particularly regarding basic safety and essential performance SOURCE: IEC 60601-1:2005, 3.4 TECHNICAL REPORT ISO/TR 13154:2017(E) ISO 2017 All rights reserved 1 PD ISO/TR 13154:2017