1、 - A-A-59355 9038054 O003973 838 = -Beneficial cmments, recommendations, additions, deletions, clarifications, etc., and any data which may improve this document should be sent to SA-ALC/TILDD, 485 Quentin Roosevelt Rd., Kelly AFB, TX 78241-6425 by using the Standardization Document Improvement Prop

2、osal (DD Form 1426) appearing at the end of this document. I INCH-POUNDI A-A-59355 19 October 1998 COMMERCIAL ITEM DESCRIPTION LEAK DETECTOR, HELIUM The General Services Administration has authorized the use of this Commercial Item Description (CID) for all federal agencies. 1. This detector will be

3、 used to calibrate mass spectrometers at military installations around the world. SCOPE. This document outlines the minimum requirements for a helium leak detector. 2. SALIENT CHARACTERISTICS. 2.1 General. This calibrated helium leak detector consists of a controlled source of helium ion gas traceab

4、le to the National Institute of Standards and Technology (NIST) external reference standards used in calibrating leak detectors. . 2.2 Leak rate. The leak rate is factor relates to the minimum detectable leak rate in std cclsec for helium molecules (1 std cdsec = 0.76 torr-litedsec). std cc/sec mini

5、mum detectable partial pressure of helium. This 2.3 Leak rate accuracv. The leak rate shall be less than or equal to 2 10 percent std cc/sec. 2.4 Media. The media shall be helium gas. 2.5 Tvpe of leak. External application. 2.6 Fittings. 1- 1/8-inch outside diameter compression fitting. 2.7 Value ty

6、pe. Manual isolation. AMSC NIA FSC 4940 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-59355 9038054 0003972 774 A-A-5 93 5 5 2.8 Outer casing material. Corrosion-resistant stainless steel. 2.9 Calibration. Each leak detector shall be calibrated

7、 and traceable to the NIST reference standard. 2.10 Marking. As a minimum, the helium leak detector should be permanently marked with the following information: type of calibration gas for the mass spectrometer, helium leak rate, date the calibration leak was filled with gas, calibration temperature

8、 at Celsius, model number, serial number of item and manufacturer name and address. 2.1 1 Instructions. An instruction and operatiodmaintenance manual shall be provided with the detector. 3. REGULATORY REQUIREMENTS. 3.1 Recvcled. recovered. or environmentallv preferable materials. Recycled, recovere

9、d, or environmentally preferable materials should be used to the maximum extent possible provided that the material meets or exceeds the operational and maintenance requirements, and promotes economically advantageous life cycle costs. 4. QUALITY ASSURANCE PROVISIONS. 4. i Product conformance. The p

10、roducts provided shall meet the salient characteristics of this CID, conform to the producers own drawings, specifications, standards, and quality assurance practices, and be the same product offered for sale in the commercial market. The Government reserves the right to require proof of such confor

11、mance. 4.2 Bid samples. Bid samples are required for examination and testing to determine whether the leak detector meets the requirements in this CID, provides calibration for helium mass spectrometers, and has acceptable form, fit, function and suitability for use. Offerors shall supply one (i) sa

12、mple of the actual item proposed to be supplied under the government contract to SA-ALCRIELA, Bldg. 320,508 Shop Lane, Kelly AFB, TX 78241-6426. 5. PACKAGING. 5.1 Packa wig. preservation and packing. Packaging, preservation, and packing shall be as specified in the contract or order. 6. NOTES. 6.1 O

13、rdering data. a. b. d. e. C. Title, number and date of this CID. Calibration report (see 2.9). Instruction and operatiodmaintenance manual. (see 2.1 i). Number of bid samples required (see 4.2). Packaging requirements (see 5.1). 2 Provided by IHSNot for ResaleNo reproduction or networking permitted

14、without license from IHS-,-,-I A-A-59355 D 9018054 0003973 600 A-A-59355 6.2 Document ordering. Copies of federal and Department of Defense specifications, standards, and handbooks are available from the Defense Automated Printing Service, 700 Robbins Avenue, Bldg. 4D, Philadelphia, PA 191 11-5094.

15、Phone (215) 697-2179. 6.3 Possible Acceptable Products. Manufacturer CAGE Model Leybold Inficon 56507 Two Technology Place East Syracuse, NY 13057-0714 MILITARY INTERESTS: Custodian: Air Force - 99 Army - AL Reviewers: Air Force - O1 DLA - CC 99077072 CIVIL AGENCY COORDINATING ACTIVITY: GSA-FSS Preparing Activity: Air Force - 82 Project 4940-0749 3 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-

展开阅读全文
相关资源
猜你喜欢
  • BS PD CEN TR 17144-2017 Resistance of metallic materials to liquid biogenic and alternative fuels and their blends《金属材料对液体生物燃料和替代燃料及其混合物的耐受性》.pdf BS PD CEN TR 17144-2017 Resistance of metallic materials to liquid biogenic and alternative fuels and their blends《金属材料对液体生物燃料和替代燃料及其混合物的耐受性》.pdf
  • BS PD CEN TR 196-4-2007 Methods of testing ncement — nPart 4 Quantitative determination of nconstituents《水泥的测试方法 第4部分 组分的定量测定》.pdf BS PD CEN TR 196-4-2007 Methods of testing ncement — nPart 4 Quantitative determination of nconstituents《水泥的测试方法 第4部分 组分的定量测定》.pdf
  • BS PD CEN TR 81-12-2014 Safety rules for the construction and installation of lifts Basics and interpretations Use of EN 81-20 and EN 81-50 in specific markets《电梯制造和安装用安全规.pdf BS PD CEN TR 81-12-2014 Safety rules for the construction and installation of lifts Basics and interpretations Use of EN 81-20 and EN 81-50 in specific markets《电梯制造和安装用安全规.pdf
  • BS PD CEN TS 115-4-2015 Safety of escalators and moving walks Interpretations related to EN 115 family of standards《自动扶梯和移动步道的安全性 有关EN 115系列标准的解释》.pdf BS PD CEN TS 115-4-2015 Safety of escalators and moving walks Interpretations related to EN 115 family of standards《自动扶梯和移动步道的安全性 有关EN 115系列标准的解释》.pdf
  • BS PD CEN TS 12200-2-2017 Plastics rainwater piping systems for above ground external use Unplasticized poly(vinyl chloride) (PVC-U) Guidance for the assessment of conform.pdf BS PD CEN TS 12200-2-2017 Plastics rainwater piping systems for above ground external use Unplasticized poly(vinyl chloride) (PVC-U) Guidance for the assessment of conform.pdf
  • BS PD CEN TS 12201-7-2014 Plastics piping systems for water supply and for drainage and sewerage under pressure Polyethylene (PE) Guidance for the assessment of conformity.pdf BS PD CEN TS 12201-7-2014 Plastics piping systems for water supply and for drainage and sewerage under pressure Polyethylene (PE) Guidance for the assessment of conformity.pdf
  • BS PD CEN TS 12390-9-2016 Testing hardened concrete Freeze-thaw resistance with de-icing salts Scaling《硬化混凝土试验 融冰盐抗冻融性 缩放比例》.pdf BS PD CEN TS 12390-9-2016 Testing hardened concrete Freeze-thaw resistance with de-icing salts Scaling《硬化混凝土试验 融冰盐抗冻融性 缩放比例》.pdf
  • BS PD CEN TS 12404-2015 Durability of wood and wood-based products Assessment of the effectiveness of a masonry fungicide to prevent growth into wood of Dry Rot Serpula la.pdf BS PD CEN TS 12404-2015 Durability of wood and wood-based products Assessment of the effectiveness of a masonry fungicide to prevent growth into wood of Dry Rot Serpula la.pdf
  • BS PD CEN TS 12633-2014 Method of polishing specimens prior to the measurement of slip and skid resistance《在抗滑移和抗滑动测量之前的试样抛光方法》.pdf BS PD CEN TS 12633-2014 Method of polishing specimens prior to the measurement of slip and skid resistance《在抗滑移和抗滑动测量之前的试样抛光方法》.pdf
  • 相关搜索

    当前位置:首页 > 标准规范 > 国际标准 > 其他

    copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
    备案/许可证编号:苏ICP备17064731号-1