1、Designation: C1000 11Standard Test Method forRadiochemical Determination of Uranium Isotopes in Soil byAlpha Spectrometry1This standard is issued under the fixed designation C1000; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, th
2、e year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of alpha-emitting uranium isotopes in soil. This test method descr
3、ibesone acceptable approach to the determination of uraniumisotopes in soil.21.2 The test method is designed to analyze 10 g of soil;however, the sample size may be varied to 50 g depending onthe activity level. This test method may not be able tocompletely dissolve all forms of uranium in the soil
4、matrix.Studies have indicated that the use of hydrofluoric acid todissolve soil has resulted in lower values than results usingtotal dissolution by fusion.1.3 The lower limit of detection is dependent on count time,sample size, detector, background, and tracer yield. Thechemical yield averaged 78 %
5、in a single laboratory evaluation,and 66 % in an interlaboratory collaborative study.1.4 The values stated in SI units are to be regarded asstandard. The values given in parentheses are for informationonly.1.5 This standard does not purport to address all of thesafety concerns, if any, associated wi
6、th its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. A specific precau-tionary statement is given in Section 10.2. Referenced Documents2.1 ASTM Standards:3C859 T
7、erminology Relating to Nuclear MaterialsC998 Practice for Sampling Surface Soil for RadionuclidesC999 Practice for Soil Sample Preparation for the Determi-nation of RadionuclidesC1163 Practice for Mounting Actinides for Alpha Spec-trometry Using Neodymium FluorideC1284 Practice for Electrodeposition
8、 of the Actinides forAlpha SpectrometryD1193 Specification for Reagent WaterD3084 Practice for Alpha-Particle Spectrometry of WaterD3648 Practices for the Measurement of RadioactivityD7282 Practice for Set-up, Calibration, and Quality Controlof Instruments Used for Radioactivity Measurements3. Summa
9、ry of Test Method3.1 A soil sample with uranium-232 tracer added is heatedto destroy organic matter and dissolved with a mixture ofhydrofluoric acid and nitric acid. The uranium is coprecipitatedwith ferric hydroxide and the precipitate is dissolved withhydrochloric acid. Iron is removed by extracti
10、on with isopropylether, and plutonium, radium, and thorium are separated fromuranium by anion exchange. Uranium is electrodeposited on astainless steel disk and determined by alpha spectrometry. Asan option, the uranium may be prepared for alpha spectromet-ric measurement by using coprecipitation wi
11、th neodymiumfluoride.4. Significance and Use4.1 This test method is used to analyze soil for alpha-emitting uranium isotopes. It can be used to establish baselineuranium levels and to monitor depositions from nuclear facili-ties.5. Interferences5.1 Protactinium-231 may not be completely separated by
12、the procedure and could interfere with the determination ofuranium-233 or uranium-234 because it has the followingalpha energies in MeV: 5.06, 5.03, 5.01, 4.95 and 4.73 (seeAppendix X1). If neptunium is present in the sample in the plusfour oxidation state, it will co-elute with the uranium.5.2 Sinc
13、e uranium-232 is added as a tracer, it can not bedetermined in soil. Uranium-232 is rarely present in soilsamples. If present in significant quantities relative to the1This test method is under the jurisdiction of ASTM Committee C26 on NuclearFuel Cycle and is the direct responsibility of Subcommitt
14、ee C26.05 on Methods ofTest.Current edition approved Feb. 1, 2011. Published February 2011. Originallyapproved in 1983. Last previous edition approved in 2005 as C1000 05. DOI:10.1520/C1000-11.2Casella, V. A., Bishop, C. T., and Glosby, A. A., “Radiometric Method for theDetermination of Uranium in S
15、oil and Air,” U.S. Environmental Protection Agency,EPA-600/7-80-019, Las Vegas, NV, February 1980; and in Practices D3084 andD3648.3For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume informa
16、tion, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.activity of uranium-232 tracer added, uranium-232 will lead toan overestimation of the chemical yield and a low bia
17、s inuranium results.6. Apparatus6.1 Alpha Pulse Height Analysis System:26.1.1 A system consisting of a charged particle detectorcapable of 50 keV or less resolution on samples electrodepos-ited on a flat mirror-finished stainless steel disk is required.6.1.2 The resolution is defined as the width of
18、 an alpha peakwhen the counts on either side of the peak are equal to one-halfof the counts at the maximum of the peak (full width at halfmaximum height (FWHM).6.1.3 The counting efficiency of the system should begreater than 15 % and the background in the energy region ofeach peak should be less th
19、an 0.010 cpm.6.1.4 A regular program of measurement control operationsshould be conducted for the alpha spectrometry system such asregular background checks, daily source check to determinesystem stability, control charting, and careful handling ofsamples during changing. See Practice D7282.6.2 Beak
20、ers and Covers (TFE-fluorocarbon), 250 mL.6.3 Porcelain Crucible,60mL.6.4 Centrifuge and Bottles, 250-mL capacity.6.5 Ion Exchange Columns, 1.3 cm inside diameter by 15cm long with 100 mL reservoir.6.6 Automatic Pipettes, 0.1 to 1.0 mL with disposable tips.6.7 Furnace, able to maintain 600C.7. Reage
21、nts7.1 Purity of ReagentsReagent grade chemicals shall beused in all tests. Unless otherwise indicated, it is intended thatall reagents conform to the specifications of the Committee onAnalytical Reagents of the American Chemical Society, wheresuch specifications are available.4Other grades may be u
22、sed,provided it is first ascertained that the reagent is of sufficientlyhigh purity to permit its use without lessening the accuracy ofthe determination.7.2 Purity of WaterUnless otherwise indicated, referencesto water shall be understood to mean reagent water as definedin Specification D1193, Type
23、III.7.3 Reagent purity shall be such that the measured radioac-tivity of blank samples does not exceed the calculated probableuncertainty of the measurement.7.4 Ammonium Hydroxide (0.15 M)Mix 10 mL of con-centrated ammonium hydroxide with water and dilute to 1 L.7.5 Ammonium Hydroxide (sp gr 0.90)Co
24、ncentrated am-monium hydroxide (NH4OH). Keep tightly capped to mini-mize the uptake of carbon dioxide.7.6 Ammonium Sulfate Solution (1 M)Dissolve 132 g of(NH4)2SO4in water and dilute to 1 L.7.7 Anion Exchange Resin5Type 1 anion exchange resin,8 % cross-linked, 100 to 200 mesh, chloride form. Prepare
25、 aresin slurry by soaking the resin in 8 M HCl and transfer theslurry to an ion exchange column so that the resin column isapproximately 10 cm high.7.8 Ferric Chloride Solution (0.18 M in 0.5 M HCl)Dissolve 48 g of FeCl3.6H2Oin0.5M HCl and dilute to 1 L.7.9 Hydriodic Acid (48 %)Concentrated hydriodi
26、c acid(HI).7.10 Hydrochloric Acid (0.5 M)Mix 42 mL of concen-trated HCl with water and dilute to 1 L.7.11 Hydrochloric Acid (1 M)Mix 83 mL of concentratedHCl with water and dilute to 1 L.7.12 Hydrochloric Acid (6 M)Mix 500 mL of concen-trated HCl with water and dilute to 1 L.7.13 Hydrochloric Acid (
27、8 M)Mix 667 mL of concen-trated HCl with water and dilute to 1 L.7.14 Hydrochloric Acid (sp gr 1.19)Concentrated hydro-chloric acid (HCl).7.15 Hydrochloric Acid-Hydriodic Acid Solution (HCl-HI)Mix 1 mL of concentrated HI with 50 mL of 6 M HCl.Prepare immediately before use.7.16 Hydrofluoric Acid (48
28、 %)Concentrated hydrofluoricacid (HF).7.17 Nitric Acid (sp gr 1.42)Concentrated nitric acid(HNO3).7.18 Uranium-232, Standard Solution.67.19 Boric AcidSolid.7.20 Isopropyl Ether.8. Sampling8.1 Collect the sample in accordance with Practice C998.8.2 Prepare the sample for analysis in accordance withPr
29、actice C999.9. Calibration and Standardization9.1 If a standard uranium-232 solution is not available foruse as a tracer, standardize a freshly prepared sample ofuranium-232; for guidance refer to Practices D3648. Thisstandard may also be used to establish the counting efficiencyof the alpha spectro
30、meter which then can be used to calculatethe chemical yield for each sample.10. Precautions10.1 Adequate laboratory facilities, such as fume hoods andcontrolled ventilation, along with safe techniques, must be usedin this procedure. Extreme care should be exercised in usinghydrofluoric and other hot
31、, concentrated acids. Use of propergloves is recommended. Refer to the laboratorys chemicalhygiene plan and other applicable guidance for handlingchemical and radioactive materials and for the management ofradioactive, mixed, and hazardous waste.4Reagent Chemicals, American Chemical Society Specific
32、ations, AmericanChemical Society, Washington, DC. For suggestions on the testing of reagents notlisted by the American Chemical Society, see Analar Standards for LaboratoryChemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeiaand National Formulary, U.S. Pharmacopeial Conventio
33、n, Inc. (USPC), Rockville,MD.5Ag1-X8 from BioRad Laboratories (Hercules, CA) and A8-B500-M-Cl fromEichrom Technoloigies, Inc.(Lisle, IL) have been found to be satisifactory.6Uranium-232 is available from the National Institute of Standards andTechnology, Gaithersburg, MD 20899, as a Standard Referen
34、ce Material.C1000 11210.2 Hydrofluoric acid is a highly corrosive acid that canseverely burn skin, eyes, and mucous membranes. Hydrofluo-ric acid is similar to other acids in that the initial extent of aburn depends on the concentration, the temperature, and theduration of contact with the acid. Hyd
35、rofluoric acid differsfrom other acids because the fluoride ion readily penetrates theskin, causing destruction of deep tissue layers. Unlike otheracids that are rapidly neutralized, hydrofluoric acid reactionswith tissue may continue for days if left untreated. Due to theserious consequences of hyd
36、rofluoric acid burns, prevention ofexposure or injury of personnel is the primary goal. Utilizationof appropriate laboratory controls (hoods) and wearing ad-equate personal protective equipment to protect from skin andeye contact is essential.11. Procedure11.1 Acid Dissolution:11.1.1 Weigh a 10.0 6
37、0.1 g soil sample into a porcelaincrucible.11.1.2 Add an appropriate amount of uranium-232 tracer tothe sample. (If the activity of the sample is expected to be lessthan 0.01 Bq/g or is unknown, add 0.1 Bq of tracer. For higherlevels add uranium-232 tracer which is equivalent to theestimated activit
38、y of uranium in the sample.)11.1.3 Heat the porcelain crucible containing the soilsample in a muffle furnace at 600C for 4 h, remove, and cool.11.1.4 Transfer the sample to a 250-mL TFE-fluorocarbonbeaker and rinse the porcelain crucible several times withconcentrated HNO3until the final volume is 6
39、0 mL.11.1.5 Add 30 mL of 48 % HF to the beaker and cover witha TFE-fluorocarbon watchglass. Heat the sample for one hourand stir frequently with a TFE-fluorocarbon stirring rod duringthe heating. Cool the solution.11.1.6 Add 30 mL each of concentrated HNO3and 48 % HFand digest with some stirring for
40、 an additional hour.11.1.7 Add 20 mL of concentrated HCl and heat. Occasion-ally stir the solution.11.1.8 Remove the cover and evaporate the solution toapproximately 20 mL. Cool the solution.11.1.9 Add 50 mL of water and 5 g boric acid to the solutionand heat while stirring for 10 min to dissolve th
41、e soluble salts.11.1.10 Cool and transfer the sample solution to a 250-mLcentrifuge bottle and wash the beaker with a minimum amountof water and combine.11.1.11 Centrifuge and transfer the supernate to a 250-mLcentrifuge bottle.11.1.12 Wash the residue with 10 mL of 1 M HCl, centri-fuge, and add the
42、 washing to the supernate.11.2 Co-Precipitation:11.2.1 Add 2 mL of 0.18 M ferric chloride solution to thesupernate in the centrifuge bottle and stir. It may not benecessary to add the iron carrier if a sufficient amount of ironis present in the soil.11.2.2 Add concentrated NH4OH to the sample soluti
43、onwhile mixing to precipitate the iron and until the pH reaches 9to 10.11.2.3 Add 5 mL of concentrated NH4OH in excess.11.2.4 Centrifuge the sample for 5 min, discard the super-nate. Dissolve the precipitate with a minimum addition ofconcentrated HCl and bring the volume to 50 mL with 8 MHCl.11.2.5
44、Transfer the sample to a 250-mL separatory funneland rinse the centrifuge bottle with two 5 mL rinses of 8 MHCl.11.3 Ether Extraction:11.3.1 Add 60 mL of isopropyl ether to the funnel, stopper,and shake for 2 min.Allow the phases to separate and drain theaqueous phase into a second separatory funnel
45、.11.3.2 Add 5 mL of concentrated HCl to the second funneland mix. Add 60 mL of isopropyl ether and repeat theextraction twice more to remove iron. (Additional extractionsmay be necessary if the third extract is yellow, indicatingincomplete removal of the iron).11.3.3 Transfer the aqueous phase to a
46、150-mL beaker andboil the solution for 15 min.11.4 Anion Exchange Separation:11.4.1 Condition the anion exchange column (7.7) by rins-ing with four column volumes of 8 M HCl.11.4.2 Transfer the sample to the anion exchange columnand pass the sample through the column at a rate of 2 mL permin.11.4.3
47、Wash the column with six column volumes of 6 MHCl containing 1 mL of concentrated HI per 50 mL of 6 MHCl, prepared immediately before use, to remove iron andplutonium.11.4.4 Wash the column with two column volumes of 6 MHCl.11.4.5 Elute the uranium with six column volumes of 1 MHCl and collect in a
48、150-mL beaker.11.4.6 Evaporate the sample to 20 mL and add 5 mL ofconcentrated HNO3. Evaporate the sample to near dryness.11.5 Electrodeposition:11.5.1 Uranium can also be prepared for alpha spectrometryby coprecipitation with neodymium fluoride, see PracticeC1163 or by electrodeposition using Pract
49、ice C1284.11.5.2 Count the sample on a calibrated alpha pulse heightanalyses system within 1 week or sooner to prevent interfer-ences from uranium-232 daughters.11.6 Alpha Pulse Height Analysis:11.6.1 Count the sample on an alpha spectrometer for 1000min or longer to resolve the uranium isotopes.211.6.2 Determine the background and reagent blank activi-ties and correct the count for each peak.12. Calculation12.1 The activity of each uranium isotope is calculated asfollows:C1000 113Xi5 CiAt!/CtWs! (1)where:Xi= concentration of a uranium isotope in Bq/g,At= ac