ASTM D3087-1991(2004) Standard Test Method for Operating Performance of Anion-Exchange Materials for Strong Acid Removal《清除强酸用阴离子交换材料使用性能的标准试验方法》.pdf

上传人:Iclinic170 文档编号:514061 上传时间:2018-12-02 格式:PDF 页数:4 大小:83.67KB
下载 相关 举报
ASTM D3087-1991(2004) Standard Test Method for Operating Performance of Anion-Exchange Materials for Strong Acid Removal《清除强酸用阴离子交换材料使用性能的标准试验方法》.pdf_第1页
第1页 / 共4页
ASTM D3087-1991(2004) Standard Test Method for Operating Performance of Anion-Exchange Materials for Strong Acid Removal《清除强酸用阴离子交换材料使用性能的标准试验方法》.pdf_第2页
第2页 / 共4页
ASTM D3087-1991(2004) Standard Test Method for Operating Performance of Anion-Exchange Materials for Strong Acid Removal《清除强酸用阴离子交换材料使用性能的标准试验方法》.pdf_第3页
第3页 / 共4页
ASTM D3087-1991(2004) Standard Test Method for Operating Performance of Anion-Exchange Materials for Strong Acid Removal《清除强酸用阴离子交换材料使用性能的标准试验方法》.pdf_第4页
第4页 / 共4页
亲,该文档总共4页,全部预览完了,如果喜欢就下载吧!
资源描述

1、Designation: D 3087 91 (Reapproved 2004)Standard Test Method forOperating Performance of Anion-Exchange Materials forStrong Acid Removal1This standard is issued under the fixed designation D 3087; the number immediately following the designation indicates the year oforiginal adoption or, in the case

2、 of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of the oper-ating capacity of anion-exchange mate

3、rials when used for theremoval of hydrochloric and sulfuric acid from water. It isdesigned to simulate operating conditions for strong acidremoval and is intended for use in testing both new and usedmaterials.1.2 The values stated in SI units are to be regarded as thestandard. The inch-pound units g

4、iven in parentheses are forinformation only.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulat

5、ory limitations prior to use. Specific precau-tionary statements are given in Note 1.2. Referenced Documents2.1 ASTM Standards:2D 1067 Test Methods for Acidity or Alkalinity of WaterD 1125 Test Methods for Electrical Conductivity and Re-sistivity of WaterD 1129 Terminology Relating to WaterD 1193 Sp

6、ecification for Reagent WaterD 2687 Practices for Sampling Particulate Ion-ExchangeMaterials3. Terminology3.1 DefinitionsFor definitions of terms used in this testmethod, refer to Terminology D 1129.4. Summary of Test Method4.1 The test method consists of repeated cycles of back-wash, base regenerat

7、ion, rinse, and exhaustion of the sample inthe form of a bed in a transparent column. The exhaustionmedium used is an ion-exchange test water.5. Significance and Use5.1 This test method can be used for evaluating performanceof commercially available anion-exchange materials regardlessof the basic st

8、rength of the ion exchange groups. Whenprevious operating history is known, a good interpretation ofresin fouling or malfunction can be obtained by comparisonagainst a reference sample of unused ion-exchange materialevaluated in the same way.5.2 While resistivity has been chosen as the preferredanal

9、ytical method for defining the exhaustion end point, withtitration as the alternative, it is understood that observation ofpH during rinse and the service run can yield useful informa-tion. The variations in pH observed with an ion exchangematerial suspected of having degraded, can be helpful ininte

10、rpretation of performance when compared with similar datafor a reference sample of unused material exhausted in thesame way.6. Apparatus6.1 Test Assembly (see Fig. 1), consisting of the following:6.1.1 Column, transparent, vertically supported, 25.4 62.5-mm (1.0 6 0.1-in.) inside diameter and approx

11、imately 1.5m (60 in.) long. The bottom of the column shall be closed andprovided with an outlet of approximately 6-mm (14-in.) insidediameter. Connections shall be provided at the top and bottomfor the admission and removal of the solutions described inSection 7. Adequate means for measuring and reg

12、ulating theflow shall also be provided. The column shall be calibrated insuch a manner that the volume readings required by the methodcan be made (see Section 9). All measurements shall be madeat 25 6 5C.6.1.2 Support the sample at least 50 mm (2 in.) above thebottom of the column outlet using quart

13、z, gravel, glass beads orother material from 1.5 to 3.5 mm (116 to18 in.) in diameter,1This test method is under the jurisdiction of ASTM Committee D19 on Water,and is the direct responsibility of Subcommittee D19.08 on Membranes and IonExchange Materials.Current edition approved June 1, 2004. Publi

14、shed June 2004. Originallyapproved in 1978. Last previous edition approved in 1998 as D 3087 91 (1998).2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards

15、 Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.insoluble in the reagents used, and retained on a corrosion-resistant screen. However, other supports may be used at thediscretion of the inter

16、ested parties.7. Reagents7.1 Purity of ReagentsReagent grade chemicals shall beused in all tests. Unless otherwise indicated, it is intended thatall reagents shall conform to the specifications of the Commit-tee on Analytical Reagents of the American Chemical Society,where such specifications are av

17、ailable.3Other grades may beused, provided it is first ascertained that the reagent is ofsufficiently high puritity to permit its use without lessening theaccuracy of the determination.7.2 Purity of Water Unless otherwise indicated, refer-ences to water shall be understood to mean reagent waterconfo

18、rming to Specification D 1193, Type III.7.3 Acidity Test ReagentsFor reagents used in determin-ing acidity, refer to Test Methods D 1067.7.4 Anion Exchange Test Water C (10 meq/L)Carefullyadd 18.1 mL of sulfuric acid (H2SO4, sp gr 1.84) and 27.5 mLof hydrochloric acid (HCl, sp gr 1.19) to 500 mL of

19、water anddilute to 1 L. Prepare the test water by adding 1 volume of themixed acid solution to 99 volumes of water. Determine theacidity of the test water in accordance with Test MethodsD 1067, titrating to the methyl purple end point. The acidityshall be 10.0 6 0.5 meq/L (epm).7.5 Base Regenerants:

20、7.5.1 For Weak Base Ion-Exchange Materials:7.5.1.1 Ammonium Hydroxide (40 g NH3/L)Dilute 155mL of ammonium hydroxide (NH4OH sp gr 0.90) to 1 L withwater. The solution should be freshly prepared to avoidabsorption of carbon dioxide (CO2) from the air.7.5.2 For Weak, Intermediate, and Strong Base Ion-

21、Exchange Materials:7.5.2.1 Sodium Hydroxide Solution (40 g/L)Dissolve 40 gof sodium hydroxide (NaOH) in 800 mL of water and dilute to1 L. The solution should be freshly prepared to avoid absorp-tion of carbon dioxide (CO2) from the air.8. Sampling8.1 For sampling procedures refer to Practices D 2687

22、.9. Procedure9.1 Adjust temperature of the water and all solutions to beused in this procedure to 25 6 5C and maintain thistemperature throughout the test.9.2 Fill the column approximately half full of water and addsufficient sample to give a bed height of 750 6 75 mm (30 63 in.) above the top of th

23、e support. To avoid drying out of theion-exchange material, maintain a layer of liquid at least 20 to30 mm (0.8 to 1.2 in.) deep above the top of the bed at all timesduring the procedure.9.3 Backwash with water for 10 min using a flow rate thatwill maintain a 50 % expansion of the bed. If the supern

24、atantliquid is clear at this point, proceed to 9.4. If the supernatantliquid is cloudy (indicating the presence of light, insoluble,extraneous material), adjust the backwash outlet tube to aheight above the bed equal to 75 % of the bed height. Continuebackwashing at the same rate until the effluent

25、is clear.9.4 Allow the bed to settle and then drain at a rate ofapproximately 100 mL/min until the water level is 20 to 30 mm(0.8 to 1.2 in.) above the top of the bed. Record the volume, inmillilitres, of ion-exchange material for use in the followingpretreatment. Regenerate the sample with the appr

26、opriatedilute sodium hydroxide solution (see Note 2) for 90 min at aflow rate of 0.11 mL/min for each millilitre of ion exchanger inthe column. This corresponds to a regeneration level of 400 g/L(25 lb/ft3) of ion exchange material.9.5 When only a 20 to 30-mm (0.8 to 1.2-in.) layer of liquidremains

27、above the bed, rinse with water using the same flowrate, until a volume equal to the volume of resin has beendisplaced (one bed volume). Increase the rinse rate to approxi-mately 100 mL/min and continue the rinse until a total of tenbed volumes liquid have been used.9.6 Backwash with water for 10 mi

28、n using a flow rate thatwill maintain a 50 % expansion of the bed. Allow the bed tosettle and then drain at rate of approximately 100 mL/min untilthe water is 20 to 30 mm (0.8 to 1.2 in.) above the top of thebed (Cautionsee Note 1). Record the volume in millilitres,of ion-exchange material. Repeat t

29、he foregoing procedure untiltwo successive volume readings agree within 5 mL. Take the3“Reagent Chemicals, American Chemical Society Specifications,” Am. Chemi-cal Soc., Washington, DC. For suggestions on the testing of reagents not listed bythe American Chemical Society, see “Analar Standards for L

30、aboratory Chemicals,”BDH Ltd., Poole, Dorset, U.K., and the “United States Pharmacopeia.”FIG. 1 Typical Arrangement of Apparatus for Performance Testingof Ion-Exchange MaterialsD 3087 91 (2004)2average of these two readings as the sample volume used incalculating the capacity of the ion-exchange mat

31、erial undertest.NOTE 1Caution: The bed must be essentially free of vibration duringthis measurement; any shock will yield low and false bed volume data.9.7 Exhaust the ion exchange material with the anion-exchange test water C at a flow rate of 0.33 mL/min/mL of ionexchange material as measured in 9

32、.6. Maintain a head ofliquid not less than 50 mm (2 in.) above the top of the bedduring the exhaustion. During the run, test the effluent periodi-cally by one of the methods given in Table 1. Continue the rununtil the designated end point is reached, and record thevolume of test water used.9.8 Repea

33、t the backwash and drain in accordance with 9.6,omitting the determination of bed volume.9.9 Regenerate the sample with the appropriate NaOHsolution (see Note 2), maintaining a flow rate through the bedof 0.054 mL/min/mL of ion-exchange material based on thesample volume determined in 9.6. Continue

34、the flow for 30min, leaving a 20 to 30-mm (0.8 to 1.2-in.) layer of liquidabove the bed. This corresponds to a regeneration level of 64.1g/L (4 lb/ft3) of ion-exchange material.NOTE 2Alternative regenerant solutions, ammonia and soda ash, maybe used in place of caustic for evaluation of weak base io

35、n-exchangematerial; however, the operating capacity results based on these regener-ants may be different than the operating capacity results based on causticregeneration.9.10 Rinse the bed with anion-exchange test water C (seeNote 3) at the same rate until one bed volume of liquid hasbeen displaced,

36、 then increase the rate to 0.33 mL/min/mL ofion-exchange material as measured in 9.6. The liquid levelabove the top of the bed should be maintained at 50 mm (2 in.).During the rinse, determine the effluent resistivity at least every15 min, until the resistivity is greater than 20 000 V-cm andrecord

37、the volume of rinse water.NOTE 3Test water C is used for the rinse step to simulate the use ofcation bed effluent for anion bed rinse which is typical in many industrialdemineralizer systems. Alternative rinse solutions, such as water, may beused in the evaluation of other types of systems; however,

38、 the operatingcapacity results based on these rinse solutions may be different than thoseusing test water C as required in 9.10.9.11 Exhaust the ion-exchange material by maintaining theflow of test water until the effluent resistivity is less than20 000 V-cm. Record the total volume of test water.9.

39、12 Repeat the cycle, beginning with the backwash (see9.8) followed by the regeneration (see 9.9). Continue with therinse (see 9.10) and end with the service run (see 9.11). Repeatuntil each of three successive runs agrees within 6 5 % of theaverage capacity as calculated in accordance with Section 1

40、0.10. Calculation10.1 Calculate the operating capacity, C8, in milliequiva-lents per millilitre of ion-exchange material as follows:C8 5 B 2 A! 3 F/Swhere:A = litres of test water used in rinse (see 9.10),B = total litres of test water used (see 9.11),F = acidity of test water, meq/L, (see 7.4) andS

41、 = millilitres of ion-exchange material used (see 9.6).10.2 Calculate the operating capacity, C9, in kilograins ofcalcium carbonate per cubic foot, as follows:C9 5 C8 3 21.810.3 Calculate the rinse value, R, in litres of rinse water perlitre of bed volume, as follows:R 5 A 3 1000!/Swhere:A = litres

42、of rinse water (see 9.10), andS = millilitres of ion-exchange material used (see 9.6).10.4 Calculate the rinse value, R8, in gallons of rinse waterper cubic foot of bed volume, as follows:R8 5 R 3 7.4811. Report11.1 Report the capacity of test material as the average ofthree successive service runs

43、that agree within 6 5 % of theaverage.12. Precision and Bias412.1 PrecisionThe Referenced Documents show thatseven laboratories participated in a round robin test of theoperating performance of anion-exchange materials of fourresins using two different regenerants, sodium hydroxide, andammonium hydr

44、oxide. Each of the four resin samples wereevaluated by at least three different labs and in some cases fourdifferent labs. Each lab evaluated the assigned resin samples atleast once, with three labs doing the assigned samples induplicate, and one lab doing the assigned samples in triplicate.Separate

45、 evaluations of two resin samples were made by twopeople in one lab.12.1.1 The precision of the values obtained by this testmethod is dependent upon the magnitude of the observedvalues and can be estimated from the following relationships:12.1.1.1 For rinse values in litre of rinse per litre of resi

46、nbed (see 10.3):log SR5 0.0267R 2 0.523where:SR= the absolute value of the standard deviation, andR = rinse value in L of rinse/L of resin bed.12.1.1.2 For capacity values in kilograins per cubic foot asCaCO3(see 10.2):Sc5 0.0481 C9 2 0.16414Supporting data have been filed at ASTM International Head

47、quarters and maybe obtained by requesting Research Report RR: D191011.TABLE 1 Test Methods for Anion-Exchanger EffluentStatus ASTMDesignationRinse EndPointService EndPointPreferred D 1125 20 000 V-cm 20 000 V-cmAlternative D 1067 0.1 meq/L phenol-phthalein0.1 meq/L methylpurpleD 3087 91 (2004)3where

48、:Sc= the absolute value of the standard deviation, andC9 = operating capacity in kilograins/ft3.12.2 BiasBecause no materials with known operatingcapacity can be prepared, calculation of bias is impossible.13. Keywords13.1 anion resin; ion exchange; operating performanceASTM International takes no p

49、osition respecting the validity of any patent rights asserted in connection with any item mentionedin this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the riskof infringement of such rights, are entirely their own responsibility.This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years andif not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional s