ASTM D3687-2007 Standard Practice for Analysis of Organic Compound Vapors Collected by the Activated Charcoal Tube Adsorption Method《用活性炭管吸附法收集的有机化合物蒸气的分析用标准实施规程》.pdf

上传人:赵齐羽 文档编号:515267 上传时间:2018-12-02 格式:PDF 页数:6 大小:86.20KB
下载 相关 举报
ASTM D3687-2007 Standard Practice for Analysis of Organic Compound Vapors Collected by the Activated Charcoal Tube Adsorption Method《用活性炭管吸附法收集的有机化合物蒸气的分析用标准实施规程》.pdf_第1页
第1页 / 共6页
ASTM D3687-2007 Standard Practice for Analysis of Organic Compound Vapors Collected by the Activated Charcoal Tube Adsorption Method《用活性炭管吸附法收集的有机化合物蒸气的分析用标准实施规程》.pdf_第2页
第2页 / 共6页
ASTM D3687-2007 Standard Practice for Analysis of Organic Compound Vapors Collected by the Activated Charcoal Tube Adsorption Method《用活性炭管吸附法收集的有机化合物蒸气的分析用标准实施规程》.pdf_第3页
第3页 / 共6页
ASTM D3687-2007 Standard Practice for Analysis of Organic Compound Vapors Collected by the Activated Charcoal Tube Adsorption Method《用活性炭管吸附法收集的有机化合物蒸气的分析用标准实施规程》.pdf_第4页
第4页 / 共6页
ASTM D3687-2007 Standard Practice for Analysis of Organic Compound Vapors Collected by the Activated Charcoal Tube Adsorption Method《用活性炭管吸附法收集的有机化合物蒸气的分析用标准实施规程》.pdf_第5页
第5页 / 共6页
亲,该文档总共6页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、Designation: D 3687 07Standard Practice forAnalysis of Organic Compound Vapors Collected by theActivated Charcoal Tube Adsorption Method1This standard is issued under the fixed designation D 3687; the number immediately following the designation indicates the year oforiginal adoption or, in the case

2、 of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers the applications of methods for theextraction and gas chromatographic

3、determination of organicvapors that have been adsorbed from air in sampling tubespacked with activated charcoal.1.2 This practice is complementary to Practice D 3686.1.3 This practice is applicable for analysis of samples takenfrom workplace or other atmospheres provided that the con-taminant adsorb

4、s onto charcoal, that it can be adequatelyextracted from the charcoal, and that it can be analyzed by gaschromatography (GC). Other adsorbents and other extractiontechniques are described in Practice D 6196.1.4 Organic compounds of multicomponent samples maymutually interfere during analysis. Method

5、s to resolve inter-ferences are given in Section 6.1.5 The values stated in SI units are to be regarded as thestandard.1.6 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-pri

6、ate safety and health practices and determine the applica-bility of regulatory limitations prior to use. Specific precau-tions are given in 8.4, 9.2, and in A1.2.3.2. Referenced Documents2.1 ASTM Standards:2D 1356 Terminology Relating to Sampling and Analysis ofAtmospheresD 3686 Practice for Samplin

7、g Atmospheres to Collect Or-ganic Compound Vapors (Activated Charcoal Tube Ad-sorption Method)D 6196 Practice for Selection of Sorbents, Sampling, andThermal Desorption Analysis Procedures for Volatile Or-ganic Compounds in AirE 355 Practice for Gas Chromatography Terms and Rela-tionships2.2 NIOSH S

8、tandards:CDC-99-74-45 Documentation of NIOSH ValidationTests3NIOSH Manual of Analytical Methods, 4thEd.42.3 OSHA Standards:29 CFR 1910 Code of Federal Regulations, RegulationsRelating to Labor, Occupational Safety and Health Ad-ministration, Department of Labor5OSHA Sampling and Analytical Methods62

9、.4 UK Health and Safety Executive (HSE):7Methods for the Determination of Hazardous Substances(MDHS)2.5 Berufsgenossenschaftliches Institut fr Arbeitsschulz(BGIA):8GESTIS Analytical Methods3. Terminology3.1 Definitions:3.1.1 For definitions of terms used in this practice, refer tothe terminology spe

10、cified in D 1356 and E 355.4. Summary of Practice4.1 Organic vapors that have been collected on activatedcharcoal are extracted with carbon disulfide or another appro-priate solvent and are determined by GC using a flameionization detector (FID). Carbon disulfide is a relatively smallmolecule that c

11、an penetrate the “ink-bottle” shaped pores ofactivated charcoal, it has a high heat of adsorption on activatedcharcoal which helps in displacing other adsorbed molecules,and it is a reasonably good solvent for most, especially1This practice is under the jurisdiction of ASTM Committee D22 on Air Qual

12、ityand is the direct responsibility of Subcommittees D22.04 on Workplace Atmo-spheres.Current edition approved Oct. 1, 2007. Published November 2007. Originallyapproved in 1978. Last previous edition approved in 2001 as D 3687 - 01.2For referenced ASTM standards, visit the ASTM website, www.astm.org

13、, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from the U.S. Department of Commerce, National Technical Infor-mation Service, Port Royal Road, Springfield, VA 22161.

14、4NIOSH Manual of Analytical Methods (NMAM). http:/www.cdc.gov/niosh/nmam/ (accessed 1/2007).5Available from Superintendent of Documents, U.S. Government PrintingOffice, Washington, DC 20402.6OSHA Sampling and Analytical Methods, http:/www.osha.gov/dts/sltc/methods/index.html (accessed 1/2007).7HSE M

15、ethods for the Determination of Hazardous Substances (MDHS).http:/www.hse.gov.uk/pubns/mdhs/index.htm# (accessed 1/2007).8GESTIS Analytical Methods. http:/www.hvbg.de/e/bia/gestis/analytical_methods/index.html (accessed 1/2007).1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West

16、Conshohocken, PA 19428-2959, United States.non-polar, organic molecules. Polar modifiers (such as N,N-dimethylformamide) are frequently added to enhance the re-covery of polar organic compounds. Other advantages to usingcarbon disulfide include an early elution time on most GCcolumns and a small FID

17、 response.4.2 Interferences resulting from the analytes having similarretention times during GC analysis are resolved by changingthe GC column, by changing the operating parameters, or byfractionating the sample using solvent extraction as describedin Section A1.1.4.3 Peak purity and identity can be

18、 confirmed using tech-niques such as GC/MS.5. Significance and Use5.1 Promulgations by the Federal Occupational Safety andHealth Administration (OSHA) in 29 CFR 1910 designate thatcertain organic compounds must not be present in workplaceatmospheres at concentrations above specified values.5.2 This

19、practice, when used in conjunction with PracticeD 3686, will promote needed accuracy and precision in thedetermination of airborne concentrations of many of theorganic chemicals given in 29 CFR 1910, CDC-99-74-45,NIOSH Manual of Analytical Methods, OSHA Sampling andAnalytical Methods, HSE Methods fo

20、r the Determination ofHazardous Substances, and BGIA GESTIS Analytical Meth-ods. It can be used to determine worker exposures to thesechemicals, provided appropriate sampling periods are used.5.3 Most laboratories are equipped with apparatus similar tothat described in Section 7. Other apparatus can

21、 be used whenanalytical procedures suitable for that equipment are em-ployed. The analytical techniques (or variations thereof) de-scribed in Sections 9-11 are in general use to analyze volatileorganic compounds extracted from charcoal. Other procedurescan be used when appropriate.6. Interferences6.

22、1 Any chemical that produces an FID response and has asimilar retention time as the analyte is a potential interference.If potential interferences were reported when the samples werereceived they should be considered before the samples areextracted. Generally, gas chromatographic conditions such ast

23、he type of GC column (phase) or operating parameters can bechanged to resolve interferences.6.2 Selective solvent stripping techniques have been usedsuccessfully to make clean and fast separations of polar,nonpolar and oxygenated compounds. A general guideline isgiven in A1.1 and detailed procedures

24、 are given in Refs9(1and 2).6.3 When necessary, the identity or purity of an analyte peakcan be confirmed by GC/mass spectrometry.6.4 The presence of co-adsorbed chemicals can affect therecovery (extraction efficiency) of a particular analyte. Sus-pected effects can be tested by spiking the analytes

25、 on charcoalas in Section 11.7. Apparatus7.1 Gas chromatograph, equipped with a flame ionizationdetector (FID), a temperature-programmable oven, and anautomatic sample injector (autosampler). Sample injection maybe performed manually if necessary. Other detectors (such aselectron capture, flame phot

26、ometric, nitrogen phosphorousdetectors) can be used when appropriate but the extractionsolvent may have to be modified.7.2 Electronic data system, or other suitable means to recordand measure detector response, to prepare calibration curves,and to process sample results.7.3 GC columns, required to s

27、eparate the complex mixtureof possible organic chemicals. Examples of the most commonand useful GC columns are 60-m long, 0.32-mm i.d. fusedsilica capillary GC columns with 0.1 to 1-m thick (df) phasessuch as 100 % dimethyl polysiloxane, 95 % dimethyl-5%diphenyl polysiloxane, and polyethylene glycol

28、.7.4 Laboratory glassware, calibrated syringes, calibratedsolvent dispensers, assorted Class A pipets and volumetricflasks and glass automatic sample injector (autosampler) vialswith PTFE septum caps to contain analytical standards andsamples.8. Reagents8.1 Analytical standards, reagent grade or bet

29、ter, typically97-99+%.8.2 Carbon disulfide, reagent grade or better, typically99.9 % with low benzene content.8.3 Internal standard, reagent grade or better, typically99+%, p-cymene and 1-phenyl hexane are often used. Otherinternal standard reagents can be used providing that they notappear in air s

30、amples and that they are fully tested.8.4 Extraction solvent, usually consists of 0.25 microlitresof internal standard per milliliter of carbon disulfide. Otherextraction solvents can be used provided they are fully tested.(WarningCarbon disulfide is toxic and extremely flam-mable, as are many of th

31、e organic chemicals to be analyzed.Work with these chemicals in a properly functioning laboratoryhood.)9. Calibration9.1 In general, follow the manufacturers manual and safetyinstructions to set up the gas chromatograph. Always use highpurity gases and high quality gas purifiers.9.2 Install the sele

32、cted GC column and set the linearvelocity of the carrier gas following manufacturers instruc-tions. Set the injector split ratio at 10:1 or at some otherappropriate ratio. The most commonly used capillary GCcarrier gas is hydrogen. Set the injector, detector, and columnoven temperatures appropriate

33、for the selected GC column. It isoften useful to heat the GC column at 10-20C below theexpected maximum operating temperature of the column forabout two hours before any analysis is performed. Beforeanalyzing standards or samples, place a fresh septum into theinjection port of the chromatograph. Rep

34、lace the septum dailyor when necessary. Septum failure is a frequent cause ofinconsistent FID response and changes in chromatography.9The boldface numbers in parentheses refer to the list of references at the end ofthis standard.D3687072(WarningHydrogen gas is explosive and extremely flam-mable. It

35、is absolutely essential that the gas chromatograph beleak free.)9.3 Make sure that the electronic data system is properly setto collect analytical data.9.4 Prepare separate solutions containing 1 mL of eachanalyte per 1 mL of extraction solvent. These solutions areused to determine GC column retenti

36、on time of the analytes.9.5 Analyze these solutions and a reagent blank (withoutcharcoal) using an appropriate GC column and an appropriateoven temperature program to determine GC column retentiontimes for each analyte and for the internal standard. It may beuseful to create an in-house “column map”

37、 for each GC columnlisting retention times for each analyte determined using astandard temperature program and a standard carrier gas linearvelocity.9.6 Prepare analytical standards that bracket the expectedrange of sample results for each of the analytes by injection ofmicroliter amounts of the ana

38、lytes into the extraction solution.For example: if the requested analyte is toluene, the air volumesampled with a charcoal tube is 12 L, the density of toluene is0.866 g/mL, the purity of the analytical standard is 99 %, andthe exposure limit (target concentration) is 200 ppm (753mg/m3). Calculate t

39、he mass of toluene equivalent to the targetconcentration by multiplying the exposure limit by the charcoalsample air volume (753 mg/m33 0.012 m3= 9.04 mg persample). Prepare a standard at approximately the target con-centration by diluting 10.00 L of toluene to 1.00 mL withextraction solvent. This s

40、tandard will contain 8.57 mg/mLtoluene (10.00 L 3 0.866 mg/L 3 0.99 pure /1.00 mL).Prepare additional analytical standards at the reporting limit,0.253, 0.53, 1.53, and 2.03 the target concentration. Stan-dards for other analytes can be prepared similarly to tolueneusing their respective exposure li

41、mits, densities, purities ofanalytical standards, and sample air volume (or average airvolume for multiple samples). Different analytes can be pre-pared in the same solution if applicable. Prepare independentanalytical standards with material obtained from a separatevendor to test the purity of the

42、source material and the accuracyof the standard preparation.9.7 Analyze the standards using the same temperatureprogram used in 9.5. Compare the chromatograms to be certainthe analytes are resolved. Generally, chromatographic condi-tions can be altered to separate interferences.9.8 Use an internal s

43、tandard (ISTD) calibration method formost organic compounds. An internal standard calibrationfunction is incorporated with most electronic data systems.Calibration curves for each analyte can be constructed byplotting detector response of standards (y axis) against massper standard (x axis). FID res

44、ponse is usually linear; therefore,linear regression is generally appropriate to find the equation ofthe best-fit line for the calibration curve. Program the datasystem to calculate results in terms of micrograms per sample.This is appropriate because both standards and samples areprepared in 1.00 m

45、L of extraction solvent. Typically, results forstandards (other than for the reporting limit) calculated fromthe calibration curve will deviate from their theoreticalamounts by not more than 610 %. Usually, deviation for thereporting limit is no more than 625 %. Prepare and analyzefresh standards as

46、 necessary. Analyze a fresh set of calibrationstandards with each sample set, or with a days sequence ofsample sets.10. Sample Preparation10.1 Consider potential analytical interferences that werereported when the samples were received. Make certain thatthe extraction efficiency (also called desorpt

47、ion efficiency) forall requested analyses has been determined (as described inSection 11) before extracting the samples.10.2 Most charcoal tubes have two sections and eachsection is quantitively transferred to a separate labeled au-tosampler vial. Some charcoal tubes have three sections andeach of t

48、he three sections should be similarly transferred to aseparate labeled autosampler vial.10.3 Remove the plastic cap from end of the charcoal withthe back-up section(s) of the sampling tube.10.4 Remove the plug that holds the back-up section inplace and transfer the charcoal to an appropriately label

49、ed vialand close the vial. Similarly transfer the second back-upsection (if present) to a separate labeled vial and close the vial.(A small crochet hook is a convenient device for removing theplugs from the samplers, or a hook can be fashioned from a fine(18 to 20-gauge) steel wire or a 3-in. (76-mm) No. 20hypodermic needle.)10.5 Remove the plug and transfer the front section ofcharcoal to an appropriately labeled vial and close the vial.Check the plugs to make sure that no charcoal adheres to them.Discard the plugs and empty glass tube.10.6 Continue this process until al

展开阅读全文
相关资源
猜你喜欢
相关搜索

当前位置:首页 > 标准规范 > 国际标准 > ASTM

copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
备案/许可证编号:苏ICP备17064731号-1