1、Designation: D3838 05 (Reapproved 2017)Standard Test Method forpH of Activated Carbon1This standard is issued under the fixed designation D3838; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in
2、 parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers determination of the pH of awater extract of activated carbon.1.2 The values stated in SI units are to be regarded assta
3、ndard. No other units of measurement are included in thisstandard.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the app
4、lica-bility of regulatory limitations prior to use. For specific hazardstatements, see Section 6.2. Referenced Documents2.1 ASTM Standards:2D1193 Specification for Reagent WaterD1293 Test Methods for pH of WaterD2867 Test Methods for Moisture in Activated CarbonD6851 Test Method for Determination of
5、 Contact pH withActivated CarbonE300 Practice for Sampling Industrial Chemicals3. Summary of Test Method3.1 An activated carbon sample is boiled in reagent waterusing a reflux condenser to recycle water vapor. The particlesof carbon are filtered out, the filtrate cooled to 50 C and thepH of the filt
6、rate determined by electrometic measurement.4. Significance and Use4.1 When a fluid containing an adsorbate is passed througha bed of activated carbon, chemical reactions may take placebetween the activated carbon, its other noncarbonaceousconstituents, and the adsorbate-containing fluid. The pH of
7、thecarbon may be a significant parameter of such a reaction andtherefore may be an important characteristic of the carbon.5. Apparatus and Materials5.1 Analytical Balance, capacity 100 g, precision 6 0.01 g.5.2 Hot Plate.5.3 Glassware for Boiler-Reflux Condenser Apparatus (Fig.1)Items shown are for
8、guidance only, providing a convenientset of equipment available off-the-shelf from many laboratorysupply houses. The “all-glass” elements, with standard-taperand ball joints, provide freedom from contamination andmaintenance. A check valve in the position shown is essentialto relieve pressure buildu
9、p while minimizing loss of vapor.5.4 Thermometer, glass, approximately 0 to 120 C, longenough to be read at 100 C when inserted to bottom ofErlenmeyer flask in Fig. 1.5.5 Thermometer, glass, approximately 20 to 55 C.5.6 Graduated Cylinder, 100-mL.5.7 Beaker, 200-mL.5.8 Filter Funnel.5.9 Filter Paper
10、, qualitative, medium flow rate, 12.5 cm orlarger in diameter.5.10 Timer.5.11 pH Meter, in accordance with Test Methods D1293,Type II (automatic thermal compensation preferred).5.12 Reagent Water, in accordance with SpecificationD1193, Type II.6. Hazards6.1 The test method involves transfer of boili
11、ng waterbetween containers; appropriate tongs or gloves should beused. In addition, the use of an electric hot plate and pH meter(if line-powered) poses a shock hazard. This equipment mustbe grounded and insulated in accordance with UL standardsand electrical codes.7. Sampling7.1 Guidance in samplin
12、g activated carbon is given inPractice E300.1This test method is under the jurisdiction of ASTM Committee D28 onActivated Carbon and is the direct responsibility of Subcommittee D28.02 on LiquidPhase Evaluation.Current edition approved March 1, 2017. Published March 2017. Originallyapproved in 1980.
13、 Last previous edition approved in 2011 as D3838 05 (2011).DOI: 10.1520/D3838-05R17.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary p
14、age onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United StatesThis international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for
15、theDevelopment of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.18. Calibration and Standardization8.1 Use the procedure of Test Methods D1293 to standard-ize the pH meter.9. Procedure9.1 Determine the moisture
16、content of the carbon in accor-dance with Test Methods D2867. Calculate the weight of moistcarbon equivalent to 10 g on a dry basis.9.2 Weigh out a sample of carbon equivalent to 10.00 60.01 g on a dry basis. Remove boiler flask from apparatus andadd carbon sample.9.3 Bring approximately 110 mL of r
17、eagent water to agentle boil. Measure 100.0 6 0.1 mL in graduated cylinderwhile water is hot. Immediately add this water to the carbon inthe flask. Reconnect joints to flask seated on hot plate.9.4 Bring water in flask to a boil, using thermometer readingto assure that no false boiling, due to gases
18、 trapped in carbon,occurs.9.5 Boil gently for 900 6 10 s.9.6 Remove the flask from the hot plate and filter itscontents immediately through the filter paper premoistenedwith the distilled water used for the test. Catch the filtrate in a500-mL vacuum filter flask, being careful to prevent carbon-fine
19、s from overrunning into the flask.9.7 Cool the filtrate to 50 6 5 C.9.8 Measure pH as described in Test Methods D1293.10. Calculation10.1 If the pH meter is calibrated in pH units, read valuefrom pH meter. If the meter reading is in voltage units, use theequation from the Terminology section of Test
20、 MethodsD1293.11. Report11.1 Report the following information:11.1.1 Source of sample,11.1.2 Type or designation of activated carbon,11.1.3 Supplier name,11.1.4 Supplier grade designation,11.1.5 Supplier lot and batch number,11.1.6 Moisture content in accordance with Test MethodsD2867,11.1.7 pH to t
21、he nearest 0.1 unit,11.1.8 Temperature of measurement to the nearest 1 C,11.1.9 Date of test,11.1.10 Name and signature of technician performing test,and11.1.11 Name and signature of supervisor approving test.12. Precision and Bias12.1 An interlaboratory comparison of the pH of four typesof activate
22、d carbons using essentially the above procedure wascarried out by four laboratories (see Test Method D6851). Theresults were as follows:Sample Number of Tests Average pH Standard Deviation2N 11 3.65 0.1811N 12 3.64 0.1912N 11 3.56 0.141D 11 5.57 0.674D 12 5.62 0.609D 12 5.65 0.603B 11 8.82 0.385B 12
23、 8.93 0.3010B 12 8.90 0.296P 11 7.91 0.177P 12 8.07 0.418P 12 7.85 0.2312.2 The measured pH values are in agreement with theactivated carbon types (N, wood-base; D, lignite; B and P,bone-char). The standard deviations also appear in agreementwith other work, in that they are higher for neutral condi
24、tions(near pH 7) than for strongly acidic or basic conditions. Someof the observed variation is in the samples themselves. Astandard deviation of 0.6 for pH values near 7 and 0.3 forvalues at the ends of the pH scale appears readily achievable bythis test method.FIG. 1 Boiler-Reflux Condenser Appara
25、tusD3838 05 (2017)2ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentionedin this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the riskof infringement
26、of such rights, are entirely their own responsibility.This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years andif not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for
27、 additional standardsand should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of theresponsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you shouldmake your views known
28、to the ASTM Committee on Standards, at the address shown below.This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM a
29、t the aboveaddress or at 610-832-9585 (phone), 610-832-9555 (fax), or serviceastm.org (e-mail); or through the ASTM website(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http:/ 05 (2017)3