ASTM D4230-2002e1 Standard Test Method of Measuring Humidity with Cooled-Surface Condensation (Dew-Point) Hygrometer《用冷却表面冷凝(露点)湿度计测量湿度的标准试验方法》.pdf

上传人:cleanass300 文档编号:516595 上传时间:2018-12-02 格式:PDF 页数:6 大小:99.70KB
下载 相关 举报
ASTM D4230-2002e1 Standard Test Method of Measuring Humidity with Cooled-Surface Condensation (Dew-Point) Hygrometer《用冷却表面冷凝(露点)湿度计测量湿度的标准试验方法》.pdf_第1页
第1页 / 共6页
ASTM D4230-2002e1 Standard Test Method of Measuring Humidity with Cooled-Surface Condensation (Dew-Point) Hygrometer《用冷却表面冷凝(露点)湿度计测量湿度的标准试验方法》.pdf_第2页
第2页 / 共6页
ASTM D4230-2002e1 Standard Test Method of Measuring Humidity with Cooled-Surface Condensation (Dew-Point) Hygrometer《用冷却表面冷凝(露点)湿度计测量湿度的标准试验方法》.pdf_第3页
第3页 / 共6页
ASTM D4230-2002e1 Standard Test Method of Measuring Humidity with Cooled-Surface Condensation (Dew-Point) Hygrometer《用冷却表面冷凝(露点)湿度计测量湿度的标准试验方法》.pdf_第4页
第4页 / 共6页
ASTM D4230-2002e1 Standard Test Method of Measuring Humidity with Cooled-Surface Condensation (Dew-Point) Hygrometer《用冷却表面冷凝(露点)湿度计测量湿度的标准试验方法》.pdf_第5页
第5页 / 共6页
亲,该文档总共6页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、Designation: D 4230 02e1Standard Test Method ofMeasuring Humidity with Cooled-Surface Condensation(Dew-Point) Hygrometer1This standard is issued under the fixed designation D 4230; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, th

2、e year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.e1NOTEEditorial corrections were made throughout the test method in July 2003.1. Scope1.1 This test method covers the d

3、etermination of the ther-modynamic dew- or frost-point temperature of ambient air bythe condensation of water vapor on a cooled surface. Forbrevity this is referred to in this method as the condensationtemperature.1.2 This test method is applicable for the range of conden-sation temperatures from 60

4、C to 70C.1.3 This test method includes a general description of theinstrumentation and operational procedures, including siteselection, to be used for obtaining the measurements and adescription of the procedures to be used for calculating theresults.1.4 This test method is applicable for the contin

5、uous mea-surement of ambient humidity in the natural atmosphere on astationary platform.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices

6、 and determine the applica-bility of regulatory limitations prior to use. For specificprecautionary statements, see Section 8.2. Referenced Documents2.1 ASTM Standards:2D 1356 Terminology Relating to Sampling and Analysis ofAtmospheresD 3631 Test Methods for Measuring Surface AtmosphericPressure3. T

7、erminology3.1 Definitions:3.1.1 For definitions of other terms in this method, refer toTerminology D 1356.3.2 Definitions of Terms Specific to This Standard:3.2.1 nonhygroscopic materialmaterial that neither ab-sorbs nor retains water vapor.3.2.2 mirror (front surface)a polished surface, usually ame

8、tallic surface, on which condensates are deposited.3.3 Symbols:e = vapor pressure of water vapor in moist air.ei= saturation pressure of water vapor in equilibriumwith the plane surface of ice.ew= saturation pressure of water vapor in equilibriumwith the plane surface of water.P = ambient pressure.r

9、 = mixing ratio.T = ambient air temperature.Td= thermodynamic dew- or frost-point temperature.RHi= relative humidity with respect to ice.RHw= relative humidity with respect to water.4. Summary of Test Method4.1 The ambient humidity is measured with a dew- andfrost-point hygrometer.4.2 The mirror or

10、some other surface on which the conden-sate is deposited is provided with the means for cooling andheating, detection of condensate, and the measurement of thetemperature of the mirror surface.4.3 Calculations of saturation vapor pressure over water andice as functions of temperature are provided.5.

11、 Significance and Use5.1 Humidity information is important for the understandingof atmospheric phenomena and industrial processes. Measure-ments of the dew-point and calculations of related vaporpressures are important to quantify the humidity information.1This test method is under the jurisdiction

12、of ASTM Committee D22 on AirQuality and is the direct responsibility of Subcommittee D22.11 on Meteorology.Current edition approved October 10, 2002. Published December 2002. Origi-nally published as D 4230 83. Last previous edition D 4230 83(96)e1.2For referenced ASTM standards, visit the ASTM webs

13、ite, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United Stat

14、es.6. Interferences6.1 This method is not applicable if other constituents in theatmosphere condense before water vapor.7. Apparatus7.1 Dew-point hygrometers, specifically designed for me-teorological observations are available commercially. A sche-matic arrangement of a typical optical dew-point hy

15、grometer isshown in Fig. 1.7.1.1 The sample air flows through a small chamber.7.1.2 Within the chamber is a mirror or surface on which thecondensate can be deposited.7.1.3 A beam of light from an incandescent lamp, lightemitting diode or other suitable light source shines on themirror.7.1.4 Dew or f

16、rost is detected with an electro-optic device.7.1.5 The mirror is cooled by a Peltier thermoelectricelement. Peltier cooling is a convenient method for unattendedand automatic instruments.7.1.6 Preferred devices of sensing mirror temperature areresistance thermometers, thermistors, and thermocouples

17、.7.1.6.1 The temperature sensors shall be attached to orembedded in the mirror to measure the temperature of thesurface of the mirror.7.1.7 Suitable control circuitry shall be provided to maintaina constant quantity of condensate on the mirror.7.1.8 Suitable provisions shall be provided to compensat

18、efor the contamination of the surface of the mirror.7.2 Auxiliary Equipment:7.2.1 Provision shall be provided for assuring air flow pastthe dewpoint mirror without changing the pressure in themirror chamber more than 0.5 % from the ambient pressuresurrounding the sensor. An air flow of approximately

19、 1.1 litresper minute is recommended for typical chambers.7.2.2 Readout instrumentation is available with the dew-point hygrometer.8. Precautions8.1 Safety Precautions:8.1.1 The hygrometer shall be packaged in a suitableenclosure for application in industrial or outdoor environment.8.1.2 Electrical

20、connectors and cables shall be suitable forthe industrial or outdoor environment.8.1.3 Appropriate voltage surge protection circuitry must beincorporated.8.2 Technical Precautions:8.2.1 The accuracy of a cooled-surface condensation hy-grometer is degraded by the presence of water-soluble materi-als.

21、 A mirror-cleaning schedule, consistent with the contami-nation rate, is necessary to maintain the initial calibrationaccuracy. The user must determine the required maintenanceschedule for the specific site, by comparison of calibrationsmade before and after cleaning.8.2.2 Caution in performing this

22、 method should be taken ifthe indicated mirror temperature is between 0C and 30C.Below freezing, the initial formation of the condensate on thesurface of a mirror may be either dew or frost. In the case ofnonfiltered atmospheric air, the supercooled water usually doesnot persist long on a mirror sur

23、face and quickly changes tofrost. The only positive method for determining the state of thecondensate is by visual observation of the mirror surface.8.2.2.1 The following illustrates the magnitude of the errorinvolved when dew or frost is not differentiated: The saturationvapor pressure of supercool

24、ed water at 30C corresponds tosaturation vapor pressure of ice at 27.2C; dew pointof 20C corresponds to frost point of 18.0C; 10C dewpoint corresponds to frost point of 8.9C. (The frost pointtemperature is approximately 90 % of the dew-point tempera-ture in degrees Celsius.)8.2.3 A positive method f

25、or identifying the state of thecondensate is to visually observe the condensate on the mirrorwith the aid of a microscope or other optical magnifier.8.2.4 Afinite length of time is required for the condensate todeposit on the mirror and for the hygrometer to reach equilib-rium with the ambient humid

26、ity. The response of the hygrom-eter depends on the humidity of the ambient air, and on suchfactors as the ventilation rate of the ambient air past the mirror,the sensitivity of the condensate detector, and the maximumcooling rate of the hygrometer. The worst case occurs duringthe initial dew-point

27、reading after clearing the mirror of allcondensates. The time it takes the hygrometer to reach equi-librium after clearing the mirror will vary from instrument toinstrument.As an illustration of the magnitude of this time, thefollowing are approximate times required by a hygrometer toreach equilibri

28、um after clearing the mirror.8.2.4.1 For dew points warmer than 0C: 5 min afterclearing.8.2.4.2 For dew points 0C to 20C: 5 to 20 min afterclearing.8.2.4.3 For dew points 20C to 40C: 20 min to 1 h afterclearing.8.2.4.4 For dew points 40C to 60C:1hto2hafterclearing.8.2.4.5 For dew points 60C to 70C:2

29、hto6hafterclearing.8.2.5 The pressure differential between the mirror chamberand the ambient shall not be greater than 0.5 % of the ambientpressure. For example, the difference shall not exceed 5 hPa atan ambient pressure of 1000 hPa.FIG. 1 Schematic of a Thermoelectric Cooled CondensationHygrometer

30、D423002e128.2.6 The thermometer must measure the temperature of themirror surface and not be influenced by the ambient airtemperature.8.2.7 All materials, which come into contact with thesample air before it reaches the dew-point mirror, shall benonhygroscopic. Metal, glass, polytetrafluoroethylene,

31、 or sta-bilized polypropylene are examples of suitable materials.Polyvinyl chloride tubing must be avoided.9. Sampling9.1 Automatic dew-point hygrometers provide an outputwhich may be recorded continuously. Modern data loggerssample temperature-sensor output periodically, convert theanalog sensor si

32、gnal to a digital form, and store the data. Theproper sampling interval depends on the data application (see13.2).9.2 Locate a blower or pump, which can be used to movethe air sample through the mirror chamber, downstream of thedew-point mirror. The airflow rate also depends on the dataapplication a

33、nd sampling environment.9.3 Select the site or location so that the measurement datarepresents the water vapor content of the ambient atmosphereor industrial environment being sampled. Local water vaporsources, including ponds, wet roads, and structures can influ-ence the ambient humidity.Avoid sour

34、ces of airborne contami-nants that can influence to condensation process on the mirror.9.4 The successful application of this method requires thatall the materials which come in contact with the sample airupstream of the dew-point mirror be nonhygroscopic.9.5 The materials which come in contact with

35、 the sample airupstream of the dew-point mirror might be wetted by rain, dew,or frost; for example, dew forming on a surface in the earlymorning. Design the sampling system to minimize thesedeleterious effects.10. Calibration10.1 Provide the calibration data for the thermometer, usedfor measuring th

36、e condensation temperature with the hygrom-eter. Consult the manufacturers operating manual for calibrat-ing the thermometer readout instrumentation.10.2 The cooled-surface condensation (dew-point) methodis considered to be an absolute or fundamental method formeasuring humidity. This method require

37、s an accurate mea-surement of the temperature of the surface of the dew-pointmirror. It is not uncommon for the dew-point temperature to bemore than 35 C colder than the ambient air temperature. Tomeasure this temperature accurately, without being influencedby the warmer ambient and the colder heat-

38、sink temperature,requires careful placement of the dew-point thermometer.10.3 Therefore, in addition to the temperature calibration ofthe thermometer, (see 10.1), a humidity calibration must alsobe performed to verify the proper operation of the hygrometer(see Annex A1). The following are additional

39、 examples offactors that can affect the accuracy of the measurement:extraneous thermally-induced voltage (emf), heat leakagethrough the thermometer leads, self-heating of the thermom-eter, poor thermal contact, temperature gradient across themirror, etc.11. Procedure11.1 Selection of Sampling SiteSe

40、lect sampling site asindicated in 9.3 and also in 1.3.2 of the World MeteorologicalOrganization, Guide to Meteorological Instrument and Observ-ing Practices (1).311.2 Consult the manufacturers operating manual forstart-up procedures.11.3 Perform necessary calibration as indicated in Section10. The d

41、ew-point thermometer will not undergo large shifts(0.05C) in calibration unless it is subjected to physical shock.If the thermometer read-out instrumentation is subjected tovarying ambient temperatures, the read-out instrumentationchecks must be over the expected range of ambient tempera-tures. The

42、frequency with which these checks are required willbe determined by the stability of the readout instrumentation.11.4 Check and verify that all necessary variables aremeasured and recorded to compute the humidity in the desiredunit(s) see also 12.NOTE 1In general, it is recommended that ambient temp

43、erature andpressure (the pressure in the mirror chamber should not differ from theambient pressure by more than 0.5 %) and the dew-point temperature bemeasured and recorded. The ambient pressure is to be measured accordingto Test Methods D 3631. This will enable other users of the data tocalculate i

44、n the different units of humidity.12. Calculations12.1 In the meteorological range of pressure and tempera-ture, the saturation vapor pressure of the pure water phase andof the moist air will be assumed to be equal. This assumptionwill introduce an error of approximately 0.5 % of reading orless.12.2

45、 Calculate the ambient relative humidity with respect towater using the following approximation.RHw!T5eTd!ewT!100 % (1)where:(RHw)T= relative humidity with respect to water, %, attemperature T (C),e(Td) = saturation vapor pressure, Pa, at condensationtemperature Td, C, where Tdis the averagevalue du

46、ring the sampling period, see Note 2,andew(T) = saturation vapor pressure, Pa, over water atambient temperature T, C, where T is theaverage value during the sampling period.NOTE 2If the condensate on the mirror is water (dew), use thesaturation vapor pressure over water corresponding to the condensa

47、tiontemperature Td. If the condensate is ice (frost), use the saturation vaporpressure over ice corresponding to the condensation temperature Td.Equations for saturation vapor pressure are provided in 12.5.12.3 Calculate the relative humidity with respect to ice asfollows:RHi!T5eTd!eiT!100 % (2)3The

48、 boldface numbers in parentheses refer to the references at the end of thismethod.D423002e13where:(RHi)T= relative humidity with respect to ice, %, at tem-perature T (C),ei(T) = saturation vapor pressure, Pa, over ice at ambienttemperature T, C, where T is the average valueduring the sampling period

49、, ande(Td) = see 12.2.12.4 Saturation vapor pressure calculations are provided inthis section. The equations provided in 12.4.1-12.4.4 are recentformulations (2) of the Magnus form of exponential expres-sions for saturation vapor pressure. Using similar constantsfrom other reliable resources can yield reasonably accurateresults as well. Using the appropriate equation depending if thecondensate is water or ice will improve the accuracy of thecalculation. The calculations for moist air rather than the purewater vap

展开阅读全文
相关资源
猜你喜欢
相关搜索

当前位置:首页 > 标准规范 > 国际标准 > ASTM

copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
备案/许可证编号:苏ICP备17064731号-1