ASTM D4794-1994(2009) Standard Test Method for Determination of Ethoxyl or Hydroxyethoxyl Substitution in Cellulose Ether Products by Gas Chromatography《用气相色谱法测定纤维素醚产品中乙氧基或羟基乙氧基代用品.pdf

上传人:王申宇 文档编号:517999 上传时间:2018-12-03 格式:PDF 页数:3 大小:66.23KB
下载 相关 举报
ASTM D4794-1994(2009) Standard Test Method for Determination of Ethoxyl or Hydroxyethoxyl Substitution in Cellulose Ether Products by Gas Chromatography《用气相色谱法测定纤维素醚产品中乙氧基或羟基乙氧基代用品.pdf_第1页
第1页 / 共3页
ASTM D4794-1994(2009) Standard Test Method for Determination of Ethoxyl or Hydroxyethoxyl Substitution in Cellulose Ether Products by Gas Chromatography《用气相色谱法测定纤维素醚产品中乙氧基或羟基乙氧基代用品.pdf_第2页
第2页 / 共3页
ASTM D4794-1994(2009) Standard Test Method for Determination of Ethoxyl or Hydroxyethoxyl Substitution in Cellulose Ether Products by Gas Chromatography《用气相色谱法测定纤维素醚产品中乙氧基或羟基乙氧基代用品.pdf_第3页
第3页 / 共3页
亲,该文档总共3页,全部预览完了,如果喜欢就下载吧!
资源描述

1、Designation: D 4794 94 (Reapproved 2009)Standard Test Method forDetermination of Ethoxyl or Hydroxyethoxyl Substitution inCellulose Ether Products by Gas Chromatography1This standard is issued under the fixed designation D 4794; the number immediately following the designation indicates the year ofo

2、riginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method is applicable to the determination ofethoxy

3、l or hydroxyethoxyl substitution in cellulose ether prod-ucts by a Zeisel gas chromatographic technique.1.2 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.3 This standard does not purport to address all of thesafety concerns,

4、 if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. For specific hazardstatements, see Section 5 and Note 3.2. Summary of Test Method2.1 W

5、hen a cellulose ether containing ethyl or hydroxyethylsubstitutes is allowed to react with hydriodic acid, one mole ofiodoethane is liberated for each mole of ethoxyl or hydroxy-ethoxyl ether substituted on the cellulose chain. The iodoet-hane is extracted in situ with o-xylene and quantitated by ga

6、schromatography using an internal standard technique. It isrecommended to run duplicate samples.3. Apparatus3.1 Gas Chromatograph with thermal conductivity detectorand heated injection port.3.2 Electronic Integrator.3.3 Columnstainless steel, 1829 mm in length, 3.2 mm inoutside diameter, packed with

7、 the reagent in 4.6, coiled to fitthe chromatograph used; or equivalent column and packing asappropriate.3.4 Syringe,10L.3.5 Reaction Vials, Caps, and Heating Block.23.6 Syringe,100L.3.7 Cover, stainless steel, fabricated to cover the heatingblock.4. Reagents and Materials4.1 Purity of ReagentsReage

8、nt grade chemicals shall beused in all tests. Unless otherwise indicated, it is intended thatall reagents shall conform to the specifications of the Commit-tee on Analytical Reagents of the American Chemical Society,where such specifications are available.3Other grades may beused, provided it is fir

9、st ascertained that the reagent is ofsufficiently high purity to permit its use without lessening theaccuracy of the determination.4.2 Iodoethane, 99 % minimum.4.3 O-xylene.4.4 Toluene.4.5 Hydriodic Acid, 57 % (sp gr 1.69 to 1.70).4.6 Column Packing, 10 % methyl silicone stationary phaseUSP code G1

10、coated on inert diatomite solid support USPcode S1A, 100/120 mesh.5. Hazards5.1 Safety precautions must be taken for handling of hydri-odic acid.5.2 During the reaction, the glass vials are under pressure.Exercise precaution in handling the hot vials.6. Apparatus Preparation and Conditioning6.1 Colu

11、mn (Note 1)Columns are packed with reagentunder vacuum and mechanical vibration using silanized glasswool to contain the packing. Install each in the gas chromato-graph to facilitate on column injection. Condition the columnat 200C for 12 h, then allow it to come to equilibrium underthe conditions d

12、escribed below:Oven temperature, C 130 isothermalInjection port temperature, C 200Detector temperature, C 2501This test method is under the jurisdiction of ASTM Committee D01 on Paintand Related Coatings, Materials, and Applications and is the direct responsibility ofSubcommittee D01.36 on Cellulose

13、 and Cellulose Derivatives.Current edition approved Feb. 1, 2009. Published February 2009. Originallyapproved in 1988. Last previous edition approved in 2003 as D 4794 94 (2003).2The sole source of supply of a heating block, Silli-Therm Heating Module,Reacti-Block 21, Reacti-Vials, and Minnert valve

14、 tops, known to the committee atthis time is the Pierce Chemical Company P.O. Box 117, Rockford, IL 611059976.If you are aware of alternative suppliers, please provide this information to ASTMInternational Headquarters. Your comments will receive careful consideration at ameeting of the responsible

15、technical committee,1which you may attend.3Reagent Chemicals, American Chemical Society Specifications, AmericanChemical Society, Washington, DC. For suggestions on the testing of reagents notlisted by the American Chemical Society, see Analar Standards for LaboratoryChemicals, BDH Ltd., Poole, Dors

16、et, U.K., and the United States Pharmacopeiaand National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville,MD.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.Detector current, ma 175 maAttenuation 1Polarity ACarrier ga

17、s heliumColumn A, mL/min 30Column B, mL/min 30NOTE 1The conditions used here were determined to be optimum forthe column. Optimum conditions should be determined for each columnon an individual basis.6.2 Integrator:6.2.1 Settings (Note 2):Attenuation 3Chart speed 1.0Peak width 0.04Threshold 4NOTE 2T

18、hese settings were used with a Hewlett-Packard 3390AIntegrator. Other units may require different settings.6.2.2 Approximate Component Retention Times:Relative Retention Time Component0.38 iodoethane0.59 toluene (internal standard)1.00 o-xylene7. Preparation of Standard Solutions7.1 Internal Standar

19、d Solution (25 mg toluene/mLo-xylene):7.1.1 Weigh 25.00 6 0.01 g of toluene into a 1000-mLvolumetric flask.7.1.2 Dilute with o-xylene to 1000 mL.7.2 Calibration Standard Solution:7.2.1 Pipet 2.0 mL of 57 % hydriodic acid into a 5-mL vial.7.2.2 Pipet 2.0 mL of the internal standard solution (pre-pare

20、d in 7.1) into the vial and seal with the mininert valve.7.2.3 Weigh the vial and contents to the nearest 0.1 mg.7.2.4 Add 50 L of iodoethane to the vial with a syringe.7.2.5 Weigh and record the amount of iodoethane added.7.2.6 Shake for 30 s and allow to stand for 20 minutes.8. Calibration of Elec

21、tronic Integrator8.1 Inject 1.0 L of the upper solvent layer of the standardsolution prepared in 7.2 into the gas chromatograph and recordthe chromatogram.8.1.1 Calibrate the integrator according to the manufactur-ers instructions.8.2 If an integrator is not available peak areas and responsefactors

22、can be determined manually as follows:RF 5A 3 w 3 P13 F0.05 3 B 3 P2(1)where:RF = response factor,A = peak height of internal standard (toluene),w = weight of iodoethane, g,P1= purity of iodoethane,F =factor of 0.289 for ethoxylS45156D, 0.391 forhydroxyethoxyl,0.05 = weight of toluene, g,B = peak he

23、ight of iodoethane, andP2= purity of toluene.9. Procedure9.1 Specimen Preparation:9.1.1 Dry the specimen at 105C for 1 h and store in adessicator until cool.9.1.2 Weigh 60 to 80 mg of the specimen prepared in dry9.1.1 into a clean, tared 5-mL reaction vial (see 3.5).9.1.3 For a cellulose ether that

24、is not soluble in o-xylene,add 2.0 mL of the internal standard solution prepared in 7.1 tothe vial.9.1.4 Add 2.0 mL of hydriodic acid to the vial.9.1.5 For a cellulose ether that is soluble in o-xylene,reverse the order of addition of 9.1.3 and 9.1.4.9.1.6 Cap tightly and weigh the vial.9.1.7 Shake

25、the specimen for 30 s.9.1.8 Place the vial in a heated block at 180 6 5C for 2 h.NOTE 3Warning: Vials contain a hot corrosive acid under pressure.9.1.9 Remove the vial and cool in the hood to roomtemperature. The specimen will separate into two layers.9.1.10 Reweigh to determine any loss due to leak

26、age.Discard any specimen with loss greater than 25 mg.9.1.11 Shake the sample vigorously and allow to stand for20 minutes.9.2 Analysis:9.2.1 Enter the weight of the toluene (internal standard) andthe weight of the specimen into the integrator if an integratoris being used.9.2.2 Inject 1.0 L of the u

27、pper layer of the specimen intothe gas chromatograph.10. Calculation10.1 If an integrator is not available, calculate the peak areasand concentrations using the following equation:ethoxyl, % or hydroxyethoxyl, % 5B 3 0.05 3 P23 RFA 3 sample weight g!(2)where:B = peak height of iodoethane,0.05 = weig

28、ht of toluene, g,P2= purity of toluene,RF = response factor determined for calibration standardsolution, andA = peak height of toluene.11. Precision and Bias411.1 PrecisionThe relative precision was found to be0.64 % at the 95 % confidence level.11.2 BiasNo justifiable statement can be made on the b

29、iasof the procedure in this test method because no suitablereference material for determining the bias exists.12. Keywords12.1 cellulose ether; ethoxyl or hydroxyethoxyl substitu-tion; gas chromatography4Supporting data is available from ASTM International Headquarters. RequestRR: D01-1058.D 4794 94

30、 (2009)2ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentionedin this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the riskof infringement of such rig

31、hts, are entirely their own responsibility.This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years andif not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional

32、 standardsand should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of theresponsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you shouldmake your views known to the ASTM

33、 Committee on Standards, at the address shown below.This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the aboveaddress or at 610-832-9585 (phone), 610-832-9555 (fax), or serviceastm.org (e-mail); or through the ASTM website(www.astm.org).D 4794 94 (2009)3