ASTM D5303-1992(2012) Standard Test Method for Trace Carbonyl Sulfide in Propylene by Gas Chromatography《气相色谱法对丙烯中痕量羰基硫化物的标准试验方法》.pdf

上传人:吴艺期 文档编号:519424 上传时间:2018-12-03 格式:PDF 页数:6 大小:99.94KB
下载 相关 举报
ASTM D5303-1992(2012) Standard Test Method for Trace Carbonyl Sulfide in Propylene by Gas Chromatography《气相色谱法对丙烯中痕量羰基硫化物的标准试验方法》.pdf_第1页
第1页 / 共6页
ASTM D5303-1992(2012) Standard Test Method for Trace Carbonyl Sulfide in Propylene by Gas Chromatography《气相色谱法对丙烯中痕量羰基硫化物的标准试验方法》.pdf_第2页
第2页 / 共6页
ASTM D5303-1992(2012) Standard Test Method for Trace Carbonyl Sulfide in Propylene by Gas Chromatography《气相色谱法对丙烯中痕量羰基硫化物的标准试验方法》.pdf_第3页
第3页 / 共6页
ASTM D5303-1992(2012) Standard Test Method for Trace Carbonyl Sulfide in Propylene by Gas Chromatography《气相色谱法对丙烯中痕量羰基硫化物的标准试验方法》.pdf_第4页
第4页 / 共6页
ASTM D5303-1992(2012) Standard Test Method for Trace Carbonyl Sulfide in Propylene by Gas Chromatography《气相色谱法对丙烯中痕量羰基硫化物的标准试验方法》.pdf_第5页
第5页 / 共6页
亲,该文档总共6页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、Designation: D5303 92 (Reapproved 2012)Standard Test Method forTrace Carbonyl Sulfide in Propylene by GasChromatography1This standard is issued under the fixed designation D5303; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the

2、year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of traces ofcarbonyl sulfide (COS) in propylene. It is applicable to

3、 COSconcentrations from 0.5 to 4.0 mg/kg (parts per million bymass). See Note 1.NOTE 1The lower limit of this test method is believed to be below 0.1mg/kg, depending on sample size and sensitivity of the instrumentationbeing used. However, the cooperative testing program was conducted inthe 0.5 to 4

4、.0 range due to limitations in preparing commercial testmixtures.1.2 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is

5、 theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. Specific hazardsstatements are given in Section 8.2. Referenced Documents2.1 ASTM Standards:2D3609 Practice for Calibration T

6、echniques Using Perme-ation TubesD4468 Test Method for Total Sulfur in Gaseous Fuels byHydrogenolysis and Rateometric ColorimetryE840 Practice for Using Flame Photometric Detectors in GasChromatography3. Summary of Test Method3.1 A procedure is given for removing a sample from thesample cylinder, se

7、parating COS from propylene, detectingCOS, calibrating the detector, quantitating COS content in thesample, and assaying the gas standard. General comments andrecommended techniques are given.3.2 Arelatively large volume of sample is injected into a gaschromatograph having a single packed column, op

8、erated iso-thermally at 10 to 50C, that separates COS from propylene.COS is detected with a flame photometric detector.3.3 Calibration data, based on peak areas, are obtained usinga known gas standard blend of COS in the range expected forthe sample. The COS peak area in the sample is measured andth

9、e concentration of COS calculated.3.4 The COS gas standard blend is assayed prior to use forcalibration.4. Significance and Use4.1 In processes producing propylene, COS usually remainswith the C3hydrocarbons and must be removed, since it affectsproduct quality. COS acts as a poison to commercial pol

10、ym-erization catalysts, resulting in deactivation and costly processdowntime.4.2 Accurate gas chromatographic determination of traceCOS in propylene involves unique analytical problems becauseof the chemical nature of COS and idiosyncracies of trace levelanalyses. These problems result from the reac

11、tive and absorp-tive nature of COS, the low concentration levels beingmeasured, the type of detector needed, and the interferencesfrom the propylene sample matrix. This test method addressesthese analytical problems and ways to properly handle them toassure accurate and precise analyses.4.3 This tes

12、t method provides a basis for agreement betweentwo laboratories when the determination of trace COS inpropylene is important. The test method permits severalcalibration techniques. For best agreement between two labs, itis recommended that they use the same calibration technique.5. Interferences5.1

13、Hydrogen sulfide (H2S) or sulfur dioxide (SO2) can bepresent in the propylene and must be separated from COS. (SeeNote 2.)NOTE 2H2S and SO2are separated from COS with the CarbopackBHT 100 columns or with the Chromosil 300 column.1This test method is under the jurisdiction of ASTM Committee D02 onPet

14、roleum Products and Lubricants and is the direct responsibility of SubcommitteeD02.D0.03 on Propylene.Current edition approved Dec. 1, 2012. Published December 2012. Originallyapproved in 1992. Last previous edition approved in 2007 as D530392(2007).DOI: 10.1520/D5303-92R12.2For referenced ASTM stan

15、dards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, P

16、A 19428-2959. United States16. Apparatus6.1 Gas ChromatographAny gas chromatograph (GC)equipped with a flame photometric detector/electrometer sys-tem (FPD), as described in 6.2, may be used. A GC/FPDequipped with an output signal linearizer is also permitted.6.2 Detector System, flame photometric d

17、etector, eithersingle or dual burner design. Noise level must be no more thanone recorder chart division (see 6.5). The signal for COS mustbe at least twice the noise level at the 0.1 mg/kg level. Adiscussion of this detector is presented in Practice E840. Theelectrometer used with the detector must

18、 have a sensitivity of1012A full scale ona1mVrecorder to achieve optimumdetectability at lowest levels.6.3 ColumnAny column that will effect the completeseparation of COS from propylene and other compoundsnormally present in propylene concentrates, and that is suffi-ciently inert to preclude the los

19、s of COS, may be used.Columns that meet these criteria, and that were used in thecooperative study for this test method, are listed in Table 1.6.4 Sample Inlet SystemAny gas sampling valve or gastight syringe that will permit introduction of up to 5.0 mLto thecolumn, and that will not cause any loss

20、 of COS, is suitable.6.5 RecorderAny strip chart recorder with a full scalerange of 1 mV, a maximum full scale balance time of 2 s, anda minimum chart speed of 0.5 cm/s, may be used.6.6 Data Handling SystemAny commercially availableGC integrator or GC computer system capable of accuratelyintegrating

21、 the area (uVs) of the COS peak is satisfactory. Datasystems that will linearize the logarithmic output of the FPDare also satisfactory.6.7 Sample Cylinders, 300 mL capacity or larger, fluorocar-bon lined stainless steel, Type DOT 3E, 12409 kPa (1800 psi)working pressure.7. Reagents and Materials7.1

22、 Air, zero grade.7.2 Carbonyl sulfide (COS), lecture bottle, 97.5 % min.(WarningToxic! See Section 8, Hazards.)7.3 Gas Calibration Blends, 1 to 10 mg/kg COS in eithernitrogen, argon, propylene or a propylene/argon mixture. Theycan be obtained from any commercial supplier or prepared asshown in Appen

23、dix X1 or Test Method D4468.7.4 Gas Sampling Syringe, 0.1, 1.0, and 5.0 mL.7.5 Gas Sampling Valve and Sample Loops, fluorocarbon or316 stainless steel. See Footnote B of Table 1.7.6 Glass Vials, 125 cm.7.7 Hydrogen, pure grade, 99.9 %.7.8 Isooctane (2,2,4-trimethylpentane), sulfur free, mini-mum pur

24、ity 99 mol %. (WarningFlammable! Health Haz-ard.)7.9 Nitrogen or Helium, 99.999 % min.7.10 TFE-fluorocarbon septa and aluminum seals for vials.8. Hazards8.1 Carbonyl sulfide is toxic and narcotic in highconcentrations, and upon decomposition can liberate hydrogensulfide. Exposure to dangerous concen

25、trations of COS is mostlikely when handling the pure component for preparation ofstandard blends for assaying the COS calibration gas standards.9. Sampling9.1 Supply samples to the laboratory in high pressurecylinders coated internally with TFE-fluorocarbon, or other-wise specially treated to reduce

26、 or eliminate loss of COS dueto reaction with the cylinder walls.9.2 The sample cylinder and contents should be at roomtemperature prior to sampling to the chromatograph. Testsamples as soon as possible after receipt.NOTE 3Cooperative studies indicate that the measured value for COSwill decrease wit

27、h time.9.3 Place the sample cylinder in a vertical position and useeither of the following two techniques to obtain a vaporizedsample from the container for introduction into the GC.9.3.1 Connect the sample cylinder to the sampling valve onthe chromatograph, using a minimum length of 316 ss tubing,s

28、o that sample is withdrawn from the bottom of the cylinder.Adjust the flow rate from the sample cylinder so that completevaporization of the liquid occurs at the cylinder valve. A flowrate of 5 to 10 bubbles/s through a water bubbler placed at thesample vent is sufficient (see Note 4). Turn the samp

29、ling valveto the “flush” position and flush for approximately 15 s. Shutoff the cylinder valve and allow the pressure to drop toatmospheric.NOTE 4If the flow rate is too fast, warming of the valve can berequired to avoid freezing and to ensure complete vaporization of thesample.9.3.2 Alternatively,

30、obtain a sample with a gas tight syringe.A convenient way to do this is to use flexible plastic tubing toconnect the bottom of the sample cylinder to the water bubblerand then to pierce the tubing with the syringe needle after flowis established.TABLE 1 Suitable GC Columns and TemperaturesAColumnNum

31、berSize, m mm TubingBTypePacking and OvenTemperature, C1 0.9 3.78 SS Porapak R, 80/100 Mesh; 472 1.43.78 TFECCarbopack BHT 100, 40/60Mesh; 25,40D3 1.8 3.78 TFE Carbopack BHT 100; 25,30D4 1.8 3.78 TFE Porapak Q, AW, 50/80 Mesh2.4 3.78 TFE (Above in Series); 745 2.4 3.78 SS Carbopack BHT 100; 476 2.8

32、3.78 TFE Carbopack BHT 100, 40/60Mesh; 507 3.6 3.78 TFE Carbopack BHT 100, 40/60Mesh; 508 4.3 3.78 TFE Chromosil 300; 50E9 6.1 3.78 TFE Hayes Sep Q, 80/100 Mesh; 65AThese columns have been tested cooperatively and found suitable for use withthis test method.B316 SS Tubing for columns or connection o

33、f sample cylinder to sampling systemcan be TFE lined internally to improve on system stability. This tubing iscommercially available from chromatography vendors.CTFEHomopolymer of tetrafluoroethylene.DIdentical columns used by different labs at different temperatures.EPropyne (methyl acetylene) can

34、interfere with COS using this column.D5303 92 (2012)210. Preparation of Apparatus10.1 Install in the GC according to the manufacturersinstructions any of the columns that meet the criteria in 6.3. Setthe instrument conditions as follows:10.1.1 Oven Temperature, as determined by column used,10.1.2 De

35、tector, 100 to 200C, and10.1.3 Injector, 100 to 150C.11. Calibration11.1 Three methods of calibration are permitted. These arethe Standard Sample Method (see 11.2), the Permeation TubeMethod (see 11.3) and Blend Preparation Techniques (see11.4). Obtain a calibration standard according to one of thes

36、emethods, which are described below. Then follow the proce-dure in 12.1-12.4 and the calculations described in 11.5.11.2 Standard Sample MethodPurchase a certified com-mercial calibration sample of 10 mg/kg COS in propylene, orother suitable matrix gas such as nitrogen, argon, or apropylene/argon mi

37、xture. If an inert gas is chosen, the usermust ensure that the column is actually effecting a separation ofCOS and propylene. Establish a calibration curve with thestandard sample using either a gas syringe or different sizesample loops. For example, assume the normal sample size forthe analysis is

38、1.0 mL and the calibration range to beestablished is 0.5 to 5 mg/kg of COS. Establish a calibrationcurve by injecting the volumes of a 10 mg/kg standard sampleshown in the first column of the table below. The equivalentconcentration of COS in a 1.0 mL sample would be that shownin the second column:S

39、tandard Sample Equivalent Concentration, COS mg/kg0.05 0.50.10 1.00.20 2.00.30 3.00.40 4.00.50 5.011.3 Permeation Tube MethodRefer to Practice D3609 fordirections on using permeation tubes.11.4 Blend Preparation TechniquesTechniques for thepreparation and assay verification of calibration blends in

40、thelaboratory are described in Appendix X1 and Appendix X2.Also, a technique using a moving piston graduated cylinderapparatus, that is described in the calibration section of TestMethod D4468, can be used. However, some laboratories havefound that the preparation of such blends is far from easy, an

41、dsuccessful efforts require considerable knowledge and experi-ence.11.5 QuantitationThe flame photometric detector re-sponds logarithmically to the mass of the sulfur present in theflame. Some GC/FPD systems are programmed to linearizelogarithmic data, and with such systems the output can becorrelat

42、ed directly with the COS concentration, using a singlepoint calibration. Calculate a calibration factor, F, in accor-dance with (Eq 1) below:F 5 C/A (1)where:F = calibration factor,C = concentration, mg/kg, of COS in this test method, andA = area (uVs) of the COS peak in this test method.F will be u

43、sed in (Eq 2)in13.1.1. However, if a linearizer isnot used, or if the data system does not have a provision tohandle logarithmic output, use the method in 11.5.1 or thealternate in 11.5.2, below:11.5.1 Calculate the nanogram (ng) amounts of sulfur, asdescribed in Appendix X3, for each injection of t

44、he standard,and plot the natural logarithm (1n) of peak area versus the1n(ng) of sulfur, as illustrated in Table 2 and Fig. 1. The plotshould be a straight line.11.5.2 Alternatively, plot the concentration of COS inmg/kg versus the square root of the peak area. This plot shouldalso be a straight lin

45、e.12. Procedure12.1 Using either the gas sampling valve or a gas tightsyringe, as described in 9.3, inject the sample into the gaschromatograph.12.2 Record the response of the FPD on the strip chartrecorder as the COS elutes from the column.12.3 Alternatively, obtain the computer or integrator outpu

46、tof COS retention time and peak area.12.4 Obtain duplicate chromatograms of the sample. Fig. 2illustrates a typical analysis using a Carbopack BHT-100column.13. Calculation13.1 Depending on the method of calibration used (seeSection 11), determine the concentration of COS in the sample.13.1.1 If the

47、 system provides a linearized output, determineCOS concentration according to (Eq 2), below:COS, mg/kg 5 F 3 S (2)where:F = calibration factor from (Eq 1), andS = area (uVs) of the COS peak from the sample.TABLE 2 Example of COS Calibration DataANOTE 1COS Standard (3.00 ng S/cm3).Amount of StandardI

48、njected (cm3)Amount ofStandard Injected (ng S)yB(peak area units)3.0 9.0 533.0 9.0 512.5 7.5 392.5 7.5 382.0 6.0 252.0 6.0 231.5 4.5 181.5 4.5 161.0 3.0 101.0 3.0 90.5 1.5 30.5 1.5 4ACorrelation coefficient of fit (r) = 0.9952:m = slope (detector response factor) = 1.4920,y = peak area units,z = nan

49、ograms of sulfur as COS injected, andb = intercept = 1.8394.BCalibration equation: y = bzm.D5303 92 (2012)313.1.2 If a calibration curve of 1n peak area versus 1n(ng)sulfur was used (see 11.5.1), then determine the concentrationof COS as shown in Appendix X1.13.1.3 If a calibration curve of concentration versus logpeak area was used (see 11.5.2), then determine the COSconcentration as follows:13.1.3.1 Calculate the log of the area of the COS peak ofsample.13.1.3.2 Take the COS concentration directly from thecurve using the log value from

展开阅读全文
相关资源
猜你喜欢
相关搜索

当前位置:首页 > 标准规范 > 国际标准 > ASTM

copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
备案/许可证编号:苏ICP备17064731号-1