ASTM D5348-1995(2012) Standard Test Method for Determination of the Moisture Content of Sulfonated and Sulfated Oils by Distillation with Xylene 《用二甲苯蒸馏测定磺化油和硫酸化油中水分含量的标准试验方法》.pdf

上传人:registerpick115 文档编号:519541 上传时间:2018-12-03 格式:PDF 页数:3 大小:68.51KB
下载 相关 举报
ASTM D5348-1995(2012) Standard Test Method for Determination of the Moisture Content of Sulfonated and Sulfated Oils by Distillation with Xylene 《用二甲苯蒸馏测定磺化油和硫酸化油中水分含量的标准试验方法》.pdf_第1页
第1页 / 共3页
ASTM D5348-1995(2012) Standard Test Method for Determination of the Moisture Content of Sulfonated and Sulfated Oils by Distillation with Xylene 《用二甲苯蒸馏测定磺化油和硫酸化油中水分含量的标准试验方法》.pdf_第2页
第2页 / 共3页
ASTM D5348-1995(2012) Standard Test Method for Determination of the Moisture Content of Sulfonated and Sulfated Oils by Distillation with Xylene 《用二甲苯蒸馏测定磺化油和硫酸化油中水分含量的标准试验方法》.pdf_第3页
第3页 / 共3页
亲,该文档总共3页,全部预览完了,如果喜欢就下载吧!
资源描述

1、Designation: D5348 95 (Reapproved 2012)Standard Test Method forDetermination of the Moisture Content of Sulfonated andSulfated Oils by Distillation with Xylene1This standard is issued under the fixed designation D5348; the number immediately following the designation indicates the year oforiginal ad

2、option or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of waterexisting in a sample

3、 of sulfonated or sulfated oil, or both, bydistilling the sample with a volatile solvent. The test method isapplicable only to sulfonated and sulfated oils that do notcontain the following: mineral acids, free sulfonic acids, or freesulfuric acid esters; or alkali hydroxides, carbonates or ac-etates

4、: or alcohol, glycerin, diethylene glycol, acetone, or otherwater-miscible volatile compounds. This test method wasderived from Test Methods D500, Sections 4 through 9.1.2 The values stated in SI units are to be regarded as thestandard. The values given in parentheses are provided forinformation onl

5、y.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced

6、 Documents2.1 ASTM Standards:2D500 Test Methods of ChemicalAnalysis of Sulfonated andSulfated Oils3. Significance and Use3.1 This test method is intended to determine the moisturecontent of fats, oils, and fatliquors used in the softening andstuffing of leather. The moisture content is measured for

7、thepurpose of quality assurance.4. Apparatus4.1 The apparatus required consists of a glass flask heatedby suitable means and provided with a reflux condenserdischarging into a trap and connected to the flask. Theconnections between the trap and the condenser and flask shallbe interchangeable ground

8、joints. The trap serves to collect andmeasure the condensed water and to return the solvent to theflask. A suitable assembly of the apparatus is illustrated in Fig.1.4.1.1 Flask, 500-mL, of either the short-neck, round-bottomtype or the Erlenmeyer type.4.1.2 Heat SourceThe source of heat may be eith

9、er an oilbath (stearic acid, paraffin wax, etc.), or an electric heaterprovided with a sliding rheostat or other means of heat control.4.1.3 Condenser, a water-cooled glass reflux condenser(Fig. 1), having a jacket approximately 400 mm (1534 in.) inlength with an inner tube 9.5 to 12.7 mm (38 to12 i

10、n.) inoutside diameter. The end of the condenser to be inserted in thetrap shall be ground off at an angle of 30 from the vertical axisof the condenser. When inserted into the trap, the tip of thecondenser shall be about 7 mm (14 in.) above the surface of theliquid in the trap after the distillation

11、 conditions have beenestablished. Fig. 1 shows a conventional sealed-in type ofcondenser, but any other condenser fulfilling the detailedrequirements above may be used.4.1.4 Trap, a trap made of well-annealed glass constructedin accordance with Fig. 1 and graduated as shown to contain 5mL at 20C. It

12、 shall be subdivided into 0.1-mL divisions, witheach 1-mL line numbered (5 mL at top). The error in anyindicated capacity may not be greater than 0.05 mL.5. Purity of Reagents5.1 Purity of ReagentsReagent grade chemicals shall beused in all tests. Unless otherwise indicated, it is intended thatall r

13、eagents shall conform to the specifications of the Commit-tee onAnalytical Reagents of theAmerican Chemical Society.36. Reagents6.1 Oleic Acid, heated previous to use for 5 to 10 min overa free flame at a temperature of 130 to 135C.1This test method is under the jurisdiction of ASTM Committee D31 on

14、 Leatherand is the direct responsibility of Subcommittee D31.08 on Fats and Oils. This testmethod was developed in cooperation with the American Leather Chemists Assn.(Method H 40-1957).Current edition approved April 1, 2012. Published April 2012. Originallyapproved in 1993. Last previous edition ap

15、proved in 2006 as D5348 95(2006).DOI: 10.1520/D5348-95R12.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Re

16、agent Chemicals, American Chemical Society Specifications , AmericanChemical Society, Washington, DC. For suggestions on the testing of reagents notlisted by the American Chemical Society, see Analar Standards for LaboratoryChemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia

17、and National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville,MD.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.6.2 Xylene.7. Calibration7.1 To calibrate the apparatus add approximately1gofwater to a mixture of 80 g

18、of xylene and 10 g of oleic acid.Conduct the distillation as described in 8.2, 8.3, and 8.4. Whenall the water has distilled, cool the apparatus, add another g ofwater, and repeat the distillation. Continue the calibration up tothe capacity of the receiving tube.8. Procedure8.1 Clean the condenser a

19、nd the receiving tube thoroughlywith soap and warm water before using. Rinse well, then treatwith hot cleaning solution (a mixture of 10 mL of saturatedpotassium dichromate (K2Cr2O7) and 990 mL of sulfuric acid(H2SO4, sp gr 1.84), and finally thoroughly wash and dry.8.2 Take enough of the sample to

20、be tested for analysis toyield about 4 mL of water. Introduce the approximate quantityinto a weighing bottle and make the weighings from the bottleinto the flask, taking care that after removal of the sample nodrops of oil are left on the outside of the weighing bottle. Add80 g of xylene and oleic a

21、cid equivalent to about two andone-half times the weight of the bone-dry sample to preventfoaming and jellying of the contents of the flask. Introduceglass beads to prevent bumping and mix the contents of theflask thoroughly by swirling, taking care to avoid any loss ofmaterial. Fill the trap with x

22、ylene and immediately connect theflask with the distillation apparatus. Insert a loose cotton plugin the top of the condenser tube to prevent condensation ofatmospheric moisture in the condenser tube.8.3 Heat the flask and regulate the heating so that thecondenser tube immediately below the water ja

23、cket is justbarely hot. In this way a minimum of water will condensefarther up the condenser where it may be difficult to volatilizeany moisture condensed on the walls.8.4 Continue the distillation at the specified rate untilpractically no water is visible on any part of the apparatusexcept within t

24、he graduations of the trap. This operationusually requires less than 1 h. Increase the rate of distillation inorder to remove all traces of condensed water in the condensertube, and continue the distillation until the water level in thetrap remains unchanged after a 10-min interval. Dislodge anydrop

25、lets adhering to the side of the receiver with a thin copperwire twisted into a loop. Immerse the receiving tube in warmwater at about 40C for 15 min or until the xylene layerbecomes clear, then read and record the temperature and theexact volume of the water in the trap.9. Calibration9.1 The volume

26、 of condensed water measured in the trapmay be converted into its equivalent weight in grams by meansof Table 1. Calculate the percentage of water as follows:Water, % 5 A/B! 3 100 (1)where:A = weight of water, g, andB = weight of sample, g.10. Precision and Bias10.1 Although this test method is wide

27、ly used, precision andbias information is not available at this time.11. Keywords11.1 distillation; leather; moisture; sulfonated and sulfatedoils; xyleneA =45to55mmB = 22 to 24 mm in inside diameterC = 9 to 11 mm in inside diameterD = 235 to 240 mmE = 146 to 156 mmF and G are interchangeable joints

28、, standard taper 24/40.FIG. 1 Apparatus for Water Determination by Distillation withVolatile Solvent, Method ATABLE 1 Specific Gravity of WaterATemperature, C Specific Gravity4 1.0000035 0.9940636 0.9937137 0.9933638 0.9929939 0.9926240 0.9922441 0.9918642 0.9914743 0.9910744 0.9906645 0.99025AThis

29、table is taken from Smithsonian Tables, compiled from various authors.D5348 95 (2012)2ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentionedin this standard. Users of this standard are expressly advised that determination of t

30、he validity of any such patent rights, and the riskof infringement of such rights, are entirely their own responsibility.This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years andif not revised, either reapproved or withdrawn. Yo

31、ur comments are invited either for revision of this standard or for additional standardsand should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of theresponsible technical committee, which you may attend. If you feel that your comment

32、s have not received a fair hearing you shouldmake your views known to the ASTM Committee on Standards, at the address shown below.This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,United States. Individual reprints (single or mul

33、tiple copies) of this standard may be obtained by contacting ASTM at the aboveaddress or at 610-832-9585 (phone), 610-832-9555 (fax), or serviceastm.org (e-mail); or through the ASTM website(www.astm.org). Permission rights to photocopy the standard may also be secured from the ASTM website (www.astm.org/COPYRIGHT/).D5348 95 (2012)3

展开阅读全文
相关资源
猜你喜欢
  • STAS 11237 1-1980 MEDICAT ELECTRICAL EQUIPMENT FOR USE IN THE PATIENT ENVIRONMENT Safety General prescription and rules《医学 用于在病人环境安全的医用电气设备 通用处方和规则》.pdf STAS 11237 1-1980 MEDICAT ELECTRICAL EQUIPMENT FOR USE IN THE PATIENT ENVIRONMENT Safety General prescription and rules《医学 用于在病人环境安全的医用电气设备 通用处方和规则》.pdf
  • STAS 11237 2-1980 MEDICAL ELECTRICAL EQUIPMENT FOR USE IN THE PATIENT ENVIRONMENT Safety Protection against electric shock《用于在病人环境的医用电气设备 安全 防触电保护》.pdf STAS 11237 2-1980 MEDICAL ELECTRICAL EQUIPMENT FOR USE IN THE PATIENT ENVIRONMENT Safety Protection against electric shock《用于在病人环境的医用电气设备 安全 防触电保护》.pdf
  • STAS 11237 22-1988 MEDICAL ELECTRICAL IGH FREQUENCY SURGICAL QUIPMENT eneral technical requirements for safety《医学的电子好频率外壳设备通用安全的技术要求 》.pdf STAS 11237 22-1988 MEDICAL ELECTRICAL IGH FREQUENCY SURGICAL QUIPMENT eneral technical requirements for safety《医学的电子好频率外壳设备通用安全的技术要求 》.pdf
  • STAS 11237 3-1980 MEDICAL ELECTRICAL EQUIPMENT FOR USE IN THE PATIENT ENVIRONMENT Safety Protection against mechanical hazards fault conditions and endurance《用于在病人环境的医用电气设备 安全 防止机械.pdf STAS 11237 3-1980 MEDICAL ELECTRICAL EQUIPMENT FOR USE IN THE PATIENT ENVIRONMENT Safety Protection against mechanical hazards fault conditions and endurance《用于在病人环境的医用电气设备 安全 防止机械.pdf
  • STAS 11237 4-1980 MEDICAL ELECTRICAL EQUIPMENT FOR USE IN THE PATIENT ENVIRONMENT Safety Construction《用于在病人环境安全的医用电气设备 建造》.pdf STAS 11237 4-1980 MEDICAL ELECTRICAL EQUIPMENT FOR USE IN THE PATIENT ENVIRONMENT Safety Construction《用于在病人环境安全的医用电气设备 建造》.pdf
  • STAS 11237 5-1980 MEDICAL ELECTRICAL EQUIPMENT FOR USE IN THE PATIENT= ENVIRONMENT Safety Protection against explosions《医学电气设备用于病人 防爆环境安全 》.pdf STAS 11237 5-1980 MEDICAL ELECTRICAL EQUIPMENT FOR USE IN THE PATIENT= ENVIRONMENT Safety Protection against explosions《医学电气设备用于病人 防爆环境安全 》.pdf
  • STAS 11238-1983 BAST FIBRES AND BAST TYPE FIBRES WOVEN FABRICS Widths《韧皮纤维和韧皮型纤维机织物 宽度》.pdf STAS 11238-1983 BAST FIBRES AND BAST TYPE FIBRES WOVEN FABRICS Widths《韧皮纤维和韧皮型纤维机织物 宽度》.pdf
  • STAS 11239-1989 Metallic coatings AYER THICKNESS DETERMINATION BY MAGNETIC METHOD《金属涂层的层厚度通过磁法测定 》.pdf STAS 11239-1989 Metallic coatings AYER THICKNESS DETERMINATION BY MAGNETIC METHOD《金属涂层的层厚度通过磁法测定 》.pdf
  • STAS 11240-1988 MODYIFING ALLOYS FOR CAST IRONS《铸铁用变质合金 》.pdf STAS 11240-1988 MODYIFING ALLOYS FOR CAST IRONS《铸铁用变质合金 》.pdf
  • 相关搜索

    当前位置:首页 > 标准规范 > 国际标准 > ASTM

    copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
    备案/许可证编号:苏ICP备17064731号-1