1、Designation: D 5666 95 (Reapproved 2009)Standard Test Method forRubber Chemical AntidegradantsPurity ofp-Phenylenediamine (PPD) Antidegradants by HighPerformance Liquid Chromatography1This standard is issued under the fixed designation D 5666; the number immediately following the designation indicat

2、es the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the purity of Type I,

3、 II, and IIIp-phenylenediamine (PPD) antidegradants as described inClassification D 4676 by high performance liquid chromatog-raphy (HPLC) using ultraviolet detection and external standardcalculations.1.2 Expertise in HPLC is necessary to the successful appli-cation of this test method.1.3 The value

4、s stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safe

5、ty and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D 3853 Terminology Relating to Rubber and RubberLaticesAbbreviations for Chemicals Used in Com-poundingD 4483 Practice for Evaluating Precision for Test MethodSt

6、andards in the Rubber and Carbon Black ManufacturingIndustriesD 4676 Classification for Rubber CompoundingMaterialsAntidegradants2.2 ISO Standards:3ISO 5725 Precision of Test MethodsISO 6472 Rubber Compounding IngredientsAbbreviations3. Terminology3.1 Definitions:3.1.1 external standard calculationa

7、 method of calculat-ing the percent composition by measuring the area of theanalyte peak, multiplying by a response factor, and dividing bythe sample concentration. All components are assumed to beresolved from the component of interest.3.1.2 lot samplea production sample representative of astandard

8、 production unit, normally referred to as the sample.3.1.3 specimenthe actual material used in the analysis,also known as the test portion. It must be representative of thelot sample.3.2 Abbreviations: The following abbreviations are in ac-cordance with Terminology D 3853 and ISO 6472:3.2.1 77PDN,N8

9、 bis-(1,4-dimethylpentyl)-p-phenylene-diamine.3.2.2 DTPDN,N8-ditolyl-p-phenylenediamine.3.2.3 IPPDN-isopropyl-N8-phenyl- p-phenylenediamine.3.2.4 PPDp-phenylenediamine.3.2.5 6PPDN-(1,3 dimethylbutyl)-N8-phenyl-p-phenyl-enediamine.4. Summary of Test Method4.1 A specimen is dissolved in acetonitrile a

10、nd a fixed loopvolume is analyzed by isocratic HPLC using a thermostatedC18 reversed phase column and an ultraviolet (UV) detector.Peak areas are determined using a chromatographic integratoror laboratory data system with the amount of analyte beingdetermined by external calibration.5. Significance

11、and Use5.1 This test method is designed to determine the purity ofp-phenylenediamine antidegradants.5.2 Since the results of this test method are based on anintegrated peak area as determined by HPLC, it is assumed thatall analytes of interest are resolved from interfering peaks.1This test method is

12、 under the jurisdiction of ASTM Committee D11 on Rubberand is the direct responsibility of Subcommittee D11.11 on Chemical Analysis.Current edition approved July 1, 2009. Published September 2009. Originallyapproved in 1995. Last previous edition approved in 2004 as D 5666 95 (2004)1.2For referenced

13、 ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from American National Standards Institute (ANSI), 25 W. 43rd St.

14、,4th Floor, New York, NY 10036.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.6. Interferences6.1 Components co-eluting with the analyte of interest willcause erroneous results; thus it is required that the system becapable of provi

15、ding a minimum of 10 000 theoretical plates.7. Apparatus7.1 Liquid Chromatograph, consisting of the following:7.1.1 Precision chromatographic pump,7.1.2 Variable wavelength UV detector,7.1.3 A method for thermostating the column at 35 6 1C,for example, a column oven or water jacket,7.1.4 A fixed inj

16、ector made of either a 20 mm3(L) rheo-dyne loop or an automatic sampler.7.2 HPLC Columns, consisting of:7.2.1 A precolumn packed with C18 grafted silica withparticle size of 35 to 40 m (100 to 150 mm), and7.2.2 A column of 10- to 15-cm length packed with C18grafted silica with particle size of 3 to

17、5 m.7.3 Integrator/Data System, capable of determining abso-lute amounts of analyte of interest by means of integration ofdetector output versus time.7.4 Analytical Balance, capable of measuring within 60.01mg.7.5 Shaking Machine, or ultrasonic tank.7.6 Volumetric Flask, 100 cm3.7.7 Syringes, with r

18、heodyne loop, 2 cm3.7.8 Clear Screw-Top Vials, with suitable septa, 125 cm3.8. Reagents and Materials8.1 Acetonitrile, HPLC grade.8.2 Ethanolamine, HPLC grade.8.3 Water, HPLC grade or double distilled water or water ofresistivity greater than 2 megohms/cm.9. Calibration and Standardization9.1 Aprima

19、ry standard of known purity is used to determinethe response factor for each analyte.10. Procedure10.1 Chromatographic Conditions:10.1.1 Determine the eluant phase composition and the flowrate by adjusting the chromatographic parameters for theparticular column chosen. The eluant phase consists of t

20、heappropriate mixture of HPLC grade acetonitrile and HPLCgrade or equivalent water, both containing 0.2 kg/m3(g/L)ethanolamine or less according to the product to be tested.NOTE 1Different liquid chromatography columns may exhibit differ-ent elution characteristics. See Table 1 for suggested chromat

21、ographicstarting parameters for analysis.10.2 DetectorMonitor the absorbance of the sample at theprescribed wavelength. The detector should be set to 1 absor-bance unit full scale (AUFS).10.3 Integrator/Data SystemThe integrator settingsshould be adjusted to give a full-scale response to 1 absorbanc

22、eunit (AU).10.4 Sample Storage Before AnalysisSamples must al-ways be stored in a refrigerator.10.5 Standard PreparationWeigh the clear vial to thenearest 0.1 mg, introduce approximately 20 mg of the standardusing a spatula and weigh the standard and vial to the nearest0.1 mg. Using a volumetric fla

23、sk, add 100 cm3of acetonitrileto the vial. Stopper the vial so that it is hermetically sealedimmediately after adding the solvent. Dissolve the product at23 6 3C in the ultrasonic bath tank or on the shakingmachine. The standard must be analyzed within4hofbeingprepared.NOTE 2Preparation of Standards

24、The analytical standards are pre-pared by multiple recrystallizations or distillations of the paraphenylene-diamines. The procedure can be repeated until the desired purity isobtained. The purity of the standard is estimated by gradient HPLCanalysis of the impurities and differential thermal analysi

25、s (DTA). Theimpurities in the standard should be reestimated every 90 days by HPLC.The standard should be stored at 5C or lower.10.6 Sample PreparationTo ensure sample homogeneity,5 g of sample should be ground with a mortar and pestle.10.7 Sample Analysis:10.7.1 Weigh at least 20 mg to the nearest

26、0.1 mg of thesample in a clear vial and dissolve it in 100 cm3acetonitrilefollowing the procedure in 10.5. The sample must be analyzedwithin4hofbeing prepared.10.7.2 Injection of the Solutions:10.7.2.1 Manual MethodTake approximately 100 L ofthe solution using a syringe and inject a quantity greater

27、 thanthe volume of the rheodyne loop, that is, approximately 60 lfor a 20-l loop. Rinse the syringe with solvent and dry.10.7.2.2 Automatic MethodPut the flasks containing thesample and standard solutions in place and program theautomatic sampler.10.7.3 Chromatograph the samples using parameters asp

28、rescribed in 10.1.1.11. Calculation11.1 Response FactorCalculate the response factor for thestandard by dividing the concentration of the standard by themeasured area count and multiplying this by the purity of thestandard:RF 5concentration/area count!3% purity (1)NOTE 3Throughout the calculation th

29、e units of concentration must beconsistent (that is, kg/m3(mg/cm3).11.2 Product PurityTo determine the purity of the prod-uct, multiply the response factor by the measured area count ofthe analyte and divide by the sample concentration:% purity 5 RF 3 area count/sample concentration (2)TABLE 1 Sugge

30、sted Chromatographic Starting ParametersN,N8-dialkyl-paraphenylene-diamine77-PPDN-aryl-N8-alkyl-paraphenylene-diamineIPPD6-PPDN,N8-diaryl-paraphenylene-diamineDPPDDTPDEluant phase% acetonitrile 85 65 70% Water 15 35 30Ethanolamine (g/L) 0.2 0.1 0.1Flow rate (cm3/min) 1 1 1Wavelength (nm) 260 290 280

31、D 5666 95 (2009)212. Report12.1 Report percent paraphenylenediamine to the nearest0.1 %.13. Precision and Bias413.1 This precision and bias section has been prepared inaccordance with Practice D 4483. Refer to Practice D 4483 forterminology and other statistical details.13.1.1 The precision results

32、in this precision and biassection give an estimate of the precision of this test methodwith the materials used in the particular interlaboratory pro-grams as described below. The precision parameters should notbe used for acceptance/rejection testing of any group ofmaterials without documentation th

33、at they are applicable tothose particular materials and the specific testing protocols thatinclude this test method.13.2 These precision and bias data were obtained in aninterlaboratory test organized in France in 1992. In thisprogram one material was analyzed by 13 different laborato-ries. Six meas

34、urements were taken over six days by one tothree operators. Statistical evaluation was carried out in accor-dance with ISO 5725-1986, which is equivalent to the calcu-lation algorithms of Practice D 4483. The results from thisprecision and bias study are given in Table 2.13.3 RepeatabilityThe differ

35、ence between two single testresults (or determinations) found on identical test materialunder the repeatability conditions prescribed for a particulartest will exceed the repeatability (r), as given in Table 2,onanaverage of not more than once in twenty cases in the normaland correct operation of th

36、e test method.13.4 ReproducibilityThe difference between two singleand independent test results found by two operators workingunder prescribed reproducibility (R) conditions in differentlaboratories on identical test material will exceed the repro-ducibility (R), as given in Table 2, on an average o

37、f not morethan once in twenty cases in the normal and correct operationof the test method.13.5 BiasSample impurities that are not resolved from theanalyte of interest will produce a falsely high result. There maybe other sources of bias that have not been determined.14. Keywords14.1 antidegradant; h

38、igh performance liquid chromatogra-phy; N-isopropl-N8-phenyl-p-phenylenediamine (IPPD); N-(1,3 dimethylbutyl)-N8-phenyl-p-phenylenediamine (6PPD); N,N8 bis-(1, 4-dimethylpentyl)-p-phenylenediamine (77PD); N,N8-ditolyl-p-phenylenediamine (DTPD); p-phenylenediamine(PPD)APPENDIX(Nonmandatory Informatio

39、n)X1. RECOMMENDATIONSX1.1 De-gas the eluents.X1.2 Use an appropriate guard column.X1.3 Acid-clean the glassware.X1.4 Keep the temperature of the samples and standard thesame.ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mention

40、edin this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the riskof infringement of such rights, are entirely their own responsibility.This standard is subject to revision at any time by the responsible technical committee and

41、 must be reviewed every five years andif not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standardsand should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of

42、 theresponsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you shouldmake your views known to the ASTM Committee on Standards, at the address shown below.This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box

43、C700, West Conshohocken, PA 19428-2959,United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the aboveaddress or at 610-832-9585 (phone), 610-832-9555 (fax), or serviceastm.org (e-mail); or through the ASTM website(www.astm.org).4Suppor

44、ting data have been filed at ASTM International Headquarters and maybe obtained by requesting Research Report RR: D11-1063.TABLE 2 Precision (Type 1)Paraphenylenediamine PurityMaterialMeanAlevelWithin laboratoriesBBetween laboratoriesBSrr (r) SRR (R)6PPD 95.27 0.335 0.948 0.995 0.441 1.248 1.31AMean

45、 level values (in percent).BSymbols are defined as follows:Sr= within laboratory standard deviation,r = repeatability (in measurement units),(r) = repeatability (in percent),SR= between laboratory standard deviation,R = reproducibility (in measurement units),(R) = reproducibility (in percent)D 5666 95 (2009)3