ASTM D6052-1997(2016) 7966 Standard Test Method for Preparation and Elemental Analysis of Liquid Hazardous Waste by Energy-Dispersive X-Ray Fluorescence《采用能量发散X射线荧光法对危险废液进行制备和元素分析的.pdf

上传人:李朗 文档编号:521403 上传时间:2018-12-03 格式:PDF 页数:7 大小:96.56KB
下载 相关 举报
ASTM D6052-1997(2016) 7966 Standard Test Method for Preparation and Elemental Analysis of Liquid Hazardous Waste by Energy-Dispersive X-Ray Fluorescence《采用能量发散X射线荧光法对危险废液进行制备和元素分析的.pdf_第1页
第1页 / 共7页
ASTM D6052-1997(2016) 7966 Standard Test Method for Preparation and Elemental Analysis of Liquid Hazardous Waste by Energy-Dispersive X-Ray Fluorescence《采用能量发散X射线荧光法对危险废液进行制备和元素分析的.pdf_第2页
第2页 / 共7页
ASTM D6052-1997(2016) 7966 Standard Test Method for Preparation and Elemental Analysis of Liquid Hazardous Waste by Energy-Dispersive X-Ray Fluorescence《采用能量发散X射线荧光法对危险废液进行制备和元素分析的.pdf_第3页
第3页 / 共7页
ASTM D6052-1997(2016) 7966 Standard Test Method for Preparation and Elemental Analysis of Liquid Hazardous Waste by Energy-Dispersive X-Ray Fluorescence《采用能量发散X射线荧光法对危险废液进行制备和元素分析的.pdf_第4页
第4页 / 共7页
ASTM D6052-1997(2016) 7966 Standard Test Method for Preparation and Elemental Analysis of Liquid Hazardous Waste by Energy-Dispersive X-Ray Fluorescence《采用能量发散X射线荧光法对危险废液进行制备和元素分析的.pdf_第5页
第5页 / 共7页
亲,该文档总共7页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、Designation: D6052 97 (Reapproved 2016)Standard Test Method forPreparation and Elemental Analysis of Liquid HazardousWaste by Energy-Dispersive X-Ray Fluorescence1This standard is issued under the fixed designation D6052; the number immediately following the designation indicates the year oforiginal

2、 adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of trace andmajor element

3、 concentrations by energy-dispersive X-ray fluo-rescence spectrometry (EDXRF) in liquid hazardous waste(LHW).1.2 This test method has been used successfully on numer-ous samples of aqueous and organic-based LHW for thedetermination of the following elements: Ag, As, Ba, Br, Cd,Cl, Cr, Cu, Fe, Hg, I,

4、 K, Ni, P, Pb, S, Sb, Se, Sn, Tl, V, and Zn.1.3 This test method is applicable for other elements (Si-U)not listed in 1.2.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-

5、priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2C982 Guide for Selecting Components for Energy-Dispersive X-Ray Fluorescence (XRF) Systems (With-drawn 2008)3D1193 Specification for Reagent Water2.2

6、Other ASTM Documents:ASTM Data Series DS 46 X-ray Emission Wavelengths andKeV Tables for Nondiffractive Analysis43. Summary of Test Method3.1 A weighed portion of activated alumina and sample arecombined in a mixing vessel and shaken until well mixed. Thesample mixture is transferred into a disposab

7、le sample cup andplaced in the spectrometer for analysis.3.2 The K spectral emission lines are used for elementsSi-Ba.3.3 The L spectral emission lines are used for elements withatomic numbers greater than Ba.4. Significance and Use4.1 The elemental analysis of liquid hazardous waste isoften importa

8、nt for regulatory and process specific require-ments. This test method provides the user an accurate, rapidmethod for trace and major element determinations.5. Interferences5.1 Spectral Overlaps (Deconvolution):5.1.1 Samples containing a mixture of elements often ex-hibit X-ray emission line overlap

9、. Modern Si (Li) detectorsgenerally provide adequate resolution to minimize the effectsof spectral overlap. In cases where emission line overlap exists,techniques of peak fitting exist for extracting corrected analyteemission line intensities. For example, the PbL “line overlapswith the AsK.” The Pb

10、L line can be used to avoid thisoverlap and the AsK lines can then be resolved from the PbLoverlap.The actual lines used for any particular element shouldbe such that overlaps are minimized. Follow the EDXRFmanufacturers recommendation concerning spectral deconvo-lution. Reference should be made to

11、ASTM Data Series DS 46for detailed information on potential line overlaps.5.2 Matrix Interferences (Regression):5.2.1 Matrix interference in the measurement of “as re-ceived” LHW samples using EDXRF has been the principlelimitation in the development and expanding use of thisinstrumental technique.

12、Using well understood XRF principlesfor controlling matrix effects, for example, dilution and matrixmodification using lithium borate fusion and addition of heavyabsorbers, a matrix can be stabilized. Using calcined aluminaand the above principles matrices are stabilized for quantitativeEDXRF analys

13、is.5.2.2 The response range of this test method should belinear with respect to the elements of interest and theirregulatory or process control, or both, action thresholds. Large1This test method is under the jurisdiction of ASTM Committee D34 on WasteManagement and is the direct responsibility of S

14、ubcommittee D34.01.06 onAnalytical Methods.Current edition approved Sept. 1, 2016. Published September 2016. Originallyapproved in 1997. Last previous edition approved in 2008 as D6052 97 (2008).DOI: 10.1520/D6052-97R16.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact

15、ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3The last approved version of this historical standard is referenced onwww.astm.org.4Available from ASTM Headquarters, Customer Service.Copyr

16、ight ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1concentration variations of element or matrix, or both, compo-nents in LHW samples can result in non-linear X-ray intensityresponse at increasing element concentrations.6. Apparatus6.1 Energy

17、-dispersive X-ray Fluorescence Spectrometer, ca-pable of measuring the wavelengths of the elements listed in1.2. Refer to Guide C982 for system specifications.6.2 Analytical Balance, capable of weighing to 0.001 g.7. Reagents and Materials7.1 Purity of ReagentsReagent grade chemicals shall beused in

18、 all tests. Unless otherwise indicated, it is intended thatall reagents conform to the specifications of the Committee onAnalytical Reagents of the American Chemical Society, wheresuch specifications are available.5Other grades may be used,provided it is first ascertained that the reagent is of suff

19、icientlyhigh purity to permit its use without lessening the accuracy ofthe determination.7.2 Purity of WaterUnless otherwise indicated, referencesto water shall be understood to mean meeting the numericalrequirements of Type II water as defined by SpecificationD1193.7.3 Aluminum Oxide, Al2O3pre-calc

20、ined at 1500C, ap-proximately 100 to 125 mesh.7.4 Aqueous or organic-based Atomic Absorption Standards(AAS), 1000 mg/L for the elementsAg,As, Ba, Cd, Cr, Cu, Fe,Hg, K, Ni, Pb, Sb, Se, Sn, Tl, V, and Zn. Standard solutions forelements not listed are also available.NOTE 1AAS standards are typically pr

21、esented in mass/vol units. Thedensity of these solutions can be considered as unity (that is, 1) thus theycan be considered as % mass/mass (m/m).7.5 1-bromonaphthalene, trichlorobenzene, iodobenzoicacid, triethyl phosphate and dithiodiglycol are the recom-mended standards for the elements Br, Cl, I,

22、 P and S,respectively.7.6 Low Molecular Weight Polyethylene Glycol (PEG 400,or equivalent) or Water is used for producing method blank.7.7 High-Density Polyethylene (HDPE) Wide-mouth,Round, Screw-Cap Bottles, 50 to 60 mL capacity.7.8 Mixing Balls, approximately 1 cm diameter, stainlesssteel or equiv

23、alent.NOTE 2Potential low level Cr, Fe or Ni (20 mg/kg1) contaminationdue to the use of stainless steel may exist. Other suitable materials wouldbe tungsten carbide, Zr or Ta.7.9 Thin-film Support.NOTE 3The user should select a thin-film support that provides formaximum transmittance and is resistan

24、t to typical components in LHW.The thin-film supports used in the development of this test method werea polypropylene base and a high-purity, 4 m polyester film.7.10 Sample Cups, vented.7.11 Helium, Heminimum 99.99 purity for use as achamber purge gas for the analysis of Cl, P and S. Thisnumerical p

25、urity is intended to specify a general grade ofhelium. Ultra-high purity helium is not required for this testmethod.8. Sample8.1 Because of the potential heterogeneous nature of LHW,all possible efforts should be made to ensure that representativesamples are taken.9. Preparation of Apparatus9.1 Foll

26、ow the manufacturers instructions for set-up,conditioning, preparation and maintenance of the XRF spec-trometer.9.2 When required, reference spectra should be obtainedfrom pure element standards for all deconvoluted elements.9.3 Spectral and matrix interferences as listed in the Inter-ferences secti

27、on must be addressed per the manufacturersrecommendations.10. Calibration and Standardization10.1 The spectrometer must be calibrated using an appro-priate reference element(s) at a minimum frequency as recom-mended by the manufacturer.10.2 Analytical standards should be prepared gravimetri-cally by

28、 blending the solution or pure element standards withAl2O3to suitable standard concentrations as determined by theusers analytical requirements. Table 1 gives recommendedconcentration ranges for regression. Standards can be single ormulti-element mixtures. Standard solutions are generallymixed with

29、Al2O3at a ratio of 3:1.NOTE 4More than one standard element(s) solution can be added to asingle 15 gAl2O3mass provided the total mass of standard is 5 g.This willmaintain the proper 3:1 ratio while allowing mixtures of potentiallyincompatible elements to be combined in a single standard.10.2.1 The n

30、umber of standards required to produce cali-brations is dependent on the number of elements to bedetermined. Generally, two calibrations are produced, the firstis to determine potentially major elements such as halogens, S5Reagent Chemicals, American Chemical Society Specifications, AmericanChemical

31、 Society, Washington, DC. For suggestions on the testing of reagents notlisted by the American Chemical Society, see Analar Standards for LaboratoryChemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeiaand National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockvill

32、e,MD.TABLE 1 Recommended Standards RangesAnalyteLow Con-centrationRange,mg/kgHigh Con-centrationRange,mg/kgAnalyteLow Con-centrationRange, mg/kgHigh Con-centrationRange,mg/kgAg 5 600 Zn 5 600Ba 5 600 As 5 600P 0.1 % 5 % Se 5 600S 0.05 % 5 % Br 10 5000Cl 0.05 % 5 % Cd 5 600K 0.1 % 5 % Sb 5 600V 5 600

33、 Sn 5 600Cr 5 600 I 5 600Fe 5 600 Hg 5 600Ni 5 600 Tl 5 600Cu 5 600 Pb 5 600D6052 97 (2016)2halides; liquid hazardous waste; spectrom-etry; spectroscopy; toxic metals; trace elements; waste-derivedfuel; XRFD6052 97 (2016)4APPENDIX(Nonmandatory Information)X1. BLANK, ACCURACY AND MATRIX SPIKE RESULTS

34、X1.1 Errors shown in Tables X1.1-X1.3 are taken from theresults output of the instrument and nominally 6 2 sigma.7These represent the total error attributed to spectrum process-ing and counting statistics. See Stratham12for details of thecalculations used in the error calculations.X1.2 Table X1.4 gi

35、ves the lower limit of detection for eachanalyte. This is based on the following equation:LLD 53x=bgnet peak31=T3concentration (X1.1)where:bg = background intensity under analyte peak cps,net peak = fitted peak intensity of analyte cps,T = count time, s, andconc. = concentration of analyte.X1.3 Matr

36、ix/Matrix Spike Recoveries:X1.3.1 A series of experiments was conducted to test theperformance of the alumina method. Three types of actualwaste solutions were selected from routine test samples takenat an incineration plant. These samples were:X1.3.1.1 Clear solutions,X1.3.1.2 Turbid solutions, tha

37、t is, contained significantsolids not in suspension, andX1.3.1.3 Biphasal solutions, that is, contained two distinctlyimmiscible liquid phases.X1.3.2 For each matrix type a sub-sample was spiked witha known concentration of analyte. The spiked sample wasprepared using the alumina technique and measu

38、red. Using EqX1.2, a recovery figure for each analyte in each matrix typewas determined. The results, referred to as a matrix spike/matrix spike duplicate (MS/MSD), are shown below.% MS/MSD recovery 2 C2 D1 3 C1!/C3! 3100(X1.2)where:D1 = dilution factor due to matrix spike addi-tion=1(mass of spike)

39、/(total mass ofsample + spike),C1 = calculation concentration of matrix without spike,C2 = calculation concentration of matrix + spike, andC3 = given concentration of matrix spike.X1.4 Precision of measurement:X1.4.1 A number of repeat measurements were made on awaste sample.X1.4.1.1 A single measur

40、ement from each of ten repeatsample preparations was made for Cl content. A repeat of thisprocess was made on newly prepared samples 48 h later.X1.4.1.2 A single measurement from each of ten repeatsample preparations was made for a waste sample spiked with54.3 mg kg1Cd.X1.4.2 A single Cl analysis of

41、 the same waste sample usedin X1.4.1.1 was measured by a second laboratory using thealumina sample preparation technique and is shown in the lastcolumn of Table X1.6.7Stratham, P., Analytical Chemistry, 1977, Vol 49, pp. 2149.TABLE X1.1 BlankNOTE 1Asample containing only PEG was used to check for an

42、y biasin the calibrations at the zero concentration level.Element V Cr Fe Ni Cu Zn As Se Cd Sn Sb l Hg Tl Pbmg kg11.2 0.0 17.3 0.0 0.0 1.6 1.9 2.6 0.0 1.3 0.0 3.9 1.9 2.2 0.0mg kg1error0.5 0.7 0.2 3.6 0.04 0.4 7.6 2.5 2.5 0.4 0.9 10 2.9 3.9 1.0D6052 97 (2016)5TABLE X1.2 AccuracyNOTE 1Standards were

43、run against the calibrations to assess accuracy and to check for bias in the calibrations due to either matrix or spectral effects.The % accuracy figure shows the match of given versus calculated concentration for each element.ElementP%m/mS%m/mCl%m/mSemg/kgAsmg/kgBrmg/kgCdmg/kgSnmg/kgSbmg/kglmg/kgCa

44、lculation concentration 0.96 2.09 0.33 93 393 4779 179 47 201 587Error 0.02 0.02 0.005 4 10 15 4 3 7 13Given concentration 1.024 2.095 0.598 100 400 4838 200 50 200 600% accuracy 96 100 89 93 98 99 98 94 100 98Element Hg Tl Pb V Cr Fe Ni Cu ZnCalculation concentration, mg kg199 96 94 109 6 589 53 55

45、 95Error 9 13 7 7 3 0.2 8 3 5Given concentration 100 100 100 100 10 600 50 50 100% accuracy 99 96 94 109 60 98 106 110 93TABLE X1.3 Light Elements and HalidesSingle Phase Solvent P S Cl Br IMS/MSD recovery (%) 111.3 120 110 95 100error, % m/m 0.02 0.01 0.007 0.0007 0.0032spike concentration, % m/m 0

46、.9156 0.926 1.279 0.0599 0.856Turbid WasteMS/MSD recovery, % 98 91 89 104 100error, % m/m 0.01 0.009 0.005 0.0006 0.0034spike concentration, % m/m 0.849 0.805 0.858 0.042 0.845Biphasal WasteMS/MSD recovery, % 98error, % m/m 0.008spike concentration, % m/m 1.654TABLE X1.4 3 Sigma Lower Limits of Dete

47、ctionElement LLD mg/kg Element LLD mg/kg Element LLD mg/kgPn/aACu 17 Se 6SnAZn 11 I 7Cl n/aAAs 5 Hg 7V8Se Tl4Cr 8 Br 5 Pb 4Fe 14 Cd 3Ni 16 Sn 5An a = these elements were calibrated at concentrations significantly higher thantheir respective detection limits.TABLE X1.5 Toxic ElementsBiphasal Waste Cd

48、 Hg Tl Se As Pb V Cr Fe Ni Cu Zn Sb SnMs/MSD recovery,% 108 103 98 109 109 99 96 111 84 86 73 105 84 103error mg kg13 9 104 9765 1 9 4 5 6 3spike concentration mgkg147.1 101.8 102.5 104.6 109.7 183.7 97.4 98.3 97.1 98.2 99.3 102.0 33.3 53.5Turbid Waste Cd Hg Tl Se As Pb V Cr Fe Ni Cu Zu Sb SnMS/MSD

49、recovery,% 92 114 43 90 121 108 102 110 104 118 94 105 108 103error mg kg13 11 13 5 10 10 6 6 2 11 5 5 6 3spike concentration mgkg144.8 96.8 47.7 93.6 94.8 177 103.4 102 106.4 104.5 103.5 106.6 53.2 53.2Single Phase Solvent Cd Hg Tl Se As Pb V Cr Fe Ni Cu Zn Sb SnMs/MSD recovery,% 100 98 77 108 108 errorA110 114 109 104 105 90 112 95error mg kg13 91049 662105563spike concentration mgkg145.5 97.0 96 103.8 105 116.8 109.5 100 105.7 109.9 105.3 54.6 55.5ASpike not added.D6052 97 (2016)6ASTM International takes no position respecting the validity o

展开阅读全文
相关资源
猜你喜欢
相关搜索

当前位置:首页 > 标准规范 > 国际标准 > ASTM

copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
备案/许可证编号:苏ICP备17064731号-1