ASTM D6348-2012 5000 Standard Test Method for Determination of Gaseous Compounds by Extractive Direct Interface Fourier Transform Infrared (FTIR) Spectroscopy《通过可提取的直接界面傅里叶转换红外度谱仪对.pdf

上传人:wealthynice100 文档编号:522245 上传时间:2018-12-03 格式:PDF 页数:15 大小:227.23KB
下载 相关 举报
ASTM D6348-2012 5000 Standard Test Method for Determination of Gaseous Compounds by Extractive Direct Interface Fourier Transform Infrared (FTIR) Spectroscopy《通过可提取的直接界面傅里叶转换红外度谱仪对.pdf_第1页
第1页 / 共15页
ASTM D6348-2012 5000 Standard Test Method for Determination of Gaseous Compounds by Extractive Direct Interface Fourier Transform Infrared (FTIR) Spectroscopy《通过可提取的直接界面傅里叶转换红外度谱仪对.pdf_第2页
第2页 / 共15页
ASTM D6348-2012 5000 Standard Test Method for Determination of Gaseous Compounds by Extractive Direct Interface Fourier Transform Infrared (FTIR) Spectroscopy《通过可提取的直接界面傅里叶转换红外度谱仪对.pdf_第3页
第3页 / 共15页
ASTM D6348-2012 5000 Standard Test Method for Determination of Gaseous Compounds by Extractive Direct Interface Fourier Transform Infrared (FTIR) Spectroscopy《通过可提取的直接界面傅里叶转换红外度谱仪对.pdf_第4页
第4页 / 共15页
ASTM D6348-2012 5000 Standard Test Method for Determination of Gaseous Compounds by Extractive Direct Interface Fourier Transform Infrared (FTIR) Spectroscopy《通过可提取的直接界面傅里叶转换红外度谱仪对.pdf_第5页
第5页 / 共15页
亲,该文档总共15页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、Designation: D6348 12Standard Test Method forDetermination of Gaseous Compounds by Extractive DirectInterface Fourier Transform Infrared (FTIR) Spectroscopy1This standard is issued under the fixed designation D6348; the number immediately following the designation indicates the year oforiginal adopt

2、ion or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.INTRODUCTIONThis extractive FTIR based field test method is used to quantify gas phas

3、e concentrations of multipletarget analytes from stationary source effluent. Because an FTIR analyzer is potentially capable ofanalyzing hundreds of compounds, this test method is not analyte or source specific. The analytes,detection levels, and data quality objectives are expected to change for an

4、y particular testing situation.It is the responsibility of the tester to define the target analytes, the associated detection limits for thoseanalytes in the particular source effluent, and the required data quality objectives for each specific testprogram. Provisions are included in this test metho

5、d that require the tester to determine criticalsampling system and instrument operational parameters, and for the conduct of QA/QC procedures.Testers following this test method will generate data that will allow an independent observer to verifythe valid collection, identification, and quantificatio

6、n of the subject target analytes.1. Scope1.1 This field test method employs an extractive samplingsystem to direct stationary source effluent to an FTIR spec-trometer for the identification and quantification of gaseouscompounds. Concentration results are provided. This testmethod is potentially app

7、licable for the determination ofcompounds that (1) have sufficient vapor pressure to betransported to the FTIR spectrometer and (2) absorb a sufficientamount of infrared radiation to be detected.1.2 This field test method provides near real time analysis ofextracted gas samples from stationary sourc

8、es. Gas streamswith high moisture content may require conditioning to mini-mize the excessive spectral absorption features imposed bywater vapor.1.3 This field test method requires the preparation of asource specific field test plan. The test plan must include thefollowing: (1) the identification of

9、 the specific target analytes(2) the known analytical interferents specific to the test facilitysource effluent (3) the test data quality necessary to meet thespecific test requirements and (4) the results obtained from thelaboratory testing (see Annex A1 for test plan requirements).1.4 The FTIR ins

10、trument range should be sufficient tomeasure from high ppm(v) to ppb(v) and may be extended tohigher or lower concentrations using any or all of the followingprocedures:1.4.1 The gas absorption cell path length may be eitherincreased or decreased,1.4.2 The sample conditioning system may be modified

11、toreduce the water vapor, CO2, and other interfering compoundsto levels that allow for quantification of the target com-pound(s), and1.4.3 The analytical algorithm may be modified such thatinterfering absorbance bands are minimized or stronger/weakerabsorbance bands are employed for the target analy

12、tes.1.5 The practical minimum detectable concentration is in-strument, compound, and interference specific (see Annex A2for procedures to estimate the achievable minimum detectableconcentrations (MDCs). The actual sensitivity of the FTIRmeasurement system for the individual target analytes dependsup

13、on the following:1.5.1 The specific infrared absorptivity (signal) and wave-length analysis region for each target analyte,1.5.2 The amount of instrument noise (see Annex A6), and1.5.3 The concentration of interfering compounds in thesample gas (in particular, percent moisture and CO2), and theamoun

14、t of spectral overlap imparted by these compounds in thewavelength region(s) used for the quantification of the targetanalytes.1This test method is under the jurisdiction of Committee D22 onAir Quality andis the direct responsibility of Subcommittee D22.03 on Ambient Atmospheres andSource Emissions.

15、Current edition approved Feb. 1, 2012. Published February 2012. Originallyapproved in 1998. Last previous edition approved in 2010 as D6348 - 03(2010).DOI: 10.1520/D6348-12.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.1.5.4 Any sa

16、mpling system interferences such as adsorptionor outgassing.1.6 Practices E168 and E1252 are suggested for additionalreading.1.7 This standard does not purport to address all of thesafety concerns associated with its use. It is the responsibilityof the user of this standard to establish appropriate

17、safety andhealth practices and to determine the applicability of regula-tory limitations prior to use. Additional safety precautions aredescribed in Section 9.2. Referenced Documents2.1 ASTM Standards:2D1356 Terminology Relating to Sampling and Analysis ofAtmospheresD3195 Practice for Rotameter Cali

18、brationE168 Practices for General Techniques of Infrared Quanti-tative AnalysisE1252 Practice for General Techniques for Obtaining Infra-red Spectra for Qualitative Analysis2.2 EPA Methods (40 CFR Part 60 Appendix A)3Method 1 Sample and Velocity Traverses for StationarySourcesMethod 2 Series Determi

19、nation of Stack Gas Velocity andVolumetric Flow Rate (Type S Pitot Tube)Method 3 Series GasAnalysis for Carbon Dioxide, Oxygen,Excess Air, and Dry Molecular WeightMethod 4 Series Determination of Moisture Content inStack Gases3. Terminology3.1 See Terminology D1356 for definition of terms relatedto

20、sampling and analysis of atmospheres.3.2 Definitions of Terms Specific to This Standard: Thissection contains the terms and definitions used in this testmethod and those that are relevant to extractive FTIR basedsampling and analysis of stationary source effluent. Whenpossible, definitions of terms

21、have been drawn from authori-tative texts or manuscripts in the fields of air pollutionmonitoring, spectroscopy, optics, and analytical chemistry.3.2.1 absorbance, nthe negative logarithm of the trans-mission, A = -log (I/I0), where I is the transmitted intensity ofthe light and I0is the incident in

22、tensity.3.2.2 absorptivity, adjthe amount of infrared radiationthat is absorbed by each molecule.3.2.3 analyte spiking, nthe process of quantitatively co-adding calibration standards with source effluent to determinethe effectiveness of the FTIR measurement system to quantifythe target analytes.3.2.

23、4 analytical algorithm, nthe method used to quantifythe concentration of the target analytes and interferences ineach FTIR Spectrum. The analytical algorithm should accountfor the analytical interferences by conducting the analysis in aportion of the infrared spectrum that is the most unique for tha

24、tparticular compound.3.2.5 analytical interference, nthe physical effects ofsuperimposing two or more light waves. Analytical interfer-ences occur when two or more compounds have overlappingabsorbance bands in their infrared spectra.3.2.6 apodization, va mathematical transformation car-ried out on d

25、ata received from an interferometer to reduce theside lobes of the measured peaks. This procedure alters theinstruments response function. There are various types oftransformation; the most common forms are boxcar, triangular,Happ-Genzel, and Beer-Norton functions.3.2.7 background spectrum, nthe spe

26、ctrum taken in theabsence of absorbing species or sample gas, typically con-ducted using dry nitrogen or zero air in the gas cell.3.2.8 bandwidth, adjthe width of a spectral feature asrecorded by a spectroscopic instrument. This width is listed asthe full width at the half maximum of the feature or

27、as the halfwidth at the half maximum of the spectral feature. This is alsoreferred to as the line width (1).43.2.9 beam splitter, na device located in the interferom-eter that splits the incoming infrared radiation into two separatebeams that travel two separate paths before recombination.3.2.10 Bee

28、rs law, nthe principal by which FTIR spectraare quantified. Beers law states that the intensity of a mono-chromatic plane wave incident on an absorbing medium ofconstant thickness diminishes exponentially with the numberof absorbers in the beam. Strictly speaking, Beers law holdsonly if the followin

29、g conditions are met: (1) perfectly mono-chromatic radiation (2) no scattering (3) a beam that is strictlycollimated (4) negligible pressure-broadening effects (2, 3).For an excellent discussion of the derivation of Beers law, see(4).3.2.11 calibration transfer standard, na certified calibra-tion st

30、andard that is used to verify the instrument stability on adaily basis when conducting sampling.3.2.12 classical least squares, na common method ofanalyzing multicomponent infrared spectra by scaled absor-bance subtraction.3.2.13 condenser system,(dryer), na moisture removalsystem that condenses wat

31、er vapor from the source effluent toprovide a dry sample to the FTIR gas cell. Part of the sampleconditioning system.3.2.14 cooler, na device into which a quantum detector isplaced for maintaining it at a low temperature in an IR system.At a low temperature, the detector provides the high sensitivit

32、ythat is required for the IR system. The two primary types ofcoolers are a liquid nitrogen Dewar and a closed-cycle Stirlingcycle refrigerator.3.2.15 electromagnetic spectrum, nthe total set of allpossible frequencies of electromagnetic radiation. Differentsources may emit over different frequency r

33、egions. All elec-tromagnetic waves travel at the same speed in free space (5).2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page on

34、the ASTM website.3Available from Superintendent of Documents, U. G. Government PrintingOffice, Washington, DC 20402.4The boldface numbers in parentheses refer to the list of references at the end ofthe standard.D6348 1223.2.16 extractive FTIR, na means of employing FTIR toquantify concentrations of

35、gaseous components in stationarysource effluent. It consists of directing gas samples to the FTIRcell without collection on sample media.3.2.17 fingerprint region, nthe region of the absorptionspectrum of a molecule that essentially allows its unequivocalidentification. For example, the organic fing

36、erprint regioncovers the wave number range from 650 to 1300 cm1(6).3.2.18 Fourier transform, va mathematical transform thatallows an aperiodic function to be expressed as an integral sumover a continuous range of frequencies (7). The interferogramrepresents the detector response (intensity) versus t

37、ime, theFourier transform function produces intensity as a function offrequency.3.2.19 frequency position, nthe accepted exact spectralline position for a specific analyte.Awave number or fractionalwavenumber is used to determine whether spectral shifts haveoccurred with time.3.2.20 FTIR, nan abbrev

38、iation for Fourier transforminfrared. A spectroscopic instrument using the infrared portionof the electromagnetic spectrum. The working component ofthis system is an interferometer. To obtain the absorptionspectrum as a function of frequency, a Fourier transform of theoutput of the interferometer mu

39、st be performed. For anin-depth description of the FTIR, see (8).3.2.21 fundamental CTS, na NIST traceable referencespectrum with known temperature and pressure, that has beenrecorded with an absorption cell that has been measured usingeither a laser or other suitably accurate physical measurementde

40、vice.3.2.22 infrared spectrum, nthat portion of the electromag-netic spectrum that spans the region from about 10 cm1toabout 12 500 cm1. It is divided (6) into (1) the near-infraredregion (from 12 500 to 4000 cm1), (2) the mid-infrared region(from 4000 to 650 cm1), and (3) the far-infrared region (f

41、rom650 to 10 cm1).3.2.23 instrument function, nthe function superimposedon the actual absorption line shape by the instrument. This issometimes referred to as the slit function; a term taken frominstruments that use slits to obtain resolution.3.2.24 instrument specific reference spectra, nreferences

42、pectra collected on the instrument that collects the actualsample spectra. The instrument specific reference spectra areused in the analytical algorithm.3.2.25 intensity, nthe radiant power per unit solid angle.When the term spectral intensity is used, the units are watts persteradian per nanometre.

43、 In most spectroscopic literature, theterm intensity is used to describe the power in a collimatedbeam of light in terms of power per unit area per unitwavelength. However, in the general literature, this definitionis more often used for the term irradiance,ornormal irradi-ance (9, 10).3.2.26 interf

44、erogram, nthe effects of interference that aredetected and recorded by an interferometer, the output of theFTIR and the primary data are collected and stored (8, 10).3.2.27 interferometer, nany of several kinds of instru-ments used to produce interference effects. The Michelsoninterferometer used in

45、 FTIR instruments is the most famous ofa class of interferometers that produce interference by thedivision of amplitude (11).3.2.28 irradiance, nradiant power per unit projected areaof a specified surface. This has units of watts per squarecentimetre. The term spectral irradiance is used to describe

46、 theirradiance as a function of wavelength. It has units of watts persquare centimetre per nanometre (9).3.2.29 laser, nan acronym for the term light amplificationby stimulated emission of radiation. A source of light that ishighly coherent, both spatially and temporally (1).3.2.30 light, nstrictly,

47、 light is defined as that portion of theelectromagnetic spectrum that causes the sensation of vision. Itextends from about 25 000 cm1to about 14 300 cm1(5).3.2.31 minimum detectable concentration, nthe minimumconcentration of a compound that can be detected by aninstrument with a given statistical p

48、robability. Usually thedetection limit is given as three times the standard deviation ofthe noise in the system. In this case, the minimum concentra-tion can be detected with a probability of 99.7 % (9, 12). SeeAnnexA2 of this standard for a series of procedures to measureMDC.3.2.32 native effluent

49、concentration, nthe underlying ef-fluent concentration of the target analytes.3.2.33 noise equivalent absorbance (NEA), nthe peak-to-peak noise in the spectrum resulting from the acquisition oftwo successive background spectra.3.2.34 path length, nthe distance that the sample gasinteracts with the infrared radiation.3.2.35 peak-to-peak noise, nthe absolute difference fromthe highest positive peak to the lowest negative peak in adefined spectral region.3.2.36 primary particulate matter filter, nfilter of 0.3microns or less to remove par