ASTM D5293-2017a Standard Test Method for Apparent Viscosity of Engine Oils and Base Stocks Between –10 C and –35 C Using Cold-Cranking Simulator.pdf

上传人:李朗 文档编号:522697 上传时间:2018-12-04 格式:PDF 页数:12 大小:459.56KB
下载 相关 举报
ASTM D5293-2017a Standard Test Method for Apparent Viscosity of Engine Oils and Base Stocks Between –10 C and –35 C Using Cold-Cranking Simulator.pdf_第1页
第1页 / 共12页
ASTM D5293-2017a Standard Test Method for Apparent Viscosity of Engine Oils and Base Stocks Between –10 C and –35 C Using Cold-Cranking Simulator.pdf_第2页
第2页 / 共12页
ASTM D5293-2017a Standard Test Method for Apparent Viscosity of Engine Oils and Base Stocks Between –10 C and –35 C Using Cold-Cranking Simulator.pdf_第3页
第3页 / 共12页
ASTM D5293-2017a Standard Test Method for Apparent Viscosity of Engine Oils and Base Stocks Between –10 C and –35 C Using Cold-Cranking Simulator.pdf_第4页
第4页 / 共12页
ASTM D5293-2017a Standard Test Method for Apparent Viscosity of Engine Oils and Base Stocks Between –10 C and –35 C Using Cold-Cranking Simulator.pdf_第5页
第5页 / 共12页
亲,该文档总共12页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、Designation: D5293 17aStandard Test Method forApparent Viscosity of Engine Oils and Base StocksBetween 10 C and 35 C Using Cold-Cranking Simulator1This standard is issued under the fixed designation D5293; the number immediately following the designation indicates the year oforiginal adoption or, in

2、 the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method covers the laboratory determination ofapparent viscosity of engin

3、e oils and base stocks by coldcranking simulator (CCS) at temperatures between 10 C and35 C at shear stresses of approximately 50 000 Pa to100 000 Pa and shear rates of approximately 105to 104s1forviscosities of approximately 900 mPas to 25 000 mPas. Therange of an instrument is dependent on the ins

4、trument modeland software version installed. Apparent Cranking Viscosityresults by this method are related to engine-cranking charac-teristics of engine oils.1.2 A special procedure is provided for measurement ofhighly viscoelastic oils in manual instruments. See AppendixX2.1.3 Procedures are provid

5、ed for both manual and automateddetermination of the apparent viscosity of engine oils using thecold-cranking simulator.1.4 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.5 This standard does not purport to address all of the

6、safety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety, health, and environmental practices and deter-mine the applicability of regulatory limitations prior to use.Specific warning statements are given in Section 8.1.6

7、This international standard was developed in accor-dance with internationally recognized principles on standard-ization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recom-mendations issued by the World Trade Organization TechnicalBarriers to Tra

8、de (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:2D2162 Practice for Basic Calibration of Master Viscometersand Viscosity Oil StandardsD2602 Test Method for Apparent Viscosity of Engine OilsAt Low Temperature Using the Cold-Cranking Simulator(Withdrawn 1993)3D4057 Practice for Manual Sam

9、pling of Petroleum andPetroleum Products2.2 ISO Standard:ISO 17025 General Requirements for the Competence ofTesting and Calibration Laboratories43. Terminology3.1 Definitions:3.1.1 Newtonian oil or fluid, none that exhibits a constantviscosity at all shear rates.3.1.2 non-Newtonian oil or fluid, no

10、ne that exhibits aviscosity that varies with changing shear stress or shear rate.3.1.3 viscosity, ,nthe property of a fluid that determinesits internal resistance to flow under stress, expressed by: 5(1)where: = the stress per unit area, and = the rate of shear.3.1.3.1 DiscussionIt is sometimes call

11、ed the coefficient ofdynamic viscosity. This coefficient is thus a measure of theresistance to flow of the liquid. In the SI, the unit of viscosityis the pascal-second; for practical use, a submultiple(millipascal-second) is more convenient and is customarilyused. The millipascal second is 1 cP (cen

12、tipoise).1This test method is under the jurisdiction of ASTM Committee D02 onPetroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility ofSubcommittee D02.07 on Flow Properties.Current edition approved Oct. 1, 2017. Published October 2017. Originallyapproved in 1991. Last prev

13、ious edition approved in 2017 as D5293 171. DOI:10.1520/D5293-17A.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM web

14、site.3The last approved version of this historical standard is referenced onwww.astm.org.4Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036, http:/www.ansi.org.*A Summary of Changes section appears at the end of this standardCopyright ASTM Inte

15、rnational, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United StatesThis international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for theDevelopment of International Standards,

16、 Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.13.2 Definitions of Terms Specific to This Standard:3.2.1 apparent viscosity, nthe viscosity obtained by use ofthis test method.3.2.1.1 DiscussionSince many engine oils are non-Newtonian at

17、 low temperature, apparent viscosity can varywith shear rate.3.2.2 calibration oils, noils with known viscosity andviscosity/temperature functionality that are used to define thecalibration relationship between viscosity and cold-crankingsimulator rotor speed.3.2.3 check oil, na batch of test oil us

18、ed to monitormeasurement performance.3.2.4 test oil, nany oil for which the apparent viscosity isto be determined by use of this test method.3.2.5 viscoelastic oil, na non-Newtonian oil or fluid thatclimbs up the rotor shaft during rotation.4. Summary of Test Method4.1 An electric motor drives a rot

19、or that is closely fittedinside a stator. The space between the rotor and stator is filledwith oil. Test temperature is measured near the stator inner walland maintained by removing heat with a controlled process tomaintain a constant stator temperature during test. The speed ofthe rotor is calibrat

20、ed as a function of viscosity. Test oilviscosity is determined from this calibration and the measuredrotor speed.5. Significance and Use5.1 The CCS apparent viscosity of automotive engine oilscorrelates with low temperature engine cranking. CCS appar-ent viscosity is not suitable for predicting low

21、temperature flowto the engine oil pump and oil distribution system. Enginecranking data were measured by the Coordinating ResearchCouncil (CRC) L-495test with reference oils that had viscosi-ties between 600 mPas and 8400 mPas (cP) at 17.8 C andbetween 2000 mPas and 20 000 mPas (cP) at 28.9 C. Thede

22、tailed relationship between this engine cranking data andCCS apparent viscosities is in Appendixes X1 and X2 of the1967 T edition of Test Method D26026and CRC Report 409.5Because the CRC L-49 test is much less precise and standard-ized than the CCS procedures, CCS apparent viscosity need notaccurate

23、ly predict the engine cranking behavior of an oil in aspecific engine. However, the correlation of CCS apparentviscosity with average CRC L-49 engine cranking results issatisfactory.5.2 The correlation between CCS and apparent viscosityand engine cranking was confirmed at temperatures between1 C and

24、 40 C by work on 17 commercial engine oils (SAEgrades 5W, 10W, 15W, and 20W). Both synthetic and mineraloil based products were evaluated. See ASTM STP 621.75.3 A correlation was established in a low temperatureengine performance study between light duty engine startabil-ity and CCS measured apparen

25、t viscosity. This study used ten1990s engines at temperatures ranging from 5 C down to40 C with six commercial engine oils (SAE 0W, 5W, 10W,15W, 20W, and 25W).85.4 The measurement of the cranking viscosity of basestocks is typically done to determine their suitability for use inengine oil formulatio

26、ns.Asignificant number of the calibrationoils for this method are base stocks that could be used in engineoil formulations.6. Apparatus6.1 Two types of apparatus are described for use in this testmethod: the manual cold-cranking simulator (see AppendixX1) and the automated CCS (see 6.2 and 6.3).6.2

27、Automated CCS,9consisting of a direct current (dc)electric motor that drives a rotor inside a stator; a rotor speedsensor or tachometer that measures rotor speed; a dc ammeterand fine current-control adjust dial; a stator temperature controlsystem that maintains temperature within 0.05 C of set poin

28、t;and a heat removal system with a temperature control system,a computer, computer interface, and test sample injectionpump.6.3 Automatic Automated CCS,9as described in 6.2 with theaddition of an automated sample table allowing multiple testsamples to be run sequentially under computer control witho

29、utoperator attention.6.4 Calibrated Thermistor,9sensor for insertion in a wellnear the inside surface of the stator to indicate the testtemperature.6.4.1 There must be good thermal contact between thetemperature sensor and the thermal well in the stator; clean thisthermal well periodically and repla

30、ce the small drop ofhigh-silver-containing heat transfer medium.6.5 Heat Removal System:6.5.1 For stators with coolant contact, a refrigerator for theliquid coolant is needed to maintain coolant temperature atleast 10 C below the test temperature. When the coolanttemperature is below 30 C a two-stag

31、e refrigeration systemis likely needed. The length of the tubing connections betweenthe CCS and the refrigerator should be as short as possible (lessthan 1 m) and well insulated.5CRC Report No. 409 “Evaluation of Laboratory Viscometers for PredictingCranking Characteristics of Engine Oils at -0F and

32、 -20F,” April 1968 availablefrom the Coordinating Research Council, 5755 North Point Pkwy, Suite 265,Alpharetta, GA 30022.6Supporting data have been filed at ASTM International Headquarters and maybe obtained by requesting Research Report RR:D02-1402. ContactASTM CustomerService at serviceastm.org.7

33、Stewart, R. M., “Engine Pumpability and Crankability Tests on Commercial“W” Grade Engine Oils Compared to Bench Test Results,” ASTM STP 621 ASTM1967, 1968. 1969 Annual Book of ASTM Standards , Part 17 (Also published as SAEPaper 780369 in SAE Publication SP-429.).8Supporting data have been filed at

34、ASTM International Headquarters and maybe obtained by requesting Research Report RR:D02-1442. ContactASTM CustomerService at serviceastm.org.9The sole source of supply of the apparatus known to the committee at this timeis Cannon Instrument Co., State College, PA 16804. Website: www.cannoninstru-. I

35、f you are aware of alternative suppliers, please provide this informationto ASTM International Headquarters. Your comments will receive careful consid-eration at a meeting of the responsible technical committee,1which you may attend.D5293 17a26.5.1.1 Coolant, Dry MethanolIf contaminated with waterfr

36、om operating under high humidity conditions, replace it withdry methanol to ensure consistent temperature control.6.5.2 For thermoelectric cooled stators, the liquid coolingtemperature of the water or other appropriate liquid used in therefrigeration system (chiller) should be set to approximately5

37、C in order to maintain the sample test temperature. Toprevent ice formation that can block the coolant flow path, it issuggested that a solution be used which contains approximately10 % glycol or that which conforms to the manufacturersguidelines.6.6 Ultrasonic Bath, Unheated(optional)with an oper-a

38、ting frequency between 25 kHz to 60 kHz and a typical poweroutput of 100 W, of suitable dimensions to hold container(s)placed inside of bath, for use in effectively dissipating andremoving air or gas bubbles that can be entrained in viscoussample types prior to analysis. It is permissible to use ult

39、ra-sonic baths with operating frequencies and power outputsoutside this range, however it is the responsibility of thelaboratory to conduct a data comparison study to confirm thatresults determined with and without the use of such ultrasonicbaths does not materially impact results.7. Reagents and Ma

40、terials7.1 Calibration OilsLow-cloud point Newtonian oils shallbe certified by a laboratory that has been shown to meet therequirements of ISO 17025 by independent assessment. Thecalibration oils shall be traceable to master viscometer proce-dures described in Test Method D2162. Table 1 shows the se

41、tsof possible test oils to be used for each test temperature.Approximate viscosities at certain temperatures are listed inAppendix X5, whereas exact viscosities are supplied with eachstandard.8. Hazards8.1 Observe both toxicity and flammability warnings thatapply to the use of methanol or glycol.8.2

42、 If methanol is leaking from the apparatus, repair the leakbefore continuing the test.TABLE 1 Calibration Oil Sets by Test TemperatureTest Temp 35 C 30 C 25 C 20 C 15 C 10 CCL080 A . . . . .CL090 A . . . . .CL100 A A . . . .CL110 B A . . . .CL120 B A A . . .CL130 B B A . . .CL140 B B A A . .CL150 B

43、B B A . .CL160 B B B A . .CL170 B B B B A .CL190 B B B B A .CL200 B B BBAACL220 C B B B B ACL240 C B BBBACL250 C B BBBBCL260 . BCL280 . CCL300 . C BBBBCL320 . C C B B BCL340 . . C B B BCL380 . . C B B BCL420 . . . C B BCL480 . . . C B BCL530 . . . C C BCL600 . . . . C CCL680 . . . . C CGroup A Inclu

44、de at least one Preferred (bold)or one Alternate.Nominal Values35 C to 25 C; 800 mPas to 1500 mPas20 C to 10 C; 800 mPas to 1400 mPasGroup B Include at least 3. The selection is to be uniformly distributed over the range.Nominal Values35 C to 20 C; 1000 mPas to 15 000 mPas15 C; 1000 mPas to 13 000 m

45、Pas10 C; 1000 mPas to 9000 mPasGroup C Include at least one.Nominal Values35 C to 20 C; 13500 mPas15 C; 11 500 mPas10 C; 9000 mPasD5293 17a39. Sampling9.1 To obtain valid results, use an appropriate means of bulksampling (see Practice D4057) to obtain a representativesample of test oil free from sus

46、pended solid material and water.When the sample in its container is received below the dewpoint temperature of the room, allow the sample to warm toroom temperature before opening its container. When thesample contains suspended solid material, use centrifuge toremove particles greater than 5 m in s

47、ize and decant off thesupernate. Filtering is not recommended. DO NOT shake thesample of test oil. This leads to entrainment of air, and a falseviscosity reading.9.2 For some sample types, such as viscous lube oils that areprone to having entrained air or gas bubbles present in thesample, the use of

48、 an ultrasonic bath (see 6.6) without theheater turned on (if so equipped), has been found effective indissipating bubbles typically within 5 min.10. Calibration10.1 On installation of a new instrument or when any part ofthe viscometric cell or drive component (motor, belt, and soforth) is replaced,

49、 set the motor current as described below.Recheck the motor current (as described in 10.3) monthly untilthe change in motor current in consecutive months is less than0.005 A and every three months thereafter.NOTE 1See Appendix X4 for a flowchart for calibration.10.2 Temperature VerificationUsing the temperature veri-fication plugs, verify that the instrument is accurately comput-ing the correct temperature. (Only available on newer modelinstruments.)10.2.1 Unplug

展开阅读全文
相关资源
猜你喜欢
相关搜索

当前位置:首页 > 标准规范 > 国际标准 > ASTM

copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
备案/许可证编号:苏ICP备17064731号-1