ASTM D7042-2004 Standard Test Method for Dynamic Viscosity and Density of Liquids by Stabinger Viscometer (and the Calculation of Kinematic Viscosity)《使用Stabinger粘度计测定液体的动态粘度和密度的标准.pdf

上传人:eveningprove235 文档编号:524343 上传时间:2018-12-04 格式:PDF 页数:7 大小:122.33KB
下载 相关 举报
ASTM D7042-2004 Standard Test Method for Dynamic Viscosity and Density of Liquids by Stabinger Viscometer (and the Calculation of Kinematic Viscosity)《使用Stabinger粘度计测定液体的动态粘度和密度的标准.pdf_第1页
第1页 / 共7页
ASTM D7042-2004 Standard Test Method for Dynamic Viscosity and Density of Liquids by Stabinger Viscometer (and the Calculation of Kinematic Viscosity)《使用Stabinger粘度计测定液体的动态粘度和密度的标准.pdf_第2页
第2页 / 共7页
ASTM D7042-2004 Standard Test Method for Dynamic Viscosity and Density of Liquids by Stabinger Viscometer (and the Calculation of Kinematic Viscosity)《使用Stabinger粘度计测定液体的动态粘度和密度的标准.pdf_第3页
第3页 / 共7页
ASTM D7042-2004 Standard Test Method for Dynamic Viscosity and Density of Liquids by Stabinger Viscometer (and the Calculation of Kinematic Viscosity)《使用Stabinger粘度计测定液体的动态粘度和密度的标准.pdf_第4页
第4页 / 共7页
ASTM D7042-2004 Standard Test Method for Dynamic Viscosity and Density of Liquids by Stabinger Viscometer (and the Calculation of Kinematic Viscosity)《使用Stabinger粘度计测定液体的动态粘度和密度的标准.pdf_第5页
第5页 / 共7页
亲,该文档总共7页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、Designation: D 7042 04An American National StandardStandard Test Method forDynamic Viscosity and Density of Liquids by StabingerViscometer (and the Calculation of Kinematic Viscosity)1This standard is issued under the fixed designation D 7042; the number immediately following the designation indicat

2、es the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method specifies a procedure for t

3、he concurrentmeasurement of both the dynamic viscosity, h, and the density,r, of liquid petroleum products and crude oils, both transparentand opaque. The kinematic viscosity, n, can be obtained bydividing the dynamic viscosity, h, by the density, r, obtained atthe same test temperature.1.2 The resu

4、lt obtained from this test method is dependentupon the behavior of the sample and is intended for applicationto liquids for which primarily the shear stress and shear rate areproportional (Newtonian flow behavior).1.3 While the precision has only been determined for baseoils in the viscosity range f

5、rom 2.05 to 456 mPas at 40C andfrom 0.83 to 31.6 mPas at 100C and in the density range from0.82 to 0.92 g/mL at 15C (see 15.4), the test method can beapplied to a wider range of materials, viscosity, density, andtemperature. For materials not listed in Precision and Bias(Section 15), the precision a

6、nd bias may not be applicable.1.4 The values stated in SI units are to be regarded as thestandard. No other units of measurement are included in thisstandard. The accepted units of measure for density are gramsper millilitre (g/mL) or kilograms per cubic metre (kg/m3).1.5 This standard does not purp

7、ort to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and to determine theapplicability of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D

8、445 Test Method for Kinematic Viscosity of Transparentand Opaque Liquids (the Calculation of Dynamic Viscos-ity)D 2162 Test Method for Basic Calibration of Master Vis-cometers and Viscosity Oil StandardsD 2270 Practice for Calculating Viscosity Index From Ki-nematic Viscosity at 40 and 100CD 4052 Te

9、st Method for Density and Relative Density ofLiquids by Digital Density MeterD 6299 Practice for Applying Statistical Quality AssuranceTechniques to Evaluate Analytical Measurement SystemPerformanceD 6300 Practice for Determination of Precision and BiasData for Use in Test Methods for Petroleum Prod

10、ucts andLubricantsD 6617 Practice for Laboratory Bias Detection UsingSingle Test Result from Standard MaterialD 6708 Practice for Statistical Assessment and Improve-ment of the Expected Agreement Between Two TestMethods that Purport the Same Property of a Material2.2 ISO Standards:3ISO 5725 Accuracy

11、 (trueness and precision) of measure-ment methods and resultsISO/IEC 17025 General Requirements for the Competenceof Testing and Calibration Laboratories2.3 Other Documents:4NIST Technical Note 1297, Guideline for Evaluating andExpressing the Uncertainty of NIST Measurement Results3. Terminology3.1

12、Definitions:3.1.1 dynamic viscosity (h), nthe ratio between the ap-plied shear stress and rate of shear of a liquid.3.1.1.1 DiscussionIt is sometimes called the coefficient ofdynamic viscosity or, simply, viscosity. Thus, dynamic viscos-ity is a measure of the resistance to flow or to deformation of

13、a liquid under external shear forces.3.1.1.2 DiscussionThe term dynamic viscosity can alsobe used in a different context to denote a frequency-dependentquantity in which shear stress and shear rate have a sinusoidaltime dependence.3.1.2 kinematic viscosity (n), nthe ratio of the dynamicviscosity (h)

14、 to the density (r) of a liquid.1This test method is under the jurisdiction of ASTM Committee D02 onPetroleum Products and Lubricants and is the direct responsibility of SubcommitteeD02.07 on Flow Properties.Current edition approved July 1, 2004. Published July 2004.2For referenced ASTM standards, v

15、isit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New Yo

16、rk, NY 10036.4Available from National Institute of Standards and Technology (NIST), 100Bureau Dr., Stop 3460, Gaithersburg, MD 20899-3460.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.3.1.2.1 DiscussionFor gravity flow under a give

17、n hydro-static head, the pressure head of a liquid is proportional to itsdensity (r). Therefore the kinematic viscosity (n) is a measureof the resistance to flow of a liquid under gravity.3.1.3 density (r), nmass per unit volume.3.1.4 Relative density (also called specific gravity (SG),nthe ratio of

18、 the density of a material at a stated temperatureto the density of a reference material (usually water) at a statedtemperature.4. Summary of Test Method4.1 The test specimen is introduced into the measuring cells,which are at a closely controlled and known temperature. Themeasuring cells consist of

19、 a pair of rotating concentric cylin-ders and an oscillating U-tube. The dynamic viscosity isdetermined from the equilibrium rotational speed of the innercylinder under the influence of the shear stress of the testspecimen and an eddy current brake in conjunction withadjustment data. The density is

20、determined by the oscillationfrequency of the U-tube in conjunction with adjustment data.The kinematic viscosity is calculated by dividing the dynamicviscosity by the density.5. Significance and Use5.1 Many petroleum products, and some non-petroleummaterials, are used as lubricants and the correct o

21、peration ofthe equipment depends upon the appropriate viscosity of theliquid being used. In addition, the viscosity of many petroleumfuels is important for the estimation of optimum storage,handling, and operational conditions. Thus, the accurate deter-mination of viscosity is essential to many prod

22、uct specifica-tions.5.2 Density is a fundamental physical property that can beused in conjunction with other properties to characterize boththe light and heavy fractions of petroleum and petroleumproducts.5.3 Determination of the density or relative density ofpetroleum and its products is necessary

23、for the conversion ofmeasured volumes to volumes at the standard temperature of15C.6. Apparatus6.1 Stabinger Viscometer5,66.1.1 Viscosity measurementThe Stabinger viscometeruses a rotational coaxial cylinder measuring system. The outercylinder (tube) is driven by a motor at a constant and knownrotat

24、ional speed. The low-density inner cylinder (rotor) is heldin the axis of rotation by the centrifugal forces of the higherdensity sample and in its longitudinal position by the magnetand the soft iron ring. Consequently, the system works free ofbearing friction as found in rotational viscometers. A

25、perma-nent magnet in the inner cylinder induces eddy currents in thesurrounding copper casing. The rotational speed of the innercylinder establishes itself as the result of the equilibriumbetween the driving torque of the viscous forces and theretarding eddy current torque. This rotational speed is

26、mea-sured by an electronic system (Hall effect sensor) by countingthe frequency of the rotating magnetic field (see Fig. 1 and Fig.2, No. 2).6.1.2 Density MeasurementThe digital density analyzeruses a U-shaped oscillating sample tube and a system forelectronic excitation and frequency counting (see

27、Fig. 2, No.3).6.1.3 Temperature ControlThe copper block surroundsboth the viscosity and the density measuring cell in a way thatboth cells are held at the same temperature. A thermoelectricheating and cooling system (see Fig. 2, No. 1) ensures thetemperature stability of the copper block within 60.0

28、05Cfrom the set temperature at the position of the viscosity cellover the whole temperature range. The uncertainty (k = 2; 95 %confidence level) of the temperature calibration shall be nomore than 60.03C over the range from 15 to 100C. Outsidethis range the calibration uncertainty shall be no more t

29、han60.05C.6.1.4 The thermal equilibration time depends on the heatcapacity and conductivity of the liquid and on the differencebetween injection temperature and test temperature. Adequatetemperature equilibration of the test specimen is automaticallydetermined when successive viscosity values are co

30、nstantwithin 60.07 % over 1 min and successive density values areconstant within 60.00003 g/mL over 1 min.6.2 Syringes, commercially available, at least 5 mL involume, with a Luer tip.All construction materials for syringesshall be fully compatible with all sample liquids and cleaningagents, which c

31、ontact them.6.3 Flow-Through or Pressure Adapter, for use as analternative means of introducing the test specimen into themeasuring cells either by pressure or by suction, provided thatsufficient care and control is used to avoid any bubbleformation in the test specimen. All construction materials f

32、oradaptors shall be fully compatible with all sample liquids andcleaning agents, which contact them.5The Stabinger viscometer is covered by a patent. Interested parties are invitedto submit information regarding the identification of an alternative to this patenteditem to the ASTM International head

33、quarters. Your comments will receive carefulconsideration at a meeting of the responsible technical committee, which you mayattend.6The sole source of supply of the apparatus known to the committee at this timeis Anton Paar GmbH, Anton-Paar-Str. 20, A-8054 Graz, Austria. If you are aware ofalternati

34、ve suppliers, please provide this information to ASTM Internationalheadquarters. Your comments will receive careful consideration at a meeting of theresponsible technical committee, which you may attend. FIG. 1 Viscosity CellD70420426.4 Screen, with an aperture of 75 m, to remove particlesfrom the s

35、ample.6.5 Magnet, strong enough to remove iron fillings from thesample. Magnetic stirring rods are suitable.7. Reagents and Materials7.1 Sample Solvent, completely miscible with the sample.7.1.1 For most samples, a volatile petroleum spirit ornaphtha is suitable. If the solvent dries up without resi

36、dues inan applicable time frame, the use of a separate drying solventis not required.7.1.2 For residual fuels, a prewash with an aromatic solventsuch as toluene or xylene may be necessary to remove asphalticmaterial.7.2 Drying Solvent, a volatile solvent miscible with thesample solvent (see 7.1).7.2

37、.1 Highly concentrated ethanol (96 % or higher) is suit-able.7.3 Dry Air or Nitrogen, for blowing the measuring cells.7.3.1 If the measuring cell temperature is below or near thedew point temperature of the ambient air, the use of anappropriate desiccator is required.8. Sampling, Test Specimens, and

38、 Test Units8.1 Sampling is defined as all the steps required to obtain analiquot of the contents of any pipe, tank, or other system, andto place the sample into the laboratory test container. Thelaboratory test container and sample volume shall be ofsufficient capacity to mix the sample and obtain a

39、 homoge-neous sample for analysis.8.2 ParticlesFor samples that are likely to contain par-ticles (for example, used oils or crude oils) pass the samplethrough a 75-m screen to remove the particles. For theremoval of iron filings the use of a magnet is appropriate. Waxysamples must be heated to disso

40、lve the wax crystals prior tofiltration and a preheated filter shall be used.8.3 Test SpecimenAportion or volume of sample obtainedfrom the laboratory sample and delivered to the measuringcells. The test specimen is obtained as follows:8.3.1 Mix the sample, if required, to homogenize. Mixing atroom

41、temperature in an open container can result in the loss ofvolatile material; mixing in closed, pressurized containers, or atsub-ambient temperatures is recommended.8.3.2 Draw the test specimen from a properly mixed labo-ratory sample using an appropriate syringe.Alternatively, if theproper attachmen

42、ts and connecting tubes are used, the testspecimen can be delivered directly to the measuring cells fromthe mixing container. For waxy or other samples with a highpour point, before drawing the test specimen, heat the labora-tory sample to the desired test temperature, which has to behigh enough to

43、dissolve the wax crystals.9. Calibration and Verification9.1 Use only a calibrated apparatus as described in 6.1. Thecalibration shall be checked periodically using certified refer-ence standards as described in 9.2 and 9.3. The recommendedinterval for viscosity and density calibration is once a mon

44、th,for temperature control once a year. For the calibrationprocedure follow the instructions of the manufacturer of theapparatus.9.2 Certified Viscosity and Density Reference StandardsThese are for use as confirmatory checks on the procedure inthe laboratory. Certified viscosity and density referenc

45、e stan-dards shall be certified by a laboratory, which has shown tomeet the requirements of ISO/IEC 17025 or a correspondingnational standard by independent assessment. Viscosity stan-dards shall be traceable to master viscometer proceduresdescribed in Test Method D 2162. Density standards shall hav

46、ea certified uncertainty of the density values of 0.0001 g/mL.The uncertainty of the reference standards shall be stated foreach certified value (k = 2; 95 % confidence level). SeeISO 5725 or NIST 1297.9.3 ThermometerFor calibration and adjustment of thetemperature control, a digital thermometer wit

47、h a probe diam-eter of 6.25 mm and a maximal length of 80 mm shall be used.For smaller probes the use of an adapter is suitable. Theuncertainty (k = 2; 95 % confidence level) of this thermometermust be no more than 60.01C and has to be certified by alaboratory which has shown to meet the requirement

48、s ofISO/IEC 17025 or a corresponding national standard byindependent assessment. A suitable thermometer is availablefrom the manufacturer of the apparatus.9.4 Acceptable ToleranceIf the determined values of acalibration check measurement do not agree within the accept-able tolerance band of the cert

49、ified values, as calculated fromAnnex A1, re-check each step in the procedure, including thespecial cleaning procedure from 12.2, to locate the source oferror.NOTE 1Values exceeding the acceptable tolerance are generallyattributable to deposits in the measuring cells that are not removed by theroutine flushing procedure.10. Adjustment10.1 An adjustment has to be carried out when repeatedcalibration check measurements do not agree with the Accept-able Tolerance Band as stated in 9.4 and the error cannot belocated elsewhere. For the adjustment procedure foll

展开阅读全文
相关资源
猜你喜欢
相关搜索

当前位置:首页 > 标准规范 > 国际标准 > ASTM

copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
备案/许可证编号:苏ICP备17064731号-1