ASTM D7304-2006 Standard Test Method for Determination of Denatonium Ion in Engine Coolant by HPLC《用高效液相色谱法(HPLC)测定发动机冷却液中苯酸苄铵酰胺离子的标准试验方法》.pdf

上传人:twoload295 文档编号:525085 上传时间:2018-12-04 格式:PDF 页数:5 大小:112.75KB
下载 相关 举报
ASTM D7304-2006 Standard Test Method for Determination of Denatonium Ion in Engine Coolant by HPLC《用高效液相色谱法(HPLC)测定发动机冷却液中苯酸苄铵酰胺离子的标准试验方法》.pdf_第1页
第1页 / 共5页
ASTM D7304-2006 Standard Test Method for Determination of Denatonium Ion in Engine Coolant by HPLC《用高效液相色谱法(HPLC)测定发动机冷却液中苯酸苄铵酰胺离子的标准试验方法》.pdf_第2页
第2页 / 共5页
ASTM D7304-2006 Standard Test Method for Determination of Denatonium Ion in Engine Coolant by HPLC《用高效液相色谱法(HPLC)测定发动机冷却液中苯酸苄铵酰胺离子的标准试验方法》.pdf_第3页
第3页 / 共5页
ASTM D7304-2006 Standard Test Method for Determination of Denatonium Ion in Engine Coolant by HPLC《用高效液相色谱法(HPLC)测定发动机冷却液中苯酸苄铵酰胺离子的标准试验方法》.pdf_第4页
第4页 / 共5页
ASTM D7304-2006 Standard Test Method for Determination of Denatonium Ion in Engine Coolant by HPLC《用高效液相色谱法(HPLC)测定发动机冷却液中苯酸苄铵酰胺离子的标准试验方法》.pdf_第5页
第5页 / 共5页
亲,该文档总共5页,全部预览完了,如果喜欢就下载吧!
资源描述

1、Designation: D 7304 06Standard Test Method forDetermination of Denatonium Ion in Engine Coolant byHPLC1This standard is issued under the fixed designation D 7304; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last rev

2、ision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the chemical analysis of enginecoolant for denatonium ion by high-performance liquid chro-matography

3、(HPLC).1.2 This test method is applicable to both new and usedcoolants.1.3 Coelution of other ions may cause interferences in thedetection of the denatonium cation. In the case of unfamiliarformulations, identification verification should be performedby either or both fortification and dilution of t

4、he sample matrixwith denatonium ion1.4 The values stated in SI units are to be regarded as thestandard. The values given in parentheses are for informationonly.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the use

5、r of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D 1193 Specification for Reagent WaterD 1176 Practice for Sampling and Preparing Aqueous So-lutions of Engine Co

6、olants or Antirusts for Testing Pur-poses3. Summary of Test Method3.1 A sample volume of working sample is prepared bydilution of the sample with water. A high-pressure pump forcesthe mobile phase, eluant, through the HPLC columns (a guardand analytical column) at typical flow rates of 0.1- 2 mL/min

7、.A sample to be separated is introduced in the mobile phase byan injection device prior to the column. The analytes areseparated as they pass through the column. An optical sensordetects the changes in characteristics of the eluant stream andconverts the signal into an absorbance spectrum. The datas

8、ystem compares this response with an external calibrationcurve and the results of the concentration of analyte reported asppm or milligrams per litre (mg/L). Refer to Appendix X1 fora HPLC flow diagram4. Significance and Use4.1 This test method provides for the qualitative and quan-titative determin

9、ation of denatonium ion in engine coolant inmilligrams per liter to low percent range and requires approxi-mately 100 mL per test, with results available in less than 30min. Acceptable levels of denatonium vary with manufactur-ers blending specifications and applicable minimum or maxi-mum industry a

10、nd state specifications. Typically specificationranges from 30 to 50 ppm denatonium ion.5. Interferences5.1 Interferences can be caused by substances with similarretention times, especially if they are in high concentrationcompared to the analyte of interest, denatonium ion. Sampledilution and optim

11、ized gradient elution can be used to mini-mize or solve most interference problems.5.2 A water dip (solvent system peak) can cause interfer-ence with some integrators. This is eliminated by dilution withthe eluant if the sample dilution factor is 4+1 (v/v) or greater.5.3 Method interferences can be

12、caused by the contamina-tion of glassware, eluant, reagents, etc. Great care must betaken to ensure that contamination is kept at the lowest possiblelevel6. Apparatus6.1 HPLC SystemAnalytical system equipped with allrequired accessories including syringes, gasses, columns,pumps, detectors and softwa

13、re.6.1.1 Gradient Pump.6.1.2 UV/VIS-Mulitple Wavelength Detector.1This test method is under the jurisdiction of ASTM Committee D15 on EngineCoolants and is the direct responsibility of Subcommittee D15.04 on ChemicalProperties.Current edition approved Nov. 1, 2006. Published January 2007.2For refere

14、nced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West

15、Conshohocken, PA 19428-2959, United States.6.1.3 Eluant Degas System.6.1.4 Analytical Column ZORBAX RX-C8 or equivalentcolumn, capable of producing analyte separation equivalent toor better than that shown in Fig. 1.6.1.5 Guard Column for ZORBAX RX-C8 or equivalent,for protection of the analytical c

16、olumn from strongly retainedconstituents.6.1.6 Integrator or Chromatography Data System Softwarecapable of obtaining approximately the same detection limitsas listed in Table 1.6.2 Analytical Balance, 0.0001 g precision.6.3 Volumetric Flasks, 50 mL, 100 mL, 200 mL.7. Reagents and Materials7.1 Aceton

17、itrile, (ACN), Reagent Grade.7.2 Phosphoric acid, H3PO4, 85 %, Reagent Grade.7.3 Potassium Phosphate, Monobasic, KH2PO4, 99.995 %purity.7.4 Denatonium Benzoate, Reagent Grade.7.5 Ethylene Glycol, Reagent Grade.7.6 Deionized water, Typee II water.7.7 Purity of ReagentsReagent grade chemicals shall be

18、used in all tests. Unless otherwise indicated, it is intended thatall reagents shall conform to the specifications of the Commit-tee on Analytical Reagents of the American Chemical Society,where such specifications are available. Other grades may beused, provided it is first ascertained that the rea

19、gent is ofsufficiently high purity to permit its use without lessening theaccuracy of the determination.7.8 Unless otherwise indicated, references to water shall beunderstood to mean reagent water conforming to SpecificationD 1193, Type II. It is recommended that all water be filteredthrough a 0.45

20、m filter. For eluant preparation, degas thewater by sparging with helium or vacuum degassing andsonication.7.9 Phosphate Buffer Stock SolutionWeigh 650 mg ofMonobasic Potassium Phosphate and transfer it intoa1Lvolumetric flask, add 0.5 mL Phosphoric acid and fill to themark with Type II water. Mix t

21、his solution thoroughly. Filterthe solution with 0.45 m filter before use.7.10 Denatonium Ion Standard Stock SolutionWeigh0.1372 g of Denatonium Benzoate into 100-mL volumetricflask and dilute with ethylene glycol to 100 g. This stocksolution is 1000 ppm of denatonium and is stable for onemonth. Mak

22、e-up working standard solutions in 25% v/vethylene glycol/Type II water to desired concentrations. Ex-ample: 5 ppm denatonium standard solution - weigh 0.5000 gof stock denatonium standard solution into a 100 mL volumet-ric flask and fill to 100 mL mark with 25 % v/v ethylene glycolwater solution.8.

23、 Hazards8.1 Personnel protective equipment (such as eye protec-tions, gloves, laboratory coat, etc) should be used in thehandling of all chemicals. Special care should be taken whenhandling solutions around electrical equipment. All solutionshould be prepared in a hood.8.2 Read all equipment manuals

24、 before attempting to oper-ate HPLC instrumentation. Special attention should be given toall warnings.8.3 Be familiar with the MSDS for all chemicals used in thisprocedure. A dust mask is recommended for the handling ofDenatonium Benzoate. Review your companys policy con-cerning the use of a dust ma

25、sk. Prepare standards in a hood.9. Sampling, Test Specimens and Test Units9.1 Collect the sample in a scrupulously clean glass orpolyethylene bottle in accordance with Test Method D 1176.Collect at least 100 mL of sample.FIG. 1 (a) Denatonium Benzoate Standard (b) Denatonium Benzoate in an Engine Co

26、olantTABLE 1 Minimum Detection LimitsAnalyte Detection Limits, ppmADenatonium ion 1.0ADetermination using 100-L sample volume. Sample diluted 1+3 (wt/v) withtype II water, full scale, UV/Vis detector set at 210 nm wavelength, Zorbax RX-C8analytical column with Zorbax RX-C8 guard columns. Other syste

27、ms will requireMDL determination using chosen dilution factors, eluants, columns and detectors.D730406210. Preparation of Apparatus10.1 Over the course of the HPLC run, the concentration ofthe eluant will change to improve the separation of theanalytes. The HPLC system is programmed to handle thecha

28、nging mix of eluant. Two different eluant gradient programsare given below, for a borate-silicate based engine coolant anda silicate free, organic acid based engine coolant. The program-ming of the eluant gradient will vary with the HPLC systemused and the inhibitor package utilized in the engine co

29、olant.10.2 Program Gradient 1: (for a borate, silicate based enginecoolant)Time Flow, mL Acetonitrile, Vol % Phosphate Buffer, Vol % CurveAInitial 1.9 20 80 0.1 1.9 20 80 110.0 1.9 45 55 620.0 1.9 90 10 623.0 1.9 20 80 627.0 1.9 20 80 1AThe Curve Number entered here specifies the manner in which suc

30、cessivegradient steps are connected.Curve 1 Go to end condition immediately2-5 Convex6 Linear7-10 Noncave11 Maintain start conditions until end of segment10.3 Program Gradient 2: (for a silicate free, organic acidbased engine coolant)Time Flow, mL Acetonitrile, Vol % Phosphate Buffer, Vol % CurveAIn

31、itial 1.9 20 80 0.1 1.9 20 80 110.0 1.9 40 60 615.0 1.9 80 20 623.0 1.9 20 80 627.0 1.9 20 80 1AThe Curve Number entered here specifies the manner in which successivegradient steps are connected.Curve 1 Go to end condition immediately2-5 Convex6 Linear7-10 Noncave11 Maintain start conditions until e

32、nd of segment11. Calibration and Standardization11.1 Analyze each standard solution separately to determinethe analytes retention time.11.2 Set the chromatograph up in accordance with theconditions specified in 11.9. The use of other equipment,eluants, or flows requires calculation of suitable dilut

33、ionfactors and instrument settings that permit the analyst to obtainthe resolution and detection limits given in Fig. 1 and Table 1,respectively.11.3 Prepare concentrations of denatonium ion at 0.005,0.025 and 0.055 g/L (5, 25 and 55 ppm) from the standardstock solution. Additional concentrations ca

34、n be prepared for amore comprehensive calibration curve. All final solutionsshould be made with Denatonium Ion Standard Stock solutionas described in 7.10. Calibrate chromatograph with a minimumof at least three levels of the analyte, starting near but above theminimum detection limit (MDL) and furt

35、her defining the workrange in samples subsequent to dilution. These denatoniumanalyte examples reflect a dilution of 3+1 (wt/v%) with Type IIwater.NOTE 1HPLC equipment other than that described in this test methodmay require that standards be prepared at higher or lower levels.11.4 Analyze a blank c

36、ontaining only the eluant as de-scribed in Section 12.11.5 A sample of known concentration should be run every10 injections to make sure the instrument is working correctly.11.6 The analytical calibration curve must be established atonly one detector scale setting in order to prevent a change ofslop

37、e affecting the analytical curve.11.7 The analytical calibration curve and an eluant blankshall be verified daily prior to the analysis of samples to verifythe system resolution, calibration and sensitivity.11.8 The analytical calibration curve, analyte retentiontimes and resolution, and an eluant b

38、lank shall be verifiedsubsequent to a change of the system eluant.11.9 Conditions (or equivalent conditions to meet require-ments of Table 1 and Fig. 1):(a) Detector Setting: 210 nm wave length(b) Analytical Column: Zorbax RX-C8 with Guard Col-umn: Zorbax RX-C8(c) Pump: High Pressure limits 3000 psi

39、; Low Pressurelimits 100 psi(d) Flow Rate: 1.9 mL/min(e) Gradient Program: See Section 10(f) Retention Time: 10 to 12 min depending on test sampleand gradient programNOTE 2These conditions may vary depending on the dimensions ofthe column used.12. Procedure12.1 Set up the HPLC (Waters Chromatographi

40、c System) inaccordance with the manufacturers instructions.12.2 Check the instruments setting: see 11.9.12.3 Fill eluant containers (acetonitrile and phosphatebuffer), sparge and pressurize with helium to 5 psi.12.4 Prime the pump, initiate flow and equilibrate thecolumn for 20-25 min with initial s

41、ettings in program gradient,Section 10.12.5 Turn on HPLC and detector. Allow the instrument toequilibrate fro 20-25 min before starting a run.12.6 At least one duplicate and one spiked coolant samplemust be analyzed with each batch of ten or fewer samples. It isrecommended that all test samples be f

42、iltered through a 0.45m filter prior to injection. Results must be recorded in thelaboratory manual or be electronically maintained. The spikedsample is to be prepared by addition of a mixed analytemid-range standard.NOTE 3Once the phosphate buffer has run through the system, do notpurge with 100 %

43、acetonitrile. This will precipitate buffer in the pump,column or detector and result in total blockage and pressure overload.12.7 The column and entire system should be flushed with50-50 %(v/v) water/acetonitrile solution after an analysis tofree from phosphate salts.13. Calculation or Interpretatio

44、n of Results13.1 Integrate the peak area for determination of the con-centrations. Plot the peak area against concentration. A linearD7304063calibration curve is generated for the denatonium ion. Calibra-tion is in the concentration range of interest (5 to 60 ppm) andmust have a linear least squares

45、 correlation coefficient of0.9990 or greater.13.2 The concentration of analyte is given by the concen-tration read from the calibration curve multiplied by thedilution factor used.DB concentration 5 DB concentration from curve dilution factorDB 5 denatonium ion (1)13.3 Relative percent difference (R

46、PD) of duplicate runs:RPD 5 Absolute Value A B!/A! 100 (2)where:A = concentration found in Analysis Run 1, andB = concentration found in Analysis Run 2.Results must be within 20 %.Spike recovery (SR) as percent:SR 5 C 1 B!/A! 100 (3)where:A = concentration found in analysis of spiked sample,B = conc

47、entration found in analysis of sample, andC = concentration of analyte (denatonium ion) added tosample.Spike recovery must be within a 80 to 120 % range.14. Report14.1 Convert and report the concentration of denatoniumion, ppm into the concentration of the denatonium benzoate inppm units found in th

48、e sample.Denatonium Benzoate, ppm 5 denationium, ppm 0.7287 (4)14.2 Include in the report: the sample analysis date, analyst,and the lot number or other identification number for thesample.15. Precision and Bias15.1 The precision and bias of this test method have notbeen determined.16. Keywords16.1

49、analytical method; antifreeze; Bitterant; chromatogra-phy; denatonium ion; denatonium benzoate; engine coolant;HPLCAPPENDIXX1. HPLC SystemNonmandatory InformationX1.1 See Fig. X1.1.3Solvent Delivery Systempushes the solvent(s) stream through the instrument at a controlled flowrate.X1.2 Autosamplerintroduces the sample into the liquidstream of the instrument.3http:/ X1.1 HPLC Flow DiagramD7304064X1.3 Columna stainless steel tube packed with siliconbeads that separates the denato

展开阅读全文
相关资源
猜你喜欢
  • AECMA PREN 3077-1998 Aerospace Series Clamps Worm Drive Technical Specification Edition P 1《航空航天系列夹持器蜗杆驱动 技术规范》.pdf AECMA PREN 3077-1998 Aerospace Series Clamps Worm Drive Technical Specification Edition P 1《航空航天系列夹持器蜗杆驱动 技术规范》.pdf
  • AECMA PREN 3078-1998 Aerospace Series P Q and Saddle Clamps with Rubber Cushion Technical Specification Edition P 1《航空航天系列带橡胶垫的P、Q和鞍形夹持器 技术规范》.pdf AECMA PREN 3078-1998 Aerospace Series P Q and Saddle Clamps with Rubber Cushion Technical Specification Edition P 1《航空航天系列带橡胶垫的P、Q和鞍形夹持器 技术规范》.pdf
  • AECMA PREN 3079-1999 Aerospace Series Pipe Coupling 8 Degree 30' Up to 28 000 kPa Adaptors Metric Series Technical Specification Edition P 1《航空航天系列管连接件8°30′ 直到 28000 kPa 异形接头米制系列 技.pdf AECMA PREN 3079-1999 Aerospace Series Pipe Coupling 8 Degree 30' Up to 28 000 kPa Adaptors Metric Series Technical Specification Edition P 1《航空航天系列管连接件8°30′ 直到 28000 kPa 异形接头米制系列 技.pdf
  • AECMA PREN 3080-2000 Aerospace Series Pipe Coupling 8 Degrees 30' in Titanium Alloy Tees Reduced Branch with Thrust Wire Nut Edition P 1《航空航天系列钛合金管连接件8°30′带可靠金属丝螺母的异形分支三通管》.pdf AECMA PREN 3080-2000 Aerospace Series Pipe Coupling 8 Degrees 30' in Titanium Alloy Tees Reduced Branch with Thrust Wire Nut Edition P 1《航空航天系列钛合金管连接件8°30′带可靠金属丝螺母的异形分支三通管》.pdf
  • AECMA PREN 3081-1997 Aerospace Series Pipe Coupling 8 Degrees 30' in Titanium Alloy Ferrules Welded with Dynamic Beam Seal End for Repair Edition P 1《航空航天系列钛合金瘔和焊接连接件8°30′带动态密封可靠金属.pdf AECMA PREN 3081-1997 Aerospace Series Pipe Coupling 8 Degrees 30' in Titanium Alloy Ferrules Welded with Dynamic Beam Seal End for Repair Edition P 1《航空航天系列钛合金瘔和焊接连接件8°30′带动态密封可靠金属.pdf
  • AECMA PREN 3082-1997 Aerospace Series Pipe Coupling 8 Degrees 30' in Titanium Alloy Unions Bulkhead with Weld End for Repair Edition P 1《航空航天系列钛合金连接件8°30′带焊接缝用于重修.P1版》.pdf AECMA PREN 3082-1997 Aerospace Series Pipe Coupling 8 Degrees 30' in Titanium Alloy Unions Bulkhead with Weld End for Repair Edition P 1《航空航天系列钛合金连接件8°30′带焊接缝用于重修.P1版》.pdf
  • AECMA PREN 3083-1997 Aerospace Series Pipe Coupling 8 Degrees 30' in Titanium Alloy Unions Bulkhead Long Welded for Repair Edition P 1《航空航天系列钛合金连接件8°30′带焊接缝用于重修.P1版》.pdf AECMA PREN 3083-1997 Aerospace Series Pipe Coupling 8 Degrees 30' in Titanium Alloy Unions Bulkhead Long Welded for Repair Edition P 1《航空航天系列钛合金连接件8°30′带焊接缝用于重修.P1版》.pdf
  • AECMA PREN 3084-1997 Aerospace Series Pipe Coupling 8 Degrees 30' in Titanium Alloy Unions Long Welded for Repair Edition P 1《航空航天系列钛合金连接件8°30′长焊接用于重修.P1版》.pdf AECMA PREN 3084-1997 Aerospace Series Pipe Coupling 8 Degrees 30' in Titanium Alloy Unions Long Welded for Repair Edition P 1《航空航天系列钛合金连接件8°30′长焊接用于重修.P1版》.pdf
  • AECMA PREN 3086-1998 Aerospace Series Hose Assemblies Designation Limited to 15 Digits Edition P 1《航空航天系列.软管系统 符号表示到15位》.pdf AECMA PREN 3086-1998 Aerospace Series Hose Assemblies Designation Limited to 15 Digits Edition P 1《航空航天系列.软管系统 符号表示到15位》.pdf
  • 相关搜索

    当前位置:首页 > 标准规范 > 国际标准 > ASTM

    copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
    备案/许可证编号:苏ICP备17064731号-1