ASTM D7754-2016 7004 Standard Test Method for Determination of Trace Oxygenates in Automotive Spark-Ignition Engine Fuel by Multidimensional Gas Chromatography《采用多维气相色谱法测定汽车火花点火发动机.pdf

上传人:fatcommittee260 文档编号:526258 上传时间:2018-12-04 格式:PDF 页数:18 大小:603.06KB
下载 相关 举报
ASTM D7754-2016 7004 Standard Test Method for Determination of Trace Oxygenates in Automotive Spark-Ignition Engine Fuel by Multidimensional Gas Chromatography《采用多维气相色谱法测定汽车火花点火发动机.pdf_第1页
第1页 / 共18页
ASTM D7754-2016 7004 Standard Test Method for Determination of Trace Oxygenates in Automotive Spark-Ignition Engine Fuel by Multidimensional Gas Chromatography《采用多维气相色谱法测定汽车火花点火发动机.pdf_第2页
第2页 / 共18页
ASTM D7754-2016 7004 Standard Test Method for Determination of Trace Oxygenates in Automotive Spark-Ignition Engine Fuel by Multidimensional Gas Chromatography《采用多维气相色谱法测定汽车火花点火发动机.pdf_第3页
第3页 / 共18页
ASTM D7754-2016 7004 Standard Test Method for Determination of Trace Oxygenates in Automotive Spark-Ignition Engine Fuel by Multidimensional Gas Chromatography《采用多维气相色谱法测定汽车火花点火发动机.pdf_第4页
第4页 / 共18页
ASTM D7754-2016 7004 Standard Test Method for Determination of Trace Oxygenates in Automotive Spark-Ignition Engine Fuel by Multidimensional Gas Chromatography《采用多维气相色谱法测定汽车火花点火发动机.pdf_第5页
第5页 / 共18页
亲,该文档总共18页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、Designation: D7754 16Standard Test Method forDetermination of Trace Oxygenates in Automotive Spark-Ignition Engine Fuel by Multidimensional GasChromatography1This standard is issued under the fixed designation D7754; the number immediately following the designation indicates the year oforiginal adop

2、tion or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method covers the determination of traceoxygenates in automot

3、ive spark-ignition engine fuel. Themethod used is a multidimensional gas chromatographicmethod using 1,2-dimethoxy ethane as the internal standard.The oxygenates that are analyzed are: methyl-tertiary butylether (MTBE), ethyl-tertiary butyl ether (ETBE), diisopropylether (DIPE), methanol, tertiary-a

4、myl methyl ether (TAME),n-propanol, i-propanol, n-butanol, i-butanol, tert-butyl alcohol,sec-butyl alcohol, and tert-pentanol. Ethanol is usually notmeasured as a trace oxygenate since ethanol can be used as themain oxygenate compound in finished automotive spark-ignition fuels such as reformulated

5、automotive spark-ignitionfuels. The concentration range of the oxygenates covered in theILS study was from 10 mgkg to 2000 mgkg. In addition thismethod is also suitable for the measurement of the C5 isomericalcohols (2-methyl-1-butanol, 2-methyl-2-butanol) presentfrom the fermentation of ethanol.1.2

6、 The ethanol blending concentration for which this testmethod applies ranges from 1 % to 15% by volume. Higherconcentrations of ethanol coelute with methanol in the analyti-cal column. Lower levels of ethanol, similar to the otheroxygenate, can be calibrated and analyzed also. If higherethanol conce

7、ntrations are expected, the window cutting tech-nique can be used to avoid ethanol from entering the analyticalcolumn and interfere with the determination of the otheroxygenates of interest. Refer to Appendix X1 for details.1.3 The values stated in SI units are to be regarded asstandard. No other un

8、its of measurement are included in thisstandard.1.3.1 Alternative units, in common usage, are also providedto increase clarity and aid the users of this test method.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of th

9、e user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D4057 Practice for Manual Sampling of Petroleum andPetroleum ProductsD4307 Practice for Preparation of Liqu

10、id Blends for Use asAnalytical StandardsD4815 Test Method for Determination of MTBE, ETBE,TAME, DIPE, tertiary-Amyl Alcohol and C1to C4Alco-hols in Gasoline by Gas ChromatographyD6304 Test Method for Determination of Water in Petro-leum Products, Lubricating Oils, and Additives by Cou-lometric Karl

11、Fischer Titration3. Terminology3.1 Definitions of Terms Specific to This Standard:3.1.1 electronic pressure control, nelectronic pneumaticcontrol of carrier gas flows. Can be flow or pressure pro-grammed to speed up elution of components.3.1.2 flame ionization detector (FID), ndetector used toanalyz

12、e the components eluting from the column.3.1.3 fluidic switch, ndevice that reverses the directionalflow in a union T altering the pressure at the midpoint. In itssimplest design it is also known as a Dean Switch.3.1.4 inlet, ncapillary split/splitless inlet system operatedin the split mode is recom

13、mended. Operate the inlet within itslinear range.3.1.4.1 split ratio, n in capillary gas chromatography, theratio of the total flow of carrier gas to the sample inlet versusthe flow of the carrier gas to the capillary column is expressedby:Split ratio 5 S1C!/C (1)1This test method is under the juris

14、diction of ASTM Committee D02 onPetroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility ofSubcommittee D02.04.0L on Gas Chromatography Methods.Current edition approved Jan. 1, 2016. Published February 2016. Originallyapproved in 2011. Last previous edition approved in 2011

15、as D7754 111.DOI:10.1520/D7754-16.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.*A Summary of Changes secti

16、on appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1where:S = flow rate at the splitter vent, andC = flow rate at the column outlet.3.1.5 low volume connector, nspecial union for connect-ing two lengt

17、hs of tubing 1.6 mm inside diameter and smaller.Sometimes this is referred to as zero dead volume union.3.1.6 multidimensional gas chromatography, ngas chro-matographic technique where using hardware (valves, pressureswitches, etc.) in which selected components from one column(primary column) are tr

18、ansferred to a secondary columndiffering in characteristics (film thickness, polarity, capacity,etc.) from the first column.3.1.7 WCOT column, nwall-coated open tubular, a type ofcapillary gas chromatographic column prepared by coating theinside of the capillary wall with a specified thin film ofsta

19、tionary phase. The coatings used are either 100 % polydim-ethyl siloxane or 5 % phenyl-polydimethylsiloxane.3.1.7.1 apolar column, npolydimethylsiloxane nonpolarcolumn used as a pre-column.3.1.7.2 PLOT column-oxygen selective, nporous-layeropen tubular which is an oxygenate selective capillary gas-s

20、olid chromatographic column. It is used as an analyticalcolumn.4. Summary of Test Method4.1 An appropriate internal standard of a product that is notpresent in refinery streams, such as 1,2-dimethoxy ethane(1,2-DME), is added to the sample, which is then introducedinto a gas chromatograph equipped w

21、ith two columns and a4-port switching valve. The sample first passes onto an apolar(non-polar) polydimethylsiloxane WCOT column that per-forms a pre-separation of the trace oxygenates and elutesunwanted high boiling hydrocarbons to vent. The oxygenatesand the DME are transferred to the analytical ox

22、ygen selectivecolumn by the switching valve. While the oxygenates and theDME are eluting from the analytical column, the inlets carriergas is used to elute the hydrocarbons from the pre-column toyield a stable baseline for the next analysis. The auxiliarypressure controller is used to provide carrie

23、r gas to theanalytical column during the analysis.4.2 The eluted components Table 1 are detected by one ortwo flame ionization detectors. In the single detector Configu-ration A (Fig. 1), only the components eluting from theanalytical column are analyzed. In the two detector Configu-ration B (Fig. 2

24、), detector one is used to monitor the apolarelution and aid in setting “heart-cut” times for specific oxy-genates while the second detector is used to monitor theanalytical column elution and also for the quantitation of theoxygenates.The second detector response is proportional to theoxygenates an

25、d DME components concentration. The signal isrecorded, the peak areas are measured, and the concentration ofeach oxygenate is calculated with reference to the internalstandard.4.3 Alternatively, a fluidic switching system, ConfigurationC(Figs. 3 and 4) may be used instead of valve switching. Inthis

26、system, the two columns are joined by a zero dead volume(ZDV) tee purged by an auxiliary carrier source. At injection,the auxiliary flow is low, and the inlet flow is sufficient so thatat the midpoint where the two columns join, the flow is therequired flow to transfer the oxygenates to the PLOT col

27、umn.Thus, there is forward flow through the pre-column and theanalytical column. Once the oxygenates have passed through tothe analytical column, the inlet flow is decreased and theauxiliary flow is increased, which results in backflushing thepre-column through the split vent of the front inlet whil

28、e theanalytical column continues the separation.5. Significance and Use5.1 The analysis of trace oxygenates in automotive spark-ignition engine fuel has become routine in certain areas toensure compliance whenever oxygenated fuels are used. Inaddition, test methods to measure trace levels of oxygena

29、tes inautomotive spark-ignition fuel are necessary to assess productquality.6. Apparatus6.1 ChromatographA multidimensional gas chromato-graphic system, which is able to adequately resolve oxygenatesand an internal standard and to eliminate hydrocarbon, as wellas other interferences, is used for the

30、se analyses. The instru-ment is to be configured to operate using the approximateconditions listed in Table 2. The system requires a columnswitching mechanism equivalent to Fig. 1 or Fig. 2 if using aTABLE 1 Component List with Retention and Calibration Characteristics for WCOT/PLOT Column Set Using

31、 Conditions of Table 2AComponent RT (min) Mol Wt BP (C) Slope y-Int Corr. Coef.ETBE 12.7 102.2 70 to 72 1.919 0.02 0.999MTBE 12.8 88.2 55 to 56 1.689 0.01 0.999DIPE 12.9 102.2 68 to 69 2.124 0.06 0.999TAME 13.6 102.2 85 to 86 2.023 0.02 0.999Methanol 15.6 32.0 65 0.779 0.09 0.997Ethanol 18.7 46.1 78

32、 1.352 0.19 0.999iso-Propanol 22.2 60.1 81 to 83 1.504 0.06 0.999n-Propanol 22.2 60.1 97 . . .t-Butanol 23.8 74.1 82 1.951 0.12 0.999s-Butanol 23.8 74.1 98 . . .iso-Butanol 23.8 74.1 117 . . .n-Butanol 24.4 74.1 118 1.906 0.05 0.999tert-Pentanol 25.1 88.1 102 2.148 0.04 0.9981,2-DME 26.0 90.1 85 . .

33、 .AFor coeluting compounds the response is assigned to the first peak listed. Values may be different for different instruments.D7754 162valve system. If using a fluidic system then the fluidic switchand auxiliary flow control are required as shown in Fig. 3 andFig. 4. Carrier gas flow controllers (

34、EPC) shall be capable ofprecise control where the flow rates are low. Pressure controldevices and gages shall be capable of precise control for thetypical pressures required.6.1.1 DetectorTwo-flame ionization detectors are prefer-ably used (Configuration B), although the analysis can beperformed usi

35、ng only one detector (Configuration A and C).The system shall have sufficient sensitivity and stability toobtain a signal-to-noise ratio of at least 5 to 1 for a 1 mgkgconcentration of any oxygenate. In the fluidic system only onedetector is used.6.1.2 Switching ValveA switching valve, to be located

36、within the gas chromatographic column oven or separate oven,capable of performing the functions described in 9.2 andFIG. 1 Schematic of Chromatographic SystemConfiguration AFIG. 2 Schematic of Chromatographic SystemConfiguration BFIG. 3 Fluidic Switch SchematicOxygenate TransferD7754 163illustrated

37、in Fig. 1. The valve shall be of low volume designand not contribute significantly to chromatographic deteriora-tion. Alternatively a Deans switching arrangement can also beused as shown in Fig. 3 and Fig. 4.6.1.2.1 A commercially available valve: 1.6 mm fittings,0.75 mm ports was used in the method

38、 development. Anequivalent valve may be used.6.1.2.2 Fluidic Switch, as an option to the two-positionswitching valve. See 4.3, Table 2, and Figs. 3 and 4 for adescription. Additional flow source is required as well ashardware, which is located in the oven for the columnconnection.6.1.2.3 Some gas ch

39、romatographs are equipped with anauxiliary oven, which can be used to contain the valve at anFIG. 4 Fluidic Switch SchematicPre-column BackflushTABLE 2 Chromatographic ConditionsConfiguration Switching Valve Fluidic SwitchCarrier Gas Helium HeliumInjection Volume 1.0 L 1.0 LInlet: Split/Splitless (S

40、plit mode) Split/Splitless (Split mode)Temperature 250 C 250 CSplit Ratio 10:1 10:1PressureA51.7 kPa, Constant Pressure 17.8 kPa, Flow Program ModeColumns and FlowsPre-column 30 m by 0.53 mm by 5.0 m PDMS7.5 mL/min 60 C15 m by 0.53 mm by 1.5 m 5 % phenyl PDMSInitial Flow: 5.4 mL/min 60 CHold for 1.5

41、 minRamp: 90 mL/min to -5 mL/minHold until end of runAnalytical Column 10 m by 0.53 mm by 10 m Oxygen Selective7.5 mL/min 60 C10 m by 0.53 mm by 10 m Oxygen Selective7.0 mL/min constant flowOven:Initial Temperature 60 C 60 CInitial Hold 6.0 min 6.0 minRamp 1 10 C min 10 C minFinal Temperature 150 C

42、150 CFinal Hold 5.0 min 5.0 minRamp 2 10 C min 10 C minFinal Temperature 220 C 220 CFinal Hold 3.0 min 3.0 minTotal Time 30 min 30 minDetector: FID FIDTemperature 275 C 275 CHydrogen 40 mL/min 40 mL/minAir 450 mL/min 450 mL/minMake-up (N2) 10 mL/min 10 mL/minValve Temperature 150 C N/AAuxiliary Pres

43、sure 73.1 kPa 10.3 kPaVent Restrictor 76 cm in length0.16 cm (O.D.) and 0.25 mm (I.D.) SSN/ADefault Valve Times (for complete analysis):Initial OFF N/A0.50 min ON N/A4.50 min OFF N/AAFor Configuration A valve timing determination, 9.2, set the inlet pressure to 34.5 kPa.D7754 164isothermal temperatu

44、re. In such a configuration, the twocapillary columns are located in the main oven and connectedto the valve by using low dead volume and inert stainless steeltubing terminated in the GC oven.6.1.2.4 An automatic valve switching device is used toensure repeatable switching times. Such a device is sy

45、nchro-nized with injection and data collection times. For the pressureswitching approach, automatic precise and stable pressurecontrol shall be used. Fluidic systems require both a fluidicswitch and a programmable auxiliary pressure source tomaintain and program flows.6.1.3 Injection SystemThe chrom

46、atograph is to beequipped with a heated splitting-type inlet device containing areplaceable glass deactivated liner (single-taper style) withdeactivated glass wool at the bottom to retain non-vaporizedcomponents). Split injection is necessary to maintain the actualchromatographed sample size within

47、the limits of column anddetector efficiency and linearity.6.1.3.1 A microliter automatic syringe injector is used forintroducing representative samples into the gas chromato-graphic inlet and for adequate repeatability.6.2 Data Acquisition System:6.2.1 ComputerA data acquisition system containing ac

48、omputer and data acquisition software is required.6.2.2 IntegratorAlternatively, an integrator can be used tomeasure peak areas and to perform the analytical calculations.6.3 Column Class:6.3.1 Apolar (Non-polar) Pre-ColumnThis column per-forms a pre-separation of the oxygenates and internal standar

49、dfrom hydrocarbons in the same boiling point range. Unless aseparate auxiliary oven is provided for it, the apolar columnshall perform at the same temperature as the polar columndoes.6.3.1.1 WCOT Methyl Silicone Pre-Column30 m long by0.53 mm inside diameter fused silica column with a 5 m filmthickness of cross-linked polydimethylsiloxane. With fluidicswitch (Configuration C) a 30 m long by 0.53 mm with a1.5 m 5 % phenyl polydimethyl siloxane is recommended.6.3.2 Polar Analytical ColumnAny column with eq