1、Designation: D7962 16Standard Practice forDetermination of Minimum Immersion Depth andAssessment of Temperature Sensor Measurement Drift1This standard is issued under the fixed designation D7962; the number immediately following the designation indicates the year oforiginal adoption or, in the case
2、of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This practice describes a two procedures for use withtemperature measurement devices. Method
3、ology is describedfor determining minimum immersion depth for thermalsensors, in particular RTDs or similar temperature sensors.Included is a procedure for consistently preparing a referencebath for the purpose of monitoring measurement drift ofthermal sensors such as liquid-in-glass or digital cont
4、actthermometers.1.2 This practice focuses on temperature measurement driftin a laboratory. If the user requires greater measurementaccuracy, then they should follow the instructions in PracticeE563.1.3 The values stated in SI units are to be regarded asstandard. No other units of measurement are inc
5、luded in thisstandard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior
6、to use.2. Referenced Documents2.1 ASTM Standards:2E563 Practice for Preparation and Use of an Ice-Point Bathas a Reference Temperature3. Terminology3.1 Definitions:3.1.1 digital contact thermometer (DCT), nan electronicdevice consisting of a digital display and associated tempera-ture sensing probe.
7、3.1.1.1 DiscussionThis device consists of a temperaturesensor connected to a measuring instrument; this instrumentmeasures the temperature-dependent quantity of the sensor,computes the temperature from the measured quantity, andprovides a digital output. This digital output goes to a digitaldisplay
8、and/or recording device that may be internal or externalto the device. These devices are sometimes referred to as a“digital thermometer”.3.1.1.2 DiscussionPortable electronic thermometers(PET) is an acronym sometimes used to refer to a subset of thedevices covered by this definition.3.1.2 ice-point
9、bath, nphysical system containing ice andwater assembled to realize the ice point as a referencetemperature, or to establish a constant temperature near 0 C.3.1.3 minimum immersion depth, ndepth that a thermom-eter should be immersed, in a uniform temperatureenvironment, such that further immersion
10、does not produce achange in indicated temperature greater than the specifiedtolerance.4. Summary of Practice4.1 This practice describes a procedure for consistentlypreparing an ice bath that is an intimate mixture of crushed iceor ice particles and water in a thermally insulating vessel opento the a
11、tmosphere.4.1.1 Cautionwhen the ice bath is not made from distilledwater, its temperature will differ from the natural fixed-pointtemperature by a consistent amount, typically less than0.02 C. If the user needs a more accurate ice point, then theyshould use Practice E563 to prepare the ice bath.4.2
12、This practice includes a procedure for determining theminimum immersion depth of the temperature sensor using anice bath.4.2.1 This procedure determines minimum immersion depthwith a 25 C differential between sensed temperature andambient temperature. If the probe is subjected to a greaterdifferenti
13、al temperature, a larger immersion depth may berequired to correctly measure the temperature.5. Significance and Use5.1 This practice provides a means for the users of ASTMCommittee D02 standards to monitor the drift in sensed1This practice is under the jurisdiction of ASTM Committee D02 on Petroleu
14、mProducts, Liquid Fuels, and Lubricants and is the direct responsibility of Subcom-mittee D02.91 on Coordinating Subcommittee on Thermometry.Current edition approved Dec. 1, 2016. Published January 2017. Originallyapproved in 2015. Last previous edition approved in 2015 as D7962 15a. DOI:10.1520/D79
15、62-16.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.*A Summary of Changes section appears at the end of thi
16、s standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United StatesThis international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for theDevelopmen
17、t of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.1temperature of liquid-in-glass thermometer (LiG), and digitalcontact thermometers (DCT). Digital contact thermometers aresometimes referred to as portable elec
18、tronic thermometers(PET) or simply digital thermometers.5.2 This practice is not suitable for determining the accu-racy or calibration of a temperature measuring device as theerror in the ice bath temperature can be greater than 0.02 C.For greater accuracy, the user should use Practice E563 toprepar
19、e the ice bath.5.3 The ice point is a common practical industrial referencepoint of thermometry. The ice point is relatively simple torealize and provides a readily available natural fixed-pointreference temperature.5.4 This practice only checks the measurement drift at asingle temperature. It will
20、not detect a change in measurementresponse with change in temperature. Temperature measuringdevices should be recalibrated at set intervals. See devicesupplier for recommendations.5.5 This practice provides a technique to determine mini-mum immersion depth of the sensing probe of the thermometerusin
21、g an ice bath. The minimum immersion depth determinedby this practice may change when the differential temperaturediffers significantly from the conditions described. A greaterdifferential will likely increase the minimum immersion depth.6. Reagents and Equipment6.1 WaterGood quality drinking water
22、from a consistentsource.6.2 IceCrushed ice prepared from the water in 6.1.6.3 Insulated containeran insulated container largeenough to hold the thermometer and provide several centime-ters of water-ice slush around the measuring portion of thethermometer.7. ProcedureDrift in Temperature Measurement7
23、.1 Preparation of the Ice-Point Bath:7.1.1 Determine the water source to be used each time thispractice is followed in your laboratory to determine measure-ment drift and note the water source with the measurement.Altering the water source between uses of the practice whenmonitoring a temperature se
24、nsor may result in observing a falsedrift.7.1.2 Ensure that all equipment is clean and free of any oilyresidue. Rinse the equipment with the type of water used forthe ice-point bath medium. Use clean, “powder free” labora-tory gloves to handle the ice and equipment.7.1.3 Prepare ice by using the wat
25、er from the sourcedetermined in 7.1.1. Chill a quantity of the water to near 0 Cin a flask then shake vigorously to aerate the water beforefreezing in small cubes or thin sheets.7.1.3.1 Prepare finely divided ice by shaving or crushing.Shaved ice resembling snow is preferred, but crushed ice isaccep
26、table if the particles are small (approximately 2 mm to6 mm in diameter) and there is a large distribution in size.7.1.4 Prepare the bath in a clean, thermally insulated vessel,preferably a wide-mouthed Dewar vacuum flask fitted with aninsulating closure such as a stopper. The vessel should be large
27、enough that its size does not affect the water-ice equilibriumtemperature and of such diameter and depth that in thermalequilibrium the test objects will not significantly modify thetemperature of the bath over the region to which the ice pointis to be applied. For usual applications, a diameter of
28、70 mmand a depth of 300 mm may be adequate.7.1.5 Alternately, add shaved ice and chilled water to thevessel, using enough water to saturate the ice but not enough tofloat it. As the vessel fills, compress the ice-water mixture toforce out excess water. The objective is to surround eachparticle of ic
29、e with water, filling all voids, but to keep the iceparticles as close together as possible. Continue adding ice andwater and compressing until the vessel is filled to the requiredlevel. Decant or siphon off excess water.7.1.6 Cover the ice-point bath leaving just enough openspace to access the ther
30、mometer probe and agitate the mix.Allow the bath and vessel to equilibrate for at least 30 minbefore using.7.2 Determination of Measurement Drift:7.2.1 Form a well in the ice-point bath that has the diameterand intended immersion depth of the test object.7.2.2 It is preferable to cool the test objec
31、t to less than 3 Cbefore immersing it in the bath. This reduces the time to reachequilibrium at the ice point. Pre-cooling the sensor helps topreserve the bath at the ice point for a prolonged time and helpsensure that the water-ice interface will be in contact with thethermometer because negligible
32、 melting will occur to increasethe water film thickness.7.2.3 Insert the test object to the appropriate depth. Thiscould be (1) the scribed immersion line, (2) the depth of theminimum immersion depth determined by this practice, or (3)the same depth as it is used. For total immersion liquid-in-glass
33、thermometers, immerse to the 0 C (32 F) mark. For partialimmersion liquid-in-glass thermometers, immerse to the im-mersion line or stated immersion depth. Keep the sensorportion of the object several centimeters above the bottom ofthe flask to avoid the zone at the bottom where denser meltwater tend
34、s to accumulate.7.2.4 Cover the top of the vessel around the test object butleave sufficient access to periodically agitate the ice-waterslush.7.2.5 Allow the bath and test object to come to thermalequilibrium.7.3 Maintaining the Bath:7.3.1 As ice particles in the bath melt, excess water beginsto ac
35、cumulate. This melt water has a temperature slightlywarmer than 0 C. Since the density of water is at a maximumat 4 C, the slightly warm melt water will collect at the bottomof the bath and around the test object if its tip is placed near thebottom of the bath. For this reason, surplus water should
36、beremoved, as it accumulates, from the bottom of the bath bydecanting or siphoning. The presence of excess water can bedetected if water overspill occurs when the ice is depressed.Add ice particles, and chilled water, as necessary so that the iceslush column always extends to at least 30 mm below th
37、elowest point of the test object.D7962 1627.3.2 In order to sustain the ice point over prolongedperiods, the ice-point bath may be immersed in another baththat is kept near 0 C.7.4 When the determination is complete, record the date,sensor identifier, immersion depth, observed temperature, roomtempe
38、rature, water source, and any other relevant observations.8. Procedure for Minimum Immersion Depth8.1 Measure the distance from the sensors tip to a referencemark on the sheath or other reference point.8.2 Using either an ice bath prepared in 7.1 or a constanttemperature bath, slowly insert the test
39、 sensor into the bathuntil no further insertion causes a significant change inobserved temperature. The portion inserted may include themounting flange, threads, etc.NOTE 1Environmental conditions will have a greater impact on theminimum immersion depth of “metal sheathed” sensors than is observedfo
40、r “glass sheathed” sensors.8.3 Wait until equilibrium is achieved. Ten minutes shouldbe sufficient, but the user will need to determine if more timeis required.8.4 Record the distance between reference mark on sensorand top of ice slush.8.5 Slowly withdraw the sensor from the bath by a smallincremen
41、t.8.6 Repeat 8.3 through 8.6 until a change in temperature isobserved.8.7 Subtract final value obtained in 8.4 from the referencemark value established in 8.1. The result is the minimumdistance the sensing probe shall be immersed in order to obtainan accurate temperature measurement.8.8 Record the d
42、ate, temperature sensor ID, minimumimmersion depth, ambient temperature, and constant tempera-ture bath used. If the constant temperature bath was an ice bath,then record the water source.9. Keywords9.1 fixed-point temperature references; ice bath; minimumimmersion depth; temperature measurement dri
43、ftSUMMARY OF CHANGESSubcommittee D02.91 has identified the location of selected changes to this standard since the last issue(D7962 15a) that may impact the use of this standard. (Approved Dec. 1, 2016.)(1) Revised 3.1.3 for minimum immersion depth.Subcommittee D02.91 has identified the location of
44、selected changes to this standard since the last issue(D7962 15) that may impact the use of this standard. (Approved Dec. 1, 2015.)(1) Added definition for DCT as 3.1.1. (2) Reworded portions of Section 8.ASTM International takes no position respecting the validity of any patent rights asserted in c
45、onnection with any item mentionedin this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the riskof infringement of such rights, are entirely their own responsibility.This standard is subject to revision at any time by the resp
46、onsible technical committee and must be reviewed every five years andif not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standardsand should be addressed to ASTM International Headquarters. Your comments will receive carefu
47、l consideration at a meeting of theresponsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you shouldmake your views known to the ASTM Committee on Standards, at the address shown below.This standard is copyrighted by ASTM International,
48、 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the aboveaddress or at 610-832-9585 (phone), 610-832-9555 (fax), or serviceastm.org (e-mail); or through the ASTM website(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http:/ 163