ASTM E1256-2010 Standard Test Methods for Radiation Thermometers (Single Waveband Type)《辐射式温度计的标准试验方法(单波段型)》.pdf

上传人:赵齐羽 文档编号:528274 上传时间:2018-12-05 格式:PDF 页数:7 大小:250.08KB
下载 相关 举报
ASTM E1256-2010 Standard Test Methods for Radiation Thermometers (Single Waveband Type)《辐射式温度计的标准试验方法(单波段型)》.pdf_第1页
第1页 / 共7页
ASTM E1256-2010 Standard Test Methods for Radiation Thermometers (Single Waveband Type)《辐射式温度计的标准试验方法(单波段型)》.pdf_第2页
第2页 / 共7页
ASTM E1256-2010 Standard Test Methods for Radiation Thermometers (Single Waveband Type)《辐射式温度计的标准试验方法(单波段型)》.pdf_第3页
第3页 / 共7页
ASTM E1256-2010 Standard Test Methods for Radiation Thermometers (Single Waveband Type)《辐射式温度计的标准试验方法(单波段型)》.pdf_第4页
第4页 / 共7页
ASTM E1256-2010 Standard Test Methods for Radiation Thermometers (Single Waveband Type)《辐射式温度计的标准试验方法(单波段型)》.pdf_第5页
第5页 / 共7页
亲,该文档总共7页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、Designation: E1256 10Standard Test Methods forRadiation Thermometers (Single Waveband Type)1This standard is issued under the fixed designation E1256; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A num

2、ber in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 The test methods described in these test methods can beutilized to evaluate the following six basic operational param-eters of a radiation

3、 thermometer (single waveband type):SectionCalibration Accuracy 7Repeatability 8Target Size 9Response Time 10Warm-Up Time 11Long-Term Stability 121.2 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this stan

4、dard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Terminology2.1 Definitions:2.1.1 blackbody, nthe perfect or ideal source of thermalradiant power having a spectral distribution described by thePlanck equation.2.1.1.

5、1 DiscussionThe term blackbody is often used todescribe a furnace or other source of radiant power whichapproximates the ideal.2.1.2 center wavelength, na wavelength, usually near themiddle of the band of radiant power over which a radiationthermometer responds, that is used to characterize its perf

6、or-mance.2.1.2.1 DiscussionThe value of the center wavelength isusually specified by the manufacturer of the instrument.2.1.3 radiation thermometer, na radiometer calibrated toindicate the temperature of a blackbody.2.1.4 radiometer, na device for measuring radiant powerthat has an output proportion

7、al to the intensity of the inputpower.2.1.5 target plane, nthe plane, perpendicular to the line ofsight of a radiation thermometer, that is in focus for thatinstrument.2.2 Definitions of Terms Specific to This Standard:2.2.1 reference temperature source, na source of thermalradiant power of known te

8、mperature or emissivity, or both,used in the testing of radiation thermometers.2.2.2 target size, nthe diameter of a circle in the targetplane of a radiation thermometer that is centered on its line ofsight and contains 99 % of the input radiant power received bythat instrument.2.2.3 temperature res

9、olution, nthe minimum simulated oractual change in target temperature that gives a usable changein output or indication, or both.3. Significance and Use3.1 The purpose of these test methods is to establishconsensus test methods by which both manufacturers and endusers may make tests to establish the

10、 validity of the readings oftheir radiation thermometers. The test results can also serve asstandard performance criteria for instrument evaluation orselection, or both.3.2 The goal is to provide test methods that are reliable andcan be performed by a sufficiently skilled end user or manu-facturer i

11、n the hope that it will result in a better understandingof the operation of radiation thermometers and also promoteimproved communication between the manufacturers and theend users. A user without sufficient knowledge and experienceshould seek assistance from the equipment makers or otherexpert sour

12、ces, such as those found at the National Institute ofStandards and Technology in Gaithersburg, Maryland.3.3 Use these test methods with the awareness that there areother parameters, particularly spectral range limits and tem-perature resolution, which impact the use and characterizationof radiation

13、thermometers for which test methods have not yetbeen developed.3.3.1 Temperature resolution is the minimum simulated oractual change in target temperature that results in a usablechange in output or indication, or both. It is usually expressedas a temperature differential or a percent of full-scale

14、value, orboth, and usually applies to value measured. The magnitude ofthe temperature resolution depends upon a combination of fourfactors: detector noise equivalent temperature difference(NETD), electronic signal processing, signal-to-noise charac-teristics (including amplification noise), and anal

15、og-to-digitalconversion “granularity.”1These test methods are under the jurisdiction of ASTM Committee E20 onTemperature Measurement and are the direct responsibility of Subcommittee E20.02on Radiation Thermometry.Current edition approved Dec. 1, 2010. Published January 2011. Originallyapproved in 1

16、988. Last previous edition approved in 2007 as E1256 95 (2007).DOI: 10.1520/E1256-10.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.3.3.2 Spectral range limits are the upper and lower limits tothe wavelength band of radiant energy t

17、o which the instrumentresponds. These limits are generally expressed in micrometers(m) and include the effects of all elements in the measuringoptical path.At the spectral response limits, the transmission ofthe measuring optics is 5 % of peak transmission (see Fig. 1).4. Apparatus4.1 The following

18、apparatus, set up as illustrated in Fig. 2,can be used to perform the standard tests for all six parameters.4.1.1 Reference Temperature SourceA blackbody (orother stable isothermal radiant source of high and knownemissivity) with an opening diameter at least as large as thatspecified in these test m

19、ethods.NOTE 1Typical examples include nearly isothermal furnaces withinternal geometries, such as a sphere with an opening small relative to itsradius, or a right circular cylinder with one end closed having a radiussmall relative to its length.24.1.2 Temperature IndicatorEither contact or radiometr

20、ic,which accurately displays the temperature of the referencetemperature source.4.1.3 Shutter MechanismOf sufficient size so as to com-pletely block the opening of the reference temperature sourcefrom the field of view of the test instrument. The shuttermechanism shall activate in a time interval th

21、at is short whencompared with the response time of the test instrument.4.1.4 Iris DiaphragmOf sufficient size so that when fullyopen the iris diameter is greater than the opening of thereference temperature source. It shall be located with itsopening concentric with and perpendicular to the line of

22、sightof the radiation thermometer.4.1.4.1 The side of the diaphragm facing the radiationthermometer should be blackened (nearly nonreflective) so asto minimize the effect of radiation reflected from the surround-ing environment. In addition the iris should be shaded fromsources of intense extraneous

23、 radiation. (See Note 9).4.1.5 Aperture SetIf an iris diaphragm is not available, anaperture disc set of appropriate diameters can be used. Eachaperture should be blackened and also mounted and protectedfrom extraneous sources of radiation as discussed in 4.1.4.1.4.1.6 Data Acquisition SystemsOf app

24、ropriate speed andstorage capacity to measure and record the output signal of theradiation thermometer in the Response Time Test Method,Section 9.4.1.7 Power SupplyCapable of supplying the proper volt-age and frequency, if necessary, to the radiation thermometer.5. Calibration Accuracy Test Method5.

25、1 SummaryThis test method outlines the procedure tobe used to evaluate the maximum deviation between thetemperature indicated by the radiation thermometer and theknown temperature of a reference temperature source, includ-ing the uncertainty of the reference temperature source relativeto the current

26、 International Temperature Scale.5.2 Test Conditions:5.2.1 Rated supply voltage and frequency.5.2.2 Prescribed warm-up period.5.2.3 After execution of internal standardization check (ifavailable).5.2.4 Emissivity compensation set to one (1).5.2.5 Minimum opening of the reference temperature sourcesh

27、all not obstruct the field of view of the radiation thermometerwith the test aperture as specified by the manufacturer.5.2.6 Laboratory ambient temperature range (20 to 25C).5.2.7 Manufacturer shall specify any special conditionssuch as atmospheric absorption effects, target distance, etc.5.2.8 Manu

28、facturer shall specify the output for determiningthe indicated temperature.5.3 Test Method:5.3.1 The radiation thermometer is sighted at the referencetemperature source whose temperature is sequentially stabi-lized at three calibration points distributed uniformly over themeasurement range of the in

29、strument.5.3.2 The temperature of the reference temperature sourceand the temperature indicated by the radiation thermometer arerecorded, then the difference between the two values iscalculated and recorded (see Fig. 3).5.3.3 The test sequence is repeated twice for the same threecalibration points,

30、and an average temperature difference iscalculated and recorded for each calibration point.5.4 Test ResultThe value for the calibration accuracy ofthe temperature indication of the radiation thermometer istaken to be the largest of the three average temperaturedifferences determined in 5.3.2 plus or

31、 minus the uncertainty ofthe temperature of the reference temperature source relative tothe current International Temperature Scale.NOTE 2The calibration accuracy is generally expressed as a tempera-ture difference or a percent of full-scale value, or both.NOTE 3The value applies across the entire m

32、easurement range.NOTE 4If the reference temperature source is measured with otherthan a calibrated reference or secondary standard radiation thermometer,then the emissivity of the source enters into the calibration of the testradiation thermometer.2DeWitt, D. P., and Nutter, G. D., eds., “Theory and

33、 Practice of RadiationThermometry,” John Wiley and Sons, New York, NY.FIG. 1 Spectral Range LimitsE1256 1026. Procedure6.1 Detailed directions for evaluation of each parameterlisted in 1.1 are included in each parameter test method.6.2 Each parameter test method is organized by: parameterterm, summa

34、ry, test conditions, test method, test result, andapplicable notes.7. Repeatability Test Method7.1 SummaryThis test method outlines the procedure tobe used to evaluate the repeatability of the temperatureindication of a radiation thermometer for a number of consecu-tive measurements made under the s

35、ame conditions over aspecified interval of time.7.2 Test Conditions:7.2.1 Rated supply of voltage and frequency.7.2.2 Prescribed warm-up period.7.2.3 After execution of internal standardization check (ifavailable).7.2.4 Diameter of the reference temperature source openingshall be greater than the ra

36、diation thermometer target size, asspecified by the manufacturer.7.2.5 Laboratory ambient temperature range (20 to 25C).FIG. 2 Test Method ApparatusFIG. 3 Worksheet for Calibration Accuracy Test MethodE1256 1037.2.6 Emissivity compensation, if any, set to one (1).7.2.7 Manufacturer shall specify any

37、 special conditionssuch as response time, atmospheric absorption effects, targetdistance, etc.7.3 Test Method:7.3.1 Once a day for twelve consecutive working days, theradiation thermometer is sighted at the reference temperaturesource whose temperature is stabilized at the approximatemidpoint of the

38、 radiation thermometer calibration range.NOTE 5The selected reference temperature source temperature shallbe reproduced for each of the twelve consecutive tests.7.3.2 The temperature of the reference temperature sourceand the temperature(s) indicated by the radiation thermometerduring each days test

39、 are recorded.7.3.3 The radiation thermometer shall be switched off aftereach series of measurements.7.4 Test ResultThe value for the repeatability of thereadings of the radiation thermometer is taken to be thestandard deviation of the twelve recorded readings.S.D. 5(i 5 1NXi2 X!2N 2 15 !(i 5 1NXi22

40、(i 5 1NXi!2NN 2 1where:S.D. = standard deviation,N = number of measurements,Xi= value of the ith measurement, andX=average of the twelve measurements =(i 5 1NXiN.NOTE 6The repeatability of the temperature indication is generallyexpressed as a temperature difference or a percent of full-scale value,

41、orboth.NOTE 7The value for the repeatability can be applied across the entiremeasuring temperature range, or, the same test can be performed at otherselected temperatures across the measuring temperature range in order toassess the repeatability of the radiation thermometer at those temperatures.8.

42、Target Size Test Method8.1 SummaryThis test method outlines the procedure tobe used to evaluate the diameter of the circle located in thetarget plane of the reference temperature source, at a knowndistance along and perpendicular to a radiation thermometersline of sight, and from which 99 % of the r

43、adiant powerreceived by the radiation thermometer is collected (see Figs. 3and 4).8.2 Test Conditions:8.2.1 Rated supply voltage and frequency.8.2.2 Prescribed warm-up period.8.2.3 After execution of internal standardization check (ifapplicable).8.2.4 Laboratory ambient temperature range (20 to 25C)

44、.8.2.5 Minimum opening of the reference temperature sourceshall be large enough so as to not obstruct the optical path ofthe radiation thermometer, as specified by the manufacturer,when it is sighted through an aperture that is twice the diameterof the instruments target size at the plane of the ape

45、rture.NOTE 8Some radiation thermometers have a target size so large thata commercially available reference temperature source cannot be used; aseparate test method is under preparation for use in such cases.8.2.6 Manufacturer shall specify any special conditionssuch as atmospheric absorption effects

46、, distance, how andwhen to clean the radiation thermometer lens, etc.8.3 Test Method:8.3.1 The temperature of the reference temperature source isstabilized at a value near the top of the calibration range of theradiation thermometer.8.3.2 The iris is positioned in the front of and concentricwith the

47、 opening of the reference temperature source (asillustrated in Fig. 2). The iris is then adjusted to a diameterslightly smaller (typically 10 %) than the expected targetdiameter.NOTE 9The iris should be kept cool enough so that its thermalemission does not contribute significantly to the output sign

48、al. Uncoveringthe iris quickly can minimize heating, but this requires care. Evaluation ofthe error from this source requires computational procedures beyond thescope of this test method.2In most cases, however, the error is insignifi-cant if the iris is maintained near room temperature (20C) and th

49、ereference temperature source temperature is at or above 200C.8.3.3 The position of the radiation thermometer is thenadjusted vertically and horizontally and focused to producemaximum output while also maintaining the line of sightperpendicular to the iris.8.3.4 The iris is then opened to the point where thetemperature indicated by the radiation thermometer stopsincreasing, but its diameter is still smaller than the referencetemperature source opening.8.3.5 The temperature indicated by the radiation thermom-eter (in degrees) is recorded.F

展开阅读全文
相关资源
猜你喜欢
相关搜索

当前位置:首页 > 标准规范 > 国际标准 > ASTM

copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
备案/许可证编号:苏ICP备17064731号-1