1、Designation: F1375 92 (Reapproved 2012)Standard Test Method forEnergy Dispersive X-Ray Spectrometer (EDX) Analysis ofMetallic Surface Condition for Gas Distribution SystemComponents1This standard is issued under the fixed designation F1375; the number immediately following the designation indicates
2、the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.INTRODUCTIONSemiconductor clean rooms are serviced by high-p
3、urity gas distribution systems. This test methodpresents a procedure that may be applied for the evaluation of one or more components considered foruse in such systems.1. Scope1.1 This test method establishes a procedure for comparingthe elemental composition of normal surfaces with any defectsfound
4、 on the surfaces of metal tubing, fittings, valves, or anymetal component.1.2 This test method applies to all steel surfaces of compo-nents such as tubings, connectors, regulators, and valves,regardless of size, style, or type.1.3 Limitations:1.3.1 This test method is intended for use by scanningele
5、ctron microscope/energy dispersive x-ray spectrometer(SEM/EDX) operators with skill level typically achieved overa twelve-month period.1.3.2 SEM used for this study should conform to thoselimitations outlined in Test Method F1372 and should have aminimum point-to-point resolution of 30 nm.1.4 The va
6、lues stated in SI units are to be regarded as thestandard. The inch-pound units given in parentheses are forinformation only.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish app
7、ro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. Specific hazardstatements are given in Section 6.2. Referenced Documents2.1 ASTM Standards:2F1372 Test Method for Scanning Electron Microscope(SEM) Analysis of Metallic Surface Condition fo
8、r GasDistribution System Components3. Terminology3.1 Definitions of Terms Specific to This Standard:3.1.1 normal surfacean area of the sample that does notexhibit any visible defect when viewed under the stipulatedSEM magnification. Normal surface is used to provide abaseline for comparison with any
9、 area exhibiting a defect.3.1.2 sample anglethe angle measured normal to theincoming electron beam.3.1.3 standard conditions101.3 kPa, 0.0C (14.73 psia,32.0F).3.1.4 working distancethe measured distance from thebottom of the objective lens to the sample.4. Significance and Use4.1 The purpose of this
10、 test method is to define a procedurefor testing components being considered for installation into ahigh-purity gas distribution system. Application of this testmethod is expected to yield comparable data among compo-nents tested for purposes of qualification for this installation.5. Apparatus5.1 Ma
11、terials:1This test method is under the jurisdiction of ASTM Committee F01 onElectronics and is the direct responsibility of Subcommittee F01.10 on Contamina-tion Control.Current edition approved July 1, 2012. Published August 2012. Originallyapproved in 1992. Last previous edition approved in 2005 a
12、s F137592(2005). DOI:10.1520/F1375-92R12.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM Inter
13、national, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States15.1.1 Mounting Stubs, specific to the instrument used arerequired.5.1.1.1 Samples shall not be coated with a conductive thinlayer (for example, gold or carbon).5.1.2 Conductive Paste/Tape, that will provide
14、 a conductivepath. Use any means of fixing the sample to a stub. Care shouldbe taken not to contaminate the area of interest.5.1.3 Adhesives, used to attach samples to sample stubs areto be vacuum stable.5.2 Instrumentation:5.2.1 Scanning Electron Microscope (SEM)Any highresolution commercially avai
15、lable SEM with photographiccapabilities of a 100 cm2image may be used for theseanalyses.5.2.2 Instrument Operating Parameters , shall be as fol-lows: accelerating voltage, 20 KeV; final aperture size nominal200 m or less; and working distance and sample tilt, asappropriate to the EDX detector.5.2.2.
16、1 SEM instrument operating parameters shall be suchthat collection efficiency for the EDX spectrometer is opti-mized.5.2.3 EDX Spectrometer, capable of full width half maxi-mum (FWHM) resolution of 170 eV or less (for MnKa), andcapable of detecting all elements with an atomic numbergreater than or e
17、qual to that of sodium (Na).5.2.4 Printer or Plotter, capable of accurate spectral repro-duction (linear-linear) is required.6. Hazards6.1 Observe all normal and acceptable precautions regard-ing use of high voltage, X-ray producing equipment. Observestandard and routine cryogenic handling procedure
18、s.6.2 Use adhesives in such a manner that they do notcontaminate the area of interest.7. Sampling, Test Specimens, and Test Units7.1 Sample Cutting and Mounting :7.1.1 Use any mechanical cutting method that minimizesalteration of the surface. A dry, clean hacksaw is preferred.7.1.2 After cutting, cl
19、ean samples with a reagent gradesolvent and rinse with a reagent grade isopropyl alcohol (IPA).Place prepared samples in a resealable non-outgassing con-tainer under nitrogen.7.1.3 Mount the samples on the instrument stub.7.2 Conduct sample preparation to ensure that the tempera-ture of the sample d
20、oes not exceed 90C (194F).7.3 Mount the samples onto SEM compatible mounts in amanner to avoid contamination of the surface to be analyzed.Non-X-ray generating substrates, such as graphite, are pre-ferred as mounting stubs.8. Calibration8.1 Calibrate and maintain instruments using standard labo-rato
21、ry practices and manufacturers recommendations. Cali-brate EDX spectrometers according to the manufacturersspecifications so that the energy calibration falls within 6 1channel.8.2 Magnification for qualitative and quantitative analysisshall result in incident beam concentration on the surfaceanomal
22、y, minimizing stray X-ray signal from the background.9. Procedure9.1 Follow sample preparation of this test method (7.1)toexpose the surface.9.2 Introduce the sample stub into the SEM vacuum cham-ber.9.3 Activate the electron beam when vacuum conditionsmeet those recommended by the manufacturer:9.4
23、Move the sample until an area of interest on thesamples surface comes into focus. The area of interest shouldbe representative of a normal surface, avoiding gross deformi-ties.9.5 Orient the sample (with respect to working distance,sample tilt, etc) to maximize X-ray collection efficiency of theEDX
24、detector.9.6 Adjust accelerating voltage to provide maximum exci-tation for the element of interest. Typically, this is 20 KeV forall elements having an atomic number greater than or equal toeleven (the atomic number of sodium) and 10 KeV for thoseelements with atomic numbers between boron and sodiu
25、m.9.7 Collect X-ray signals for a minimum of 100 s from acontrol area.9.8 Move sample to that area showing surface anomaly andacquire X-ray signals for a minimum of 100 s.9.9 Identify peaks and label the spectrum appropriately.9.10 Print or plot the spectrum (see Fig. 1).9.11 Photograph the surface
26、anomaly at a magnification bestsuited to document the anomalys physical characteristics.9.12 Repeat 9.5-9.11 for all areas of interest.9.13 Turn off the SEM electron beam and remove thesample from the vacuum chamber.10. Report10.1 Report the following information:10.1.1 Data Analysis The first sampl
27、e data collectionmust be taken from a selected site that is representative of thebest normal surface available for the sample.10.1.2 Data Presentation:10.1.2.1 All elements not of the base metal will be consid-ered unusual and shall be listed in tabular form with thenumber of particles demonstrating
28、 the presence of thoseelements being recorded.10.1.2.2 Data shall be presented in the form of linear-linearprinted or plotted spectra (see Fig. 1). The plotted spectral scaleshall be such that the smallest peak can easily be discerned. Ifpeak height differences are such that adequate representation
29、ofall peaks cannot be made from the same plot, two spectral plotsshall be made using different scale factors.F1375 92 (2012)210.1.2.3 The EDX spectra and corresponding photographsshould be appropriately labeled so that the elemental compo-sition of any specific defect, particle, or anomaly is readil
30、yapparent to any third party.FIG. 1 Relative Abundance of Elements of an InclusionF1375 92 (2012)310.1.2.4 EDX spectra and related photomicrographs mustinclude the following information: sample identification, date,peak identification, tilt angle, and voltage.10.1.3 All data reported must identify t
31、he SEM and EDXequipment manufacturer and model number.10.1.4 Any special modifications in equipment or procedurenecessary to acquire data must be documented and fullydescribed.11. Precision and Bias11.1 Precision and bias for this test method are beingdetermined.12. Keywords12.1 components; connecto
32、rs; contamination; EDX; gasdistribution components; SEM/EDX; semiconductor process-ing; surface condition; tubingASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentionedin this standard. Users of this standard are expressly advi
33、sed that determination of the validity of any such patent rights, and the riskof infringement of such rights, are entirely their own responsibility.This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years andif not revised, either
34、reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standardsand should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of theresponsible technical committee, which you may attend. If
35、 you feel that your comments have not received a fair hearing you shouldmake your views known to the ASTM Committee on Standards, at the address shown below.This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,United States. Individ
36、ual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the aboveaddress or at 610-832-9585 (phone), 610-832-9555 (fax), or serviceastm.org (e-mail); or through the ASTM website(www.astm.org). Permission rights to photocopy the standard may also be secured from the ASTM website (www.astm.org/COPYRIGHT/).F1375 92 (2012)4