ASTM UOP1025-2018 Density of Viscous Liquids and Semi-Solid Hydrocarbons by Nitrogen Displacement.pdf

上传人:sumcourage256 文档编号:540824 上传时间:2018-12-07 格式:PDF 页数:4 大小:258.75KB
下载 相关 举报
ASTM UOP1025-2018 Density of Viscous Liquids and Semi-Solid Hydrocarbons by Nitrogen Displacement.pdf_第1页
第1页 / 共4页
ASTM UOP1025-2018 Density of Viscous Liquids and Semi-Solid Hydrocarbons by Nitrogen Displacement.pdf_第2页
第2页 / 共4页
ASTM UOP1025-2018 Density of Viscous Liquids and Semi-Solid Hydrocarbons by Nitrogen Displacement.pdf_第3页
第3页 / 共4页
ASTM UOP1025-2018 Density of Viscous Liquids and Semi-Solid Hydrocarbons by Nitrogen Displacement.pdf_第4页
第4页 / 共4页
亲,该文档总共4页,全部预览完了,如果喜欢就下载吧!
资源描述

1、 t IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCR

2、IBED HEREIN CAN BE HAZARDOUS. SAFETY DATA SHEETS (SDS) OR EXPERIMENTAL SAFETY DATA SHEETS (ESDS) FOR ALL THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 2018 UOP LLC. All rights reserved. Nonconfidential UOP Meth

3、ods are available from ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or 610.832.9585 PHONE. Densi

4、ty of Viscous Liquids and Semi-Solid Hydrocarbons by Nitrogen Displacement UOP Method 1025-18 Scope This method is for determining the true density of viscous liquids and semisolid hydrocarbons by displacement of nitrogen gas. This method is written for use with a Micromeritics AccuPyc II 1340 Pycno

5、meter; however, equivalent equipment may be used. References Owners Manual and Quick Reference Guide, AccuPyc II 1340 Pycnometer, Micromeritics Corporation UOP Method 851, “Density of Powders and Solids by Helium Displacement,” www.astm.org UOP Method 999, “Precision Statements in UOP Methods,” www.

6、astm.org Outline of Method The pycnometer works by measuring the pressure drop of a gas as it is expanded into a known volume. The observed pressure change is used to compute the volume of the sample. The measured sample volume excludes all open pore structure as the gas molecules will rapidly fill

7、the pores of the sample; only the truly solid phase of the sample displaces the gas. Apparatus References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Balance, readability, 0.1-mg Gas Pycnometer, Micromeritics, AccuPyc II 1340 Oven,

8、capable of operation up to 150C Regulator, nitrogen, two-stage, high purity, delivery pressure range 15-200 kPa (2-30 psi), Matheson Tri-Gas, Model 3121-580 Tongs or Tweezers, suitable for removing sample cups from the oven 2 of 4 1025-18 Reagents and Materials Cloth, lint-free Nitrogen, gas, cylind

9、er, 99.995% minimum purity, local supply Sample Cup, Aluminum, 3.3 cc, Micromeritics Cat no. 134-25600-00 Procedure The analyst is expected to be familiar with general laboratory procedures, the technique of pycnometer measurement, and the equipment being used. Set up the Micromeritics AccuPyc II 13

10、40 Gas Pycnometer as outlined in the manufacturers instruction manual. Calibration volumes are included with the instrument. Be certain that gauges are calibrated and properly zeroed before proceeding. Specific Instructions for Micromeritics AccuPyc II 1340 Nitrogen Pycnometer The following detailed

11、 procedure for using the Micromeritics AccuPyc II 1340 Pycnometer describes the technique for making measurements and is used for the empty cell as well as the cell containing the sample within a disposable cup. If a different instrument is used, follow instructions supplied with that instrument. 1.

12、 Ensure that the instrument is turned on and that the nitrogen is flowing to it. 2. Ensure that the cell cooling is turned on and temperature is stable. Depending on instrument configuration, this could be from a circulating water bath or direct thermoelectric cooling. 3. Place samples in an oven to

13、 melt, if solid at room temperature. Sample should be liquid and pourable after heating. Stir thoroughly to mix before taking a portion of the sample. 4. Obtain a tare weight from an empty, disposable sample cup. 5. Fill sample cup approximately full with sample and reweigh. 6. Subtract the tare wei

14、ght from the filled sample cup and record the sample weight to the nearest 0.1 mg. 7. If the sample was solid at room temperature, then place the cup back into the oven and re-melt for 15-20 minutes to allow any bubbles or voids to fill. 8. Cool the sample to room temperature before placing it into

15、the measurement cell. 9. Remove the cell chamber cap. Do not remove the cell chamber cap when the pycnometer is pressurized. 10. Insert the sample cup into the cell chamber. 11. Replace the cell chamber cap. 12. On the PC, select the instrument unit and click sample analysis. Select the correct file

16、 in the directory column or create a new file. 13. Enter necessary sample information and correct sample mass. Ensure the 3.5 cc chamber insert option is selected. 14. Click start to begin analysis. 15. Obtain data from the report printed on the screen. 3 of 21 1025-18 16. The instrument will beep o

17、nce it has depressurized itself and is safe to open. Remove the cell chamber cap. 17. Remove the sample cup from the chamber and dispose of appropriately. 18. For further analyses, repeat procedure starting with Step 3, above. Calculations All calculations are performed by the instrument using the s

18、ample weight as entered. Report results to the nearest 0.001 g/cm3. Vsamp = Vcel - VexpP1P2-1Dsamp= MsampVsampwhere Vcel = sample chamber volume Vexp = expansion chamber volume Vsamp = sample volume Msamp = sample mass Dsamp = sample density P1 = gauge pressure after fill P2 = gauge pressure after e

19、xpansion Notes 1. Before setting regulator pressure, make sure the tank pressure for the gas regulator is at least 200 psig. Pressures less than 200 psig may cause the sample to be inadequately saturated, resulting in inaccurate data or termination of analysis. 2. The cell chamber and cap must be ke

20、pt clean and lightly greased at all times. Particles on the cap seating surface, in the sample cup, under the sample, or clinging to the sample chamber wall may cause inaccurate results. Inspect the cell and cap before each use. Use a lint-free cloth to remove any dust or particles. Very lightly gre

21、ase O-ring on cell chamber cap. 3. Keep the cap on the cell chamber except when inserting or removing a sample. If the chamber remains uncapped, temperature instability will occur which could affect analysis results. Precision Precision statements were determined using UOP Method 999. Repeatability

22、and Site Precision A nested design was carried out for determining the density of one sample with two analysts in one laboratory. Each analyst carried out tests on two separate days, performing two tests on each sample each day. The precision data are summarized in Table 1. Two tests performed by th

23、e same analyst on the same day should not differ by more than the repeatability allowable difference with 95% confidence. Two tests performed in one laboratory by different analysts on different days should not differ by more than the site precision allowable difference with 95% confidence. 4 of 4 1

24、025-18 Table 1 Repeatability and Site Precision, g/mL Repeatability Site Precision Sample Mean Within Day esd Allowable Difference Within Lab esd Allowable Difference A 1.0037 0.00008 0.0003 0.00034 0.0021 The data in Table 1 is a short-term estimate of repeatability. When the test is run routinely,

25、 a control standard and chart should be used to develop a better estimate of the long-term repeatability. Reproducibility There is insufficient data to calculate reproducibility of the test at this time. Time for Analysis The elapsed time and labor requirement for a single analysis are identical, 1 hour. Suggested Suppliers Matheson Tri-Gas, 166 Keystone Drive, Montgomeryville, PA 18936 (215-648-4000) www.matheson- Micromeritics Instrument Corporation, One Micromeritics Drive, Norcross, GA 30093-1877 (770-662-3620)

展开阅读全文
相关资源
猜你喜欢
相关搜索

当前位置:首页 > 标准规范 > 国际标准 > ASTM

copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
备案/许可证编号:苏ICP备17064731号-1