1、-. :s *- A-A-5L308 b5 m 7779974 0057Ll4b T m C-9b-3+5 A-A-51 3 O 8 12 March 1986 COMMERCIAL ITEM DESCRIPTION SPLINT, ARM and SPLINT, LEG (Pneumatic) The General Services Administration has authorized the use of this commercial item description in preference to Military Specification MIL-S-37025A. Th

2、is commercial item description covers transparent, plastic, mouth inflatable, pneumatic splints, with full length zipper closure, for the emergency treatment of fractures or superficial bleeding of the full adult arm or leg. Salient characteristics. The splint shall be adult size, double-wall plasti

3、c construction, with full length zipper closure, and a valve with mouthpiece for inflation by mouth. The splint shall be one of the following types: Type I - Splint, Arm, Pneumatic, Adult, Zipper Closure. Type II - Splint, Leg, Open Toe, Pneumatic, Adult, Zipper Closure. The splint shall be construc

4、ted in accordance.with DPSC Drawing No. 23187 to the revision in effect on the date of solicitation. Materials used shall contain no ingredients known to be skin sensitizers or irritants. The splint shal be constructed of clear, transparent film conforming to the requirements of Federal Speciication

5、 L-P- 375 Type II Class 1, except for the following paragraphs of L-P-375: 3.3.2 Type II, Table II. Delete the requirements for Stiffness and Blocking. Cold Crack Test delete -40 degrees and substitute O degrees. AMSC N/A FSC 6515 DISTRIBUTION STATEMENT A. Approved for public release; distribution i

6、s unlimited. I- Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,- A-A-51308 65 7799774 0057447 1 COMMERCIAL ITEM DESCRIPTION A-A-51308 3.7 Finish. Delete entirely and- substitute: “3.7 Finish. The film shall be double polished, conforming to a 60 degr

7、ee specular gloss of 140 units on each surface when tested on a dull background as specified in ASTM-D-2457.“ The outer wall shall be 12 gage (0.012 inch nominal), the inner wall shall be 8 gage (0.008 inch nominal), and the zipper reinforcing strips shall be between 0.008 and 0.020 inch material th

8、ickness. The zipper closure shall conform to the requirements o Federal Specification V-F-106 except it shall be fabricated of a polyester coil with white enameled die cast zinc slide and end stops, size L, LS or LM. The arm splint shall be Type I Style 3, nonseparating, nonreversible. The leg splin

9、t shall be Type IV Style 8, separating, nonreversible. Tape shall be 9/16 inch wide, white open weave (Leno) cotton or white cotton impregnated with a clear vinyl plastic. The leg splint zipper pull shall include a thong fabricated of nylon cord conforming to Type IA (coreless) of MIL-C-5040. The Co

10、re may be removed. The ends of the cord shall be tied by means of a non-slippable knot to form a loop 2 inches 5 1/4 inch in length. Strands of cord ends shall be fused. Grommets on the leg splint shall be fabricated from polyvinyl chloride. Shall be Halkey-Roberts Inc. No. 44GF or No. 44GH, or equa

11、l. The valve for opening and closing the air inflation passage shall be a screw type with knurled external member to which a plastic mouthpiece is assembled. The valve housing shall be fabricated from a polyvinyl chloride compatible with the splint plastic. The valves shall be Halkey-Roberts No. 40

12、ASE or equal for the arm splint and No. 240 ASE or equal for the leg splint. The inner and outer walls, zipper reinforcing strips, zipper tapes and valve assembly shall be heat sealed to make an air tight end item. 2 O Provided by IHSNot for ResaleNo reproduction or networking permitted without lice

13、nse from IHS-,-,-T A-A-51308 65 U 9979974 QI357448 3 i COMMERCIAL ITEM DESCRITPTION A-A-51308 Each splint shall be loosely folded in half lengthwise in such a manner that the fold is not creased. A cushioning medium such as plastic air bubbles or other suitable material may be placed between the fol

14、ds of the splint to prevent the folded edge from being creased. With the zipper in the closed position, the folded splint shall be inserted into the immediate container which shall be an air-tight bag fabricated of polyethylene having a minimum thickness of 0.002 inch. The bag shall be heat sealed a

15、fter excess air has been evacuated. Instructions for use shall be permanently marked on the immediate container. In addition, the following statement shall be located below the instructions for use in 3/16 minimum capital letters: “CAUTION - USE IN AIRCRAFT REQUIRES MONITORING THE SPLINT PRESSURE TO

16、 PREVENT NEUROVASCULAR INJURY. LIMIT INFLATION PRESSURE TO ADEQUATE FRACTURE SUPPORT ONLY.“ The following information shall be printed in 1/8 inch minimum letters: “The aneroid pressure gage of a standard sphygmomanometer should be available for monitoring the splint pressure during ascent and desce

17、nt. Splints air transported should be inflated to provide adequate support of the fractured limb (below 50 mm Hg). Over-inflation is to be specifically avoided. I As an alternate, the above markings shall be legibly printed on a separate sheet of paper which shall be inserted in the immediate contai

18、ner in such a manner as to permit ready identification. Leakage Test. The splint shall be tested for leakage by closing the zipper inflating the splint to a pressure of 60 mm Hg, closing the valve and immersing in water. There shall be no air bubble escape from the immersed splint. Reference specifi

19、cation. Where Military or Federal Specifications, ASTM Stnadards or drawings are referenced herein, the issue of that specification standard or drawing in effect on the date of solicitation shall be used to determine compliance with stated requirements. Workmanship. The splint shall be free from def

20、ects which detract from its appearance or impair its serviceability. - Unit. Each.-(EA) . One splint, as specified, constitutes one unit. - - 3 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-51308 65 m 777777i.1 00579i.17 5 m COMMERCIAL ITEM DES

21、CRIPTION A-A-51308 Quality Assurance Provisions. Responsibility for inspection. Unless otherwise specified in the contract or purchase order, the contractor is responsible for the performance of all inspection requirements as specified herein. Except as otherwise specified in the contract or purchas

22、e order, the contractor may use his own or any facilities suitable for the performance of the inspection requirements specified herein, unless disapproved by the Government. The Government reserves the right to perform any of the inspections set forth in the specification where such inspections are

23、deemed necessary to assure supplies and services conform to prescribed requirements. Records. Records of examinations and tests performed by or for the contractor shall be maintained. by the contractor and made available to the Government upon the Governments request, at any time, or from time to ti

24、me, during the performance of the contract and for a period of three years after delivery of the supplies to which such records relate. i Inspection. Inspection, as used herein, is defined as both examination (such as visual or auditory investigation without the use of special laboratory appliances

25、or procedures) and testing (determination by technical means of physical and chemical properties) of the item. accordance with commercial practice. The unit of product for tests purposes shall be one splint. Samplinq for tests. Sampling for tests shall be conducted in Inspections and tests. Inspecti

26、ons and tests shall be conducted in accordance with commercial practice to determine compliance with specification requirements. Where feasible, the same sample shall be used for the determination of two or more test characteristics. Samplinq for inspections and tests. Sampling for inspection and te

27、sts shall be conducted in accordance with commercial practice. The unit of product for test purposes shall be one splint. Contractor, certification. The contractor shall certify that the product offered meets the salient characteristics of this description and conforms to the producers own drawings,

28、 specifications, standards, and quality assurance practices. The Government reserves the right to require proof of such conformance prior to first delivery and thereafter as may be otherwise provided for under the provisions of the contract. 4 Provided by IHSNot for ResaleNo reproduction or networki

29、ng permitted without license from IHS-,-,-A-A-5130 65 7779774 0057i.150 1 m COMMERCIAL ITEM DESCRIPTION A-A-51308 Regulatory requirements. Federal Food, Druq and Cosmetic Act. If the product covered by this document has been determined by the U. S. Food and Drug Administration to be under its jurisd

30、iction, the offeror/ contractor shall comply, and be responsible for compliance by its subcontractors/suppliers, with the requirements of the Federal Food, Drug and Cosmetic-Act, as amended, and regulations promulgated thereunder. In addition, the offeror/contractor shall comply, and be responsible

31、for compliance by its subcontractors/suppliers, with the requirements of all other applicable Federal, State and local statutes, ordinances, and regulations. Recovered materials. The offeror/contractor is encouraged to use recovered material in accordance with Public Law 94-580, as amended, and as i

32、mplemented by Federal Acquisition Regulation Subpart 23.4 to the maximum extent practical. Preservation, packaqing, packinq, labeling, and Markinq. Unless otherwise specified, preservation, packaging, and packing shall be to a degree of protection to preclude damage to containers and/or contents the

33、reof under normal shipping conditions, handling, etc., involving shipment from the supply source to the receiving activity, plus reshipment from receiving activity, and shall conform to applicable carriers rules and regulations. Intermediate and exterior package quantities and labeling and marking s

34、hall be as specified in the contract and/or order. NOTE: The following National Stock Number is covered by this document: 6515-00-935-6592 SPLINT, ARM, Pneumatic, Adult, Zipper Closure 6515-00-935-6593 SPLINT, LEG, Pneumatic, Adult, Zipper Closure Orderinq data. (Intermediate/exterior package quanti

35、ties and labeling and marking must-be specified in the contract and/or order). 5 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-COMMERCIAL ITEM DESCRIPTION A-A-51308 MILITARY INTERESTS: PREPARING ACTIVITY: DoD - ME Custodians : Army - MD Navy - MS Air Force - 03 Agent: DLA-DM CIVIL AGENCY COORDINATING ACTIVITIES: VA-OSS PHS FDA-MPQAS Project No. 6515-3526 Location: CID/SPLINT/17 &UsS. GOVERNMENT PRINTING OFFICE: 19119 - 604-033/126r) 6 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-