DOD A-A-54016-1993 REFRIGERATOR SOLID STATE BIOLOGICALS《固态生物学冷藏库》.pdf

上传人:ownview251 文档编号:701702 上传时间:2019-01-01 格式:PDF 页数:6 大小:255.64KB
下载 相关 举报
DOD A-A-54016-1993 REFRIGERATOR SOLID STATE BIOLOGICALS《固态生物学冷藏库》.pdf_第1页
第1页 / 共6页
DOD A-A-54016-1993 REFRIGERATOR SOLID STATE BIOLOGICALS《固态生物学冷藏库》.pdf_第2页
第2页 / 共6页
DOD A-A-54016-1993 REFRIGERATOR SOLID STATE BIOLOGICALS《固态生物学冷藏库》.pdf_第3页
第3页 / 共6页
DOD A-A-54016-1993 REFRIGERATOR SOLID STATE BIOLOGICALS《固态生物学冷藏库》.pdf_第4页
第4页 / 共6页
DOD A-A-54016-1993 REFRIGERATOR SOLID STATE BIOLOGICALS《固态生物学冷藏库》.pdf_第5页
第5页 / 共6页
点击查看更多>>
资源描述

1、A-A-54OLb 9999974 0142235 435 A-A-5 4 O 16 19 March 1993 COMMERCIAL ITEM DESCRIPTION REFRIGERATOR, SOLID STATE BIOLOGICALS The General Services Administration has authorized the use of this Commercial Item Description. This Commercial Item Description covers a portable solid state biologicals refrig

2、erator. Salient characteristics: The refrigerator shall be capable of maintaining operating temperatures between 33.8-42.8O, inclusive within all areas of the storage compartment. The refrigerator shall be capable of cooling the minimum gayload of whole blood units described below from 50F to 43 F,

3、at an ambient temperature of 81F, within seven hours. The refrigerator shall be capable of maintaining the payload temperature below 10C following power loss at 77OF ambient temperature for a minimum of six hours providing the appropriate 34OF-43OF storage temperature existed prior to power failure.

4、 It shall be suited for use in all rugged field environments within temperatures ranging from -25OF through 120F, inclusive. The unit shall pull down to 6OC within 7 hours at an ambient temperature of 27.2OC (after being grecooled to 6OC, with the full load of blood pre-cooled to 10 C). The refriger

5、ator with all accessories in place and containing a minimum payload of 30 whole blood units of 450 ml each or 50 packed red blood cell units of 270 ml each or the equivalent of either of the previous shall not exceed 125 pounds in weight and 8.5 cubic feet in external volume. Outside dimensions of u

6、nit shall not exceed 22.0 inches in depth or 32.0 inches in width (length). Volumes and dimensions do not include recording thermometer. Legs are not required on the refrigerator. AMSC N/A FSC 4110 DISTRIBUTION STATEMENT A: Approved for public release; distribution is unlimited. Provided by IHSNot f

7、or ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-54 O 16 The units internal storage capacity shall be capable of storing the minimum payload as described above, It shall have an easily removable basket(s) to facilitate loading and unloading, The refrigerator shall be

8、 equipped with a temperature indicator, and an audible and visual alarm system which will alert the operator that an abnormal temperature condition exists or existed. The temperature indicator and alarm system shall be flush mounted as an integral part of the unit. The alarm shall be activated when

9、the temperature falls outside the acceptable 34OF-43OF temperature range or in the event of electrical power loss. The alarm shall have a minimum dB rating of 60. The power source for the alarm system shall be separate from that of the refrigerator and shall incorporate a continually recharging batt

10、ery power source or utilize a standard commercially available battery in the Federal Supply System. The capability to measure battery voltage or indicate low battery voltage shall be included in the alarm system. The unit shall be certified by an independent testing laboratory to be in compliance wi

11、th UL 471, Section 49, Rain Test; Section 37, Leakage Current Test; and Section 44, Dielectric Voltage Withstand Test. The unit shall be supplied with a recording thermometer to provide a continuous hard copy recording of at least one week of storage compartment temperatures under non-transit condit

12、ions. The recording thermometer shall record all temperatures between minus 10C to plus 20C, in increments of 2OC as a minimum. supplied as an integral component of the thermometer alarm system or as a separate accessory item (packaged with the refrigerator), The maximum interval between points of r

13、ecord shall be 5 minutes. If the temperature recorder is supplied as an accessory, it shall have a carrying handle, The recording thermometer shall be The unit shall be capable of operating from 110/220 VAC, 50/60 Hz and 12/24/28 VDC power supplies. AC and DC input power circuits shall be protected

14、via circuit breakers. Control and power circuits shall be isolated from input line power circuit. A minimum of 2 carrying handles shall be affixed to the refrigerator to facilitate loading and unloading the unit from vehicles and carrying the unit to the operational site. 2 Provided by IHSNot for Re

15、saleNo reproduction or networking permitted without license from IHS-,-,-A-A-5401b m 9999974 O142237 208 m A-A-54 O 16 The refrigerator shall be durably constructed to withstand normal field handling and vibration peculiar to wheeled vehicle movement over roads, trails, and cross-country. The unit s

16、hall be nonreflective “olive drab“ in color. The units are not subject to biological and chemical warfare cleaning. Workmanship. The solid state biologicals refrigerator shall be free from defects which detract from its appearance or impair its serviceability. Unit. Each (EA). One refrigerator, as s

17、pecified, constitutes one unit. Quality Assurance Provisions. Responsibility for inspection. Unless otherwise specified in the contract or purchase order, the contractor is responsible for the performance of all inspection requirements as specified herein. Except as otherwise specified in the contra

18、ct or purchase order, the contractor may use his own or any facilities suitable for the performance of the inspection requirements specified herein, unless disapproved by the Government. The Government reserves the right to perform any of the inspections set forth in the specification where such ins

19、pections are deemed necessary to assure supplies and services conform to prescribed requirements. Records. Records of examinations and tests performed by or for the contractor shall be maintained by the contractor and made available to the Government upon the Governments request, at any time, or fro

20、m time to time, during the performance of the contract and for a period of three years after delivery of the supplies to which such records relate. Inspection. Inspection, as used herein, is defined as both examination (such as visual or auditory investigation without the use of special laboratory a

21、ppliances or procedures) and testing (determination by technical means of physical and chemical properties) of the item. Tests. Tests shall be performed on a first article sample, as follows, to determine compliance with specification requirements. Where feasible, the same sample shall be used for t

22、he determination of two or more test characteristics. 3 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-5401b 4 OLlr2238 144 A-A-54 O 16 1 unit w racks . Examine the unit for removable storage racks. Load the pith thirty blood bags of 450 ml each

23、. Examine the storage for bending or breaking. 2. Weigh the unit loaded with thirty blood bags of 450 ml each. The loaded unit shall weigh no more than 125 pounds. 3. Measure the external volume of the unit. The unit shall have an external volume of 8.5 cubic feet. 4. Examine the unit for power requ

24、irements. The unit shall operate on all of the following power sources: 120 VAC, 50 Hz 120 VAC, 60 Hz 220 VAC, 50 Hz 220 VAC, 60 Hz 12 VDC 24 VDC 5. Load the unit with 30 blood bags of 450 ml each filled with fluid. Precondition the unit to approximately 39OF. Vary the ambient temperature as follows

25、 and record the internal temperature via thermocouple each hour. The unit internal temperature shall stay within the range 34OF to 43OF. unit from 120 VAC, 60 Hz. Power the Time Ambient Time Ambient Lg= 0600 0700 0800 O900 1000 1100 1200 1300 1400 1500 1600 1700 1800 90 91 95 101 106 110 112 116 118

26、 119 120 119 118 1900 2000 2100 2200 2300 2400 O100 0200 0300 0400 0500 0600 114 108 105 102 100 98 95 94 93 92 91 90 6. Precondition the loaded unit to within the 34OF to 43OF range. Remove power from the unit. The internal temperature of the unit shall not exceed 50F for a minimum of 6 hours with

27、an ambient temperature of approximately 76OF. 4 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-5403b 9999974 0342239 O80 W A-A-5 4 O 16 7. Load the refrigerator to specified capacity with blood bags filled with blood simulant. the blood simulant

28、 is within the l0C to 6OC temperature range. Secure the refrigerator to a vibration platform. Excite the refrigerator at a 10 to 50 Hz frequency, with power spectral density of 0.15 g2/Hz for 20 minutes. excitation, examine refrigerator for satisfactory operation. Operate the refrigerator until Upon

29、 completion of Samplinq for examination. Sampling for examination shall be conducted in accordance with MIL-STD-105. Failure of one unit to pass inspection shall result in failure of the lot. product for examination purposes shall be one solid state biologicals refrigerator as specified in this desc

30、ription. Metric products. Products manufactured to metric dimensions will be considered on an equal basis with those manufactured using inch-pound units, providing they fall within the tolerances specified using conversion tables contained in the latest revision of Federal Standard 376, and all othe

31、r requirements of this document are met. The unit of If a product is manufactured to metric dimensions and those dimensions exceed the tolerances specified in the inch/pound units, a request should be made to the contracting officer to determine if the product is acceptable. The contracting officer

32、has the option of accepting or rejecting the product. Contractor certification. The contractor shall certify and maintain substantiating evidence that the product offered meets the salient characteristics of this Commercial Item Description, and that the product conforms to the producers own drawing

33、s, specifications, standards, and quality assurance practices. The government reserves the right to require proof of such conformance prior to first delivery and thereafter as may be otherwise provided for under the provisions of the contract. Requlatorv requirements. Federal Food, Drus and Cosmetic

34、 Act. If the product covered by this document has been determined by the U.S. Food and Drug Administration to be under its jurisdiction, the offeror/contractor shall comply, and be responsible for compliance by its subcontractors/suppliers, with the requirements of the Federal Food, Drug and Cosmeti

35、c Act, as amended, and 5 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-54016 regulations promulgated thereunder. In addition, the offeror/contractor shall comply, and be responsible for compliance by its subcontractors/suppliers, with the requi

36、rements of all other applicable Federal, State, and local statutes, ordinances, and regulations. Recovered materials. The offeror/contractor is encouraged to use recovered material in accordance with Federal Acquisition Regulation Subpart 23.4 to the maximum extent practical. Preservation, Packaqinq

37、, P ackinq, labelins, and markins. Unless otherwise specified, preservation, packaging, and packing shall be to a degree of protection to preclude damage to containers and/or contents thereof under normal shipping conditions, handling, etc:, involving shipment from the supply source to the receiving

38、 activity, plus reshipment from receiving activity, and shall conform to applicable carriers rules and regulations. Intermediate and exterior package quantities and labeling and marking shall be as specified in the contract and/or order. Orderins data. Intermediate/exterior package quantities, label

39、ing, and marking must be specified in the contract and/or order. NOTE: The following National Stock Number is covered by this document. 4110-01-287-7111 MILITARY INTERESTS: PREPARING ACTIVITY: Custodians: DoD-MB Air Force - 03 Army - MD Navy - MS CIVIL AGENCY COORDINATING Agent: DLA-DM ACTIVITIES: FDA-MPQAS USPHS VA - OSS Project Number: 4110-0432 Location: REFRIG17.DOC 6 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-