1、A-A-54386 22 April 1991 COMMERCIAL ITEM DESCRIPTION MICROSCOPE, OPTICAL (Binocular, Brightfield) The General Services Administration has authorized the use of this CommercialItem Description. This Comme.rcia1 Item Description covers a brightfield binocular microscope for histopathological and pathol

2、ogical studies for use inside medical laboratories and for outside field use. Salient Characteristics: Shall be an inclined binocular brightfield microscope for histopathological and pathological studies. Shall be a professional model that is suitable for diagnostic and research applications. A stud

3、ent grade microscope is not acceptable for use under this specification document. Shall be suitable for use in laboratories with adequate AC supply lines. Shall also be adaptable for outside field use where AC facilities are not available by the adaption of a mirror and fork assembly that shall be s

4、upplied as an accessory. Shall be supplied with the following accessories: A carrying case with a handle, a protective dust cover, and a mirror and fork assembly. Shall be suitable for operation from 50/60 Hz AC supply lines. The applicable AC line voltage shall be as indicated in the contract and/o

5、r purchase order (see “Ordering data paragraph). As a minimum, the microscope shall consist of a base with variable light source,stand with supporting arm and focusing controls, inclined binocular body (head), revolving nosepiece assembly with three planachromatic objectives and dust plug(s1, specim

6、en stage, mechanical stage, and substage condenser. The focusing of the microscope shall be controlled by the vertical movement of the specimen stage or the objective. AMSC N/A FSC 6650 DISTRIBUTION STATEMENT-hi Approved for public release; distribution is unlimited. Provided by IHSNot for ResaleNo

7、reproduction or networking permitted without license from IHS-,-,-A-A-5438b bb 9999974 OObO9bY 4 A-A-54386 The base shall have built-in a variable light source (illuminator) for operation from AC supply lines through a power cord. The light source shall be a 6 volt, 20 or 30 watt tungsten halogen la

8、mp. The circuitry for the light source shall include a continuous voltage (intensity) control, on-off switch and transformer. The lamp 8hall be supported in a lamp socket which has the facility for easy replacement of the bulb. The power cord shall be a flexible 3-conductor cord that terminates with

9、 a non-detachable 3-blade male plug. The illuminator shall have a filter holder with a removable blue filter and a built-in field diaphragm for Koehler illumination. The mirror and fork assembly shall be attachable to the base, for outside field use where AC facilities are not being used. The stand

10、shall house the focusing mechanism, and shall be attached to the base. The supporting arm of the stand shall support the binocular body. The focusing mechanism shall raise and lower the nosepiece assembly OP stage. The focusing mechanism shall include coarse and fine controls that are manually contr

11、olled by coarse and fine knob (dials) located in the lower position of both sides of the stand. For purposes of measuring specimen thickness, the fine adjustment shall be graduated in one (1) or two (2) micrometer divisions with one complete revolution representing 100 micrometers as a minimum. For

12、protecting an objective and specimen (slide), an adjustable mechanical stop shall be included in the stand to limit the travel of the coarse control. In addition, a safety clutch shall be included to stop the travel of the fine control mechanism in the event the operator continues to turn the fine c

13、ontrol when the objective makes contact with a slide or its top mounted cover slip. The binocular body (head) shall have eyepiece tubes with 1OX wide field eyepieces and eyeguard relief. The eyepiece tubes shall be inclined 30-40 degrees from the horizontal. The interpupillary distance (between eyep

14、iece tubes) shall be adjustable within a range that includes 56 to 72 millimeters (mm). The interpupillary adjustment shall include a millimeter scale for easy reference. Automatic tube length compensation shall be included if the tube length is affected by changes in the interpupillary settings. Th

15、e eyepiece tubes shall include provisions for installing micrometer discs. One eyepiece tube shall be focusable for acuity equalization of a users eye. The eypieces shall have a minimum 18 mm diameter field view and a minimum eye relief , distance 18.7 mm. The binocular body shall be 360 degrees rot

16、atable on the supporting arm (stand). The binocular body shall also be detachable for interchangeability with heads (e.g. a trinocular body for photomicrographic procedures). 2 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,- .f * A-A-543b bb m 99999

17、74 OObO9b5 b m A-A-54386 Three objectives shall be supplied with a quadruple or quintuple 360 degree revolving nosepiece having a ball detent index. Dust plugs shall be included in the unused objective hole(s) of the nosepiece. The objectives furnished shall be lox, 40X or 45X and 1OOX oil immersion

18、 planachromatics having numerical apertures of 0.25, 0.65 or 0.66 and 1.25 respectively. Each planachromatic objective shall have an image flatness over the entire field of view without any loss of sharpness or any loss of focus around the periplery of the field of view. The objectives shall be colo

19、r coded and marked for identification. In the nosepiece, the objectives shall be parfocal and parcentered so they are not affected by switching from one objective to another. The 40X, 45X and 100X objectives shall be retractable (i.e. spring loaded) when contact is made with a microscope slide or it

20、s cover slip. As an alternate to retractable objectives, the nosepiece assembly shall be retractable when an objective makes contact with a microscope slide or its cover slip. The specimen stage shall accommodate microscope slides that include 25 by 75 mm and 50 by 75 mm sizes. The specimen stage sh

21、all include a slide holder that is detachable by releasing a thumbscrew hold. The slide holder shall be in the form of a finger assembly that can grasp the sides of a microscope slide instead of its corners. The finger assembly shall also be retractable to allow more space on the specimen stage for

22、manual scanning of a microscope slide. The aperture width of the stage mount shall be not less than 25 mm. A right hand ungraduated mechanical stage, as a minimum, shall control the X-Y horizontal movement of the slide holder. The mechanical stage shall have ball bearing slideways in two perpendicul

23、ar directions that can give complete coverage under an objective of both size miscrscope slides. The substage condenser shall be located below the mechanical stage in a fork-type mount which is moved vertically by rack and pinion control for focusing purposes. The substage condenser shall have a num

24、erical aperture (N.A.) of 1.25 or 1.30, and shall include an integral iris diaphragm and shall provide an even full field illumination and a variable aperture control for all the objectives, i.e. from 1OX to 1OOX (oil immersion). The fork mount shall contain two knurled centering screws. The condens

25、er shall be supported by a snap action in the fork-type mount. The condenser shall also be removable from the fork-type mount without having to remove the centering screws. 3 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,- A-A-54386 66 m 9999974 006

26、09bb 8 m L A-A-54386 Provisions shall be made in the microscope to control the vertical movement of the condenser so that it does not make contact with the stage mounted microscope slide or the base mounted light source. In addition, a quick set.condenser shall be supplied which allows the condenser

27、 to be lowered from a focused position and then returned to its original position by being raised to a positive stop. The microscope shall be capable of accepting the following accessories which shall be commercially available from the manufacturer, but are not required to be supplied with the micro

28、scope: 1. 2. 3. 4. 5. A rotatable polarizer to fit the light source and an associated slide attachment to fit below the binocular body (head). A paraboloid or oil immersion dark field substage condenser that is interchangeable with the substage condenser that is supplied with the microscope. A dual

29、viewing adapter with a built-in pointer, for use with the binocular body (head) that is supplied with the microscope. A trinocular head for photomicroscopy, that is interchangeable with the binocular body (head) that is supplied with the microscope. A 4X planachromatic objective for the nosepiece, w

30、ith an auxiliary swing-in condenser for attachment to the substage condenser. To eliminate vibrations during photomicroscopy, the fully assembled microscope (i.e., stand, accessory trinocular body with objectives, stage mount and substage condenser) shall not weigh less than 19 pounds. The microscop

31、e shall be legibly and permanently marked with the name or registered trademark of the manufacturer, serial number and model number. Workmanship. The microscope shall be free from defects which detract from its appearance or impair its serviceability. Unit. Each (EA). One microscope, as specified, c

32、onstitutes one unit. 4 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-54386 66 m 9999974 O060967 T m A-A-54386 hgmysibility for insEgction. Unless otherwise specified in the contract or purchase order, the contractor is responsible for the perfo

33、rmance of all inspection requirements as specified herein. Except as otherwise specified in the contract or purchase order, the contractor may use his own or any facilities suitable for the performance of the inspection requirements specified herein, unless disapproved by the Government. The Governm

34、ent reserves the right to perform any of the inspections set forth in the specification where such inspections are deemed necessary to assure supplies and services conform to prescribed requirements. Recoyis- Records of examinations and tests performed by or for the contractor shall be maintained by

35、 the contractor and made available to the Government upon the Governments request, at any time, or from time to time, during the performance of the contract and for a period of three years after delivery of the supplies to which such records relate. - InsEection. _- Inspection, as used herein, is de

36、fined as both examination (such as visual or auditory investigation without the use of special laboratory appliances or procedures) and testing (determination by technical means of physical and chemical properties) of the item. Testing and examination. Testing and examination shall be conducted in a

37、ccordance with commercaial practice to determine compliance of the microscope with specification requirements. In addition, each microscope shall be tested to demonstrate its capability to operate properly. O - Sampling _- - for testing - and examination. Sampling for testing and examination shall b

38、e conducted in accordance with commercial practice. _- Contractor certification. The contractor shall certify that the product offered meets the salient characteristics of this description and conforms to the producers own drawings, specifications, standards, and quality assurance practices. The Gov

39、ernment reserves the right to require proof of such conformance prior to first delivery and thereafter as may be otherwise provided for under the provisions of the contract. 5 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-543b bb 9999974 00b09b

40、 L A-A-54386 Fgdgyal Fgd,-Drggand Cosmg&c-Ag&L If the product covered by this document has been determined by the U.S. Food and Drug Administration to be under its jurisdiction, the offeror/contractor shall comply, and be responsible for compliance by its subcontractors/suppliers, with the requireme

41、nts of the Federal Food, Drug and Cosmetic Act, as amended, and regulations promulgated thereunder. In addition, the offeror/contractor shall comply, and be responsible for compliance by its subcontractors/suppliers, with the requirements of all other applicable Federal, State, and local statutes, o

42、rdinances and regulations. Recovered materials. The offeror/contractor is encouraged to use recovered material in accordance with Federal Acquisition Regulation Subpart 23.4 to the maximum extent practical. oreiera4on,Eac5ag,-a5n,-ae-ar5n Unless otherwise specified, pregervation, packaging, and pack

43、ing shall be to a degree of protection to preclude damage to containers and/or contents thereof under normal shipping conditions, handling, etc., involving shipment from the supply source to the receiving activity, plus reshipment from receiving activity, and shall conform to applicable carriers rul

44、es and regulations. Intermediate and exterior package quantities and labeling and marking shall be as specified in the contract and/or order. NOTE: The following National Stock Numbers are covered by this document. 6650-01-207-0829 6650-01-315-3745 ITEM IDENTIFICATION - MICROSCOPE, OPTICAL, Binocula

45、r, 120/230 Volt, 50/60 Hz, AC MICROSCOPE, OPTICAL, Binocular, 120/230 Volt, 50/60 Hz, AC Ordering data. (Applicable line voltage requirement(s1, intermediate/exterior package quantities, labeling, and marking must be specified in the contract and/or order.) MILITARY INTERESTS: PREPARING ACTIVITY: Cu

46、stodians: DoD-MB 6 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-. II A-A-5Y3b bb m 7999974 OObO7b7 3 m A-A-54386 Army - MD Navy - MS Air Force - 03 Agent: DLA-DM CIVIL AGENCY COORDINATING ACTIVITIES: VA-OSS PHS FDA-MPQAS Project Number: 6650-0186 Location: ENABLE/MICOPTCD/S31 7 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-

展开阅读全文
相关资源
猜你喜欢
  • BS DD CEN ISO TS 14822-1-2006 Traffic and travel information - General specifications for medium-range pre-information via dedicated short-range communication - Downlink《交通和旅行信息 借助.pdf BS DD CEN ISO TS 14822-1-2006 Traffic and travel information - General specifications for medium-range pre-information via dedicated short-range communication - Downlink《交通和旅行信息 借助.pdf
  • BS DD CEN ISO TS 15876-7-2006 Plastics piping systems for hot and cold water installations - Polybutylene (PB) - Guidance for the assessment of conformity《冷热水装置用塑料管道系统 聚丁烯(PB) 合格评定.pdf BS DD CEN ISO TS 15876-7-2006 Plastics piping systems for hot and cold water installations - Polybutylene (PB) - Guidance for the assessment of conformity《冷热水装置用塑料管道系统 聚丁烯(PB) 合格评定.pdf
  • BS DD CEN ISO TS 16181-2011 Footwear Critical substances potentially present in footwear and footwear components Determination of phthalates in footwear materials《鞋类 在鞋类和鞋类部件中可能存在的.pdf BS DD CEN ISO TS 16181-2011 Footwear Critical substances potentially present in footwear and footwear components Determination of phthalates in footwear materials《鞋类 在鞋类和鞋类部件中可能存在的.pdf
  • BS DD CEN ISO TS 17665-2-2009 Sterilization of health care products - Moist heat - Guidance on the application of ISO 17665-1《保健产品的消毒 潮热 ISO 17665-1应用指南》.pdf BS DD CEN ISO TS 17665-2-2009 Sterilization of health care products - Moist heat - Guidance on the application of ISO 17665-1《保健产品的消毒 潮热 ISO 17665-1应用指南》.pdf
  • BS DD CEN ISO TS 17764-1-2002 Animal feeding stuffs - Determination of the content of fatty acids - Preparation of methyl esters《动物饲料 脂肪酸含量的测定 甲酯的制备》.pdf BS DD CEN ISO TS 17764-1-2002 Animal feeding stuffs - Determination of the content of fatty acids - Preparation of methyl esters《动物饲料 脂肪酸含量的测定 甲酯的制备》.pdf
  • BS DD CEN ISO TS 17764-2-2007 Animal feeding stuffs - Determination of the content of fatty acids - Gas chromatographic method《牲畜饲料 脂肪酸含量的测定 气相色谱法》.pdf BS DD CEN ISO TS 17764-2-2007 Animal feeding stuffs - Determination of the content of fatty acids - Gas chromatographic method《牲畜饲料 脂肪酸含量的测定 气相色谱法》.pdf
  • BS DD CEN ISO TS 17845-2004 Welding and allied processes - Designation system for imperfections《焊接和相关程序 缺陷的名称与符号系统》.pdf BS DD CEN ISO TS 17845-2004 Welding and allied processes - Designation system for imperfections《焊接和相关程序 缺陷的名称与符号系统》.pdf
  • BS DD CEN ISO TS 18234-4-2006 Traffic and travel information (TTI) - TTI via transport protocol expert group (TPEG) data-streams - Road traffic message (RTM) application《交通和旅行信息(TT.pdf BS DD CEN ISO TS 18234-4-2006 Traffic and travel information (TTI) - TTI via transport protocol expert group (TPEG) data-streams - Road traffic message (RTM) application《交通和旅行信息(TT.pdf
  • BS DD CEN ISO TS 18234-5-2006 Traffic and travel information (TTI) - TTI via transport protocol expert group (TPEG) data-streams - Public transport information (PTI) application《交通.pdf BS DD CEN ISO TS 18234-5-2006 Traffic and travel information (TTI) - TTI via transport protocol expert group (TPEG) data-streams - Public transport information (PTI) application《交通.pdf
  • 相关搜索

    当前位置:首页 > 标准规范 > 国际标准 > 其他

    copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
    备案/许可证编号:苏ICP备17064731号-1