资源描述

1、JEDEC PUBLICATION Characterization and Monitoring of Thermal Stress Test Oven Temperatures JEP153A (Revision of JEP153, January 2008) MARCH 2014 JEDEC SOLID STATE TECHNOLOGY ASSOCIATION NOTICE JEDEC standards and publications contain material that has been prepared, reviewed, and approved through th

2、e JEDEC Board of Directors level and subsequently reviewed and approved by the JEDEC legal counsel. JEDEC standards and publications are designed to serve the public interest through eliminating misunderstandings between manufacturers and purchasers, facilitating interchangeability and improvement o

3、f products, and assisting the purchaser in selecting and obtaining with minimum delay the proper product for use by those other than JEDEC members, whether the standard is to be used either domestically or internationally. JEDEC standards and publications are adopted without regard to whether or not

4、 their adoption may involve patents or articles, materials, or processes. By such action JEDEC does not assume any liability to any patent owner, nor does it assume any obligation whatever to parties adopting the JEDEC standards or publications. The information included in JEDEC standards and public

5、ations represents a sound approach to product specification and application, principally from the solid state device manufacturer viewpoint. Within the JEDEC organization there are procedures whereby a JEDEC standard or publication may be further processed and ultimately become an ANSI standard. No

6、claims to be in conformance with this standard may be made unless all requirements stated in the standard are met. Inquiries, comments, and suggestions relative to the content of this JEDEC standard or publication should be addressed to JEDEC at the address below, or refer to www.jedec.org under Sta

7、ndards and Documents for alternative contact information. Published by JEDEC Solid State Technology Association 2014 3103 North 10th Street Suite 240 South Arlington, VA 22201-2107 This document may be downloaded free of charge; however JEDEC retains the copyright on this material. By downloading th

8、is file the individual agrees not to charge for or resell the resulting material. PRICE: Contact JEDEC Printed in the U.S.A. All rights reserved PLEASE! DONT VIOLATE THE LAW! This document is copyrighted by JEDEC and may not be reproduced without permission. For information, contact: JEDEC Solid Sta

9、te Technology Association 3103 North 10th Street Suite 240 South Arlington, VA 22201-2107 or refer to www.jedec.org under Standards-Documents/Copyright Information. JEDEC Publication No. 153A -i- CHARACTERIZATION AND MONITORING OF THERMAL STRESS TEST OVEN TEMPERATURES Contents Foreword ii Introducti

10、on ii 1 Scope 1 2 Terms and Definitions 1 3 References 2 4 Characterization Requirements 2 4.1 General requirements 2 4.2 Characterization Requirements for fixed temperature convection ovens including freezers 4 4.3 Characterization Requirements for temperature cycle chamber single and dual zone 6 4

11、.4 Requirements for temperature chambers with bias 7 4.5 Characterization Requirements for belt reflow systems 7 5 Monitoring Requirements 9 6 Summary / Reporting 9 Annex A Monitoring 10 Annex B Reference to other documents 11 Annex C Differences between revisions 12 JEDEC Publication No. 153A -ii-

12、CHARACTERIZATION AND MONITORING OF THERMAL STRESS TEST OVEN TEMPERATURES Foreword This document provides an industry standard method for characterization and monitoring thermal stress test oven temperatures. The procedures described in this document should be used to insure thermal stress test condi

13、tions are being achieved and maintained during various test procedures. Introduction JEDEC has a series of thermal stress tests and procedures that require specific temperatures be met and controlled within specified limits. The characterization and monitoring methods described in this test procedur

14、e should insure that the devices under test (DUT) achieve the correct temperatures and tolerances while being stressed within the specific thermal test oven or chamber. JEDEC Publication No. 153A Page 1 CHARACTERIZATION AND MONITORING OF THERMAL STRESS TEST OVEN TEMPERATURES (From JEDEC Board ballot

15、 JCB-13-67, formulated under the cognizance of the JC-14.1 Subcommittee on Reliability Test Methods for Packaged Devices.) 1 Scope This document specifies a uniform and reproducible method of confirming that temperature test chambers meet the requirements specified in thermal stress test procedures.

16、 This publication is intended for users when conducting one of the following procedures: monitoring (in situ and non in situ), characterization (e.g., for evaluation, release/acceptance and during development as well as for test set up). The document is applicable to the following type of equipment:

17、 temperature cycle chamber single zone, temperature cycle chamber dual zone, fixed convection oven with or without bias, fixed freezer, belt reflow ovens (used for preconditioning). Hot-plate systems and temperature-humidity chambers or HAST chambers are not covered by this document. Some remarks ma

18、y highlight the differences or challenges associated with these systems. 2 Terms and definitions characterization: A description of the characteristics of a product or process by mathematical modeling, design of experiments, or statistical data evaluation. NOTE Methods of statistical data evaluation

19、 are described in EIA557 and JEP132. convection oven: A controlled-temperature chamber in which the heat is transferred by air flow, rather than by conduction or radiation. NOTE The chamber must be capable of maintaining specified temperatures over the entire working area. fixed freezer: A controlle

20、d-temperature chamber capable of maintaining specified subfreezing temperatures over the entire working area. load map: A map of the positions where devices are placed for testing within the working area. NOTE See Figure 1 for an example. JEDEC Publication No. 153A Page 2 2 Terms and definitions (co

21、ntd) monitoring: Using a program for observing, and often controlling, a device or equipment. temperature profile: (1) A graphical portrayal of temperature experienced by an assembly as it passes through a soldering process. (2) A graphical portrayal of the spatial temperature distribution in an ove

22、n. working area: The portion of a chamber, used for stress testing, that meets the calibration requirements. 3 References IEC 60068-3-5, Environmental testing. Guidance. Confirmation of the performance of temperature chambers JEP140, Beaded Thermocouple Temperature Measurement of Semiconductor Packa

23、ges J-STD-020, JOINT IPC/JEDEC Standard For Moisture/Reflow Sensitivity Classification for Nonhermetic Solid State Surface-Mount Devices JESD47, Stress-Test-Driven Qualification of Integrated Circuits JESD22A104, Temperature Cycling JESD22A105; Power and Temperature Cycling JESD22A113, Preconditioni

24、ng of Plastic Surface Mount Devices Prior to Reliability Testing 4 Characterization requirements 4.1 General requirements When chambers are used, the following requirements are examples of best practice and should be used as guideline: a) specimens under test shall be located only within the working

25、 area (see Figure 1); b) the controls for the chamber shall be capable of maintaining the temperature of a reference point within 2C of the set point in an unloaded condition. This requirement does not apply for T/C chambers; c) permissible temperature variation in environmental chambers: temperatur

26、e on the test specimens within the working area must be maintained per the requirement of the stress test; d) the working space has to be defined depending on the requirements of the stress test; e) sensors have to be chosen according to the required resolution. JEDEC Publication No. 153A Page 3 4 C

27、haracterization requirements (contd) 4.1 General requirements (contd) Thermocouples can be reused, if desired, after profile development by removing them from the test parts. Verify the thermocouple operation and calibration status before reuse. For details on calibration see reference 1. ChamberWor

28、king areawithin chamberParts within working areaFigure 1 Parts within working area JEDEC Publication No. 153A Page 4 4.1 General requirements (contd) 4.1.1 Typical profile characterization procedure Heat the oven until the complete working area meets oven set-up parameters (e.g., target temperature)

29、. The procedure may require several trials of locating the DUTs until all devices can be brought to the required temperature. It should be taken into account that the thermal response of parts on the inner trays of a stack of trays could be slow. Especially for temperature cycle stress this could re

30、sult in too low Tmax values. Run oven using test load with thermocouples. Compare the actual data for each of the thermocouples to the profile requirements. If the data from all thermocouples meets all the profile requirements then record/store the oven set-up parameters along with the actual profil

31、e data and the characterization is complete. If the data from any of the thermocouples does not meet any of the profile requirements then the oven set-up parameters must be adjusted and the characterization repeated until all profile requirements are met. For positions of the sensors (thermo element

32、) see 4.2 4.5. Re-characterization is recommended a) in case of equipment changes affecting the process, b) after one year as part preventive maintenance, c) in case of deviations detected by monitoring, d) when new test sample configuration or load is changed from which has been characterized (char

33、acterization for test set up) unless, when applicable, the initial test set up covers the worst case of maximum thermal load allowed for the test oven. 4.2 Characterization requirements for fixed temperature convection ovens including freezers The specified test temperature is the ambient temperatur

34、e to which all devices in the working area of the chamber shall be exposed. The temperature range is based on the test methods tolerance requirements. This shall be assured by making whatever adjustments are necessary in the chamber profile, loading, location of control or monitoring instruments, an

35、d the flow of air or other suitable gas or liquid chamber medium. Therefore, characterization shall be accomplished in the chamber in a loaded condition as used (any characterization mode must be representative of chamber loading for the specific test), and in a non powered configuration. The charac

36、terization sensors located on the part or board within the working area (JEP140) has to represent the temperature range of the working area. A load map for the characterization must be generated upon completion. JEDEC Publication No. 153A Page 5 4.2 Characterization requirements for fixed temperatur

37、e convection ovens including freezers (contd) 4.2.1 Parameter to be measured is the temperature stability over time. The output of the temperature measuring system determine, after chamber stabilization, the achieved temperature, temperature fluctuation and temperature gradient of the working space.

38、 Uncertainty of measurement of the temperature measuring system should be evaluated. 4.2.2 Verification of test condition requirements Test temperatures: at test condition Loading condition: as under test for characterization of test set up, maximum thermal load allowed for equipment characterizatio

39、n. Test time: Start when test conditions have been verified with the load under the desired configuration. A reference point of the profile should be recorded regularly. Measurement period: follow equipment manual or manufacturers recommendation, if not specified recommendation is to start with 1/we

40、ek 4.2.3 Position thermo element (sensor) The positions of the sensors should be chosen depending on volume and size of the working area. The distribution should be chosen from one of the following options: a) Temperature measuring sensors are located on a test sample in each corner and in the cente

41、r of the working space (see Figure 2, minimum 9 sensors). b) For larger temperature chambers, additional sensors should be located across the test samples within the chamber including multiple shelf (e.g., additional sensors at the center of each wall of the working area). Dependent on chamber loadi

42、ng and set up, alternative positions for sensors may be needed to best characterize the load. JEDEC Publication No. 153A Page 6 4.2 Characterization requirements for fixed temperature convection ovens including freezers (contd) 4.2.3 Position thermo element (sensor) (contd) T6 T7T8T9T5T1 T2T3T4Chamb

43、erWorking area wit h in ch am b erThe thermocouples shall be attached to the center of the test parts according to JEP140. Figure 2 Positions thermo elements option A 4.3 Characterization Requirements for temperature cycle chamber single and dual zone Parameter to be measured in addition to the para

44、meters from 4.2: heat-up-rate from Tmin+10% to Tmax-10% 1)including linearity 2)for fast and slow ramp, cool-down-rate from Tmax-10% to Tmin+10% 1)including linearity 2)for fast and slow ramp. For equipment characterization the longest soak mode should be chosen. For test set up the soak mode chosen

45、 should be representative of the soak mode of the stress test to be performed. Stability at Tmin and Tmax. 1) Remark: rates should be measured in the recommended minimum linearity range between 10% and 90% of (Tmax Tmin) to avoid that non linearity and fluctuation at the boundary values disturb the

46、measurement results. See also JESD22A104 and A105. 2) Remark: Linearity is the goodness of fit in the recommended minimum linearity range between the 10% and the 90% value for a straight line. JEDEC Publication No. 153A Page 7 4.3 Characterization Requirements for temperature cycle chamber single an

47、d dual zone (contd) 4.3.1 Verification of test condition requirements Same as 4.2 4.3.2 Position thermo element (sensor) Same as 4.2 4.4 Requirements for temperature chambers with bias For steady state temperature bias chambers the regulations of clause 4.1 and 4.2 apply. Before each test the influe

48、nce of the device under test on the temperature has to be characterized because heat dissipation can have major influence on the chamber temperature. In some cases the chamber has to cool the device to reach the planned stress temperature. To adjust the chamber temperature to the intended junction t

49、emperature the following equation could be used: Tj= Tchamber+ P Rj with Tj: junction temperature of the device (C) P power dissipation (Watt) Rj thermal resistance of the device For temperature cycling with bias the regulation of clause 4.3 applies. 4.5 Characterization Requirements for belt reflow systems Belt reflow systems, typically used during Moisture Sensitivity Level testing and preconditioning, simulate the high temperature solder reflows that surface mount devices are exposed to during board assembly. MSL testing according to test spe

展开阅读全文
相关资源
猜你喜欢
  • DIN EN 2887-1995 Aerospace series - bolts normal hexagonal head threaded to head in corrosion resisting steel passivated - Classification 600 MPa (at ambient temperature) 425 C Ger.pdf DIN EN 2887-1995 Aerospace series - bolts normal hexagonal head threaded to head in corrosion resisting steel passivated - Classification 600 MPa (at ambient temperature) 425 C Ger.pdf
  • DIN EN 2888-2007 Aerospace series - Metric bolts normal hexagon head coarse tolerance normal shank short thread in corrosion resisting steel passivated - Classification 600 MPa (at.pdf DIN EN 2888-2007 Aerospace series - Metric bolts normal hexagon head coarse tolerance normal shank short thread in corrosion resisting steel passivated - Classification 600 MPa (at.pdf
  • DIN EN 2889-2007 Aerospace series - Metric bolts normal hexagon head coarse tolerance normal shank short thread in alloy steel cadmium plated - Classification 900 MPa (at ambient t.pdf DIN EN 2889-2007 Aerospace series - Metric bolts normal hexagon head coarse tolerance normal shank short thread in alloy steel cadmium plated - Classification 900 MPa (at ambient t.pdf
  • DIN EN 289-2014 Plastics and rubber machines - Compression moulding machines and transfer moulding machines - Safety requirements German version EN 289 2014《塑料和橡胶机械 压缩成型机和压铸机 安全要求 .pdf DIN EN 289-2014 Plastics and rubber machines - Compression moulding machines and transfer moulding machines - Safety requirements German version EN 289 2014《塑料和橡胶机械 压缩成型机和压铸机 安全要求 .pdf
  • DIN EN 2897-1996 Aerospace series - Screws pan head offset cruciform recess close tolerance normal shank short thread in alloy steel cadmium plated - Classification 1100 MPa 235 C .pdf DIN EN 2897-1996 Aerospace series - Screws pan head offset cruciform recess close tolerance normal shank short thread in alloy steel cadmium plated - Classification 1100 MPa 235 C .pdf
  • DIN EN 2898-2009 Aerospace series - Corrosion and heat resisting steel rivets - Technical specification German and English version EN 2898 2008《航空航天系列 耐腐蚀和耐热钢制柳钉 技术规范》.pdf DIN EN 2898-2009 Aerospace series - Corrosion and heat resisting steel rivets - Technical specification German and English version EN 2898 2008《航空航天系列 耐腐蚀和耐热钢制柳钉 技术规范》.pdf
  • DIN EN 2906-1994 Aerospace series - Nuts bihexagonal self-locking in heat resisting steel FE-PA92HT (A286) - Classification 1100 MPa (at ambient temperature) 650 C German version E.pdf DIN EN 2906-1994 Aerospace series - Nuts bihexagonal self-locking in heat resisting steel FE-PA92HT (A286) - Classification 1100 MPa (at ambient temperature) 650 C German version E.pdf
  • DIN EN 2907-1995 Aerospace series - Nuts bihexagonal self-locking in heat resisting steel FE-PA92HT (A286) silver plated - Classification 1100 MPa (at ambient temperature) 650 C Ge.pdf DIN EN 2907-1995 Aerospace series - Nuts bihexagonal self-locking in heat resisting steel FE-PA92HT (A286) silver plated - Classification 1100 MPa (at ambient temperature) 650 C Ge.pdf
  • DIN EN 2908-1994 Aerospace series - Nuts bihexagonal self-locking deep counterbored in heat resisting steel FE-PA92HT (A286) - Classification 1100 MPa (at ambient temperature) 650 .pdf DIN EN 2908-1994 Aerospace series - Nuts bihexagonal self-locking deep counterbored in heat resisting steel FE-PA92HT (A286) - Classification 1100 MPa (at ambient temperature) 650 .pdf
  • 相关搜索

    当前位置:首页 > 标准规范 > 国际标准 > 其他

    copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
    备案/许可证编号:苏ICP备17064731号-1