IEC 61676 AMD 1-2008 Medical electrical equipment - Dosimetric instruments used for non-invasive measurement of x-ray tube voltage in diagnostic radiology Amend.pdf

上传人:lawfemale396 文档编号:1365981 上传时间:2019-11-26 格式:PDF 页数:8 大小:900.85KB
下载 相关 举报
IEC 61676 AMD 1-2008 Medical electrical equipment - Dosimetric instruments used for non-invasive measurement of x-ray tube voltage in diagnostic radiology Amend.pdf_第1页
第1页 / 共8页
IEC 61676 AMD 1-2008 Medical electrical equipment - Dosimetric instruments used for non-invasive measurement of x-ray tube voltage in diagnostic radiology Amend.pdf_第2页
第2页 / 共8页
IEC 61676 AMD 1-2008 Medical electrical equipment - Dosimetric instruments used for non-invasive measurement of x-ray tube voltage in diagnostic radiology Amend.pdf_第3页
第3页 / 共8页
IEC 61676 AMD 1-2008 Medical electrical equipment - Dosimetric instruments used for non-invasive measurement of x-ray tube voltage in diagnostic radiology Amend.pdf_第4页
第4页 / 共8页
IEC 61676 AMD 1-2008 Medical electrical equipment - Dosimetric instruments used for non-invasive measurement of x-ray tube voltage in diagnostic radiology Amend.pdf_第5页
第5页 / 共8页
点击查看更多>>
资源描述

1、 IEC 61676Edition 1.0 2008-11INTERNATIONAL STANDARD Medical electrical equipment Dosimetric instruments used for non-invasive measurement of X-ray tube voltage in diagnostic radiology IEC61676:2002/A1:2008(E) AMENDMENT 1 THIS PUBLICATION IS COPYRIGHT PROTECTED Copyright 2008 IEC, Geneva, Switzerland

2、 All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from either IEC or IECs member National Committee in the country of t

3、he requester. If you have any questions about IEC copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or your local IEC member National Committee for further information. IEC Central Office 3, rue de Varemb CH-1211 Geneva 20 Switzerlan

4、d Email: inmailiec.chWeb: www.iec.chAbout the IEC The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes International Standards for all electrical, electronic and related technologies. About IEC publications The technical content of IEC pu

5、blications is kept under constant review by the IEC. Please make sure that you have the latest edition, a corrigenda or an amendment might have been published. Catalogue of IEC publications: www.iec.ch/searchpub The IEC on-line Catalogue enables you to search by a variety of criteria (reference numb

6、er, text, technical committee,). It also gives information on projects, withdrawn and replaced publications. IEC Just Published: www.iec.ch/online_news/justpub Stay up to date on all new IEC publications. Just Published details twice a month all new publications released. Available on-line and also

7、by email. Electropedia: www.electropedia.org The worlds leading online dictionary of electronic and electrical terms containing more than 20 000 terms and definitions in English and French, with equivalent terms in additional languages. Also known as the International Electrotechnical Vocabulary onl

8、ine. Customer Service Centre: www.iec.ch/webstore/custserv If you wish to give us your feedback on this publication or need further assistance, please visit the Customer Service Centre FAQ or contact us: Email: csciec.chTel.: +41 22 919 02 11 Fax: +41 22 919 03 00 IEC 61676Edition 1.0 2008-11INTERNA

9、TIONAL STANDARD Medical electrical equipment Dosimetric instruments used for non-invasive measurement of X-ray tube voltage in diagnostic radiology INTERNATIONAL ELECTROTECHNICAL COMMISSION DICS 11.040.50; 11.040.55 PRICE CODEISBN 2-8318-1019-6AMENDMENT 1 Registered trademark of the International El

10、ectrotechnical Commission 2 61676 Amend. 1 IEC:2008 FOREWORD This amendment has been prepared by subcommittee 62C, Equipment for radiotherapy, nuclear medicine and radiation dosimetry, of IEC technical committee 62, Electrical equipment in medical practice. The text of this amendment is based on the

11、 following documents: Enquiry draft Report on voting 62C/445/CDV 62C/452/RVC Full information on the voting for the approval of this amendment can be found in the report on voting indicated in the above table. NOTE In this amendment, a new influence quantity “Additional tungsten filtration (tube agi

12、ng)” has been introduced. The committee has decided that the contents of this publication will remain unchanged until the maintenance result date indicated on the IEC web site under “http:/webstore.iec.ch“ in the data related to the specific publication. At this date, the publication will be reconfi

13、rmed, withdrawn, replaced by a revised edition, or amended. _ 4.3.5 Limits of variation Replace the existing Table 2 by the following: 61676 Amend. 1 IEC:2008 3 Table 2 Minimum RATED RANGE OF USE, REFERENCE CONDITIONS, STANDARD TEST CONDITIONS, LIMITS OF VARIATION ( L) and INTRINSIC ERROR (E) over t

14、he EFFECTIVE RANGE of use, for the pertaining INFLUENCE QUANTITY INFLUENCE QUANTITY Minimum RATED RANGE of use REFERENCE CONDITIONS STANDARD TEST CONDITIONS E kV L % Sub-clause Voltage waveform and frequency: Diagnostic Mammography Constant potential, 2-, 6-, 12-pulse and medium frequency generators

15、 aConstant potential Constant potential Constant potential, ripple less than 4 % 0,5 2,0 4.4.2 Anode angle: Diagnostic Mammography 6 to 18 15 to 24 12 20 REFERENCE VALUE 2 REFERENCE VALUE 2 0,5 0,5 4.4.3 Filtration: Diagnostic Mammography CT Dental 2,5 mm Al to 3,5 mm Al b25 mm Al to 35 m Mo c4 mm A

16、l to 8 mm Al 1 mm Al to 2 mm Al 3,0 mm Al 30 m Mo 6 mm Al 1,5 mm Al REFERENCE VALUE 5 % REFERENCE VALUE 5 % REFERENCE VALUE 5 % REFERENCE VALUE 5 % 0,5 1,5 1,5 1,5 4.4.4 Dose rate: Diagnostic MammographyCT Dental Fluoroscopic 20 mGy/s to 200 mGy/s 25 mGy/s to 150 mGy/s 20 mGy/s to 200 mGy/s 5 mGy/s

17、to 50 mGy/s 1 mGy/s to 10 mGy/s As stated by manufacturer REFERENCE VALUE 20 % 0,5 0,5 0,5 0,5 0,5 4.4.5 Irradiation time: Diagnostic Other 10 ms to 1 000 ms 200 ms to 1 000 ms 100 ms 500 ms REFERENCE VALUE 20 % REFERENCE VALUE 20 % 0,5 0,5 4.4.6 Field size: Rated Range Large Field Length and width

18、stated by manufacturer + 30 % 10 %30 cm by 30 cm As stated bymanufacturer.30 cm by 30 cm REFERENCE VALUE 2 % REFERENCE VALUE 2 % 0,5 2,0 4.4.7.1 4.4.7.2 Detector-Focal distance 32 cm to 60 cm or as stated by Mfg 40 cm or asstated by manufacturer REFERENCE VALUE 1 % 0,5 4.4.8 Angle of incidence Rotat

19、ion 5 180 0 0 REFERENCE VALUE 1 REFERENCE VALUE 1 0,5 0,5 4.4.9 4.4.10 Temperature Relative humidity 15 C to 35 C 80 % (max 20 g/m3) 20 C 50 % REFERENCE VALUE 2C 30 % TO 75 % 1,0 4.4.11 Power supply Line voltage and frequency Batteries Rechargeable batteries 115 V or 230 V + 10 % 15 %50 or 60 Hz As

20、stated by Mfg. Fresh to Low 115 V/230 V50 Hz/60 Hz as stated Fresh, mains disconnected REFERENCE VALUE 1 % REFERENCE VALUE 1 % REFERENCE VALUE 1 % 0,5 0,5 0,5 4.4.12.14.4.12.2 4.4.12.3Electromagnetic compatibility IEC 61000-4-2 IEC 61000-4-3 IEC 61000-4-4 IEC 61000-4-5 IEC 61000-4-6 IEC 61000-4-11 W

21、ithout anydisturbance Insignificant 1,0 4.4.13 Additional tungsten filtration (tube aging) 0 m to 10 m W 3 m W 0m W -3 m W 2,0 4.4.14 a Frequency range f = 50 Hz to 50 kHz, VOLTAGE RIPPLE (%) from 0 to (50-10log f), e.g. 0 % to 20 % at 1 000 Hz, 0 % to 3 % at 50 kHz. All frequencies above 50 kHz are

22、 treated as constant potential generators. b Filtration outside of MINIMUM RATED RANGE may be met by applying corrections. c X-RAY generator with a molybdenum anode, a beryllium window, and no ADDED FILTRATION other than the 30 m Mo. 4 61676 Amend. 1 IEC:2008 4.4 Performance test procedures Add a ne

23、w subclause as follows: 4.4.14 Additional tungsten filtration (tube aging) Over the RATED RANGE of additional tungsten filtration, the LIMITS OF VARIATION of RESPONSE shall not be greater than stated in Table 2. NOTE The higher the age of an X-ray tube, the anode roughens more and more depending on

24、the cumulative heat load during its total operation time. The roughening of the anode results in a hardening of the spectral photon distribution, which can be simulated by additional tungsten filtration, where zero filtration represents a new tube, and 10 m W an X-ray tube near the end of its lifeti

25、me, respectively. Before performing this test it should be proved that the X-ray tube used for this test is of moderate age, corresponding to an additional filtration of 0 m W 3 m W as required for the standard test conditions. This can be shown under the following conditions: 70 kV tube voltage and

26、 3,0 mm Al total filtration, by measuring the Al-HALF-VALUE LAYER (HVL) which should be less than the values in Table 4, depending on the anode angle of the tube: Table 4 Maximum HALF-VALUE LAYER (HVL) depending on anode angle Anode angle() 6 8 10 12 14 16 18 HVL (mm Al) 3,23 3,07 2,98 2,91 2,86 2,8

27、3 2,80 Compliance with the performance requirement shall now be checked by measuring the response of the instrument with the detector of the instrument exposed to the minimum and the maximum rated additional tungsten filtration and compared with a reference set of readings at reference filtration (w

28、ith 3 m additional tungsten filtration). Tests shall be made at the minimum test points indicated in Table 3 and in 4.4.1 to show compliance over the effective range of voltages. _ INTERNATIONAL ELECTROTECHNICAL COMMISSION 3, rue de Varemb PO Box 131 CH-1211 Geneva 20 Switzerland Tel: + 41 22 919 02 11 Fax: + 41 22 919 03 00 infoiec.ch www.iec.ch

展开阅读全文
相关资源
  • IEC TS 62492-1-2008 Industrial process control devices - Radiation thermometers - Part 1 Technical data for radiation thermometers《工业过程控制装置 辐射温度计 第1部分 辐射温度计的技术数.pdfIEC TS 62492-1-2008 Industrial process control devices - Radiation thermometers - Part 1 Technical data for radiation thermometers《工业过程控制装置 辐射温度计 第1部分 辐射温度计的技术数.pdf
  • IEC TR2 61464-1998 Insulated bushings - Guide for the interpretation of dissolved gas analysis (DGA) in bushings where oil is the impregnating medium of the mai.pdfIEC TR2 61464-1998 Insulated bushings - Guide for the interpretation of dissolved gas analysis (DGA) in bushings where oil is the impregnating medium of the mai.pdf
  • IEC TR 61241-2-2-1993 Electrical apparatus for use in the presence of combustible dust part 2 test methods section 2 method for determining the electrical resis.pdfIEC TR 61241-2-2-1993 Electrical apparatus for use in the presence of combustible dust part 2 test methods section 2 method for determining the electrical resis.pdf
  • IEC TR 60972-1989 Classification and interpretation of new lighting products《新型照明产品的分类和说明》.pdfIEC TR 60972-1989 Classification and interpretation of new lighting products《新型照明产品的分类和说明》.pdf
  • IEC TR 60943 Edition 21-2009 Guidance concerning the permissible temperature rise for parts of electrical equipment in particular for terminals《特殊终端中电气设备部件用关于允许.pdfIEC TR 60943 Edition 21-2009 Guidance concerning the permissible temperature rise for parts of electrical equipment in particular for terminals《特殊终端中电气设备部件用关于允许.pdf
  • IEC TR 60943 AMD 1-2008 Guidance concerning the permissible temperature rise for parts of electrical equipment in particular for terminals Amendment 1《电气设备部件(特别.pdfIEC TR 60943 AMD 1-2008 Guidance concerning the permissible temperature rise for parts of electrical equipment in particular for terminals Amendment 1《电气设备部件(特别.pdf
  • IEC TR 60919-2-2008 Performance of high-voltage direct current (HVDC) systems with line-communicated converters - Part 2 Faults and switching《带线性通信转换器的高压直流(HVDC.pdfIEC TR 60919-2-2008 Performance of high-voltage direct current (HVDC) systems with line-communicated converters - Part 2 Faults and switching《带线性通信转换器的高压直流(HVDC.pdf
  • IEC TR 60870-6-505 Edition 11-2006 Telecontrol equipment and systems - Part.6-505 Telecontrol protocols compatible with ISO standards and ITU-T recommendations .pdfIEC TR 60870-6-505 Edition 11-2006 Telecontrol equipment and systems - Part.6-505 Telecontrol protocols compatible with ISO standards and ITU-T recommendations .pdf
  • IEC TR 60344 CORR1-2012 Calculation of d c resistance of plain and coated copper conductors of low-frequency cables and wires - Application guide Corrigendum 1《.pdfIEC TR 60344 CORR1-2012 Calculation of d c resistance of plain and coated copper conductors of low-frequency cables and wires - Application guide Corrigendum 1《.pdf
  • IEC 62560 CORR1-2012 Self-ballasted LED-lamps for general lighting services by voltage 50 V - Safety specifications Corrigendum 1《普通照明用50 V以上自镇流LED灯 安全要求 勘误表1》.pdfIEC 62560 CORR1-2012 Self-ballasted LED-lamps for general lighting services by voltage 50 V - Safety specifications Corrigendum 1《普通照明用50 V以上自镇流LED灯 安全要求 勘误表1》.pdf