ANSI ISO ASTM 51956-2013 Standard Practice for Use of a Thermoluminescence-Dosimetry System (TLD System) for Radiation Processing.pdf

上传人:terrorscript155 文档编号:437122 上传时间:2018-11-14 格式:PDF 页数:8 大小:327.45KB
下载 相关 举报
ANSI ISO ASTM 51956-2013 Standard Practice for Use of a Thermoluminescence-Dosimetry System (TLD System) for Radiation Processing.pdf_第1页
第1页 / 共8页
ANSI ISO ASTM 51956-2013 Standard Practice for Use of a Thermoluminescence-Dosimetry System (TLD System) for Radiation Processing.pdf_第2页
第2页 / 共8页
ANSI ISO ASTM 51956-2013 Standard Practice for Use of a Thermoluminescence-Dosimetry System (TLD System) for Radiation Processing.pdf_第3页
第3页 / 共8页
ANSI ISO ASTM 51956-2013 Standard Practice for Use of a Thermoluminescence-Dosimetry System (TLD System) for Radiation Processing.pdf_第4页
第4页 / 共8页
ANSI ISO ASTM 51956-2013 Standard Practice for Use of a Thermoluminescence-Dosimetry System (TLD System) for Radiation Processing.pdf_第5页
第5页 / 共8页
亲,该文档总共8页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、ISO/ASTM 51956:2013(E)Standard Practice forUse of a Thermoluminescence-Dosimetry System (TLDSystem) for Radiation Processing1This standard is issued under the fixed designation ISO/ASTM 51956; the number immediately following the designation indicates theyear of original adoption or, in the case of

2、revision, the year of last revision.1. Scope1.1 This practice covers procedures for the use of thermolu-minescence dosimeters (TLDs) to measure the absorbed dosein materials irradiated by photons or electrons in terms ofabsorbed dose to water. Thermoluminescence-dosimetry sys-tems (TLD systems) are

3、generally used as routine dosimetrysystems.1.2 The thermoluminescence dosimeter (TLD) is classifiedas a type II dosimeter on the basis of the complex effect ofinfluence quantities on the dosimeter response. See ISO/ASTMPractice 52628.1.3 This document is one of a set of standards that providesrecomm

4、endations for properly implementing dosimetry inradiation processing, and describes a means of achievingcompliance with the requirements of ISO/ASTM 52628 “Prac-tice for Dosimetry in Radiation Processing” for a TLD system.It is intended to be read in conjunction with ISO/ASTM 52628.1.4 This practice

5、 covers the use of TLD systems under thefollowing conditions:1.4.1 The absorbed-dose range is from 1 Gy to 10 kGy.1.4.2 The absorbed-dose rate is between110-2and11010Gy s-1.1.4.3 The radiation-energy range for photons and electronsis from 0.1 to 50 MeV.1.5 This practice does not cover measurements o

6、f absorbeddose in materials subjected to neutron irradiation.1.6 This practice does not cover procedures for the use ofTLDs for determining absorbed dose in radiation-hardnesstesting of electronic devices. Procedures for the use of TLDsfor radiation-hardness testing are given in ASTM PracticeE668.1.

7、7 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced doc

8、uments2.1 ASTM Standards:2E170 Terminology Relating to Radiation Measurements andDosimetryE666 Practice for Calculating Absorbed Dose From Gammaor X RadiationE668 Practice for Application of Thermoluminescence-Dosimetry (TLD) Systems for Determining AbsorbedDose in Radiation-Hardness Testing of Elec

9、tronic Devices2.2 ISO/ASTM Standards:251261 Practice for Calibration of Routine Dosimetry Sys-tems for Radiation Processing51608 Practice for Dosimetry in an X-Ray (Bremsstrahlung)Facility for Radiation Processing51649 Practice for Dosimetry in an Electron-Beam Facilityfor Radiation Processing at En

10、ergies Between 300 keVand 25 MeV51702 Practice for Dosimetry in Gamma Irradiation Facili-ties for Radiation Processing51707 Guide for Estimating Uncertainties in Dosimetry forRadiation Processing51939 Practice for Blood Irradiation Dosimetry51940 Guide for Dosimetry for Sterile Insect Release Pro-gr

11、ams52628 Practice for Dosimetry in Radiation Processing52701 Guide for Performance Characterization of Dosim-eters and Dosimetry Systems for Use in Radiation Pro-cessing2.3 Joint Committee for Guides in Metrology (JCGM)Reports:JCGM 100:2008, GUM 1995, with minor corrections,Evaluation of measurement

12、 dataGuide to the Expres-sion of Uncertainty in Measurement31This practice is under the jurisdiction of ASTM Committee E61 on RadiationProcessing and is the direct responsibility of Subcommittee E61.02 on DosimetrySystems, and is also under the jurisdiction of ISO/TC 85/WG 3.Current edition approved

13、 Aug. 1, 2013. Published November 2013. Originallypublished as ASTM E 195698. The present International Standard ISO/ASTM51956:2013(E) is a major revision of the last previous edition ISO/ASTM51956:2005(E).2For referenced ASTM and ISO/ASTM standards, visit the ASTM website,www.astm.org, or contact A

14、STM Customer Service at serviceastm.org. ForAnnual Book of ASTM Standards volume information, refer to the standardsDocument Summary page on the ASTM website.3Document produced by Working Group 1 of the Joint Committee for Guides inMetrology (JCGM/WG 1). Available free of charge at the BIPM website

15、(http:/www.bipm.org. ISO/ASTM International 2017 All rights reservedThis international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recommendation

16、s issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.1JCGM 200:2008, VIM, International Vocabulary ofMetrologyBasis and general concepts and associatedterms42.4 ISO Standard:ISO 10012 Measurement Management SystemsRequirements for Measurement Processes and MeasuringEq

17、uipment52.5 International Commission on Radiation Units and Mea-surements (ICRU) Report:ICRU Report 85a Fundamental Quantities and Units forIonizing Radiation63. Terminology3.1 Definitions:3.1.1 annealingthermal treatment of a TLD prior to irra-diation or prior to readout.3.1.1.1 DiscussionPre-irrad

18、iation annealing of TLDs isusually done to erase the effects of previous irradiation and toreadjust the sensitivity of the phosphor; pre-readout annealingusually is done to reduce low-temperature TLD response.3.1.2 calibrationset of operations that establish, underspecified conditions, the relations

19、hip between values of quan-tities indicated by a measuring instrument or measuringsystem, or values represented by a material measure or areference material, and the corresponding values realized bystandards.3.1.2.1 DiscussionCalibration conditions include environ-mental and irradiation conditions p

20、resent during irradiation,storage and measurement of the dosimeters that are used for thegeneration of a calibration curve. To achieve stable environ-mental conditions, it may be necessary to condition thedosimeters before performing the calibration procedure.3.1.3 calibration curveexpression of the

21、 relation betweenindication and corresponding measured quantity value. (VIM)3.1.4 charged-particle equilibriumcondition in which thekinetic energy of charged particles (or electrons), excludingrest mass, entering an infinitesimal volume of the irradiatedmaterial equals the kinetic energy of charge p

22、articles (orelectrons) emerging from it.3.1.4.1 DiscussionWhen electrons are the predominantcharged particles, the term “electron equilibrium” is often usedto describe charged-particle equilibrium.3.1.5 dosimeter batchquantity of dosimeters made from aspecific mass of material with uniform compositi

23、on, fabricatedin a single production run under controlled, consistentconditions, and having a unique identification code.3.1.6 dosimeter stockpart of a dosimeter batch held by theuser.3.1.7 dosimetry systemsystem used for measuring ab-sorbed dose, consisting of dosimeters, measurement instru-ments a

24、nd their associated reference standards, and proceduresfor the systems use.3.1.8 electron equilibriumcharged-particle equilibriumfor electrons. See charged-particle equilibrium.3.1.9 measurement management systemset of interrelatedor interacting elements necessary to achieve metrologicalconfirmation

25、 and continual control of measurement processes.(ISO 10012)3.1.10 quality assuranceall systematic actions necessaryto provide adequate confidence that a calibration,measurement, or process is performed to a predefined level ofquality.3.1.11 reference standard dosimetry systemdosimetrysystem, general

26、ly having the highest metrological qualityavailable at a given location or in a given organization, fromwhich measurements made there are derived.3.1.12 routine dosimetry systemdosimetry system cali-brated against a reference standard dosimetry system and usedfor routine absorbed dose measurements,

27、including dose map-ping and process monitoring.3.1.13 thermoluminescence dosimeter (TLD)TLphosphor, alone or incorporated in a material, used for deter-mining the absorbed dose to materials.3.1.13.1 DiscussionFor example, the TL phosphor issometimes incorporated in a TFE-fluorocarbon matrix.3.1.14 t

28、hermoluminescence dosimeter reader (TLDreader)instrument used to measure the light emitted from aTLD consisting essentially of a heating element, a light-measuring device, and appropriate electronics.3.1.15 thermoluminescence dosimeter response (TLDresponse)light emitted by the TLD and read out duri

29、ng itsheating cycle consisting of one of the following: (a) the totallight output over the entire heating cycle, (b) a part of that totallight output, or (c) the peak amplitude of the light output.3.1.16 thermoluminescence phosphor (TL phosphor)material that stores, upon irradiation, a fraction of i

30、ts absorbeddose in various excited energy states and when thermallystimulated, it emits this stored energy as ultraviolet, visible,and infrared lights.3.1.17 TLD preparationprocedure of cleaning, annealing,and encapsulating the TL phosphor prior to irradiation.3.2 Definitions of other terms used in

31、this standard thatpertain to radiation measurement and dosimetry may be foundin ASTM Terminology E170. Definitions in ASTM Terminol-ogy E170 are compatible with ICRU Report 85a; thatdocument, therefore, may be used as an alternative reference.4. Significance and use4.1 In radiation processing, TLDs

32、are mainly used in theirradiation of blood products (see ISO/ASTM Practice 51939)and insects for sterile insect release programs (see ISO/ASTMGuide 51940). TLDs may also be used in other radiationprocessing applications such as the sterilization of medical4Document produced by Working Group 2 of the

33、 Joint Committee for Guides inMetrology (JCGM/WG 2). Available free of charge at the BIPM website (http:/www.bipm.org.5Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036, http:/www.ansi.org.6Available from International Commission on Radiation U

34、nits andMeasurements, 7910 Woodmont Ave., Suite 800, Bethesda, MD 20814, USA.ISO/ASTM 51956:2013(E)2 ISO/ASTM International 2017 All rights reserved products, food irradiation, modification of polymers, irradia-tion of electronic devices, and curing of inks, coatings andadhesives. (See ISO/ASTM Prac

35、tices 51608, 51649, and51702.)4.2 For radiation processing, the absorbed-dose range ofinterest is from 1 Gy to 100 kGy. Some TLDs can be used inapplications requiring much lower absorbed doses (forexample, for personnel dosimetry), but such applications areoutside the scope of this practice. Example

36、s of TLDs andapplicable dose ranges are given in Table 1. Information onvarious types of TLDs and their applications can be found inRefs (1-10).75. Overview5.1 During the irradiation of certain crystalline materials,for example, LiF, CaF2, CaSO4,Li2B4O7, andAl2O3, the fillingof electron and hole tra

37、ps between the ground state and theconduction band results in stored energy that can be released asluminescence during subsequent heating.TLD systems providea means of determining absorbed dose to materials by measur-ing this luminescence by the controlled heating of the irradi-ated crystalline mate

38、rial. The amount of luminescence emittedby the TL phosphor upon heating can be directly related toabsorbed dose by a calibration.5.2 TLDs can be reused by subjecting the irradiated TLDsto an annealing process at a higher temperature to release allthe electron and hole traps.6. Influence quantities6.

39、1 Factors other than absorbed dose which influence thedosimeter response are referred to as influence quantities andare discussed in the following sections. Examples of suchfactors are temperature, relative humidity, light and dose rate(see ISO/ASTM Guide 52701). See Refs (1-10) for examplesof the t

40、ypes and magnitudes of the effects for different TLDs.6.2 Pre-Irradiation Conditions:6.2.1 Dosimeter PackagingThe TLD response is not usu-ally influenced by the water content, so the TLDs are notusually supplied in vapor tight pouches. They may be suppliedin light tight pouches to minimize the effec

41、t of light.6.2.2 Time Since ManufactureThere is no known influ-ence of time since manufacture on TLDs when stored underrecommended conditions. However, it is recommended thatusers carry out periodic performance verification of responseover the time the dosimeter batch is used.6.2.3 TemperatureExposu

42、re to extreme temperature dur-ing shipment and storage at the users facility might affect theTLD response. Manufacturer should be consulted for specificrecommendation for dosimeter shipment and storage.6.2.4 Relative HumidityThe TLD response is not usuallyaffected by environmental changes in humidit

43、y.6.2.5 Exposure to LightTLDs with high sensitivity shouldbe packaged to protect them from light such as sunlight orfluorescent light which have an appreciable ultraviolet compo-nent. Prolonged exposure to ultraviolet light before irradiationcan cause spurious TLD response or enhanced post-irradiati

44、onfading. Incandescent lighting should be used for the TLDpreparation and readout areas. However, brief exposures of afew minutes to normal room fluorescent lighting is not likely tosignificantly affect the TLD response except for low dosemeasurements (102Gy) recalibration may be required after each

45、anneal-irradiation cycle because of possible changes inabsorbed-dose sensitivity (7). If the TLD system being used issubject to this effect, it is recommended that each TLD in thebatch be irradiated only once until the entire batch has beenused after which the entire batch can be annealed and a newc

46、alibration performed. In addition, because of possible changesin batch response uniformity due to high absorbed-doseirradiations, periodically repeat the tests.10. Routine use10.1 Before Irradiation:10.1.1 TLDs may be used either as reusable or as single-usedosimeters. Single-use dosimeters are irra

47、diated once, read out,and then discarded; they are generally used as received fromthe manufacturer. Dosimeters that are reused are cycled repeat-edly through an anneal-irradiation-readout procedure.ISO/ASTM 51956:2013(E)4 ISO/ASTM International 2017 All rights reserved 10.1.2 Preparation of the TLDs

48、 for irradiation may requirecleaning, annealing, or encapsulation, or combinations thereof,depending on the type and form of the TL phosphor.10.1.3 Reusable TLDs require careful treatment duringannealing in order to obtain reliable results in dose measure-ments. The annealing procedure should includ

49、e a reproducibletemperature cycle of the annealing oven, accurate timing of theannealing period, and a reproducible cooling rate.10.1.4 Ensure that the dosimeters are selected from anapproved batch stored according to users procedures. Theseprocedures should be based on manufacturers written recom-mendations or user specific performance characterization re-sults.10.1.5 Use only dosimeters

展开阅读全文
相关资源
  • ANSI Z97 1-2009 American National Standard for Safety Glazing Materials used in Buildings - Safety Performance Specifications and Methods of Test《建筑物中窗用玻璃材料安全性用.pdfANSI Z97 1-2009 American National Standard for Safety Glazing Materials used in Buildings - Safety Performance Specifications and Methods of Test《建筑物中窗用玻璃材料安全性用.pdf
  • ANSI Z97 1 ERTA-2010 Re ANSI Z97 1 - 2009 Errata《修订版 美国国家标准学会Z97 1-2009标准的勘误表》.pdfANSI Z97 1 ERTA-2010 Re ANSI Z97 1 - 2009 Errata《修订版 美国国家标准学会Z97 1-2009标准的勘误表》.pdf
  • ANSI Z21 40 2a-1997 Gas-Fired Work Activated Air-Conditioning and Heat Pump Appliances (Same as CGA 2 92a)《燃气、工作激活空气调节和热泵器具(同 CGA 2 92a)》.pdfANSI Z21 40 2a-1997 Gas-Fired Work Activated Air-Conditioning and Heat Pump Appliances (Same as CGA 2 92a)《燃气、工作激活空气调节和热泵器具(同 CGA 2 92a)》.pdf
  • ANSI Z124 9-2004 American National Standard for Plastic Urinal Fixtures《塑料小便器用美国国家标准》.pdfANSI Z124 9-2004 American National Standard for Plastic Urinal Fixtures《塑料小便器用美国国家标准》.pdf
  • ANSI Z124 4-2006 American National Standard for Plastic Water Closet Bowls and Tanks《塑料抽水马桶和水箱用美国国家标准》.pdfANSI Z124 4-2006 American National Standard for Plastic Water Closet Bowls and Tanks《塑料抽水马桶和水箱用美国国家标准》.pdf
  • ANSI Z124 3-2005 American National Standard for Plastic Lavatories《塑料洗脸盆用美国国家标准》.pdfANSI Z124 3-2005 American National Standard for Plastic Lavatories《塑料洗脸盆用美国国家标准》.pdf
  • ANSI T1 659-1996 Telecommunications - Mobility Management Application Protocol (MMAP) RCF-RACF Operations《电信 可移动管理应用协议(MMAP) RCF-RACF操作》.pdfANSI T1 659-1996 Telecommunications - Mobility Management Application Protocol (MMAP) RCF-RACF Operations《电信 可移动管理应用协议(MMAP) RCF-RACF操作》.pdf
  • ANSI T1 651-1996 Telecommunications – Mobility Management Application Protocol (MMAP)《电信 可移动性管理应用协议》.pdfANSI T1 651-1996 Telecommunications – Mobility Management Application Protocol (MMAP)《电信 可移动性管理应用协议》.pdf
  • ANSI T1 609-1999 Interworking between the ISDN User-Network Interface Protocol and the Signalling System Number 7 ISDN User Part《电信 ISDN用户间网络接口协议和7号信令系统ISDN用户部分.pdfANSI T1 609-1999 Interworking between the ISDN User-Network Interface Protocol and the Signalling System Number 7 ISDN User Part《电信 ISDN用户间网络接口协议和7号信令系统ISDN用户部分.pdf
  • ANSI T1 605-1991 Integrated Services Digital Network (ISDN) - Basic Access Interface for S and T Reference Points (Layer 1 Specification)《综合服务数字网络(ISDN) S和T基准点的.pdfANSI T1 605-1991 Integrated Services Digital Network (ISDN) - Basic Access Interface for S and T Reference Points (Layer 1 Specification)《综合服务数字网络(ISDN) S和T基准点的.pdf
  • 猜你喜欢
    相关搜索

    当前位置:首页 > 标准规范 > 国际标准 > ANSI

    copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
    备案/许可证编号:苏ICP备17064731号-1