ASTM D1481-2012 Standard Test Method for Density and Relative Density (Specific Gravity) of Viscous Materials by Lipkin Bicapillary Pycnometer《用利普金双毛细管比重计测定粘性材料密度和相对密度(比重)的标准试验方法》.pdf

上传人:brainfellow396 文档编号:509873 上传时间:2018-12-01 格式:PDF 页数:6 大小:166.74KB
下载 相关 举报
ASTM D1481-2012 Standard Test Method for Density and Relative Density (Specific Gravity) of Viscous Materials by Lipkin Bicapillary Pycnometer《用利普金双毛细管比重计测定粘性材料密度和相对密度(比重)的标准试验方法》.pdf_第1页
第1页 / 共6页
ASTM D1481-2012 Standard Test Method for Density and Relative Density (Specific Gravity) of Viscous Materials by Lipkin Bicapillary Pycnometer《用利普金双毛细管比重计测定粘性材料密度和相对密度(比重)的标准试验方法》.pdf_第2页
第2页 / 共6页
ASTM D1481-2012 Standard Test Method for Density and Relative Density (Specific Gravity) of Viscous Materials by Lipkin Bicapillary Pycnometer《用利普金双毛细管比重计测定粘性材料密度和相对密度(比重)的标准试验方法》.pdf_第3页
第3页 / 共6页
ASTM D1481-2012 Standard Test Method for Density and Relative Density (Specific Gravity) of Viscous Materials by Lipkin Bicapillary Pycnometer《用利普金双毛细管比重计测定粘性材料密度和相对密度(比重)的标准试验方法》.pdf_第4页
第4页 / 共6页
ASTM D1481-2012 Standard Test Method for Density and Relative Density (Specific Gravity) of Viscous Materials by Lipkin Bicapillary Pycnometer《用利普金双毛细管比重计测定粘性材料密度和相对密度(比重)的标准试验方法》.pdf_第5页
第5页 / 共6页
亲,该文档总共6页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、Designation: D1481 12Standard Test Method forDensity and Relative Density (Specific Gravity) of ViscousMaterials by Lipkin Bicapillary Pycnometer1This standard is issued under the fixed designation D1481; the number immediately following the designation indicates the year oforiginal adoption or, in

2、the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method covers the determination of the densityof oils more viscous than 1

3、5 cSt at 20C (mm2/s), and ofviscous oils and melted waxes at elevated temperatures, but notat temperatures at which the sample would have a vaporpressure of 100 mm Hg (13 kPa) or above.NOTE 1To determine the densities of less viscous liquids at 20 or25C use Test Method D1217.1.2 This test method pro

4、vides a calculation procedure forconverting density to relative density (specific gravity).1.3 WARNINGMercury has been designated by manyregulatory agencies as a hazardous material that can causecentral nervous system, kidney and liver damage. Mercury, orits vapor, may be hazardous to health and cor

5、rosive tomaterials. Caution should be taken when handling mercury andmercury containing products. See the applicable product Ma-terial Safety Data Sheet (MSDS) for details and EPAswebsitehttp:/www.epa.gov/mercury/faq.htmfor addi-tional information. Users should be aware that selling mercuryand/or me

6、rcury containing products into your state or countrymay be prohibited by law.1.4 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with it

7、s use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D1217 Test Method for Density and Relative Density (Spe-cific Gravity)

8、of Liquids by Bingham PycnometerD1250 Guide for Use of the Petroleum Measurement TablesD4052 Test Method for Density, Relative Density, and APIGravity of Liquids by Digital Density Meter3. Terminology3.1 Definitions:3.1.1 density, nmass per unit volume at a specifiedtemperature. D40523.1.2 relative

9、density (specific gravity), nthe ratio of thedensity of a material at a stated temperature to the density ofwater at a stated temperature. D40524. Summary of Test Method34.1 The liquid is drawn into the bicapillary pycnometerthrough the removable siphon arm and adjusted to volume atthe temperature o

10、f test, in such a manner that there ispractically no drainage in the unfilled tubing. After equilibra-tion at the test temperature, liquid levels are read, and thepycnometer is removed from the thermostated bath, cooled toroom temperature, and weighed.4.2 Density or relative density (specific gravit

11、y), as desired,is then calculated from the volume at the test temperature andthe weight of the sample. The effect of air buoyancy is includedin the calculations.1This test method is under the jurisdiction of ASTM Committee D02 onPetroleum Products and Lubricants and is the direct responsibility of S

12、ubcommitteeD02.04.0D on Physical and Chemical Methods.Current edition approved Nov. 1, 2012. Published December 2012. Originallyapproved in 1957. Last previous edition approved in 2007 as D148102(2007).DOI: 10.1520/D1481-12.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcont

13、act ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3For a more complete discussion of this procedure, see Lipkin, M. R., Mills, I.W., Martin, C. C., and Harvey, W. T., Analytical Chemistry

14、, ANCHA, Vol 21, 1949,p. 504.*A Summary of Changes section appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States15. Significance and Use5.1 Density is a fundamental physical property that can beused in con

15、junction with other properties to characterize boththe light and heavy fractions of petroleum and to access thequality of crude oils.5.2 Determination of the density or relative density ofpetroleum and its products is necessary for the conversion ofmeasured volumes to volumes at the standard tempera

16、tures of15C.5.3 The determination of densities at the elevated tempera-tures of 40 and 100C is particularly useful in providing thedata needed for the conversion of kinematic viscosities incentistokes (mm2/s) to the corresponding dynamic viscositiesin centipoises (mPas).6. Apparatus6.1 Pycnometer4As

17、ide-arm type of pycnometer conform-ing to the dimensions given in Fig. 1 and made of borosilicateglass. The weight shall not exceed 35 g without the side arm.6.2 RackA rack to use in filling the pycnometer (see Fig.2).6.3 Constant-Temperature OvenAn oven for use in fillingthe pycnometer. Any oven ca

18、pable of holding the filling rack,and of maintaining a temperature of approximately 100C, canbe used.6.4 Constant-Temperature BathA mixture of water andglycerin, or oil bath having a depth of at least 305 mm (12 in.)and provided with heating, stirring, and thermostating devicesadequate to maintain d

19、esired temperatures in the range from 20to 100C with an accuracy of 60.01C.6.5 Bath ThermometersThermometers graduated in 0.1Csubdivisions and standardized for the range of use to thenearest 0.01C (ASTM Saybolt Viscosity Thermometers 17Cto 22C are recommended). For most hydrocarbons, the densitycoef

20、ficient is about 0.0008 units/C, and therefore a tempera-ture error of 60.013C would cause an error of 60.000 01 indensity.6.6 Pycnometer HolderA holder, as shown in Fig. 3,isrecommended for supporting the pycnometer in the bath. Asingle clamp device may be used.6.7 BalanceAbalance able to reproduce

21、 weighings within0.1 mg when carrying a load of 35 g or less on each pan. Thebalance shall be located in a room shielded from drafts andfumes and in which the temperature changes between relatedweighings (empty and filled pycnometer) do not cause asignificant change in the ratio of the balance arms.

22、 Otherwise,weighings shall be made by the substitution method in whichthe calibrated weights and pycnometer are alternativelyweighed on the same balance pan. The same balance shall beused for all related weighings.6.8 WeightsWeights shall be used whose relative valuesare known to the nearest 0.05 mg

23、 or better. The same set ofweights shall be used for the calibration of the pycnometer andthe determination of the densities, or the sets of weights shallbe calibrated relative to each other.4The sole source of supply of the pycnometers known to the committee at thistime is Reliance Glass Co., 220 G

24、ateway Rd., Bensenville, IL 60106-0825 havebeen found satisfactory. If you are aware of alternative suppliers, please provide thisinformation to ASTM International Headquarters. Your comments will receivecareful consideration at a meeting of the responsible technical committee,1whichyou may attend.F

25、IG. 1 PycnometerFIG. 2 Rack for Filling PycnometerD1481 1227. Reagents and Materials7.1 Acetone(Warning Extremely flammable. Use ad-equate ventilation.)7.2 Isopentane( WarningExtremely flammable. Avoidbuildup of vapors and remove all sources of ignition, especiallynonexplosion-proof electrical appar

26、atus.)7.3 Chromic Acid (Potassium Dichromate/Conc. SulfuricAcid)(WarningCauses severe burns. A recognized car-cinogen. Do not get in eyes, on skin or clothing.)7.4 Xylenes(Warning Flammable liquid. Aspirationhazard. May irritate skin, eyes, respiratory tract or digestivetract, or both. May cause cen

27、tral nervous system depression,liver and kidney damage, or exhibit reproductive and fetaleffects, or both.)8. Preparation of Apparatus8.1 Thoroughly clean the pycnometer and side arm with hotchromic acid cleaning solution (WarningSee 7.4). Chromicacid solution is the most effective cleaning agent. H

28、owever,surfactant cleaning fluids have also been used successfully.Rinse well with distilled water; and dry at 105 to 110C for atleast 1 h, preferably with a slow current of filtered air passingthrough the pycnometer. Cleaning shall be done in this mannerwhenever the pycnometer is to be calibrated o

29、r whenever liquidfails to drain cleanly from the walls of the pycnometer or itscapillary. Ordinarily, the pycnometer may be cleaned betweendeterminations by washing with a suitable solvent, such asisopentane or xylenes, and vacuum drying. If acetone is used asthe wash liquid, the pycnometer should t

30、hen be rinsed withisopentane or xylenes.9. Calibration of Pycnometer9.1 Weigh the clean, dry pycnometer (without the side arm)to the nearest 0.1 mg, and record the weight.9.2 Fill the pycnometer with freshly boiled distilled water.This may be conveniently done by placing the pycnometer inthe holder

31、with the side arm dipping into a sample cupcontaining water. Allow the pycnometer to fill by siphoning.Break the siphon by removing the side arm when the liquidlevel in the bulb arm of the pycnometer reaches 6 on the scale.9.3 Remove the side arm which was used to fill thepycnometer and remove exces

32、s liquid from the capillary tip bywiping with a small piece of absorbent paper.9.4 Place the pycnometer in the holder in the constant-temperature bath at temperature t with the liquid level in thecapillaries below the liquid level in the bath. When the liquidlevel has reached equilibrium (not less t

33、han 15 min), read thescale to the nearest 0.2 small division at the liquid level in eacharm. After 5 min, read the liquid level again. If the sum of thescale readings in each reading differs by more than 60.04,repeat readings at 5-min intervals. When readings are constant,record.9.5 Remove the pycno

34、meter from the bath and allow it tocome to room temperature. Rinse the outer surface withdistilled water, with acetone, then with redistilled xylenes, anddry thoroughly with a chemically clean lint-free cloth, slightlydamp with water.Allow to stand a few minutes, and then weighto nearest 0.1 mg.NOTE

35、 2In atmospheres of low humidity (60 % or lower), drying thepycnometer by rubbing with dry cotton cloth will induce static chargesequivalent to a loss of about 1 mg or more in the weight of thepycnometer. This charge may not be completely dissipated in less than12h and can be detected by touching th

36、e pycnometer to the wire hook on thebalance and then drawing it away slowly. If the pycnometer exhibits anattraction for the wire hook, it may be considered to have a static charge.9.6 Repeat the above, but break the siphon when water hasreached the 3 mark in the bulb arm, and in the next experiment

37、,at the 0 mark in the bulb arm. Obtain the apparent volume foreach filling by dividing the weight of water held by thepycnometer in each experiment by the density of water at thecalibration temperature t. Calibration shall be made at 20, 40,and 50C. Prepare a calibration curve for 20C by plotting th

38、esum of the two scale readings versus the apparent volume at20C. If the curve is not a straight line, and future checks donot correct it, discard the pycnometer. The line shall not bemore than 0.0002 mL/unit from any one determined point.Metric Equivalentsin. mm in. mm in. mm12 3.212 12.7 158 41.314

39、 6.4916 14.3 2316 55.7516 7.934 19.1 712 19138 9.5 112 38.1FIG. 3 Pycnometer HolderD1481 1239.7 Corresponding calibration curves shall be made for 40and 50C. These calibration curves are checked using thefollowing equation:V25 V111ct! (1)where:V2= apparent volume at test temperature,V1= apparent vol

40、ume at 20C, andc = cubical coefficient of expansion of borosilicate glass(9.9 106/C).The calculated and determined curves at 40 and 50C shouldcheck to within 60.0002 mL/unit at all points. The calibrationcurves for higher temperatures shall be obtained by calculation.10. Procedure10.1 Weigh the clea

41、n, dry pycnometer, without the side arm,to 0.1 mg and record the weight.10.2 Place a 10-mL sample beaker in the wooden rack (Fig.2). Before attaching the side arm to the pycnometer, drain afew drops of sample through the side arm to wet the insidesurface and reduce the chance of trapping air bubbles

42、 in thecapillary during the filling operation. Place the side arm on thepycnometer, and place the assembly on the rack with the sidearm dipping into the sample beaker as shown in Fig. 4.10.3 In filling the pycnometer with very viscous oils orhigh-melting waxes, place the whole filling assembly in ah

43、ot-air oven to facilitate filling. An oven at approximately100C is usually hot enough for this purpose.10.4 Apply gentle suction to the bulb arm of the pycnometerto start the siphoning action. The suction must be gentle toavoid the formation of bubbles. After siphoning is started,allow filling by si

44、phoning to continue until the liquid level inthe bulb arm ceases to rise. Then remove the pycnometer fromthe rack and place in the thermostated bath, in the same tiltedposition, until the oil ceases to contract.At this point, place thepycnometer in an upright position, and allow the liquid level int

45、he bulb arm to reach the upper portion of the calibratedcapillary, but not above 6.4. Stop siphoning by removing theside arm.NOTE 3With viscous oils, it will reduce drainage errors to fill to the6.0 to 6.4 mark, and it may be necessary to apply a little suction to the longarm during cooling to preve

46、nt the meniscus in the bulb arm from falling.Maintain the meniscus at about the same level in the long arm throughoutthe whole determination.FIG. 4 Pycnometer Filling AssemblyD1481 12410.5 After removing the side-arm cap from the short arm ofthe pycnometer, wipe the tip and ground joint of thepycnom

47、eter, and adjust it to an upright position in the thermo-stated bath. The bath liquid level shall be above the 6 mark onthe pycnometer and below the ground glass tip of the pycnom-eter.10.6 Allow 15 min for equilibrium to be obtained. After thestated 15-min time for coming to equilibrium, read the m

48、enis-cus levels in both arms of the pycnometer to the nearest 0.2 ofthe smallest scale division. Wait 5 min and check readings. Ifthe sum of the readings at the two different times do not agreeto within 60.04, repeat at 5-min intervals until checks areobtained. Record the sum of these readings and a

49、lso record thecorresponding apparent volume from the calibration curve forthe same temperature.NOTE 4The final level of oil in the pycnometer should not be morethan 5 mm below the tip of the ground glass end of the pycnometer, andthe level in the long (bulb) side of the pycnometer should be no lower thanit has been at any time during the procedure. With these precautions,drainage error (which is important with very viscous samples) is entirelyeliminated.10.7 Remove the pycnometer from the bath and tilt it so thatthe