ASTM D7236-2007 459 Standard Test Method for Flash Point by Small Scale Closed Cup Tester (Ramp Method)《小型闭杯试验器对闪点的测定用标准试验方法(斜坡法)》.pdf

上传人:syndromehi216 文档编号:524901 上传时间:2018-12-04 格式:PDF 页数:6 大小:223.91KB
下载 相关 举报
ASTM D7236-2007 459 Standard Test Method for Flash Point by Small Scale Closed Cup Tester (Ramp Method)《小型闭杯试验器对闪点的测定用标准试验方法(斜坡法)》.pdf_第1页
第1页 / 共6页
ASTM D7236-2007 459 Standard Test Method for Flash Point by Small Scale Closed Cup Tester (Ramp Method)《小型闭杯试验器对闪点的测定用标准试验方法(斜坡法)》.pdf_第2页
第2页 / 共6页
ASTM D7236-2007 459 Standard Test Method for Flash Point by Small Scale Closed Cup Tester (Ramp Method)《小型闭杯试验器对闪点的测定用标准试验方法(斜坡法)》.pdf_第3页
第3页 / 共6页
ASTM D7236-2007 459 Standard Test Method for Flash Point by Small Scale Closed Cup Tester (Ramp Method)《小型闭杯试验器对闪点的测定用标准试验方法(斜坡法)》.pdf_第4页
第4页 / 共6页
ASTM D7236-2007 459 Standard Test Method for Flash Point by Small Scale Closed Cup Tester (Ramp Method)《小型闭杯试验器对闪点的测定用标准试验方法(斜坡法)》.pdf_第5页
第5页 / 共6页
亲,该文档总共6页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、Designation: D 7236 07An American National StandardStandard Test Method forFlash Point by Small Scale Closed Cup Tester (RampMethod)1, 2This standard is issued under the fixed designation D 7236; the number immediately following the designation indicates the year oforiginal adoption or, in the case

2、of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.INTRODUCTIONThis flash point test method is a dynamic test method and depends on a definite rate of temp

3、eratureincrease to control the precision of the test method. The rate of heating may not in all cases give theprecision quoted in the test method because of the low thermal conductivity of certain materials. Theuse of an equilibrium method such as Test Methods D 3828, Method B improves the predictio

4、n offlammability for such materials, as the vapors above the test specimen and the test specimen are closerto thermal equilibrium.Flash point values are a function of the apparatus design, the condition of the apparatus used, andthe operational procedure carried out. Flash point can therefore only b

5、e defined in terms of a standardtest method, and no general valid correlation can be guaranteed between results obtained by differenttest methods or with test apparatus different than that specified.1. Scope*1.1 This test method covers the determination of the flashpoint of aviation turbine fuel, di

6、esel fuel, kerosine and relatedproducts in the temperature range of 40 to 135C by a smallscale closed cup apparatus.1.2 This test method is only applicable to homogeneousmaterials that are liquid at or near ambient temperature and attemperatures required to perform the test.1.3 This test method is n

7、ot applicable to liquids contami-nated by traces of highly volatile materials.1.4 This test method is a dynamic method and depends on adefinite rate of temperature increase. It is one of many flashpoint methods available, and every flash point test method,including this one, is an empirical one.1.5

8、If the users specification requires a defined flash pointmethod, neither this test nor any other method should besubstituted for the prescribed method without obtaining com-parative data and an agreement from the specifier.1.6 The values stated in SI units are to be regarded asstandard. No other uni

9、ts of measurement are included in thisstandard.1.7 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regu

10、latory limitations prior to use. For specific hazardstatements, see Section 7 and the Material Safety Data Sheetfor the product being tested.2. Referenced Documents2.1 ASTM Standards:3D 3828 Test Methods for Flash Point by Small Scale ClosedCup TesterD 4057 Practice for Manual Sampling of Petroleum

11、andPetroleum ProductsD 4177 Practice for Automatic Sampling of Petroleum andPetroleum ProductsD 6299 Practice for Applying Statistical Quality AssuranceTechniques to Evaluate Analytical Measurement SystemPerformanceD 6300 Practice for Determination of Precision and BiasData for Use in Test Methods f

12、or Petroleum Products andLubricantsE 300 Practice for Sampling Industrial Chemicals2.2 ISO Standards:4ISO Guide 34 Quality Systems Guidelines for the Produc-tion of Reference Materials1This test method is under the jurisdiction of ASTM Committee D02 onPetroleum Products and Lubricants and is the dir

13、ect responsibility of SubcommitteeD02.08 on Volatility.Current edition approved Nov. 1, 2007. Published January 2008. Originallyapproved in 2006. Last previous edition approved in 2006 as D 723606.2This test method is being jointly developed and harmonized with the EnergyInstitute.3For referenced AS

14、TM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.4Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4t

15、h Floor, New York, NY 10036.1*A Summary of Changes section appears at the end of this standard.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.ISO Guide 35 Certification of Reference MaterialGeneraland Statistical Principles3. Termino

16、logy3.1 Definitions:3.1.1 flash point, nlowest temperature corrected to apressure of 101.3 kPa, at which application of a test flamecauses the vapors of a test specimen of the sample to ignitemomentarily under the specified conditions of the test.3.2 Definitions of Terms Specific to This Standard:3.

17、2.1 dynamic, adjthe condition where the vapor abovethe test specimen, and the test cup, are not in temperatureequilibrium at the time the test flame is applied.3.2.2 ramp method, nin flash point methods, method thathas a test temperature increase at a set rate.4. Summary of Test Method4.1 A 2 6 0.1

18、mL test specimen is introduced into the testcup that is then heated automatically at a constant rate of 2 60.5C/min. A gas test flame is directed through an openingshutter, in the test cup lid, at specified temperature intervalsuntil a flash is detected by the automatic flash detector. Theflash is r

19、eported as defined in 3.1.1.5. Significance and Use5.1 The flash point temperature is one measure of thetendency of the test specimen to form a flammable mixturewith air under controlled laboratory conditions. It is only oneof a number of properties that must be considered in assessingthe overall fl

20、ammability hazard of a material.5.2 Flash point is used in shipping and safety regulations todefine flammable and combustible materials and classify them.Consult the particular regulation involved for precise defini-tions of these classifications.5.3 This test method can be used to measure and descr

21、ibethe properties of materials in response to heat and a test flameunder controlled laboratory conditions and shall not be used todescribe or appraise the fire hazard or fire risk of materialsunder actual fire conditions. However, results of this testmethod may be used as elements of a fire risk ass

22、essment, thattakes into account all of the factors that are pertinent to anassessment of the fire hazard of a particular end use.6. Apparatus6.1 Flash Point ApparatusThe apparatus consists of anelectrically heated test cup that is controlled automatically togive a temperature increase of 2.0 6 0.5C/

23、min, a lid andshutter assembly, a pilot and test flame, an automatic flashdetector and a temperature display that displays and holds thetest temperature when a flash is detected. The key parts anddimensions are described in Annex A1 and illustrated in Fig.A1.1.6.2 Syringe, 2 mL, adjusted to deliver

24、2.00 6 0.1 mL.6.3 Barometer, accurate to 0.5 kPa. Barometers pre-corrected to give sea level readings, such as those used atweather stations and airports, shall not be used.NOTE 1Automatic barometric correction in accordance with 12.2 maybe installed in the apparatus.6.4 Cooling Bath or Freezer, for

25、 cooling the samples, ifrequired, and capable of cooling the sample to at least 10Cbelow the expected flash point. If a freezer is used, it shall beof explosion-protected design.6.5 Draft Shield, a shield fitted at the back and on two sidesof the instrument, for use in circumstances where naturalpro

26、tection from drafts does not exist.NOTE 2A shield 350-mm high, 480-mm wide, and 240-mm deep, issuitable.7. Reagents and Materials7.1 Cleaning SolventUse only noncorrosive solvents ca-pable of cleaning the test cup and lid assembly. Two commonlyused solvents are toluene and acetone. (WarningToluene,a

27、cetone and many solvents are flammable and a health hazard.Dispose of solvents and waste material in accordance withlocal regulations.)7.2 Butane or Natural GasFor use as the pilot and testflame. (WarningButane and natural gases are flammableand a health hazard.)8. Sampling8.1 Obtain at least a 25 m

28、L sample from a bulk test site inaccordance with Practices D 4057, D 4177, E 300 or othercomparable sampling practices.8.2 Store samples in clean, tightly sealed containers atnormal room temperature (20 to 25C) or colder. Ensure thatthe container is 85 to 95 % full.8.3 Do not store samples for an ex

29、tended period of time ingas permeable containers, such as plastic, because volatilematerial may diffuse through the walls of the container.Samples in leaky containers are suspect and not a source ofvalid results.8.4 Erroneously high flash points can be obtained whenprecautions are not taken to avoid

30、 loss of volatile materials. Donot open containers unnecessarily. Do not make a transferunless the sample temperature is at least 10C below theexpected flash point. Where possible perform the flash point asthe first test on the sample.8.5 Samples containing dissolved or free water may bedehydrated w

31、ith calcium chloride. Warming the sample ispermitted, but do not heat the sample above a temperature of10C below its expected flash point. (WarningIf the sampleis expected of containing volatile components, the treatment asdescribed in 8.5 should be omitted.)8.6 Cool or adjust the temperature of the

32、 sample and itscontainer to at least 10C below the expected flash point beforeopening to remove the test specimen. If an aliquot of theoriginal sample is to be stored prior to testing, ensure that thecontainer is filled to between 85 and 95 % of its capacity.Gently mix the subsample to ensure unifor

33、mity while mini-mizing the possible loss of volatile components and light ends.NOTE 3Results of flash point tests may be affected if the samplevolume falls below 85 % of the containers capacity.9. Preparation of Apparatus9.1 Support the apparatus (Annex A1) on a level and steadysurface in a draft-fr

34、ee position (see Notes 4 and 5).D7236072NOTE 4A draft shield (6.5) is recommended to be used when naturalprotection from drafts does not exist.NOTE 5When testing materials that may produce toxic vapors, theapparatus may be located in a fume hood with an individual control of airflow, adjusted such t

35、hat the toxic vapors can be withdrawn withoutcausing air currents around the test cup during the test.9.2 Clean the test cup, cover, and its accessories with anappropriate solvent (7.1) to remove traces of gum or residuefrom the previous test.NOTE 6A stream of clean dry air may be used to remove the

36、 lasttraces of solvent used.NOTE 7The filler orifice may be cleaned with a pipe cleaner.10. Verification of Apparatus10.1 Verify that the temperature measuring device is inaccordance with A1.7.10.2 Verify the performance of the apparatus at least onceper year by determining the flash point of a Cert

37、ified ReferenceMaterial (CRM) that has a certified flash point reasonably closeto the expected range of the samples to be tested. Examples ofmaterials that may be used to create CRMs are listed inAppendix X1. The flash point shall be determined using theprocedures stated in Section 11. Appendix X1 i

38、ndicates typicallimits in Table X1.1 for a typical CRM and shows how tocalculate limits for an unlisted CRM.10.3 Once the performance of the apparatus has beenverified using a CRM, the flash point of Secondary WorkingStandards (SWS) can be determined along with their controllimits. These SWS can the

39、n be utilized for more frequentperformance checks. (See Appendix X1.)10.4 When the flash point obtained is not within the limitsstated in 10.2 or 10.3 check the condition and operation of theapparatus to ensure conformance to the details listed in AnnexA1, especially with regard to the action of the

40、 shutter, theposition and size of the test flame, the heating rate and thetemperature sensor readings. After any adjustment, repeat thetest in 10.2 or 10.3 using a fresh test specimen, with specialattention to the procedural details prescribed in the test method.11. Procedure11.1 Set the test cup te

41、mperature to 15 6 5C below theexpected flash point temperature.11.2 When the test cup is at the required temperature, chargea clean and dry syringe (6.2), adjusted to a temperature at least10C below the expected flash point, witha2mLtestspecimen. Close the sample container immediately after with-dra

42、wal of the test specimen, to minimize any loss of volatilecomponents.11.3 Carefully transfer the syringe to the filler orifice anddischarge the test specimen into the test cup by fully depressingthe syringe plunger. Remove the syringe.11.4 Ignite the pilot and test flames. Adjust the test flame toco

43、nform to the size of the 4 mm gauge ring. Reset the flashdetector (A1.9). Start the test.NOTE 8The ignition of the test flame and control of the gas supplymay be automated on some instruments.11.5 The test cup is heated automatically at such a rate thatthe temperature, as indicated by the temperatur

44、e sensor, in-creases by 1.5 to 2.5C/min. The test flame is dipped into thetest cup, as the shutter is slowly and uniformly opened andclosed overa2to3-speriod, at 1C intervals for temperaturesup to and including 100C and thereafter every 2C fortemperatures over 100C.11.5.1 Observe the first two dips

45、of the flame for continuousburning in the orifice. If a continuous luminous flame burns inthe orifice when the slide is opened and the test flame isintroduced, then the flash point lies considerably below the testtemperature. In such cases, discontinue the test and repeat thetest with a fresh test s

46、pecimen using a lower expected flashpoint.11.5.2 Application of the test flame may cause a blue halo oran enlarged flame prior to the actual flash point. This is not aflash and shall be ignored by the flash detector.11.6 Record the displayed and held temperature of thetemperature sensor when the app

47、lication of the test flamecauses a flash to be automatically detected.11.7 Record the ambient barometric pressure using a barom-eter (6.3) in the vicinity of the apparatus at the time of the test.NOTE 9It is not considered necessary to correct the barometricpressure reading to 0C, although some baro

48、meters are designed to makethis correction automatically.11.8 When the temperature at which the flash is observed isless than 10C or greater than 30C from the start temperatureof the test, the result is not valid. Repeat the test using a freshtest specimen adjusting the temperature of the first appl

49、icationof the test flame until a valid determination is made.12. Calculation12.1 Barometric Pressure ConversionIf the barometricpressure reading is measured in a unit other than kilopascals,convert to kilopascals using one of the following equations:Reading in hPa 3 0.1 5 kPa (1)Reading in mbar 3 0.1 5 kPa (2)Reading in mmHg 3 0.1333 5 kPa (3)12.2 Correction of Detected Flash Point to Standard Atmo-spheric PressureCalculate the flash point corrected to astandard atmospheric pressure of 101.3 kPa, Tc, using thefollowing equation:Tc5 TD1 0.25 101

展开阅读全文
相关资源
猜你喜欢
相关搜索

当前位置:首页 > 标准规范 > 国际标准 > ASTM

copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
备案/许可证编号:苏ICP备17064731号-1