ASTM D7598-2009e1 8750 Standard Test Method for Determination of Thiodiglycol in Water by Single Reaction Monitoring Liquid Chromatography Tandem Mass Spectrometry《用单反应检测液相色谱 串联质谱法.pdf

上传人:appealoxygen216 文档编号:525918 上传时间:2018-12-04 格式:PDF 页数:8 大小:262.96KB
下载 相关 举报
ASTM D7598-2009e1 8750 Standard Test Method for Determination of Thiodiglycol in Water by Single Reaction Monitoring Liquid Chromatography Tandem Mass Spectrometry《用单反应检测液相色谱 串联质谱法.pdf_第1页
第1页 / 共8页
ASTM D7598-2009e1 8750 Standard Test Method for Determination of Thiodiglycol in Water by Single Reaction Monitoring Liquid Chromatography Tandem Mass Spectrometry《用单反应检测液相色谱 串联质谱法.pdf_第2页
第2页 / 共8页
ASTM D7598-2009e1 8750 Standard Test Method for Determination of Thiodiglycol in Water by Single Reaction Monitoring Liquid Chromatography Tandem Mass Spectrometry《用单反应检测液相色谱 串联质谱法.pdf_第3页
第3页 / 共8页
ASTM D7598-2009e1 8750 Standard Test Method for Determination of Thiodiglycol in Water by Single Reaction Monitoring Liquid Chromatography Tandem Mass Spectrometry《用单反应检测液相色谱 串联质谱法.pdf_第4页
第4页 / 共8页
ASTM D7598-2009e1 8750 Standard Test Method for Determination of Thiodiglycol in Water by Single Reaction Monitoring Liquid Chromatography Tandem Mass Spectrometry《用单反应检测液相色谱 串联质谱法.pdf_第5页
第5页 / 共8页
亲,该文档总共8页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、Designation: D7598 091Standard Test Method forDetermination of Thiodiglycol in Water by Single ReactionMonitoring Liquid Chromatography/Tandem MassSpectrometry1This standard is issued under the fixed designation D7598; the number immediately following the designation indicates the year oforiginal ad

2、option or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1NOTEThis Test Method was changed editorially in 2012.1. Scope1.1 This procedure c

3、overs the determination of thiodiglycol(TDG) in surface water by direct injection using liquidchromatography (LC) and detected with tandem mass spec-trometry (MS/MS). TDG is qualitatively and quantitativelydetermined by this method. This method adheres to singlereaction monitoring (SRM) mass spectro

4、metry.1.2 This test method has been developed by US EPARegion5 Chicago Regional Laboratory (CRL).1.3 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.4 The Detection Verification Level (DVL) and ReportingRange for TDG are liste

5、d in Table 1.1.4.1 The DVL is required to be at a concentration at least3 times below the reporting limit (RL) and have a signal/noiseratio greater than 3:1. Fig. 1 displays the signal/noise ratio atthe DVL.1.4.2 The RL is the concentration of the level 1 calibrationstandard as shown in Table 2. The

6、 reporting limit for thismethod is 100 g/L.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulato

7、ry limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D1129 Terminology Relating to WaterD1193 Specification for Reagent WaterD2777 Practice for Determination of Precision and Bias ofApplicable Test Methods of Committee D19 on WaterD3856 Guide for Management Systems in LaboratoriesE

8、ngaged in Analysis of WaterD3694 Practices for Preparation of Sample Containers andfor Preservation of Organic ConstituentsD5847 Practice for Writing Quality Control Specificationsfor Standard Test Methods for Water AnalysisE2554 Practice for Estimating and Monitoring the Uncer-tainty of Test Result

9、s of a Test Method in a SingleLaboratory Using a Control Sample Program2.2 Other Documents:EPApublication SW-846 Test Methods for Evaluating SolidWaste, Physical/Chemical Methods33. Terminology3.1 Definitions:3.1.1 detection verification level (DVL), na concentrationthat has a signal/noise ratio gre

10、ater than 3:1 and is at least 3times below the reporting limit (RL).3.1.2 reporting limit (RL), nthe concentration of thelowest-level calibration standard used for quantification.3.2 Abbreviations:3.2.1 NDnon-detect4. Summary of Test Methods4.1 This is a performance based method and modificationsare

11、 allowed to improve performance.4.2 For thiodiglycol analysis, samples are shipped to the labbetween 0C and 6C and analyzed within 7 days of collection.In the lab, the samples are spiked with surrogate, filtered using1This test method is under the jurisdiction of ASTM Committee D19 on Waterand is th

12、e direct responsibility of Subcommittee D19.06 on Methods forAnalysis forOrganic Substances in Water.Current edition approved Dec. 1, 2009. Published February 2012. DOI: 10.1520/D7598-09E01.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at servi

13、ceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from National Technical Information Service (NTIS), U.S. Depart-ment of Commerce, 5285 Port Royal Road, Springfield, VA, 22161 or at http:/www.epa.gov/epawaste/

14、hazard/testmethods/index.htm.TABLE 1 Detection Verification Level and Reporting RangeAnalyte DVL (g/L) Reporting Range (g/L)Thiodiglycol 20 10010 0001Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.a syringe driven Millex HV PVDF filt

15、er unit and analyzeddirectly by LC/MS/MS.4.3 Thiodiglycol and 3,3-thiodipropanol (surrogate) areidentified by retention time and one SRM transition. The targetanalyte and surrogate are quantitated using the SRM transitionsutilizing an external calibration. The final report issued foreach sample list

16、s the concentration of TDG and the 3,3-thiodipropanol surrogate recovery.5. Significance and Use5.1 Thiodiglycol is a Schedule 2 compound under theChemical Weapons Convention (CWC). Schedule 2 chemicalsinclude those that are precursors to chemical weapons, chemi-cal weapons agents or have a number o

17、f other commercialuses. They are used as ingredients to produce insecticides,herbicides, lubricants, and some pharmaceutical products.Schedule 2 chemicals can be found in applications unrelated tochemical weapons. Thiodiglycol is both a mustard gas precur-sor and degradant as well as an ingredient i

18、n water-based inks,ballpoint pen inks, dyes and some pesticides.45.2 This method has been investigated for use with reagentand surface water.6. Interferences6.1 Method interferences may be caused by contaminants insolvents, reagents, glassware and other apparatus producingdiscrete artifacts or eleva

19、ted baselines. All of these materialsare demonstrated to be free from interferences by analyzinglaboratory reagent blanks under the same conditions assamples.6.2 All glassware is washed in hot water with a detergent,rinsed in hot water followed by distilled water. The glasswareis then dried and heat

20、ed in an oven at 250C for 15 to 30minutes. All glassware is subsequently cleaned with acetone,then methanol.6.3 All reagents and solvents should be pesticide residuepurity or higher to minimize interference problems.6.4 Matrix interferences may be caused by contaminantsthat are co-extracted from the

21、 sample. The extent of matrixinterferences can vary considerably from sample source de-pending on variations of the sample matrix.7. Apparatus7.1 LC/MS/MS System4Additional information about CWC and thiodiglycol is available on the Internetat http:/www.opcw.org (2009)FIG. 1 Example SRM Chromatograms

22、 Signal/Noise at Detection Verification LevelTABLE 2 Concentrations of Calibration Standards (PPB)Analyte/Surrogate LV 1 LV 2 LV 3 LV 4 LV 5 LV 6 LV 7Thiodiglycol 100 250 500 1000 2500 5000 10 0003,3-Thiodipropanol 100 250 500 1000 2500 5000 10 000D7598 09127.1.1 Liquid Chromatography (LC) SystemA c

23、ompleteLC system is needed in order to analyze samples.5A systemthat is capable of performing at the flows, pressures, controlledtemperatures, sample volumes and requirements of the stan-dard may be used.7.1.2 Analytical Column-SIELCPrimesep SB 5 m, 100 particle, 150 mm 3 2.1 mm or equivalent.7.1.3

24、Tandem Mass Spectrometer (MS/MS) SystemAMS/MS system capable of MRM analysis.6A system that iscapable of performing at the requirements in this standard maybe used.7.2 Filtration Device7.2.1 Hypodermic syringeAlock tip glass syringe capableof holding a Millex HV Syringe Driven Filter Unit PVDF 0.45m

25、 (Millipore Corporation, Catalog # SLHV033NS) or similarmay be used.7.2.1.1 A 25-mL lock tip glass syringe size is recommendedsince a 25-mL sample size is used in this test method.7.2.2 FilterMillex HV Syringe Driven Filter Unit PVDF0.45 m (Millipore Corporation, Catalog # SLHV033NS) orsimilar may b

26、e used.8. Reagents and Materials8.1 Purity of ReagentsHigh-performance liquid chroma-tography (HPLC) pesticide residue analysis and spectropho-tometry grade chemicals shall be used in all tests. Unlessindicated otherwise, it is intended that all reagents shallconform to the Committee on Analytical R

27、eagents of theAmerican Chemical Society.7Other reagent grades may beused provided they are first determined they are of sufficientlyhigh purity to permit their use without affecting the accuracy ofthe measurements.8.2 Purity of WaterUnless otherwise indicated, referencesto water shall be understood

28、to mean reagent water conformingto Type 1 of Specification D1193. It must be demonstrated thatthis water does not contain contaminants at concentrationssufficient to interfere with the analysis.8.3 GasesUltrapure nitrogen and argon.8.4 Acetonitrile (CAS # 75-05-8).8.5 Methanol (CAS # 67-56-1).8.6 Ac

29、etone (CAS # 67-64-1).8.7 Ammonium formate (CAS # 540-69-2).8.8 Formic acid (64-18-6).8.9 Thiodiglycol (CAS # 111-48-8).8.10 3,3-Thiodipropanol (CAS # 10595-09-2).9. Hazards9.1 Normal laboratory safety applies to this method. Ana-lysts should wear safety glasses, gloves, and lab coats whenworking in

30、 the lab. Analysts should review the Material SafetyData Sheets (MSDS) for all reagents used in this method.10. Sampling10.1 SamplingGrab samples must be collected in$25-mL pre-cleaned amber glass bottles with Teflon-linedcaps demonstrated to be free of interferences. This test methodrequires a 25-m

31、L sample size per analysis. Conventionalsampling practices should be followed. Refer to Guide D3856and Practices D3694.10.2 PreservationStore samples between 0C and 6Cfrom the time of collection until analysis. Analyze the samplewithin 1 day of collection.11. Preparation of LC/MS/MS11.1 LC Chromatog

32、raph Operating Conditions5:11.1.1 Injection volumes of all calibration standards andsamples are 50 L. The first sample analyzed after thecalibration curve is a blank to ensure there is no carry-over. Thegradient conditions for the liquid chromatograph are shown inTable 3.11.1.2 TemperaturesColumn, 3

33、0C; Sample compart-ment, 15C.11.1.3 Seal WashSolvent: 50 % Acetonitrile/50 % Water;Time: 5 minutes.11.1.4 Needle WashSolvent: 50 % Acetonitrile/50 % Wa-ter; Normal Wash, approximately 13 second wash time.11.1.5 Autosampler PurgeThree loop volumes.11.1.6 Specific instrument manufacturer wash/purge sp

34、eci-fications should be followed in order to eliminate samplecarry-over in the analysis of TDG.11.2 Mass Spectrometer Parameters6:11.2.1 In order to acquire the maximum number of datapoints per SRM channel while maintaining adequate sensitiv-ity, the tune parameters may be optimized according to you

35、rinstrument. Each peak requires at least 10 scans per peak foradequate quantitation. This standard contains only one targetcompound and one surrogate which are in different SRMexperiment windows in order to optimize the number of scansand sensitivity. Variable parameters regarding retention times,SR

36、M Transitions and cone and collision energies are shown inTable 4.5A Waters Alliance High Performance Liquid Chromatography (HPLC) Systemwas used to develop this test method. The multi-laboratory study included Agilentand Waters LC systems.6A Waters Quattro micro API mass spectrometer was used to de

37、velop this testmethod. The multi-laboratory study included Agilent, Applied Biosystems, Varianand Waters mass spectrometers.7Reagent Chemicals, American Chemical Society Specifications, AmericanChemical Society, Washington, DC. For Suggestions on the testing of reagents notlisted by the American Che

38、mical Society, see Annual Standards for LaboratoryChemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeiaand National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville,MD.TABLE 3 Gradient Conditions for Liquid ChromatographyTime(min)Flow(L/min)PercentCH3CNPercentW

39、aterPercent500 mmolarAmmoniumFormate/2%Formic Acid0 300 0 95 52.5 300 0 95 56 300 90 5 510 300 90 5 512 300 0 95 516 300 0 95 5D7598 0913The instrument is set in the Electrospray (+) positive setting.Capillary Voltage: 3.5 kVCone: Variable depending on analyte (Table 4)Extractor: 2 VoltsRF Lens: 0.2

40、 VoltsSource Temperature: 120CDesolvation Temperature: 300CDesolvation Gas Flow: 500 L/hrCone Gas Flow: 25 L/hrLow Mass Resolution 1: 14.5High Mass Resolution 1: 14.5Ion Energy 1: 0.5Entrance Energy: 1Collision Energy: Variable depending on analyte (Table 4)Exit Energy: 2Low Mass Resolution 2: 15Hig

41、h Mass resolution 2: 15Ion Energy 2: 0.5Multiplier: 650Gas Cell Pirani Gauge: 3.3 3 103TorrInter-Channel Delay: 0.02 secondsInter-Scan Delay: 0.1 secondsRepeats: 1Span: 0 DaltonsDwell: 0.1 Seconds12. Calibration and Standardization12.1 The mass spectrometer must be calibrated per manu-facturer speci

42、fications before analysis. In order that analyticalvalues obtained using this test method are valid and accuratewithin the confidence limits of the test method, the followingprocedures must be followed when performing the test method.12.2 Calibration and StandardizationTo calibrate the in-strument,

43、analyze seven calibration standards containing theseven concentration levels of TDG and 3,3-thiodipropanolprior to analysis as shown in Table 2. A calibration stockstandard solution is prepared from standard materials orpurchased as certified solutions. Stock standard solution A(Level 7) containing

44、TDG and 3,3-thiodipropanol is preparedat Level 7 concentration and aliquots of that solution arediluted to prepare Levels 1 through 6. The following steps willproduce standards with the concentration values shown inTable 2. The analyst is responsible for recording initialcomponent weights carefully

45、when working with pure materi-als and correctly carrying the weights through the dilutioncalculations.12.2.1 Prepare stock standard solution A (Level 7) byadding to a 100-mL volumetric flask individual methanolsolutions of the following: 250 L of TDG and 3,3-thiodipropanol each at 4 g/L, dilute to 1

46、00 mL with water. Thepreparation of the Level 7 standard can be accomplished usingdifferent volumes and concentrations of stock solutions as isaccustomed in the individual laboratory. Depending on stockconcentrations prepared, the solubility at that concentrationwill have to be ensured.12.2.2 Aliquo

47、ts of Solution A are then diluted with water toprepare the desired calibration levels in 2-mL amber glass LCvials. The calibration vials must be used within 24 hours toensure optimum results. Stock calibration standards are rou-tinely replaced every six months if not previously discarded forquality

48、control failure. Calibration standards are not filtered.12.2.3 Inject each standard and obtain a chromatogram foreach one. An external calibration is used monitoring the SRMtransition of each analyte. Calibration software is utilized toconduct the quantitation of the target analyte and surrogate.The

49、 SRM transition of each analyte is used for quantitation andconfirmation. This gives confirmation by isolating the parention, fragmenting it to the product ion, and also relating it to theretention time in the calibration standard.12.2.4 The calibration software manual should be consultedto use the software correctly. The quantitation method is set asan external calibration using the peak areas in ppb or ppm unitsas long as the analyst is consistent. Concentrations may becalculated using the data system software to generate linea