ASTM E452-2002(2013) 5625 Standard Test Method for Calibration of Refractory Metal Thermocouples Using a Radiation Thermometer《用辐射高温计校准耐火金属热电偶的标准试验方法》.pdf

上传人:confusegate185 文档编号:532781 上传时间:2018-12-06 格式:PDF 页数:14 大小:265.56KB
下载 相关 举报
ASTM E452-2002(2013) 5625 Standard Test Method for Calibration of Refractory Metal Thermocouples Using a Radiation Thermometer《用辐射高温计校准耐火金属热电偶的标准试验方法》.pdf_第1页
第1页 / 共14页
ASTM E452-2002(2013) 5625 Standard Test Method for Calibration of Refractory Metal Thermocouples Using a Radiation Thermometer《用辐射高温计校准耐火金属热电偶的标准试验方法》.pdf_第2页
第2页 / 共14页
ASTM E452-2002(2013) 5625 Standard Test Method for Calibration of Refractory Metal Thermocouples Using a Radiation Thermometer《用辐射高温计校准耐火金属热电偶的标准试验方法》.pdf_第3页
第3页 / 共14页
ASTM E452-2002(2013) 5625 Standard Test Method for Calibration of Refractory Metal Thermocouples Using a Radiation Thermometer《用辐射高温计校准耐火金属热电偶的标准试验方法》.pdf_第4页
第4页 / 共14页
ASTM E452-2002(2013) 5625 Standard Test Method for Calibration of Refractory Metal Thermocouples Using a Radiation Thermometer《用辐射高温计校准耐火金属热电偶的标准试验方法》.pdf_第5页
第5页 / 共14页
亲,该文档总共14页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、Designation: E452 02 (Reapproved 2013)Standard Test Method forCalibration of Refractory Metal Thermocouples Using aRadiation Thermometer1This standard is issued under the fixed designation E452; the number immediately following the designation indicates the year oforiginal adoption or, in the case o

2、f revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the calibration of refractorymetal thermocouples using a radiation the

3、rmometer as thestandard instrument. This test method is intended for use withtypes of thermocouples that cannot be exposed to an oxidizingatmosphere. These procedures are appropriate for thermo-couple calibrations at temperatures above 800 C (1472 F).1.2 The calibration method is applicable to the f

4、ollowingthermocouple assemblies:1.2.1 Type 1Bare-wire thermocouple assemblies in whichvacuum or an inert or reducing gas is the only electricalinsulating medium between the thermoelements.1.2.2 Type 2Assemblies in which loose fitting ceramicinsulating pieces, such as single-bore or double-bore tubes

5、, areplaced over the thermoelements.1.2.3 Type 2AAssemblies in which loose fitting ceramicinsulating pieces, such as single-bore or double-bore tubes, areplaced over the thermoelements, permanently enclosed andsealed in a loose fitting metal or ceramic tube.1.2.4 Type 3Swaged assemblies in which a r

6、efractoryinsulating powder is compressed around the thermoelementsand encased in a thin-walled tube or sheath made of a highmelting point metal or alloy.1.2.5 Type 4Thermocouple assemblies in which one ther-moelement is in the shape of a closed-end protection tube andthe other thermoelement is a sol

7、id wire or rod that is coaxiallysupported inside the closed-end tube. The space between thetwo thermoelements can be filled with an inert or reducing gas,or with ceramic insulating materials, or kept under vacuum.1.3 This standard does not purport to address all of thesafety concerns, if any, associ

8、ated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E344 Terminology Relating to Thermometry and Hydrom-etryE56

9、3 Practice for Preparation and Use of an Ice-Point Bathas a Reference TemperatureE988 Temperature-Electromotive Force (EMF) Tables forTungsten-Rhenium Thermocouples (Withdrawn 2011)3E1256 Test Methods for Radiation Thermometers (SingleWaveband Type)E1751 Guide for Temperature Electromotive Force (EM

10、F)Tables for Non-Letter Designated Thermocouple Combi-nations (Withdrawn 2009)33. Terminology3.1 Definitions:3.1.1 For definitions of terms used in this test method seeTerminology E344.3.1.2 radiation thermometer, nradiometer calibrated toindicate the temperature of a blackbody.3.1.2.1 DiscussionRad

11、iation thermometers include instru-ments having the following or similar names: (1) opticalradiation thermometer, (2) photoelectric pyrometer, ( 3) singlewavelength automatic thermometer, (4) disappearing filamentpyrometer, (5) dual wavelength pyrometer or ratio radiationthermometer, (6) visual opti

12、cal thermometer, (7) infraredthermometer, (8) infrared pyrometer, and permutations on theterms above as well as some manufacturer-specific names.3.2 Definitions of Terms Specific to This Standard:3.2.1 automatic radiation thermometer, n radiation ther-mometer whose temperature reading is determined

13、by elec-tronic means.3.2.2 disappearing filament pyrometer, n radiation ther-mometer that requires an observer to match visually thebrightness of a heated filament mounted inside the radiationthermometer to that of the measured object.1This test method is under the jurisdiction of ASTM Committee E20

14、 onTemperature Measurementand is the direct responsibility of Subcommittee E20.04on Thermocouples.Current edition approved May 1, 2013. Published May 2013. Originallyapproved in 1972. Last previous edition approved in 2007 as E452 02 (2007).DOI: 10.1520/E0452-02R13.2For referenced ASTM standards, vi

15、sit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3The last approved version of this historical standard is referenced onwww.astm.org.Copyright A

16、STM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States13.2.3 equalizing block, nobject, usually metal, that whenplaced in a nonuniform temperature region, has greater tem-perature uniformity (due to its relatively high thermoconduc-tivity and mass) tha

17、n the medium surrounding the object.3.2.4 spectral emissivity, nratio of the spectral radiance ata point on a particular specimen and in a particular directionfrom that point to that emitted by a blackbody at the sametemperature.3.2.5 spectral radiance, npower radiated by a specimen ina particular d

18、irection, per unit wavelength, per unit projectedarea of the specimen, and per unit solid angle.3.2.6 spectral response, nsignal detected by a radiometerat a particular wavelength of incident radiation, per unit powerof incident radiation.3.2.7 test thermocouple, nthermocouple that is to have itstem

19、perature-emf relationship determined by reference to atemperature standard.3.2.8 thermocouple calibration point, n temperature, es-tablished by a standard, at which the emf developed by athermocouple is determined.4. Summary of Test Method4.1 The thermocouple is calibrated by determining thetemperat

20、ure of its measuring junction with a radiation ther-mometer and recording the emf of the thermocouple at thattemperature. The measuring junction of the thermocouple isplaced in an equalizing block containing a cavity whichapproximates blackbody conditions. The radiation thermom-eter is sighted on th

21、e cavity in the equalizing block and theblackbody temperature or true temperature of the block,including the measuring junction, is determined.4.2 Since the spectral emissivity of the radiation emanatingfrom a properly designed blackbody is considered unity (one)for all practical purposes, no spectr

22、al emissivity correctionsneed be applied to optical pyrometer determinations of theblackbody temperature.4.3 Although the use of a radiation thermometer (Note 1)isless may require more effort and more complex apparatus toachieve a sensitivity equivalent to that of commonly usedthermocouples, a radia

23、tion thermometer has the advantage ofbeing physically separated from the test assembly; thus, itscalibration is not influenced by the temperatures and atmo-spheres in the test chamber. By comparison, a standardthermocouple that is used to calibrate another thermocouplemust be subjected to the temper

24、atures at which the calibrationsare performed and in some cases must be exposed to theenvironment that is common to the test thermocouple. If astandard thermocouple is exposed to high temperatures orcontaminating environments, or both, for long periods of time,its calibration becomes questionable an

25、d the uncertainty in thebias of the calibration increases.NOTE 1Disappearing filament pyrometers are somewhat less sensitivethan many of the thermocouples used above 800 C (1472 F). Theadvantages of physical separation of the disappearing filament pyrometerfrom the test assembly may still justify it

26、s use over use of a standardthermocouple.5. Significance and Use5.1 This test method is intended to be used by wireproducers and thermocouple manufacturers for certification ofrefractory metal thermocouples. It is intended to provide aconsistent method for calibration of refractory metal thermo-coup

27、les referenced to a calibrated radiation thermometer.Uncertainty in calibration and operation of the radiationthermometer, and proper construction and use of the testfurnace are of primary importance.5.2 Calibration establishes the temperature-emf relationshipfor a particular thermocouple under a sp

28、ecific temperature andchemical environment. However, during high temperaturecalibration or application at elevated temperatures in vacuum,oxidizing, reducing or contaminating environments, and de-pending on temperature distribution, local irreversible changesmay occur in the Seebeck Coefficient of o

29、ne or both thermo-elements. If the introduced inhomogeneities are significant, theemf from the thermocouple will depend on the distribution oftemperature between the measuring and reference junctions.5.3 At high temperatures, the accuracy of refractory metalthermocouples may be limited by electrical

30、 shunting errorsthrough the ceramic insulators of the thermocouple assembly.This effect may be reduced by careful choice of the insulatormaterial, but above approximately 2100 C, the electricalshunting errors may be significant even for the best insulatorsavailable.6. Sources of Error6.1 The most pr

31、evalent sources of error (Note 2) in thismethod of calibration are: (1) improper design of the black-body enclosure, (2) severe temperature gradients in the vicinityof the blackbody enclosure, ( 3) heat conduction losses alongthe thermoelements, and ( 4) improper alignment of theradiation thermomete

32、r with respect to the blackbody cavity andunaccounted transmission losses along the optical path of theradiation thermometer.NOTE 2These are exclusive of any errors that are made in theradiation thermometer measurements or the thermocouple-emf measure-ments.7. Apparatus7.1 Furnace:7.1.1 The calibrat

33、ion furnace should be designed so that anytemperature within the desired calibration temperature rangecan be maintained constant within a maximum change of 1 C(1.8 F) per minute in the equalizing block over the period ofany observation. Figs. 1-3 show three types of furnaces (1 and2)4that can be use

34、d for calibrating refractory-metal thermo-couples. Fig. 4 is a detailed drawing of the upper section of thefurnace in Fig. 3. An equalizing block containing a blackbodycavity is suspended in the central region of the furnace bymeans of support rods or wires. The mass of the support rodsor wires shou

35、ld be kept to a minimum to reduce heat losses byconduction. When the furnace is in operation, a sufficiently4The boldface numbers in parentheses refer to the list of references at the end ofthis standard.E452 02 (2013)21. Caps for making vacuum tight seals around the thermoelements. A cylinder 18. F

36、urnace shell (brass).type neoprene gasket is compressed around the thermoelements. 19. First radiation shield. 0.020-in. (0.51-mm) tantalum sheet rolled into a cylinder2. Kovar metal tube. and secured with tantalum rivets.3. Dome made of No. 7052 glass providing electrical insulation for 20. Second

37、radiation shield. (0.020-in. (0.51-mm) molybdenum.)thermoelements. 21. Third radiation solid. (0.020-in. (0.51-mm) molybdenum.)4. Neoprene O-ring gasket. 22. Fourth radiation shield. (0.010-in. (0.25-mm) molybdenum.)5. Top plate extension (brass). 23. Liquid nitrogen trap.6. Aluminum oxide radiation

38、 shield. 24. Metal baffle plates at liquid nitrogen temperature.7. Ionization vacuum gage. 25. Liquid nitrogen chamber.8. Thermocouple vacuum gage. 26. Vacuum chamber.9. No. 7052 glass tube providing electrical insulation for thermoelements. 27. Borosilicate glass window.10. Chamber for water flow d

39、uring furnace operation. 28. Hole (0.045-in. (1.14-mm) diameter) for sighting with disappearing filamentpyrometer.11. Electrically insulating spacers. 29. Molybdenum blackbody.12. Power supply terminal. 30. Tantalum tube.13. Removable top plate (brass). 31. Inert gas entrance.14. Tantalum spacing ri

40、ng providing electrical contact between plate and 32. Tantalum rings for electrical contact.tantalum tube. 33. Removable copper plate for electrical contact.15. Thermal expansion joint of tantalum tube. 34. Hex-head nut for tightening copper plate against O-ring gasket.16. Copper tubing for water co

41、oling. 35. Bottom plate (brass).17. Auxiliary radiation shield.FIG. 1 High-Temperature Furnace (Example 1)E452 02 (2013)3large region in the center of the furnace should be at a uniformtemperature to ensure that the temperature throughout theequalizing block (when all test thermocouple assemblies ar

42、e inposition in the block) is uniform. At temperatures greater than2000 C, furnace parts made from tantalum may introducecontamination of exposed thermoelements. In this case, it maybe desirable to fabricate heated furnace components fromtungsten.7.1.2 The blackbody cavity in the equalizing block sh

43、ouldbe designed in accordance with established criteria set forth inthe literature (4-8). Such factors as interior surface texture,diameter-to-depth ratio of the blackbody cavity opening, andinternal geometry can have an appreciable effect on the spectralemissivity of the cavity.7.1.3 Figs. 5-7 show

44、 three typical equalizing block designsthat are used in thermocouple calibrating furnaces. The designin Fig. 5 is used in furnaces where the standard radiationthermometer is sighted horizontally into the blackbody throughthe hole in the side of the block. This design is particularlyuseful in the cal

45、ibration of bare-wire thermocouples since thelid on the blackbody (or the entire blackbody) can be anelectrically insulating material such ashafnium oxide or beryl-lium oxide. Thus, if the bare thermocouple wires should comein contact with the equalizing block, the wires will not be(a) Nylon bushing

46、, ( b) stainless steel support, (c) rectangular stainless steel shutter, (d) borosilicate glass window, (e) brass shutter support, ( f) shutter rotationmechanism, (g) copper lead, (h) steel housing, ( i) brass plate, (j) copper coil spring, (k) alumina closed-end tube, (l) port, (m) O-ring gaskets,

47、(n) copper water-cooledelectrode, (o) tantalum heater element, ( p) tantalum radiation shields, (q) water-cooling coils, (r) ceramic insulator, ( s) tantalum radiation shield, (t) adjustable clamp,(u) water out, (v) electrical leads, (w) water in, and ( x) to vacuum system.FIG. 2 High-Temperature Fu

48、rnace (Example 2)E452 02 (2013)4electrically shorted. If this design is used in the calibration ofTypes, 2, 3, or 4 thermocouple assemblies (see 1.2), theblackbody lid can be metal since electrical insulation betweenthe thermoelements is included as part of the assembly.(WarningBeryllium oxide shoul

49、d be considered a hazardousmaterial. Material Safety Data Sheets and precautions inhandling this toxic substance should be obtained from thesupplier.)7.1.4 The designs in Figs. 6 and 7 are used in furnaceswhere the standard radiation thermometer is sighted verticallyinto the blackbody cavity. In cases where it is necessary tocalibrate a number of thermocouples during one calibration runor to calibrate thermocouple assemblies that are large indiameter and mass, the equalizing block designs in Figs. 6 and7 are appropriate. If the thermocouple assemblies b