ASTM F2504-2005(2014) Standard Practice for Describing System Output of Implantable Middle Ear Hearing Devices《可植入中耳的听觉设备的描述系统输出的标准实施规程》.pdf

上传人:吴艺期 文档编号:538864 上传时间:2018-12-07 格式:PDF 页数:6 大小:193.47KB
下载 相关 举报
ASTM F2504-2005(2014) Standard Practice for Describing System Output of Implantable Middle Ear Hearing Devices《可植入中耳的听觉设备的描述系统输出的标准实施规程》.pdf_第1页
第1页 / 共6页
ASTM F2504-2005(2014) Standard Practice for Describing System Output of Implantable Middle Ear Hearing Devices《可植入中耳的听觉设备的描述系统输出的标准实施规程》.pdf_第2页
第2页 / 共6页
ASTM F2504-2005(2014) Standard Practice for Describing System Output of Implantable Middle Ear Hearing Devices《可植入中耳的听觉设备的描述系统输出的标准实施规程》.pdf_第3页
第3页 / 共6页
ASTM F2504-2005(2014) Standard Practice for Describing System Output of Implantable Middle Ear Hearing Devices《可植入中耳的听觉设备的描述系统输出的标准实施规程》.pdf_第4页
第4页 / 共6页
ASTM F2504-2005(2014) Standard Practice for Describing System Output of Implantable Middle Ear Hearing Devices《可植入中耳的听觉设备的描述系统输出的标准实施规程》.pdf_第5页
第5页 / 共6页
亲,该文档总共6页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、Designation: F2504 05 (Reapproved 2014)Standard Practice forDescribing System Output of Implantable Middle EarHearing Devices1This standard is issued under the fixed designation F2504; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision

2、, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice defines means for describing system per-formance (ex vivo) and, in particular, system outp

3、ut of animplantable middle ear hearing device (IMEHD) by measuringa physical quantity that is relevant to the insertion gain andoutput level of the IMEHD when implanted in the patient.1.2 This practice is similar to headphone calibration on anartificial ear in which the sound pressure level (in deci

4、belsound pressure level (SPL) measured in the artificial ear canbe converted to patient hearing level (in decibel hearing level(HL) using a known transfer function, as defined byANSI 3.7.These measurements can then be used to predict systemparameters relevant for patient benefit such as functional g

5、ain,maximum output, and variability. Measurements defined in thispractice should be useful for patients, clinicians,manufacturers, investigators, and regulatory agencies in mak-ing comparative evaluations of IMEHDs.1.3 The values given in SI units are to be considered thestandard.1.4 This standard d

6、oes not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ANSI St

7、andards:2ANSI 3.6 Specification for AudiometersANSI 3.7 Method for Coupler Calibration of EarphonesANSI 3.22 Specification of Hearing Aid Characteristics3. Terminology3.1 Refer to the block diagram of Fig. 1 for a clarification ofthe mathematical notations used in this section.3.2 In the following d

8、efinitions, these symbols are used forphysical quantities:3.2.1 E = electrical drive signal (voltage or current)3.2.2 p = sound pressure3.2.3 v = vibration velocity3.3 All transfer functions are denoted by the symbol H, withthe following subscripts indicative of the type of transferfunction:3.3.1 A

9、= IMEHD-aided3.3.2 E = electrical3.3.3 H = hearing level3.3.4 S = sound field sound pressure3.3.5 T = tympanic membrane (ear drum) sound pressure3.3.6 U = unimplanted3.3.7 V = vibration of stapes3.4 Definitions:3.4.1 coupling, npoints and methods of attachment.3.4.2 displacement, nintegral of veloci

10、ty measured innanometres.3.4.3 ear-canal sound pressure, pT,nsound pressure pro-duced in the ear canal, at the tympanic membrane, by a soundfield stimulus, specified in units of pascals.3.4.4 equivalent hearing level, LH,nratio of an equivalentsound pressure, pQ, relative to the sound field pressure

11、,pRETSPL, at 0 incidence that is just detectable monaurally by anormally hearing individual, as defined in ANSI S3.6, Table 9,expressed in decibels: LH= 20log10(pQ/pRETSPL).3.4.5 equivalent sound pressure, pQ,nunimplanted inputsound field pressure needed to produce a stapes velocity equalto that pro

12、duced by a specified IMEHD input in the IMEHD-aided condition: pQ= E HES.3.4.5.1 DiscussionThe equivalent sound pressure is theproduct of the equivalent sound pressure transfer function, HES,and the IMEHD output transducer electrical input E: pQ= E HES. The equivalent sound pressure can be expressed

13、 asequivalent sound pressure level in units of decibels, SPLeq,calculated as 20log10(pQ/210-5Pa).3.4.6 equivalent sound pressure level, LQ,nlogarithmicrepresentation of equivalent sound pressure, LQ= 20log10(pQ).1This practice is under the jurisdiction ofASTM Committee F04 on Medical andSurgical Mat

14、erials and Devices and is the direct responsibility of SubcommitteeF04.37 on Implantable Hearing Devices (IHDs).Current edition approved March 1, 2014. Published April 2014. Originallyapproved in 2005. Last previous edition approved in 2005 as F2504 05. DOI:10.1520/F2504-05R14.2Available from Americ

15、an National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036, http:/www.ansi.org.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States13.4.7 hearing level (HL), L, nratio of the input sound fieldpressure, pS, relativ

16、e to the sound field pressure pRETSPLat 0incidence that is just detectable monaurally by a normallyhearing individual, as defined inANSI S3.6, Table 9, expressedin decibels as: L = 20log10(pS/pRETSPL).3.4.8 IMEHD electrical input at threshold Ethreshold,nelectrical input to the IMEHD output transduc

17、er at thresh-old of audibility.3.4.9 IMEHD harmonic distortion, nharmonic distortionof the stapes velocity IMEHD-aided analogous toANSI S3.22,Section 6.11S, from sinusoidal inputs of the frequencies 500,800, and 1600 Hz; input levels shall be Emax20dB.3.4.10 IMEHD output transducer, nelectromechanic

18、aloutput transducer of the IMEHD.3.4.11 IMEHD output transducer frequency range, nusingthe equivalent sound pressure transfer function, HES, draw ahorizontal line at the average for 1000, 1600, and 2500 Hz,then subtract 20 dB, or divide by 10; the lower and the upperbounds of the frequency response

19、range are where the averageline crosses the transfer function curve.3.4.12 IMEHD output transducer input, E, n electricalinput to the IMEHD output transducer, specified in volts oramperes, as appropriate for the particular device.3.4.13 IMEHD system frequency range, nusing the inser-tion gain transf

20、er function (velocity), HVV, draw a horizontalline at the average for 1000, 1600, and 2500 Hz, then subtract20 dB, or divide by 10; the lower and the upper bounds of thefrequency response range are where the average line crossesthe transfer function curve.3.4.14 input sound field pressure, pS,nsound

21、 stimulusmeasured in the free field and presented to the listener in eitherthe IMEHD-aided or unimplanted condition, specified in unitsof pascals.3.4.15 maximum electrical transducer input, Emax,nmaximum electrical output of the sound signal processor,specified as peak-to-peak or root mean square va

22、lue, specifiedin volts or amperes, as appropriate for the particular device.3.4.16 maximum equivalent sound pressure, pE,max,nequivalent sound pressure that corresponds to the maximumelectrical output Emaxof the implant electronics, pE,max= EmaxHES.3.4.17 maximum equivalent sound pressure level, LE,

23、max,nlogarithmic representation of the maximum equivalentsound pressure LE,max= 20log10(pE,max/210-5Pa).3.4.18 sound pressure at threshold, pthreshold,nstimulussound field pressure at the threshold of audibility.3.4.19 stapes velocity (IMEHD-aided), vA,ntranslationalvelocity of the stapes when drive

24、n by the IMEHD outputtransducer, specified in units of mm/s.3.4.20 stapes velocity (unimplanted), vU,ntranslationalvelocity of the stapes when driven by sound input to the middleear specified in units of mm/s.Transfer Function3.4.21 acousto-electric transfer function, HSE,nelectricalinput to the IME

25、HD output transducer E produced by a soundfield, divided by the input sound field pressure pS: HSE= E/pS.3.4.21.1 DiscussionHSEwill depend on the particular gainsettings used, for example, full-on gain or minimal gain. Thegain should be reported whenever that transfer function is used.3.4.22 acousto

26、-vibrational transfer function (IMEHDaided), HSVAstapes velocity (IMEHD aided) divided by theinput sound field pressure: HSVA=vA/pS.3.4.22.1 DiscussionThis quantity can be measured di-rectly or computed from the product of the electro-vibrationaltransfer function, HEV, and the acousto-electric trans

27、ferfunction, HSE, measured in the IMEHD-aided condition:HSVA=vA/pS.3.4.23 acousto-vibrational transfer function (unimplanted),HSVU,nstapes velocity (unimplanted) when driven by theinput sound field, divided by the input sound field pressure:HSVU= vU/pS.3.4.23.1 DiscussionThis quantity can be measure

28、d di-rectly or computed from the product of the middle-ear transferfunction, HTV, and the ear-canal transfer function, HST, mea-sured in the unimplanted condition: HSVU= vU/pS= HST HTV.FIG. 1 Signal Flow in the Unimplanted and IMEHD-Aided Middle EarF2504 05 (2014)23.4.24 ear-canal pressure transfer

29、function, HST,nearcanal sound pressure, pT, produced by the input sound fieldpressure, pS, in the unimplanted case, divided by that inputsound field pressure: HST= pT/pS; this quantity is unitless (1,2).33.4.25 electro-vibrational transfer function, HEV,nstapesvelocity (IMEHD-aided) when driven by t

30、he IMEHD outputtransducer, divided by the transducer input: HEV= vA/E.3.4.26 equivalent sound pressure transfer function, HES,nunimplanted sound field pressure needed to produce astapes velocity equivalent to that produced by an electricalIMEHD input in the IMEHD-aided condition, divided by theIMEHD

31、 input.3.4.26.1 DiscussionIf the electrical IMEHD input pro-duces a linear change in stapes velocity with a change in inputelectrical stimulus, the equivalent sound pressure transferfunction, HES, can be computed as the quotient between thevibro-electric transfer function (IMEHD-aided), HEV, and the

32、vibro-acoustic transfer function (unimplanted), HSVU: HES=(v/E)/(v/pS)=HEV/HSVU.3.4.27 insertion gain transfer function (sound field), HSS,nratio of the equivalent sound pressure produced in theIMEHD-aided case with a given electrical input to the IMEHDoutput transducer and the input sound field pre

33、ssure used asinput in the IMEHD-aided case required to produce the sameIMEHD output transducer electrical input: HSS= pE/pS; thisratio is unitless.3.4.27.1 DiscussionWith a linear sound signal processor,the insertion gain (sound field) can be computed from theproduct of the equivalent sound pressure

34、 transfer function, HES,and the electro-acoustic transfer function, HSE: HSS= pE/pS=HSE HES. HSSwill depend on the particular gain settings used,for example, full-on gain or minimal gain. The gain should bereported whenever that transfer function is used.3.4.28 insertion gain transfer function (velo

35、city), HVv,nratio of the stapes velocity (IMEHD-aided) and the stapesvelocity (unimplanted) produced by a given input sound field:HVV= vA/vU; the ratio is unitless and can be expressed indecibels as 20log10(HVv).3.4.28.1 DiscussionWith a linear sound signal processorand IMEHD, that is, a processor w

36、hose electrical output E isproportional to the input sound field pressure, pS, and anIMEHD whose vibrational output is proportional to its electri-cal output, the insertion gain (sound field), HSS, will equal theinsertion gain transfer function (velocity), HVV.3.4.29 maximum insertion gain transfer

37、function (soundfield), HSS,max,nmaximum insertion gain transfer function(sound field) that can be achieved with the implant electronics.3.4.30 middle-ear transfer function, HTV,nstapes velocity(unimplanted) produced by an ear-canal sound pressure, di-vided by the ear-canal sound pressure, in units o

38、f mm/s/Pa:HTV= vU/pT.3.5 Acronyms:3.5.1 IHDimplantable hearing device3.5.2 IMEHDimplantable middle-ear hearing device3.5.3 LDVlaser Doppler vibrometry3.5.4 SPLsound pressure level4. Summary of Practice4.1 This practice involves the use of human temporal bonesand laser Doppler interferometry measurem

39、ents of middle earstructures velocities, to test for the ex-vivo performances ofIMEHD. Once a procedure for measuring system output hasbeen defined, several characteristics of the IMEHD can bespecified. Detailed instructions for measuring and reportingthese characteristics are given below. The impor

40、tant character-istics are:4.1.1 For Transducers:4.1.1.1 Equivalent sound pressure transfer function,4.1.1.2 IMEHD output transducer frequency range, and4.1.1.3 IMEHD harmonic distortion.4.1.2 For the System:4.1.2.1 Maximum insertion gain transfer function (soundfield) (see full-on gain in ANSI S3.22

41、, paragraph 3.7),4.1.2.2 Maximum equivalent sound pressure level (seeOSPL90 in ANSI S3.22, paragraph 3.5), and4.1.2.3 IMEHD system frequency range.5. Significance and Use5.1 IMEHDs are alternatives to air conduction hearing aids.They are similar to air conduction hearing aids in that theyprocess inc

42、oming sound by applying frequency shaping andcompression to create an analog, vibratory audio frequencyoutput. IMEHDs differ from hearing aids in that they do notcreate an airborne acoustical output signal with an electroa-coustical output transducer in the external ear canal, but rathera mechanical

43、 stimulation that results in the vibration of thecochlear fluid. Therefore, the IMEHD output signal is notreadily accessible after implantation in the way hearing aidoutput is accessible with real-ear probe microphone measure-ments. Different devices will use different methods of couplingto the ossi

44、cular chain or cochlea. This makes it difficult todesign a uniform model of the middle ear in the way the 2-cm3coupler is used as a model of the external ear canal withconventional hearing aids.5.2 This practice provides uniformity of data collectionpractices, thus allowing IMEHD in vitro performanc

45、es to beevaluated and readily compared. Once clinical data areavailable, the performance specifications can be augmentedwith corresponding transfer functions or results from measure-ments in patients.5.3 The temporal bone is a well-accepted model that relatesclosely to the biomechanics of the living

46、 middle ear, which isreadily relatable to hearing level. Laser Doppler vibrometryprovides accurate velocity measurements in the ranges requiredfor human hearing.3The boldface numbers in parentheses refer to the list of references at the end ofthis standard.F2504 05 (2014)36. Procedure6.1 Procedure S

47、etupThe basic procedure is to define amethod for the measurement of the insertion gain and maxi-mum output of the IMEHD based on temporal bone measure-ments of sound-induced stapes velocity before implantationand measurements of IMEHD-induced stapes velocity afterimplantation.6.2 Components of the T

48、est System:6.2.1 Test ModelHuman Temporal Bone:6.2.1.1 Temporal Bone Pre-SelectionFresh or fresh-frozentemporal bones shall be used. Fresh temporal bones shall bekept moist and used within seven days of harvesting or befrozen within 24 h of harvesting for later testing (3). Thawedbones shall be used

49、 within six days. Fresh or thawed temporalbones shall be stored at 5C in isotonic saline with 1:10 000thimerosal (or antibacterial equivalent). Once removed fromstorage, all measurements, including qualification, shall bemade within one 8-h period. The middle and inner ears shall beinspected to be morphologically intact and with normal-appearing tympanic membrane and middle ear. Diseased earsor overtly abnormal structures or evidence of a conductivedefect, or combination thereof, shall disqualify bones fromtesting. Tem

展开阅读全文
相关资源
猜你喜欢
  • BS PD IEC TS 62763-2013_5284 Pilot function through a control pilot circuit using PWM (pulse width modulation) and a control pilot wire《通过控制导向线使用PWM (脉冲宽度调制) 的导向功能和控制导向线》.pdf BS PD IEC TS 62763-2013_5284 Pilot function through a control pilot circuit using PWM (pulse width modulation) and a control pilot wire《通过控制导向线使用PWM (脉冲宽度调制) 的导向功能和控制导向线》.pdf
  • BS ISO 8070-2007 Milk and milk products - Determination of calcium sodium potassium and magnesium contents - Atomic absorption spectrometric method《牛奶和奶制品 钙、钠、钾和镁含量的测定 原子吸.pdf BS ISO 8070-2007 Milk and milk products - Determination of calcium sodium potassium and magnesium contents - Atomic absorption spectrometric method《牛奶和奶制品 钙、钠、钾和镁含量的测定 原子吸.pdf
  • BS ISO 8082-1-2009 Self-propelled machinery for forestry - Laboratory tests and performance requirements for roll-over protective structures - General machines《林业用自推进机械 防倾.pdf BS ISO 8082-1-2009 Self-propelled machinery for forestry - Laboratory tests and performance requirements for roll-over protective structures - General machines《林业用自推进机械 防倾.pdf
  • BS ISO 8082-2-2011 Self-propelled machinery for forestry Laboratory tests and performance requirements for roll-over protective structures Machines having a rotating platf.pdf BS ISO 8082-2-2011 Self-propelled machinery for forestry Laboratory tests and performance requirements for roll-over protective structures Machines having a rotating platf.pdf
  • BS ISO 8083-2006 Machinery for forestry - Falling-object protective structures (FOPS) - Laboratory tests and performance requirements《林业机械 落体防护装置(FOPS) 实验室试验和性能要求》.pdf BS ISO 8083-2006 Machinery for forestry - Falling-object protective structures (FOPS) - Laboratory tests and performance requirements《林业机械 落体防护装置(FOPS) 实验室试验和性能要求》.pdf
  • BS ISO 8086-2004 Dairy plant - Hygiene conditions - General guidance on inspection and sampling procedures《乳品厂 卫生条件 检验和取样程序通用指南》.pdf BS ISO 8086-2004 Dairy plant - Hygiene conditions - General guidance on inspection and sampling procedures《乳品厂 卫生条件 检验和取样程序通用指南》.pdf
  • BS ISO 8096-2005 Rubber- or plastics-coated fabrics for water resistant clothing - Specification《雨衣用橡胶或塑料涂覆织物 规范》.pdf BS ISO 8096-2005 Rubber- or plastics-coated fabrics for water resistant clothing - Specification《雨衣用橡胶或塑料涂覆织物 规范》.pdf
  • BS ISO 8097-2001 Aircraft Minimum airworthiness requirements and test conditions for certified air cargo unit load devices《航空器 经认证的航空货运集装单元装置最低适航性要求和试验条件》.pdf BS ISO 8097-2001 Aircraft Minimum airworthiness requirements and test conditions for certified air cargo unit load devices《航空器 经认证的航空货运集装单元装置最低适航性要求和试验条件》.pdf
  • BS ISO 8114-1993 Textile machinery and accessories - Spindles for ring-spinning and doubling machines - List of equivalent terms《纺织机械和附件 环锭纺纱机和并线机用锭子 同义术语表》.pdf BS ISO 8114-1993 Textile machinery and accessories - Spindles for ring-spinning and doubling machines - List of equivalent terms《纺织机械和附件 环锭纺纱机和并线机用锭子 同义术语表》.pdf
  • 相关搜索

    当前位置:首页 > 标准规范 > 国际标准 > ASTM

    copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
    备案/许可证编号:苏ICP备17064731号-1