ASTM F2778-2009(2015) Standard Test Method for Measurement of Percent Crystallinity of Polyetheretherketone (PEEK) Polymers by Means of Specular Reflectance Fourier Transform Infra.pdf

上传人:刘芸 文档编号:539444 上传时间:2018-12-07 格式:PDF 页数:4 大小:287.47KB
下载 相关 举报
ASTM F2778-2009(2015) Standard Test Method for Measurement of Percent Crystallinity of Polyetheretherketone (PEEK) Polymers by Means of Specular Reflectance Fourier Transform Infra.pdf_第1页
第1页 / 共4页
ASTM F2778-2009(2015) Standard Test Method for Measurement of Percent Crystallinity of Polyetheretherketone (PEEK) Polymers by Means of Specular Reflectance Fourier Transform Infra.pdf_第2页
第2页 / 共4页
ASTM F2778-2009(2015) Standard Test Method for Measurement of Percent Crystallinity of Polyetheretherketone (PEEK) Polymers by Means of Specular Reflectance Fourier Transform Infra.pdf_第3页
第3页 / 共4页
ASTM F2778-2009(2015) Standard Test Method for Measurement of Percent Crystallinity of Polyetheretherketone (PEEK) Polymers by Means of Specular Reflectance Fourier Transform Infra.pdf_第4页
第4页 / 共4页
亲,该文档总共4页,全部预览完了,如果喜欢就下载吧!
资源描述

1、Designation: F2778 09 (Reapproved 2015)Standard Test Method forMeasurement of Percent Crystallinity ofPolyetheretherketone (PEEK) Polymers by Means ofSpecular Reflectance Fourier Transform InfraredSpectroscopy (R-FTIR)1This standard is issued under the fixed designation F2778; the number immediately

2、 following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test m

3、ethod describes the collection of absorptionspectra of polyetheretherketone (PEEK) polymer in filled andunfilled grades, as supplied by a vendor, and the subsequentcalculation of the percent crystallinity. The material is evalu-ated by infrared spectroscopy. The intensity (height) of theabsorbance p

4、eaks is related to the amount of crystalline regionspresent in the material.1.2 This test method can be used for PEEK consolidatedforms, such as injection molded parts, as long as the samplesare optically flat and smooth.1.3 The applicability of the infrared method to industrialand medical grade PEE

5、K materials has been demonstrated byscientific studies.2,3Percentage of crystallinity is related toR-FTIR measurement by calibration through wide-angle x-rayscattering (WAXS) crystallinity measurements.2,3It is antici-pated that this test method, involving the peak heights near1305 cm-1and 1280 cm-1

6、, will be evaluated in an Interlabora-tory Study (ILS) conducted according to Test Method E691.1.4 This test method does not suggest a desired range ofcrystallinity for specific applications.1.5 The values stated in SI units are to be regarded asstandard. No other units of measurement are included i

7、n thisstandard.1.6 This standard may involve hazardous materials,operations, and equipment. This standard does not purport toaddress all of the safety concerns, if any, associated with itsuse. It is the responsibility of the user of this standard toestablish appropriate safety and health practices a

8、nd deter-mine the applicability of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:4E691 Practice for Conducting an Interlaboratory Study toDetermine the Precision of a Test Method3. Terminology3.1 Definitions of Terms Specific to This Standard:3.1.1 crystallinity index

9、 (CI), nthe ratio of the heightbetween the absorption peaks 1305 cm-1and 1280 cm-1.4. Significance and Use4.1 Mechanical properties of PEEK, such as stiffness oryield strength, are influenced by the level of crystallinity.5Thereported crystallinity index determined by this test method hasbeen correl

10、ated with percent crystallinity in PEEK by wide-angle X-ray scattering (WAXS) experiments.2,34.2 This test method may be useful for both processdevelopment, process control, product development, and re-search.5. Interferences5.1 Samples must be smooth and optically flat over the areaof investigation

11、, typical of injection-molded specimens. Theymust be sufficiently thick (for example, 1 to 2 mm) such thatthere is no detectable back surface reflected radiation.1This test method is under the jurisdiction ofASTM Committee F04 on Medicaland Surgical Materials and Devices and is the direct responsibi

12、lity of SubcommitteeF04.15 on Material Test Methods.Current edition approved May 1, 2015. Published May 2015. Originallyapproved in 2009. Last previous edition approved in 2009 as F2778 09. DOI:10.1520/F277809R15.2Chalmers J. M., Everall, N. J., Hewitson, K., Chesters, M. A., Pearson, M.,Grady,A., K

13、uzicka, B., “FourierTransform Infrared Microscopy: SomeAdvances inTechniques for Characterisation and Structure-Property Elucidations of IndustrialMaterial,” The Analyst, Vol 23, 1998, pp. 579586.3Jaekel, D. J., Medel, F. J., Kurtz, S. M., “Validation of Crystallinity Measure-ments of Medical Grade

14、PEEK Using Specular Reflectance FTIR-microscopy,”Society of Plastics Engineers Annual Technical Conference 2009, Chicago 2009,Manuscript ID ANTEC-0248-2009.4For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of AST

15、MStandards volume information, refer to the standards Document Summary page onthe ASTM website.5Kurtz, S. M., Devine, J. N., “PEEK biomaterials in trauma, orthopedic, andspinal implants,” Biomaterials , Vol 28, No. 32, 2007, pp. 48454869.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C7

16、00, West Conshohocken, PA 19428-2959. United States15.2 Samples must be sized appropriately to be accommo-dated in the FTIR apparatus.6. Apparatus6.1 Infrared Spectrometer:6.1.1 An infrared spectrometer capable of recording a re-flection absorption spectrum over the range of 500 cm-1to1500 cm-1is ne

17、cessary.6.1.1.1 A minimum scan resolution of 4 cm-1shall be used.6.1.1.2 Aminimum aperture of 360 360 m shall be used.6.1.1.3 A minimum of 100 scans shall be collected perspectrum.6.1.2 Software capable of using the Kramers-Kronig trans-form algorithm to derive the absorbance spectra.6.2 Specimen Ho

18、lderEquipment capable of accuratelypositioning the sample under the orifice and allowing thesample to be in focus.6.3 Samples Preparation EquipmentEquipment capableof producing optically flat and optically thick samples.7. Preparation of Apparatus7.1 Prior to testing, the FTIR dewer chamber should b

19、efilled with liquid nitrogen until peak-to-peak signal in trans-mission mode is over 10.8. Calibration and Standardization8.1 Validation and calibration should be conducted weeklyby running a validation check of the infrared spectrometerbased on manufacturers instructions.8.2 A background scan shoul

20、d be conducted at the start oftesting and every 30 min thereafter for the duration of testingwhile using the same scanning settings as the test parametersdictate. The background scan should be taken from a com-pletely reflective surface.9. Procedure9.1 Check settings to ensure you are in reflectance

21、 mode,collecting at 100 scans per spectrum with a resolution of4cm-1, an aperture of 360 360 m, and the spectrometer isset to use a Kramers-Kronig transform algorithm.9.2 Place and secure the sample in the holder so that itappears level and flat.9.3 Through the microscope, focus the view on the surf

22、aceof the sample.9.4 Samples can continue to be tested as long as a newbackground is collected every 30 min.10. Calculation or Interpretation of Results10.1 Crystallinity Index (CI):10.1.1 Open a spectrum data file, and restrict the absorptionband range to 900 cm-1and 1400 cm-1.10.1.2 Perform an aut

23、omatic baseline correct of this spec-trum range.10.1.2.1 Determine a reference baseline for the heightmeasurements from the zero value absorbance points on thespectrum as shown in Fig. 1. The first point for the baseline isthe lowest absorbance value point between the spectra bands,1340 to 1375 cm-1

24、, and the second point is lowest absorbancevalue point between 1000 cm-1and 1080 cm-1.10.1.3 Calculate the heights of the peaks near1305 cm-1(HA) and 1280 cm-1(HB) as shown in Fig. 1.10.1.3.1 HA is the peak height corresponding to carbonyllinkages, while HB is primarily influenced by the diphenylFIG

25、. 1 Representative Spectrum of Unfilled, Medical Grade PEEK and Example of Crystallinity Index CalculationF2778 09 (2015)2ether groups of the PEEK molecular chain. The bands influ-enced by the diphenyl ether groups (HB) are invariant to thelevel of crystallinity, whereas the absorption peak correspo

26、nd-ing to carbonyl linkages (HA) increases with the level ofcrystallinity.610.1.4 CI shall be calculated by dividing the calculatedheight HA (peak height at 1305 cm-1) by the calculated heightHB (peak height at 1280 cm-1)(Fig. 1):CI 5 HA/HB (1)10.2 Percent Crystallinity:10.2.1 Based on published dat

27、a,3the CI is related to thepercent crystallinity.10.2.1.1 Crystallinity Correlation:%Crystallinity 5CI 2 0.7281.5493100 (2)10.2.1.2 Eq 2 was derived from CI measurements in speci-mens with known crystallinity from WAXS (Fig. 2).11. Report11.1 Report the following information:11.1.1 Description of th

28、e raw material, preparation methods,or treatment protocols. This should include any information onfiller material, annealing procedures, or processing (injectionmolding, extruding, etc.).11.1.2 Material used for background collection.11.1.3 Description of instrument and software packageused.11.1.4 T

29、he heights of the peaks near 1305 cm-1and1280cm-1, as well as the subsequent Crystallinity Index calcu-lated by Eq 1.11.1.5 The percent crystallinity calculated by Eq 2.12. Precision and Bias12.1 A round robin study will be conducted on PEEKsamples to determine a precision and bias statement.13. Key

30、words13.1 crystallinity; Fourier transmission infrared spectros-copy (FTIR); polyetheretherketone (PEEK); specularreflectance6Nguyen, H. X., and Ishida, H., “MolecularAnalysis of the Melting Behavior ofPoly(Aryl-Ether-Ether-Ketone),” Polymer, Vol 27, No. 9, 1986, pp. 14001405.F2778 09 (2015)3ASTM In

31、ternational takes no position respecting the validity of any patent rights asserted in connection with any item mentionedin this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the riskof infringement of such rights, are entire

32、ly their own responsibility.This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years andif not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standardsand s

33、hould be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of theresponsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you shouldmake your views known to the ASTM Committee on S

34、tandards, at the address shown below.This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the aboveaddress or at 6

35、10-832-9585 (phone), 610-832-9555 (fax), or serviceastm.org (e-mail); or through the ASTM website(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http:/ 2 FTIR and WAXS Calibration Curves Defining the Relationship Between CI and Percent Crystallinity for Filled and Unfilled Gradesof PEEK (Adapted from Jaekel3)F2778 09 (2015)4

展开阅读全文
相关资源
猜你喜欢
相关搜索

当前位置:首页 > 标准规范 > 国际标准 > ASTM

copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
备案/许可证编号:苏ICP备17064731号-1