1、A-A-548b9 W 9999974 Oit44132 b2T W - .e A-A-54 869 17 May 1993 COMMERCIAL ITEM DESCRIPTION MICROSCOPE, SURGICAL Binocular, Multi-Disciplinary The General Services Administration has authorized the use of this Commercial Item Description. This Commercial Item Description covers a multi-disciplinary s

2、urgical microscope with interchangeable objectives. Salient Characteristics: The surgical microscope shall be a fiber optic illuminated unit with interchangeable objectives for use by the Ophthalmic surgeon, ENT surgeon, Neurosurgeon, Oral surgeon and Vascular surgeon. The microscope body shall cons

3、ist of a three step magnification changer providing settings of 0.6/1.0/1.5-1.6. The microscope shall have straight binocular tubes with a focal length of 160 mm and a 0-180 degree inclined binocular tube with a focal length of 160 mm. The surgical microscope shall have an adjustable interpupillary

4、adjustment range of at least 55-75 mm. The eyepiece shall be wide high eye-point with a diopter adjustment range of plus 8 to minus 4 diopters. easily changeable rubber cups shall be provided. magnification factor of the eye pieces shall be lox. High and low The The system shall be provided with fou

5、r interchangeable objective lenses with focal lengths of 175, 250, 300 and 400 mrn. The sysfem shall provide for fine focusing with a AMSC N/A . FSC 6515 DISTRIBUTION STATEMENT A: distribution is unlimited, Approved for public release; Provided by IHSNot for ResaleNo reproduction or networking permi

6、tted without license from IHS-,-,-A-A-54869 W 9999974 0144133 5bb A-A-5 4 86 9 travel range of at least 10 mm. layer broad band, anti-reflective coatings. Capability for a field of view of at least 40 mm with a 175 nun focal length objective and 160 mm focal length binocular tube shall be provided.

7、Sterilizable and removable maneuvering handles shall be provided. Lenaes shall have multi- The illumination shall be provided by a coaxial fiber optic illumination system with exchangeable fiber optic bundle. Light shall be provided by halogen reflector or projector lamps. Either a built-in spare la

8、mp or a spare plug-in module with bulb shall be provided. Light cohtrol shall be by two position brightness control as well as by a dimmer switch. The structural support system shall include a floor stand with lockable casters. The height of the stand shall be within a range of 135 to 150 cm. The sy

9、stem shall be provided with spring balanced double suspension arms providing an extension of at least 100 cm (plus or minus 10 cm) horizontal and vertical range of at least 67 cm. Power supply shall be usable with 110/220V/50-60 Hz. WoxkmaashiP. The surgical microscope shall be free from defects whi

10、ch detract from its appearance or impair its serviceability. Unit. Each (EA). One surgioal microscope, as specified, constitutes one unit, Qua1it.y Aseurance Provisions. Unless otherwise specified in the conttaat ar piirahase order, the aentractar is respaneible or the erfomande of all inspecttion r

11、equiremento as specified hare P n. purchaee order, the contraator may ua his own or any facilities suitable far the performance of the intapsation requirement apecifiad herein, unless i#approve$ by the Crlveiinmwnt* The Government fesretves the right ta perfam any ef the inepectione set farth in the

12、 speaification where )wh inegeetisna necessary to a$iife dupplieu and servides aonfom te pmscribod raquirsmarnte. ReBaaniibil itu for inamsction. $wept PCB otherwise specified in the aoritract ar deemed 4. d I Provided by IHSNot for ResaleNo reproduction or networking permitted without license from

13、IHS-,-,-A-A-54867 = 9999974 0144134 4T2 A-A- 5 4 8 6 9 Records. Records of examinations and tests performed by or for the contractor shall be maintained by the contractor and made available to the Government upon the Governments request, at any time, or from time to time, during the performance of t

14、he contract and for a period of three years after delivery of the supplies to which such records relate. Inspection. Inspection, as used herein, is defined as both examination (such as visual or auditory investigation without the use of special laboratory appliances or procedures) and testing (deter

15、mination by technical means of physical and chemical properties) of the item. Samplins for examination. Sampling for examination shall be conducted in accordance with MIL-STD-105, with an AQL of 2.5 (percent defective) and an inspection level of S-2. The unit of product for examination purposes shal

16、l be one microscope. Sampling for tests. Sampling for tests, including dimensional tests, shall be conducted in accordance with MIL-STD- 105, with an AQL of 2.5 (percent defective) and an inspection level of S-2. The unit of product for test purposes shall be one microscope. Tests. Tests shall be co

17、nducted to determine compliance with specification requirements . Where feasible, the same sample shall be used for the determination of two or more test characteristics. Tests shall include, but shall not be limited to, the following: Performance testins. The surgical microscope shall be completely

18、 assembled in its operating configuration and tested to determine its capability for proper operation. Metric products. Products manufactured to metric dimensions will be considered on an equal basis with those manufactured using inch-pound units, providing they fall within the tolerances specified

19、using conversion tables contained in the latest revision of Federal Standard 376, and all other requirements of this document are met. If a product is manufactured to metric dimensions and those dimensions exceed the tolerances specified in the inch/pound units, a request should be made to the contr

20、acting officer to determine if the product is acceptable. The contracting officer has the option of accepting or rejecting the product. 3 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-5 48 b9 9999974 OL44135 339 A-A-54 8 69 Contractor certifica

21、tion. The contractor shall certify and maintain substantiating evidence that the product offered meets the salient characteristics of this Commercial Item Description, and that the product conforms to the producers own drawings, specifications, standards, and quality assurance practices. The governm

22、ent reserves the right to require proof of such conformance prior to first delivery and thereafter as may be otherwise provided for under the provisions of the contract. Requlatorv reauirements. *- Federal Food, Drus and Cosmetic Act. If the product covered by this document has been determined by th

23、e U.S. Food and Drug Administration to be under its jurisdiction, the offeror/contractor shall comply, and be responsible for compliance by its subcontractors/suppliers, with the requirements of the Federal Food, Drug and Cosmetic Act, as amended, and regulations promulgated thereunder. In addition,

24、 the offeror/contractor shall comply, and be responsible for compliance by its subcontractors/suppliers, with the requirements of all other applicable Federal, State, and local statutes, ordinances, and regulations. Recovered materials. The offeror/contractor is encouraged to use recovered material

25、in accordance with Federal Acquisition Regulation Subpart 23.4 to the maximum extent practical. Preservation, Dackaqinq, P ackinq, labelina, and markina. Unless otherwise specified, preservation, packaging, and packing shall be to a degree of protection to preclude damage to containers and/or conten

26、ts thereof under normal shipping conditions, handling, etc:, involving shipment from the supply source to the receiving activity, plus reshipment from receiving activity, and shall conform to applicable carriers rules and regulations. Intermediate and exterior package quantities and labeling and mar

27、king shall be as specified in the contract and/or order. labeling, and marking must be specified in the contract and/or order . Orderina data. Intermediate/exterior package quantities, NOTE: The following National Stock Number is covered by this document. 6515-01-335-9383 4 Provided by IHSNot for Re

28、saleNo reproduction or networking permitted without license from IHS-,-,-A-A-548b9 = 9999974 Ol14413b 275 D MILITARY INTERESTS: A-A-54 8 6 9 PREPARING ACTIVITY: Custodians: DoD-MB Air Force - 03 Army - MD Navy - MS Agent: DLA-DM CIVIL AGENCY COORDINATING ACTIVITIES: FDA-MPQAS USPHS VA - OSS Project Number: 6515-5441 Location: MICR2C14.DOC 5 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-

展开阅读全文
相关资源
猜你喜欢
相关搜索

当前位置:首页 > 标准规范 > 国际标准 > 其他

copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
备案/许可证编号:苏ICP备17064731号-1