UOP 953-2013 Sulfate and Thiosulfate in Caustic Aqueous Solutions by Ion Chromatography.pdf

上传人:syndromehi216 文档编号:1074583 上传时间:2019-04-04 格式:PDF 页数:7 大小:166.73KB
下载 相关 举报
UOP 953-2013 Sulfate and Thiosulfate in Caustic Aqueous Solutions by Ion Chromatography.pdf_第1页
第1页 / 共7页
UOP 953-2013 Sulfate and Thiosulfate in Caustic Aqueous Solutions by Ion Chromatography.pdf_第2页
第2页 / 共7页
UOP 953-2013 Sulfate and Thiosulfate in Caustic Aqueous Solutions by Ion Chromatography.pdf_第3页
第3页 / 共7页
UOP 953-2013 Sulfate and Thiosulfate in Caustic Aqueous Solutions by Ion Chromatography.pdf_第4页
第4页 / 共7页
UOP 953-2013 Sulfate and Thiosulfate in Caustic Aqueous Solutions by Ion Chromatography.pdf_第5页
第5页 / 共7页
点击查看更多>>
资源描述

1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB

2、ED HEREIN CAN BE HAZARDOUS. SAFETY DATA SHEETS (SDS) OR EXPERIMENTAL SAFETY DATA SHEETS (ESDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 1997, 2013 UOP LLC. All rights reserved. Nonconfidential U

3、OP Methods are available from ASTM International, 100 Barr Harbor Drive, P.O. Box C700, West Conshohocken, PA 19428-2959, USA. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or 610.832.9585 PHONE. Sulfat

4、e and Thiosulfate in Caustic Aqueous Solutions by Ion Chromatography UOP Method 953-13 Scope This method is for determining levels of sulfate and thiosulfate in aqueous, spent caustic, process streams by ion chromatography (IC). The range of quantitation for sulfate and thiosulfate is 0.004 to 5 mas

5、s-%. Outline of Method For analysis, the sample is diluted by mass in water to an analyte concentration between 0.0004 and 0.0180 mass-%. A repeatable volume of diluted sample is injected into an ion chromatograph equipped with an anion exchange column, conductivity suppressor device and conductivit

6、y detector, see Note 1. The method is calibrated by an external standard procedure using calibration standards prepared from pure sodium sulfate and sodium thiosulfate. Apparatus References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used

7、. Balance, readable to 0.0001 g Chromatographic column, Dionex IonPac AS11-HC Analytical, 250-mm length by 4-mm ID, Thermo Scientific, Cat. No. 052960 Deionized water system, NANOpure II Water Purification System with Total Organic Carbon Analyzer, Thermo Scientific, VWR, Cat. No. 47729-610 Ion chro

8、matograph, Dionex ICS-2100, equipped with a pump, injection valve/autosampler, eluent generator, continuously regenerated trap column, suppressor, conductivity detector, solvent bottles, computer, and Chromeleon software, Thermo Scientific Pipet bulb, 1-mL, Fisher Scientific, Cat. No. 13-678-9A Refr

9、igerator, laboratory, explosion proof or flammable storage, Fisher Scientific, Cat. No. 97-950 Vortex mixer, digital, 120V, 50/60Hz, 75W, VWR, Cat. No. 14005-824 2 of 7 953-13 Reagents and Materials References to catalog numbers and suppliers are included as a convenience to the method user. Other s

10、uppliers may be used. References to water mean deionized and distilled water that is subsequently treated with a Thermo Scientific NANOpure II water purification system to produce ionically pure, 18.2 megohm-cm, organic-free (1 ppm) water. Pasteur pipet, disposable, borosilicate glass, 146-mm length

11、, Fisher Scientific, Cat. No. 13-678-20A Potassium hydroxide cartridge, RFIC EluGen, EGC II KOH, Dionex, Thermo Fisher, Cat. No. 058900 Sodium sulfate, 99.99+% pure, Sigma-Aldrich, Cat. No. 20,444-7 Sodium thiosulfate, 99% pure, Sigma-Aldrich cal, Cat. No. 21,726-3 Vials, autosampler, 2-mL, polyprop

12、ylene, for use with Dionex Thermo Scientific AS Autosampler, Fisher Scientific, Cat. No. 10-800-167 Vial caps, for autosampler vials, blue, with pre-slit PTFE/silicone septum, Fisher Scientific, Cat. No. 03-376-482 Vials, sample, 40-mL, clear VOA vials with closed cap, borosilicate glass, with white

13、 PP cap with PTFE-faced silicone liner, Fisher Scientific, Cat. No. C336-0040 Water, deionized and distilled, subsequently treated with a Thermo Scientific NANOpure II Water Purification System with Total Organic Carbon Analyzer to produce ionically pure, 18.2 megaohm-cm, organic-free (1 ppm) water

14、Procedure The analyst is expected to be familiar with general laboratory practices, the technique of ion chromatography, and the equipment being used. Dispose of used reagents, materials, and samples in an environmentally safe manner according to local regulations. Preparation of Instrument 1. Assem

15、ble and test all ion chromatographic system components per the manufacturers specifications. 2. Install the analytical column and suppressor (configured in recycle mode). 3. Turn on the IC and establish the recommended chromatographic conditions listed in Table 1. Table 1 Recommended Chromatographic

16、 Conditions Analytical column Dionex IonPac AS11-HC Suppressor current 81 mA Mobile phase 27mM KOH Mobile phase flow rate 1.2 mL/min Column temperature 35C Detector cell temperature 35C Background conductivity 1S Analysis time 18 minutes Injection loop 25L 3 of 7 953-13 4. Allow the IC to equilibrat

17、e. Typical background conductivity should be at or below 1S. Change the source of deionized water or check the functionality of the suppressor if the background conductivity is above 1S. Operating Conditions Table 1 summarizes the recommended chromatographic operating conditions for the convenience

18、of the method user. Other conditions may be used; provided they produce the required sensitivity and chromatographic separation equivalent to that shown in the Typical Chromatogram, see Figure. Preparation of Calibration Standards Calibration factors are required for both sulfate and thiosulfate tha

19、t relate the varying instrument response for each analyte, see Note 2. Calibrate with freshly prepared calibration standards each day samples are to be analyzed. 1. In one 40-mL sample vial, weigh approximately 0.2 g each of sodium sulfate and sodium thiosulfate. Record the weight of each compound t

20、o the nearest 0.0001 g. 2. Dissolve the salts with approximately 39.6 g of water and record the total weight to the nearest 0.0001 g. Cap the solution and vortex until the entire solid is completely dissolved. Label the mixture: Standard Stock Solution. Calculate the analyte concentrations in this s

21、olution using Equation 1 in Calculations. 3. Using a Pasteur pipet, transfer approximately 0.2 g of the Standard Stock Solution to a separate 40-mL vial. Weigh and record the weight of the Standard Stock Solution aliquot to the nearest 0.0001 g. 4. Dilute the aliquot with approximately 39.8 g of wat

22、er and record the total weight to the nearest 0.0001 g. Label the mixture: Standard Calibration Solution 1. Cap the vial and vortex for approximately 15 seconds. Calculate the analyte concentrations using Equation 2 in Calculations. Standard Calibration Solution 1 should contain approximately 0.0017

23、 mass-% sulfate and 0.0018 mass-% thiosulfate. 5. Using a Pasteur pipet, transfer approximately 0.2 g of the Standard Stock Solution to a separate 40-mL vial. Weigh and record the weight of the Standard Stock Solution aliquot to the nearest 0.0001 g. 6. Dilute the aliquot with approximately 19.8 g o

24、f water and record the total weight to the nearest 0.0001 g. Label the mixture: Standard Calibration Solution 2. Cap the vial and vortex for approximately 15 seconds. Calculate the analyte concentrations using Equation 2 in Calculations. Standard Calibration Solution 2 should contain approximately 0

25、.0034 mass-% sulfate and 0.0036 mass-% thiosulfate. 7. Using a Pasteur pipet, transfer approximately 0.5 g of the Standard Stock Solution to a separate 40-mL vial. Weigh and record the weight of the Standard Stock Solution aliquot to the nearest 0.0001 g. 8. Dilute the aliquot with approximately 19.

26、5 g of water and record the total weight to the nearest 0.0001 g. Label the mixture: Standard Calibration Solution 3. Cap the vial and vortex for approximately 15 seconds. Calculate the analyte concentrations using Equation 2 in Calculations. Standard Calibration Solution 3 should contain approximat

27、ely 0.0085 mass-% sulfate and 0.0090 mass-% thiosulfate. 4 of 7 953-13 9. Using a Pasteur pipet, transfer approximately 1.0 g of the Standard Stock Solution to a separate 40-mL vial. Weigh and record the weight of the Standard Stock Solution aliquot to the nearest 0.0001 g. 10. Dilute the aliquot wi

28、th approximately 19.0 g of water and record the total weight to the nearest 0.0001 g. Label the mixture: Standard Calibration Solution 4. Cap the vial and vortex for approximately 15 seconds. Standard Calibration Solution 4 should contain approximately 0.0170 mass-% sulfate and 0.0180 mass-% thiosul

29、fate. Calculate the actual concentrations using Equation 2 in Calculations. 11. Transfer each standard calibration solution to separate autosampler vials and place them in the autosampler. Set the software to inject 50 L and initiate the sequence. The chromatographic analysis should run for 20 minut

30、es. 12. Identify and integrate the peak areas of sulfate and thiosulfate by comparing the chromatogram obtained with the Typical Chromatogram, see Figure. Sample Analysis 1. Weigh approximately 1 g of the sample into a 40-mL vial and record the weight to the nearest 0.0001 g. 2. Add approximately 9

31、g of water to the solution and record the total diluted sample weight to the nearest 0.0001 g. Cap the vial and vortex the diluted sample solution for approximately 15 seconds so that the sample is thoroughly mixed. Label the finished prepared solution to identify the sample. Repeat step 1 and 2 if

32、analyzing multiple samples 3. Transfer aliquot(s) into autosampler vial(s) and place in the autosampler. Set the software to inject 50 L, enter the dilution factors, and initiate the sequence. The chromatographic analysis should run for 20 minutes per sample 4. Identify and integrate the peaks by co

33、mparing the chromatogram obtained with the Typical Chromatogram (see Figure). 5. If the values obtained for either of the analytes are above 0.0180 mass-%, prepare a quantitative serial dilution(s) of the diluted sample solution. This is done by diluting the diluted sample solution in water on a mas

34、s/mass basis until the resulting chromatographic analyte peak areas closely match the area between Standard Calibration Solution 1 and 4. Record the weights of each serial dilution. Calculations All calculations are performed by the instrument software, and results are displayed and printed in mass-

35、%. It is recommended to include the origin in the calibration and have an r2 value of at least 0.999. The calculations below are included for information, and for manual calculation, if desired. Calculate the concentration of sulfate and thiosulfate in the Standard Stock Solution to three significan

36、t figures using Equation 1. D100XCBA = (1) where: A = concentration of analyte (sulfate or thiosulfate) in Standard Stock Solution, mass-% B = mass of analyte in pure solid, g; if the purity of the C = percent sulfate (67.62) or thiosulfate (70.92) in pure solid, mass-% 5 of 7 953-13 D = total mass

37、of Standard Stock Solution, g X = purity of the pure solid, from the Certificate of Analysis, mass-% 100 = factor to convert purity from percent Calculate the concentration of each analyte in each Standard Calibration Solution to three significant figures using Equation 2. GAFE = (2) where: A = as d

38、efined in Equation 1 E = concentration of analyte in Standard Calibration Solution, mass-% F = mass of aliquot of Standard Stock Solution, g G = total mass of Standard Calibration Solution, g Calculate the response factor for each analyte in each Standard Calibration Solution to three significant fi

39、gures using Equation 3. HEM = (3) where: E = as defined in Equation 2 H = analyte area response M = response factor for each analyte Calculate each analyte concentration in the original sample to two significant figures, but not exceeding three decimal places, using Equation 4. = PQLSMNR (4) where:

40、L = mass of sample weighed (Sample Analysis, Step 1), g M = response factor (Equation 3) for each analyte N = total IC Sample Stock Solution mass, g P = mass of IC Sample Stock Solution aliquot (Sample Analysis, Step 5), g Q = final diluted sample mass (Sample Analysis, Step 5), g R = concentration

41、of each analyte in sample, mass-% S = analyte peak area of injected sample Note: Q/P = 1 if no further dilution is required. Notes 1. The analytes are separated by anion exchange with hydroxide ions supplied by the chromatographic system mobile phase. The column effluent enters a conductivity suppre

42、ssor device, where the background conductivity of the column effluent is reduced and the analytes are simultaneously converted into a more conductive form. This is accomplished by exchange of potassium ions in the column effluent with protons supplied by electrolysis of the mobile phase. The analyte

43、 effluent from the conductivity suppressor device then flows to a conductivity detector where solution conductance is continuously measured. The solution conductance at any given time is proportional to ionic concentration and the inherent mobility of each type of ion present in the detector flow ce

44、ll. 6 of 7 953-13 2. Detector response is measured in S units of conductivity. It is proportional to the ionic concentration and the inherent mobility of each analyte ion type present when measured by the conductivity detector. Precision Precision statements were determined using UOP Method 999, “Pr

45、ecision Statements in UOP Methods.” The following precision data were obtained from analysis of synthetic blends prepared in spent caustic as process samples were unavailable. Repeatability and Site Precision A nested design was carried out for determining sulfate and thiosulfate in caustic aqueous

46、solutions. The sample was analyzed by two analysts, with each analyst performing analyses on two separate days, performing two analyses each day for a total of eight analyses. Using a stepwise analysis of variance procedure, the within-day and within-lab estimated standard deviations (esd) were calc

47、ulated at the concentration means listed in Table 2. Two analyses performed in one laboratory by the same analyst on the same day should not differ by more than the repeatability allowable differences shown in Table 2 with 95% confidence. Two analyses performed in one laboratory by different analyst

48、s on different days should not differ by more than the site precision allowable differences shown in Table 2 with 95% confidence. Table 2 Repeatability and, Site Precision, mass-% Repeatability Site Precision Component Mean Within- Day esd Allowable Difference Within- Lab esd Allowable Difference Su

49、lfate 0.005 0.0001 0.001 0.0001 0.001 Sulfate 0.011 0.0001 0.001 0.0002 0.001 Thiosulfate 0.005 0.0001 0.001 0.0001 0.001 Thiosulfate 0.015 0.0001 0.001 0.0001 0.001 The data in Table 2 is a short-term estimate of the repeatability. When the test is run routinely in the field, a control standard and chart should be used to develop a better estimate of the long-term repeatability. Reproducibility There is insufficient data to calculate the reproducibility of the test at this time. Time for Analysis The elapsed time for the preparation and analysis of o

展开阅读全文
相关资源
猜你喜欢
  • DIN EN ISO 6145-9-2011 Gas analysis - Preparation of calibration gas mixtures using dynamic volumetric methods - Part 9 Saturation method (ISO 6145-9 2009) German version EN ISO 61.pdf DIN EN ISO 6145-9-2011 Gas analysis - Preparation of calibration gas mixtures using dynamic volumetric methods - Part 9 Saturation method (ISO 6145-9 2009) German version EN ISO 61.pdf
  • DIN EN ISO 6149-1-2007 Connections for hydraulic fluid power and general use - Ports and stud ends with ISO 261 metric threads and O-ring sealing - Part 1 Ports with truncated hous.pdf DIN EN ISO 6149-1-2007 Connections for hydraulic fluid power and general use - Ports and stud ends with ISO 261 metric threads and O-ring sealing - Part 1 Ports with truncated hous.pdf
  • DIN EN ISO 6149-2-2007 Connections for hydraulic fluid power and general use - Ports and stud ends with ISO 261 metric threads and O-ring sealing - Part 2 Dimensions design test me.pdf DIN EN ISO 6149-2-2007 Connections for hydraulic fluid power and general use - Ports and stud ends with ISO 261 metric threads and O-ring sealing - Part 2 Dimensions design test me.pdf
  • DIN EN ISO 6149-3-2007 Connections for hydraulic fluid power and general use - Ports and stud ends with ISO 261 metric threads and O-ring sealing - Part 3 Dimensions design test me.pdf DIN EN ISO 6149-3-2007 Connections for hydraulic fluid power and general use - Ports and stud ends with ISO 261 metric threads and O-ring sealing - Part 3 Dimensions design test me.pdf
  • DIN EN ISO 6157-2-2004 Fasteners - Surface discontinuities - Part 2 Nuts (ISO 6157-2 1995) German version EN ISO 6157-2 2004《紧固件 表面缺陷 第2部分 螺母》.pdf DIN EN ISO 6157-2-2004 Fasteners - Surface discontinuities - Part 2 Nuts (ISO 6157-2 1995) German version EN ISO 6157-2 2004《紧固件 表面缺陷 第2部分 螺母》.pdf
  • DIN EN ISO 6158-2011 Metallic and other inorganic coatings - Electrodeposited coatings of chromium for engineering purposes (ISO 6158 2011) German version EN ISO 6158 2011《金属和其他无机涂.pdf DIN EN ISO 6158-2011 Metallic and other inorganic coatings - Electrodeposited coatings of chromium for engineering purposes (ISO 6158 2011) German version EN ISO 6158 2011《金属和其他无机涂.pdf
  • DIN EN ISO 6165-2013 Earth-moving machinery - Basic types - Identification and terms and definitions (ISO 6165 2012) German version EN ISO 6165 2012《土方机械 基本类型 鉴定、术语和定义(ISO 6165-201.pdf DIN EN ISO 6165-2013 Earth-moving machinery - Basic types - Identification and terms and definitions (ISO 6165 2012) German version EN ISO 6165 2012《土方机械 基本类型 鉴定、术语和定义(ISO 6165-201.pdf
  • DIN EN ISO 6185-1-2002 Inflatable boats - Part 1 Boats with a maximum motor power rating of 4 5 kW (ISO 6185-1 2001) German version EN ISO 6185-1 2001《充气艇 第1部分 发动机最大额定功率为4 5kW的动力艇 .pdf DIN EN ISO 6185-1-2002 Inflatable boats - Part 1 Boats with a maximum motor power rating of 4 5 kW (ISO 6185-1 2001) German version EN ISO 6185-1 2001《充气艇 第1部分 发动机最大额定功率为4 5kW的动力艇 .pdf
  • DIN EN ISO 6185-2-2002 Inflatable boats - Part 2 Boats with a maximum motor power rating of 4 5 kW to 15 kW inclusive (ISO 6185-2 2001) German version EN ISO 6185-2 2001《充气艇 第2部分 发.pdf DIN EN ISO 6185-2-2002 Inflatable boats - Part 2 Boats with a maximum motor power rating of 4 5 kW to 15 kW inclusive (ISO 6185-2 2001) German version EN ISO 6185-2 2001《充气艇 第2部分 发.pdf
  • 相关搜索

    当前位置:首页 > 标准规范 > 国际标准 > 其他

    copyright@ 2008-2019 麦多课文库(www.mydoc123.com)网站版权所有
    备案/许可证编号:苏ICP备17064731号-1