ASTM B568-1998(2014) Standard Test Method for Measurement of Coating Thickness by X-Ray Spectrometry《用X射线光度法测量镀层厚度的标准试验方法》.pdf

上传人:brainfellow396 文档编号:461592 上传时间:2018-11-25 格式:PDF 页数:8 大小:124.90KB
下载 相关 举报
ASTM B568-1998(2014) Standard Test Method for Measurement of Coating Thickness by X-Ray Spectrometry《用X射线光度法测量镀层厚度的标准试验方法》.pdf_第1页
第1页 / 共8页
ASTM B568-1998(2014) Standard Test Method for Measurement of Coating Thickness by X-Ray Spectrometry《用X射线光度法测量镀层厚度的标准试验方法》.pdf_第2页
第2页 / 共8页
ASTM B568-1998(2014) Standard Test Method for Measurement of Coating Thickness by X-Ray Spectrometry《用X射线光度法测量镀层厚度的标准试验方法》.pdf_第3页
第3页 / 共8页
ASTM B568-1998(2014) Standard Test Method for Measurement of Coating Thickness by X-Ray Spectrometry《用X射线光度法测量镀层厚度的标准试验方法》.pdf_第4页
第4页 / 共8页
ASTM B568-1998(2014) Standard Test Method for Measurement of Coating Thickness by X-Ray Spectrometry《用X射线光度法测量镀层厚度的标准试验方法》.pdf_第5页
第5页 / 共8页
亲,该文档总共8页,到这儿已超出免费预览范围,如果喜欢就下载吧!
资源描述

1、Designation: B568 98 (Reapproved 2014)Standard Test Method forMeasurement of Coating Thickness by X-Ray Spectrometry1This standard is issued under the fixed designation B568; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year

2、 of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the U.S. Department of Defense.1. Scope1.1 This test method covers the

3、use of X-ray spectrometry todetermine thickness of metallic and some nonmetallic coatings.1.2 The maximum measurable thickness for a given coatingis that thickness beyond which the intensity of the character-istic secondary X radiation from the coating or the substrate isno longer sensitive to small

4、 changes in thickness.1.3 This test method measures the mass of coating per unitarea, which can also be expressed in units of linear thicknessprovided that the density of the coating is known.1.4 Problems of personnel protection against radiation gen-erated in an X-ray tube or emanating from a radio

5、isotopesource are not covered by this test method. For information onthis important aspect, reference should be made to currentdocuments of the National Committee on Radiation Protectionand Measurement, Federal Register, Nuclear RegulatoryCommission, National Institute of Standards and Technology(fo

6、rmerly the National Bureau of Standards), and to state andlocal codes if such exist.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and

7、 determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E135 Terminology Relating to Analytical Chemistry forMetals, Ores, and Related Materials2.2 International Standard:ISO 3497 Metallic CoatingsMeasurement of CoatingThicknessX-ray Spectromet

8、ric Methods3. Terminology3.1 Definitions of technical terms used in this test methodmay be found in Terminology E135.4. Summary of Test Method4.1 ExcitationThe measurement of the thickness of coat-ings by X-ray spectrometric methods is based on the combinedinteraction of the coating and substrate wi

9、th incident radiationof sufficient energy to cause the emission of secondary radia-tions characteristic of the elements composing the coating andsubstrate. The exciting radiation may be generated by an X-raytube or by certain radioisotopes.4.1.1 Excitation by an X-Ray TubeSuitable exciting radia-tio

10、n will be produced by an X-ray tube if sufficient potential isapplied to the tube. This is on the order of 35 to 50 kV for mostthickness-measurement applications. The chief advantage ofX-ray tube excitation is the high intensity provided.4.1.2 Excitation by a Radioisotope Of the many availableradioi

11、sotopes, only a few emit gamma radiations in the energyrange suitable for coating-thickness measurement. Ideally, theexciting radiation is slightly more energetic (shorter in wave-length) than the desired characteristic X rays. The advantagesof radioisotope excitation include more compact instrument

12、a-tion essentially monochromatic radiation, and very low back-ground intensity. The major disadvantage of radioisotopeexcitation is the much lower intensities available as comparedwith X-ray tube sources. X-ray tubes typically have intensitiesthat are several orders of magnitude greater than radiois

13、otopesources. Due to the low intensity of radioisotopes, they areunsuitable for measurements on small areas (less than 0.3 mmin diameter). Other disadvantages include the limited numberof suitable radioisotopes, their rather short useful lifetimes, andthe personnel protection problems associated wit

14、h high-intensity radioactive sources.4.2 DispersionThe secondary radiation resulting from theexposure of an electroplated surface to X radiation usuallycontains many components in addition to those characteristicof the coating metal(s) and the substrate. It is necessary,therefore, to have a means of

15、 separating the desired compo-nents so that their intensities can be measured. This can bedone either by diffraction (wavelength dispersion) or byelectronic discrimination (energy dispersion).1This test method is under the jurisdiction ofASTM Committee B08 on Metallicand Inorganic Coatings and is th

16、e direct responsibility of Subcommittee B08.10 onTest Methods.Current edition approved May 1, 2014. Published May 2014. Originallyapproved in 1972. Last previous edition approved in 2009 as B568 98(2009). DOI:10.1520/B0568-98R14.2For referenced ASTM standards, visit the ASTM website, www.astm.org, o

17、rcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States14.2.1 Wavelength D

18、ispersionBy means of a single-crystalspectrogoniometer, wavelengths characteristic of either thecoating or the substrate may be selected for measurement.Published data in tabular form are available that relate spec-trogoniometer settings to the characteristic emissions of ele-ments for each of the c

19、ommonly used analyzing crystals.4.2.2 Energy DispersionX-ray quanta are usually speci-fied in terms of their wavelengths, in angstroms (), or theirequivalent energies in kiloelectron volts (keV). The relation-ship between these units is as follows:keV!A! 5 12.396where:keV = the quantum energy in tho

20、usands of electron volts, and = the equivalent wavelength in angstroms (10-10m).In a suitable detector (see 4.3.2), X rays of different energieswill produce output pulses of different amplitudes. Aftersuitable amplification, these pulses can be sorted on the basisof their amplitudes and stored in ce

21、rtain designated channels ofa multichannel analyzer, each adjacent channel representing anincrement of energy. Typically, a channel may represent a spanof 20 eV for a lithium-drifted silicon detector or 150 to 200 eVfor a proportional counter. From six to sixty adjacent channelscan be used to store

22、the pulses representing a selectedcharacteristic emission of one element, the number of channelsdepending on the width of the emission peak (usually displayedon the face of a cathode ray tube). The adjacent channels usedto store the pulses from the material under analysis are calledthe “region of in

23、terest” or ROI.4.3 Detection:4.3.1 Wavelength Dispersive SystemsThe intensity of awavelength is measured by means of an appropriate radiationdetector in conjunction with electronic pulse-countingcircuitry, that is, a scaler. With wavelength dispersive systems,the types of detectors commonly used as

24、the gas-filled typesand the scintillation detector coupled to a photomultiplier tube.4.3.2 Energy-Dispersive SystemsFor the highest energyresolution with energy dispersive systems, a solid-state devicesuch as the lithium-drifted silicon detector must be used. Thistype of detector is maintained at a

25、very low temperature in aliquid-nitrogen cryostat (77K). Acceptable energy resolutionfor most thickness measurement requirements can be realizedwith proportional counters, and these detectors are being usedon most of the commercially available thickness gages basedon X-ray spectrometry. In setting u

26、p a procedure for coating-thickness measurement using an energy-dispersive system,consideration should be given to the fact that the detector“sees” and must process not only those pulses of interest butalso those emanating from the substrate and from supportingand masking materials in the excitation

27、 enclosure. Therefore,consideration should be given to restricting the radiation to thearea of interest by masking or collimation at the radiationsource. Similarly, the detector may also be masked so that itwill see only that area of the specimen on which the coatingthickness is to be determined.4.4

28、 Basic PrincipleA relationship exists between coatingthickness and secondary radiation intensity up to the limitingthickness mentioned in 1.2. Both of the techniques describedbelow are based on the use of primary standards of knowncoating thicknesses which serve to correlate quantitatively theradiat

29、ion intensity and thickness.4.5 Thickness Measurement by X-Ray EmissionIn thistechnique, the spectrogoniometer is positioned to record theintensity of a prominent wavelength characteristic of thecoating metal or, in the case of an energy-dispersive system,the multichannel analyzer is set to accept t

30、he range of energiescomprising the desired characteristic emission. The intensity ofthe coatings X-ray emission (coating ROI) will be at aminimum for a sample of the bare substrate where it willconsist of that portion of the substrate fluorescence which mayoverlap the ROI of the coating and a contri

31、bution due tobackground radiation. This background radiation is due to theportion of the X-ray tubes output which is the same energy asthe coatings X-ray emission. The sample will always scattersome of these X rays into the detector. If the characteristicemission energies of the coating and substrat

32、e are sufficientlydifferent, the only contribution of the substrate will be due tobackground. For a thick sample of the solid coating metal orfor an electroplated specimen having an “infinitely thick”coating, the intensity will have its maximum value for a givenset of conditions. For a sample having

33、 a coating of less than“infinite” thickness, the intensity will have an intermediatevalue. The intensity of the emitted secondary X radiationdepends, in general, upon the excitation energy, the atomicnumbers of the coating and substrate, the area of the specimenexposed to the primary radiation, the

34、power of the X-ray tube,and the thickness of the coating. If all of the other variables arefixed, the intensity of the characteristic secondary radiation isa function of the thickness or mass per unit area of the coating.The exact relationship between the measured intensity and thecoating thickness

35、must be established by the use of standardshaving the same coating and substrate compositions as thesamples to be measured. The maximum thickness that can bemeasured by this method is somewhat less than what is,effectively, infinite thickness. This limiting thickness depends,in general, upon the ene

36、rgy of the characteristic X-ray and thedensity and absorption properties of the material under analy-sis. The typical relationship between a coating thickness andthe intensity of a characteristic emission from the coating metalis illustrated by the curve in the Appendix, Fig. X1.1.4.6 Thickness Meas

37、urements by X-Ray AbsorptionIn thistechnique the spectrometer, in the case of a wavelength-dispersive system, is set to record the intensity of a selectedemission characteristic of the basis metal. In an energy-dispersive system, the multichannel analyzer is set to accumu-late the pulses comprising

38、the same energy peak. The intensitywill be a maximum for a sample of the uncoated basis metaland will decrease with increasing coating thickness. This isbecause both the exciting and secondary characteristic radia-tions undergo attenuation in passing through the coating.Depending upon the atomic num

39、ber of the coating, when thecoating thickness is increased to a certain value, the character-istic radiation from the substrate will disappear, although aB568 98 (2014)2certain amount of scattered radiation will still be detected. Themeasurement of a coating thickness by X-ray absorption is notappli

40、cable if an intermediate coating is present because of theindeterminate absorption effect of intermediate layer. Thetypical relationship between coating thickness and the intensityof a characteristic emission from the substrate is shown in theAppendix, see Fig. X1.2.4.7 Thickness and Composition Mea

41、surement by Simultane-ous X-ray Emission and Absorption (Ratio Method)It ispossible to combine the X-ray absorption and emission tech-niques when coating thicknesses and alloy composition aredetermined from the ratio of the respective intensities ofsubstrate and coating materials. Measurements by th

42、is ratiomethod are largely independent of the distance between testspecimen and detector.4.8 Multilayer MeasurementsMany products have multi-layer coatings in which it is possible to measure each of thecoating layers by using the multiple-energy-region capabilityof the multichannel analyzer of an en

43、ergy-dispersive system.The measuring methods permit the simultaneous measurementof coating systems with up to three layers. Or the simultaneousmeasurement of thickness and compositions of layers with upto three components. Such measurements require unique dataprocessing for each multilayer combinati

44、on to separate thevarious characteristic emissions involved, to account for theabsorption by intermediate layers, and to allow for anysecondary excitation which may occur between layers. Typicalexamples of such combinations are gold on nickel on copperand nickel on copper on steel.4.9 Mathematical D

45、econvolutionWhen using a multi-channel analyzer a mathematical deconvolution of the second-ary radiation spectra can be used to extract the intensities of thecharacteristic radiation. This method can be used when theenergies of the detected characteristic radiations do not differsufficiently (for ex

46、ample, characteristic radiation from Au andBr). This method sometimes is described as numerical filteringin order to distinguish from the technique of setting fixedRegion of Interest (ROI) channel limits in the multichannelanalyzer.5. Significance and Use5.1 This is a sensitive, noncontact, and nond

47、estructivemethod for measuring the coating thickness (and in somecases, coating composition) of metallic and some nonmetalliccoatings over a range of thicknesses from as little as 0.01 mto as much as 75 m depending on the coating and substratematerials. It can be used to measure coating and base com

48、bi-nations that are not readily measured by other techniques.5.2 The coating thickness is an important factor in theperformance of a coating in service.6. Factors Affecting Accuracy6.1 Counting StatisticsThe production of X-ray quanta israndom with respect to time. This means that during a fixedtime

49、 interval, the number of quanta emitted will not always bethe same. This gives rise to the statistical error which isinherent in all radiation measurements. In consequence, anestimate of the counting rate based on a short counting interval(for example, 1 or 2 s) may be appreciably different from anestimate based on a longer counting period, particularly if thecounting rate is low. This error is independent of other sourcesof error such as those arising from mistakes on the part of theoperator or from the use of inaccurate standards. To reduce thestatistica