1、Designation: D4600 95 (Reapproved 2016)Standard Test Method forDetermination of Benzene-Soluble Particulate Matter inWorkplace Atmospheres1This standard is issued under the fixed designation D4600; the number immediately following the designation indicates the year oforiginal adoption or, in the cas
2、e of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method describes the sampling and gravimetricdetermination of benzene-soluble par
3、ticulate matter that hasbecome airborne as a result of certain industrial processes. Thistest method can be used to determine the total weight ofbenzene-soluble materials and to provide a sample that may beused for specific and detailed analyses of the soluble compo-nents.1.2 The limit of detection
4、is 0.05 mg/m3by sampling a 1-m3volume of air.NOTE 1Other volatile organic solvents have been used for thisdetermination and whereas a less toxic solvent for this analysis might bedesirable, the substitution of a solvent other than benzene is unwise at thistime. A tremendous volume of environmental s
5、ampling data based onbenzene-soluble determinations has been accumulated over many years inseveral industries.2Some of the determinations have been used inepidemiological studies. Furthermore, the use of benzene is specified inexisting federal standards.3As a result, it appears imprudent to use adif
6、ferent solvent until the qualitative and quantitative relationship ofanalyses derived from benzene and a substitute solvent is established.With proper care, benzene can be safely used in the laboratory.1.3 The values stated in SI units are to be regarded asstandard. No other units of measurement are
7、 included in thisstandard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations pr
8、ior to use.2. Referenced Documents2.1 ASTM Standards:4D1356 Terminology Relating to Sampling and Analysis ofAtmospheres3. Terminology3.1 Definitions:3.1.1 For definitions of terms used in this test method, referto Terminology D1356.4. Summary of Test Method4.1 By the use of a personal sampling pump,
9、 air is pulledthrough a glass-fiber filter held in a sampling cassette. The filteris extracted ultrasonically with benzene. After filtration, analiquot of the total extract is evaporated to dryness and theresidue weighed.5. Significance and Use5.1 This test method provides a means of evaluating expo
10、-sures to benzene-soluble particulate matter in a concentrationrange that can be related to occupational exposures.6. Interferences6.1 This test method is, by definition, free of interferences.However, this test method is nonspecific. It supplies noinformation on the composition of the soluble mater
11、ial. Itmeasures all those substances in the sample that are soluble inbenzene. The composition must be determined by some inde-pendent means.6.2 The greatest errors most likely to be incurred in the useof this test method are associated with the sample collection orhigh blanks from solvent and filte
12、r. Avoid contamination fromextraneous material by using high-purity solvents for the finalcleaning of all apparatus. The use of plastic containers other1This test method is under the jurisdiction of ASTM Committee D22 on AirQuality and is the direct responsibility of Subcommittee D22.04 on Workplace
13、 AirQuality.Current edition approved May 15, 2016. Published May 2016. Originallyapproved in 1986. Last previous edition approved in 2010 as D4600 95 (2016).DOI: 10.1520/D4600-95R16.2“Benzene-Soluble Compounds in Air, P Gelman Acrodisc CR disposable syringefilter (0.45 to 1.0 m pore size) attached t
14、oa5to10-mLsyringe having a Luer tip; or other equivalent suitable media.NOTE 2When using a disposable syringe filter, prewash with 10 mLof acetonitrile.NOTE 3Placing a pre-extracted filter (37-mm glass fiber) ahead of thefritted glass in the Allihn tube will increase the life of the frit.7.6 Microba
15、lance, capable of weighing to 1 g. (If entiresample is to be used for the benzene-soluble determination, afive-place balance is acceptable.)7.7 Weighing Cups, aluminum or TFE-fluorocarbon of suit-able capacity with a weight not to exceed 500 mg.7.8 Vacuum Oven.8. Reagent8.1 BenzeneThe benzene used s
16、hould be of sufficientpurity so that the residue weight following evaporation of 5 mLof the solvent is less than 10 g.9. Safety Precautions9.1 To minimize the exposure hazard to benzene, analystsshould avoid inhalation of solvent vapors and, if indicated, bemonitored to determine the magnitude of th
17、eir exposures.Transfer and evaporation of the solvent should be carried out ina properly ventilated fume hood (150-fpm face velocity).Protective gloves should be used to minimize the possibility ofabsorbing benzene through the skin.9.2 The extract must be handled with proper caution be-cause some po
18、lycyclic aromatic hydrocarbons are potentialcarcinogens.10. Sampling10.1 Place the cassette assembly, in the closed face mode, inthe breathing zone of the employee and for the entire shift,draw air through the filter using a calibrated sampling pump.The pump should be operated at a rate of about 2 L
19、/min andshould be checked periodically to ensure maintenance of thisrate.10.2 Upon completion of sampling, seal the cassette withthe plugs provided and send it to the laboratory for analysis.The cassette must be protected from exposure to heat and light.To minimize deterioration of samples prior to
20、analysis, theymay be stored in a freezer.10.3 Field blanks are treated in the same manner as samples.They are opened in the environment to be sampled andimmediately closed and placed with the samples to be sent tothe laboratory for analysis. An unopened, unused cassetteassembly is used as a laborato
21、ry blank. At least one laboratoryblank and one field blank should be submitted with each set ofsamples collected.11. Calibration and Standardization11.1 No standards are required for this test method.12. Procedure12.1 Remove the filter from the cassette and, using forceps,insert into the extraction
22、vessel. Treat the blanks in the samemanner as the samples.12.2 Pipet 5.0 mLof benzene into the extraction vessel. Sealimmediately with a TFE-fluorocarbon-lined cap if an aliquot ofthe sample is to be used for analysis.12.3 Place the extraction vessel in the ultrasonic bath for 5min and then allow th
23、e vessel to stand for 30 min. If the entiresample is to be used for the benzene-soluble determination,proceed to 12.8.12.4 Decant the solution into the filter vessel. Immediatelypressure or vacuum filter the solution through the filter into aclean, dry collection vessel. The entire 5-mL volume ofben
24、zene solution need not be recovered; however, evaporationlosses must be minimized. On completion of filtration, seal thevessel immediately.12.5 Weigh the weighing cup to the nearest microgram onthe microbalance. Store the weighed cup in a covered glasspetri dish.12.6 Pipet an aliquot no less than 0.
25、5 mL of the solutionfrom 12.4 into the tared weighing cup. Evaporate the solvent ina preheated vacuum oven at 40C and 26.66 kPa and continuedrying for a total of 3 h.NOTE 4It is recommended to air evaporate benzene to dryness undera hood before transferring to the preheated vacuum oven. This willmin
26、imize benzene exposures in the laboratory.12.7 After cooling, reweigh the cup against the same tareused in 12.5. The difference between the weight obtained andthe tare obtained in 12.5 is the residue weight for the aliquot.12.8 When entire sample is to be used for the benzene-soluble determination,
27、decant the solution into the filteringvessel. Pressure filter the solution into a preweighed (60.01mg) 50-mL beaker. Rinse filtration apparatus with three 1-mLwashes of benzene, directly into the beaker.12.9 Evaporate the solvent in a preheated vacuum oven at40C and 26.66 kPa and continue drying for
28、 a total of 3 h.Reweigh the beaker after cooling. The difference between theD4600 95 (2016)2final weight and the tare weight of the beaker is the residueweight of the sample.13. Calculation13.1 Calculate the air concentration of soluble particulateorganic matter as follows:C 5A 2 B!V5F(1)where:C = c
29、oncentration, mg/m3,A = residue weight of sample, mg,B = residue weight of blank, mg,V = volume of air sample, m3, andF = volume of aliquot, mL.14. Precision and Bias514.1 Precision, Repeatability (Both Single Analyst andOverall)Ahigh-volume air sample was collected on a filter ata coke oven battery
30、 and extracted with benzene. The extractwas used to generate four levels of benzene soluble material:0.09, 0.18, 0.36, and 0.72 mg/sample. Each concentration wasanalyzed in triplicate by nine participating laboratories. Thesingle operator relative standard deviation (RSD) varied from25 % to 5 % with
31、 a pooled relative standard deviation of 13 %.The overall relative standard deviation for the nine laboratoriesvaried from 47 % to 10 % with a pooled RSD of 32 %.14.2 BiasThe bias of this method has not been determinedbecause the reference values were established by the methoditself.15. Keywords15.1
32、 air monitoring; benzene-soluble particulate matter;coal tar pitch volatiles; sampling and analysis; workplaceatmospheresASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentionedin this standard. Users of this standard are expres
33、sly advised that determination of the validity of any such patent rights, and the riskof infringement of such rights, are entirely their own responsibility.This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years andif not revised,
34、 either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standardsand should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of theresponsible technical committee, which you may at
35、tend. If you feel that your comments have not received a fair hearing you shouldmake your views known to the ASTM Committee on Standards, at the address shown below.This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,United States.
36、 Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the aboveaddress or at 610-832-9585 (phone), 610-832-9555 (fax), or serviceastm.org (e-mail); or through the ASTM website(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http:/ data have been filed at ASTM International Headquarters and maybe obtained on loan by requesting Research Report RR:D22-1016. Contact ASTMCustomer Service at serviceastm.org.D4600 95 (2016)3